← Product Code [NKB](/productcode/NKB) · K090523

# POLARIS SPINAL SYSTEM (K090523)

_Ebi, L.P. · NKB · Jul 1, 2009 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K090523

## Device Facts

- **Applicant:** Ebi, L.P.
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Jul 1, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion. The Ballista instruments are intended to be used with the 5.5 Polaris implants. The Ballista instruments when used with the Ballista cannulated screws and percutaneous rods, are indicated to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft. The Accuvision Instruments, when used with the Polaris Spinal System implants are indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.

## Device Story

Polaris Spinal System is a line extension of EBI Helical Flange System; provides non-cervical spinal fixation via bone screws, hooks, plugs, rods, and connectors. Includes Ballista and Accuvision instrument sets for percutaneous and minimally invasive posterior spinal surgery. Used by surgeons in clinical settings to immobilize and stabilize spine as adjunct to fusion. Implants constructed of stainless steel. System provides mechanical support to spinal column; aids in fusion process. Benefits include stabilization of spinal segments in patients with degenerative or traumatic conditions.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Non-cervical spinal fixation system; components include bone screws, hooks, plugs, rods, connectors. Material: stainless steel. Design/sizing similar to predicate systems. Mechanical fixation principle.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Helical Flange Spinal System ([K081952](/device/K081952.md), [K061441](/device/K061441.md), [K041449](/device/K041449.md))
- Synergy Spinal System (K01143, [K950709](/device/K950709.md), [K950099](/device/K950099.md))
- Array Spinal System ([K033312](/device/K033312.md), [K061563](/device/K061563.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K090523

Image /page/0/Picture/1 description: The image shows the logo for Biomet Spine. The word "BIOMET" is in large, block letters at the top of the image. Below that, the word "SPINE" is in smaller, block letters. The logo is simple and easy to read.

# JUL - 1 2009

## 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.

Preparation Date:

Applicant/Sponsor:

July 1, 2009 Biomet Spine 100 Interpace Parkway Parsippany, NJ 07054

Phone: 973-299-9300 x2214

Vivian Kelly

Fax: 973-257-0232

Polaris Spinal System

Contact Person:

Trade name:

Common Name:

Classification Name (Product Code):

Device Panel - Regulation No .:

### Device Description:

This submission is a line extension to the EBI Helical Flange System to add new 5.5 and 6.35 Polaris components in order to build various types of spinal constructs. The new components include various types of bone screws, bone hooks, plugs, rods, connectors and accessories in stainless steel.

Non-cervical spinal fixation system

Posterior, noncervical, nonpedicle use (KWP)

Anterior/anterolateral noncervical use (KWQ)

Noncervical pedicle applications (MNI, MNH and NKB) Orthopedic - 21 CFR 888.3050, 888.3060 and 888.3070

### Indications for Use:

The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation systerior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis), turnor, stenosis, pseudarthrosis, and failed previous fusion.

The Ballista instruments are intended to be used with the 5.5 Polaris implants. The Ballista instruments when used with the Ballista cannulated screws and percutaneous rods, are indicated to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.

The Accuvision Instruments, when used with the Polaris Spinal System implants are indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery for the following indications: 

{1}------------------------------------------------

degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.

### Summary of Technologies:

The technological characteristics (material, design and sizing) of the Polaris components are the same as, or similar to, the predicate devices.

### Substantial Equivalence:

The Polaris System is substantially equivalent to its predicate devices with respect to intended use and indications, technological characteristics, and principles of operation and do not present any new issues of safety or effectiveness. Examples of predicates include Helical Flange Spinal System (K081952, K061441and K041449), the Synergy Spinal System (K01143, K950709 & K950099) and the Array Spinal System (K033312 & K061563.) Based upon the mechanical testing, the Polaris Spinal System is substantially equivalent for its intended use to other spinal systems currently on the market.

{2}------------------------------------------------

:

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle-like symbol with three wing-like shapes.

### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL - 1 2009

## EBI LP

% Ms. Vivian Kelly Regulatory Affairs Project Manager 100 Interpace Parkway Parsippany, New Jersey 07054

Re: K090523

Trade/Device Name: Polaris Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: III Product Code: NKB, KWQ, KWP, MNH, MNI Dated: June 17, 2009 Received: June 18, 2009

Dear Ms. Kelly:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 – Ms. Vivian Kelly

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Barbara Myers

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications for Use

## 510(k) Number (if known): K090523

Device Name: Polaris Spinal System

Indications for Use:

The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion.

The Ballista instruments are intended to be used with the 5.5 Polaris implants. The Ballista instruments when used with the Ballista cannulated screws and percutaneous rods. are indicated to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.

The Accuvision Instruments, when used with the Polaris Spinal System implants are indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), tumor; stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/liac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH. Office of Device Evaluation (ODE)

---

(Division Sign-Off) (TEXT for MKM)

Page 1 of 1

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K090523

---

**Source:** [https://fda.innolitics.com/device/K090523](https://fda.innolitics.com/device/K090523)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).