← Product Code [FRC](/productcode/FRC) · K090514

# VERIFY SELF-CONTAINED BIOLOGICAL INDICATOR FOR VAPORIZED (VH2O2) STERILIZATION PROCESSES (K090514)

_STERIS Corporation · FRC · Mar 16, 2009 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K090514

## Device Facts

- **Applicant:** STERIS Corporation
- **Product Code:** [FRC](/productcode/FRC.md)
- **Decision Date:** Mar 16, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 880.2800
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Verify V24 Self-Contained Biological Indicator is intended as a standard method for frequent monitoring of the Amsco V-PRO 1 Low Temperature Sterilization System.

## Device Story

Verify V24 Self-Contained Biological Indicator (SCBI) monitors Amsco V-PRO 1 Low Temperature Sterilization System performance. User places SCBI in sterilizer with medical devices; after cycle, SCBI is activated using STERIS Verify SCBI HP activator, which seals vial and releases growth media. Activated SCBI incubated at 55-60°C. Pass indicated by orange, non-turbid media; failure indicated by yellow, turbid media. Provides healthcare providers with verification of sterilization cycle efficacy.

## Clinical Evidence

Bench testing only. Testing included growth inhibition, sterilization process effect on recovery media, incubation time validation (24 hours), media performance, and media stability. All tests passed.

## Technological Characteristics

Self-contained biological indicator vial containing growth media and biological spores. Operates via colorimetric/turbidity change upon incubation at 55-60°C. Modified media formulation for 24-hour incubation.

## Regulatory Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

## Predicate Devices

- Verify® Self-Contained Biological Indicator for Vaporized VH2O2 Sterilization Processes ([K073244](/device/K073244.md))
- STERRAD® Cyclesure™ Biological Indicator ([K994055](/device/K994055.md))
- STERRAD® Cyclesure Biological Indicator ([K031226](/device/K031226.md))
- STERRAD® Cyclesure Biological Indicator ([K071014](/device/K071014.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# MAR 1 6 2009

## 1. Device Name

| Trade Name:          | Verify® V24 Self-Contained Biological Indicator                        |
|----------------------|------------------------------------------------------------------------|
| Common/usual Name:   | Biological Indicator (BI)                                              |
| Classification Name: | Indicator, Biological Sterilization Process<br>(21 CFR 880.2800, FRC). |

## 2. Predicate Device

- Verify® Sclf-Contained Biological Indicator for Vaporized VH2O2 트 Sterilization Processes, K073244, May 28, 2008
- STERRAD® Cyclesure™ Biological Indicator, K994055, Feb 13, 2002 트
- STERRAD® Cyclesure Biological Indicator, K031226, May 2, 2003 ■
- STERRAD® Cyclesure Biological Indicator, K071014 May 24, 2007 트

## 3. Description of Device

The Verify V24 Self-Contained Biological Indicator (SCBI) is used by healthcare providers to monitor the Amsco® V-PROTM 1 Low Temperature Sterilizer. It is designed to accompany medical devices placed in the sterilizer.

The user places the packaged Verify V24 SCBI into the V-PRO 1 Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation.

The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS Verify SCBI HP activator. The activator automatically seals the SCBI vial and releases the growth media.

The activated SCBI is incubated at 55 - 60 ℃. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.

The modified Verify V24 Self-Contained Biological Indicator is identical in design to the original K073244 device with exception of the media formulation. It is identical with respect to performance specifications with exception of the incubation time.

{1}------------------------------------------------

#### Intended Use 4.

The Verify V24 Self-Contained Biological Indicator is intended as a standard method for frequent monitoring of the Amsco V-PRO 1 Low Temperature Sterilization System.

#### 5. Description of Modification

The culture media has been modified to achieve a 24 hour incubation time and the device labeling has been modified to reflect the change in incubation period as well as a change in the media color.

#### 6. Description of Safety and Substantial Equivalence

The Verify V24 Self-Contained Biological Indicator has the same or similar intended use, accessories, viable population, resistance characteristics, culture conditions, primary and secondary packaging, and storage conditions as compared to its predicate devices the Verify® Self-Contained Biological Indicator for Vaporized VH2O2 Sterilization Processes and STERRAD CycleSure Biological Indicator.

Summary of nonclinical tests performed to confirm the performance of the modified culture media:

| Test                                                                                | Result                                                                                                                                                                                                                       |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Growth Inhibition Testing & Effect of<br>Sterilization Process on Recovery<br>Media | Pass<br>There was no growth inhibition and no<br>effect of the sterilization process on the<br>media.                                                                                                                        |
| Incubation Time Validation                                                          | Pass<br>24 hour Incubation                                                                                                                                                                                                   |
| Media Performance Evaluations                                                       | Pass<br>All viability control SCBIs displayed<br>growth within 24 hrs.  Media was<br>present and the color remained yellow<br>for 7 days.<br>All uninoculated SCBIs were sterile and<br>the media remained orange for 7 days |
| Media Stability Evaluation                                                          | Pass<br>On Going Stability Evaluation                                                                                                                                                                                        |

The test results indicate that the proposed Verify V24 Self-Contained Biological Indicator is substantially equivalent to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its wings. The eagle is positioned to the right of the department's name, which is arranged in a circular fashion around the left side of the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA".

MAR 1 6 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. John Robert (Jack) Scoville, Jr Fellow, Regulatory Affairs STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060

Re: K090514

> Trade/Device Name: Verify® V24 Self-Contained Biological Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: FRC Dated: February 25, 2009 Received: February 26, 2009

Dear Mr. Scoville:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 -Mr. Scoville

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Sutte Y. Michael Dms.

Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K090514

Device Name:

Verify® V24 Self-Contained Biological Indicator

Indications For Use:

The Verify® V24 Self-Contained Biological Indicator is intended as a standard method for frequent monitoring of the Amsco® V-PRO™ 1 Low Temperature Sterilization System.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_X (21 CFR 801 Subpart C)

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Shule M. Murphy, MD

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: _

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K090514](https://fda.innolitics.com/device/K090514)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
