K090462 · Philips Medical Systems (Cleveland), Inc. · JAK · Jun 23, 2009 · Radiology
Device Facts
Record ID
K090462
Device Name
BRILLIANCE DUAL ENERGY OPTION
Applicant
Philips Medical Systems (Cleveland), Inc.
Product Code
JAK · Radiology
Decision Date
Jun 23, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Intended Use
The Brilliance CT is a Computed Tomography X-Ray System intended to produce crosssectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories. The dual energy option allows the system to acquire two CT images of the same anatomical location using two distinct tube voltages and/or tube currents during two tube rotations. The x-ray dose will be the sum of the doses of each tube rotation at its respective tube voltage and current. Information regarding the material composition of various organs, tissues, and contrast materials may be gained from the differences in x-ray attenuation between these distinct energies. This information may also be used to reconstruct images at multiple energies within the available spectrum, and to reconstruct basis images that allow the visualization and analysis of anatomical and pathological materials.
Device Story
Brilliance Dual Energy option is an add-on for Philips Brilliance CT scanners; automates sequential scanning protocols using two distinct tube voltages/currents during two rotations. Input: x-ray transmission data at two energy levels. Processing: computer reconstruction of cross-sectional images; analysis of attenuation differences between energy levels. Output: side-by-side or overlaid images; basis images for material composition visualization. Used in clinical settings by radiologists/technicians. Output assists clinicians in identifying anatomical/pathological structures and characterizing tissue/contrast material composition. Benefits: enhanced diagnostic information regarding material composition compared to single-energy CT.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Computed Tomography X-Ray System (21 CFR 892.1750). Dual-energy acquisition via sequential tube rotations at distinct voltages/currents. Software-based image reconstruction and analysis of attenuation differences. Connectivity via standard CT system architecture.
Indications for Use
Indicated for patients requiring cross-sectional imaging of the body via CT. Used for anatomical and pathological structure visualization, including material composition analysis of organs, tissues, and contrast agents using dual-energy attenuation differences.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K152036 — SOMATOM Definition Edge, SOMATOM Definition AS/AS+ · Siemens Medical Solutions USA, Inc. · Oct 9, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
K09c462
Page 1 of 2
# 510(k) Summary Philips Medical Systems (Cleveland) Inc. Brilliance Dual Energy option
This summary of this 510(k) provides safety and effectiveness information submitted in accordance with the requirements of 21 CFR 807.92.
#### Submitter 1.
Philips Medical Systems (Cleveland), Inc. 595 Miner Road Cleveland, OH 44143 (440) 483-3000
Contact:
| | Michael A. Chilbert, Ph.D., P.E. |
|------------|----------------------------------|
| Telephone: | 440 483-3284 |
| FAX: | 440 483 2989 |
| Email: | Michael.Chilbert@philips.com |
# Date of Summary: January 7, 2009
## Device Name and Classification 2.
| Device Name: | Brilliance Dual Energy option |
|----------------------|----------------------------------|
| Classification Name: | Computed Tomography X-Ray System |
The FDA has classified the Computed Tomography X-Ray System and its accessories as Class II in 21 CFR 892.1750 (Product Code 90 JAK)
#### 3. Predicate Device Information
In the opinion of Philips Medical Systems Inc., the Brilliance Dual Energy option is of comparable type and substantially equivalent to the legally marketed devices currently in commercial distribution, namely: the Philips Brilliance Volume - K060937 and the GE Lightspeed CT750 HD - K081105.
#### 4. Device Description
Philips Healthcare offers a Dual Energy scanning option on the Brilliance CT Scanner. The Brilliance Dual Energy option automates the execution of sequential scanning protocols acquired during the same episode of care using two unique tube voltages and/or currents. The acquired datasets can be displayed side-by-side or overlaid and then analyzed to augment the review of anatomical and pathological structures. Dual energy imaging, by nature of differing x-ray energy values, enables the identification of attenuation differences found in those structures between the two applied energies.
JUN 2 3 2009
{1}------------------------------------------------
K09c462
Page 2 of 2
### Intended Use of the device 5.
The Brilliance CT is a Computed Tomography X-Ray System intended to produce crosssectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
The dual energy option allows the system to acquire two CT images of the same anatomical location using two distinct tube voltages and/or tube currents during two tube rotations. The x-ray dose will be the sum of the doses of each tube rotation at its respective tube voltage and current. Information regarding the material composition of various organs, tissues, and contrast materials may be gained from the differences in xray attenuation between these distinct energies. This information may also be used to reconstruct images at multiple energies within the available spectrum, and to reconstruct basis images that allow the visualization and analysis of anatomical and pathological materials.
### Comparison to Predicate Devices 6.
In the opinion of Philips Medical Systems (Cleveland), Inc., the Brilliance CT scanner with Dual Energy option is of comparable type and substantially equivalent to the legally marketed devices because it has the similar technological characteristics and subassemblies as the current commercial distribution of Philips Brilliance Volume (K060937) and the GE Lightspeed CT750 HD (K081105).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 3 2009
Michael Chilbert, Ph.D., P.E. Quality & Regulatory Engineer Philips Medical Systems (Cleveland), Inc. 595 Miner Road CLEVELAND OH 44143
Re: K090462
Trade/Device Name: Brilliance Dual Energy option Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: June 5, 2009 Received: June 9, 2009
## Dear Dr. Chilbert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrlv/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.lttml.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
KO90462
Page 1 of 1
Device Name:
Brilliance Dual Energy option
The Brilliance CT is a Computed Tomography X-Ray Indications for Use: System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories,
The dual energy option allows the system to acquire two CT images of the same anatomical location using two distinct tube voltages and/or tube currents during two tube rotations. The x-ray dose will be the sum of the doses of each tube rotation at its respective tube voltage and current. Information regarding the material composition of various organs, tissues, and contrast materials may be gained from the differences in x-ray attenuation between these distinct energies. This information may also be used to reconstruct images at multiple energies within the available spectrum, and to reconstruct basis images that allow the visualization and analysis of anatomical and pathological materials.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)ivision of Reproductive. Abdomin: and Radiological Devi 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.