← Product Code [DTZ](/productcode/DTZ) · K090450

# MEDOS HILITE PEDIATRIC OXYGENATOR, MODELS 2800, 2400 LT (K090450)

_Gish Biomedical, Inc. · DTZ · Sep 18, 2009 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K090450

## Device Facts

- **Applicant:** Gish Biomedical, Inc.
- **Product Code:** [DTZ](/productcode/DTZ.md)
- **Decision Date:** Sep 18, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4350
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The MEDOS HILITE 2800 & 2400 LT Hollow Fiber Oxygenators are indicated for use in procedures requiring the extracorporeal oxygenation of and carbon dioxide removal from human blood. The 2800 & 2400 LT are pediatric oxygenators intended for use at blood flow rates of 0.5 to 2.8 liters per minute for the HILITE 2800 and 0.5 to 2.4 liters per minute for the HILTE 2400 LT for periods of up to six (6.0) hours. Federal Law (US) restricts this device to sale by or on the order of a physician.

## Device Story

Pediatric hollow fiber membrane oxygenator with integrated heat exchanger; used in extracorporeal circuits. HILITE 2800 uses polypropylene gas-permeable mat; HILITE 2400 LT uses polymethylpentene gas plasma-tight mat. Heat exchanger utilizes polyester non-porous hollow fibers within polycarbonate housing. Blood flows outside fibers; water flows through inner lumen for heat exchange. Designed to increase blood-gas interaction efficiency while minimizing priming volume. Used in clinical settings by medical professionals during cardiopulmonary bypass procedures to oxygenate blood and remove carbon dioxide.

## Clinical Evidence

Bench testing only. Device underwent extensive safety, performance, and validation testing per FDA guidance for cardiopulmonary bypass oxygenators to ensure functional requirements and industry safety standards were met.

## Technological Characteristics

Hollow fiber membrane oxygenator with integrated heat exchanger. Materials: polypropylene (2800) or polymethylpentene (2400 LT) gas-permeable mats; polyester non-porous hollow fibers (heat exchanger); polycarbonate housing. Dimensions/form factor: pediatric. Energy source: none (passive). Sterilization: not specified.

## Regulatory Identification

A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”

## Predicate Devices

- CAPIOX® RX15 Hollow Fiber Oxygenator ([K051997](/device/K051997.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MEDOS Medizintechnik AG

# ્વિક્તિત્વવાયુ

Traditional 510(k) Section 5, 510(k) Summary

# SEP 1 8 2009

Summary

This summary of 510(k)-safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.

# 1. Company making the submission:

| Name:      | GISH Biomedical, Inc.<br>A member of the MEDOS group                |
|------------|---------------------------------------------------------------------|
| Address:   | 22942 Arroyo Vista<br>Rancho Santa Margarita, CA 92688-<br>2600     |
| Telephone: | 949-635-6200 voice<br>949-635-6299 fax<br>janetp@gishbiomedical.com |
| Contact:   | Janet Peets<br>Regulatory & Clinical Affairs                        |

- 2. Device:
Proprietary Name: Common Name: Classification Name:

MEDOS HILITE 2800 & 2400 LT Oxygenator Blood Oxygenator Oxygenator, Cardiopulmonary Bypass

### 3. Predicate Devices:

CAPIOX® RX15 Hollow Fiber Oxygenator, K051997, Manufactured by Terumo

#### Classifications Names & Citations: 4. 21 CFR 870.4350, Oxygenator, Cardiopulmonary Bypass, Class II, DTZ, Cardiovascular

### 5. Description:

The MEDOS HILITE 2800 & 2400 LT Hollow Fiber Oxygenators consist of a hollow fiber membrane oxygenator and extracorporeal heat exchanger. The MEDOS HILITE 2800 hollow fiber membrane consists of a polypropylene gas permeable mat. The MEDOS HILITE 2400 LT hollow fiber membrane consists of a polymethylpentene gas plasma tight mat. The unique mat design increases the interaction between blood and gas, creating a highly efficient blood oxygenator. The heat exchanger consists of a polyester non-porous hollow fiber configured heat exchanger as the primary element to affect heat exchange. This element is encased by a polycarbonate housing, which directs the blood around the outside of the fibers while water flows through the inner lumen of the fibers and therefore effects heat exchange while minimizing priming volume.

# 6. Indications for use:

The MEDOS HILITE 2800 & 2400 LT Hollow Fiber Oxygenators are indicated for use in procedures requiring the extracorporeal oxygenation of and carbon dioxide removal from human blood. The 2800 & 2400 LT are pediatric oxygenators intended for use at blood flow rates of 0.5 to 2.8 liters per minute for the HILITE 2800 and 0.5 to 2.4 liters per minute for the HILTE 2400 LT for periods of up to six (6.0) hours.

{1}------------------------------------------------

#### 7. Contra-indications:

For oxygenators, no contra-indications have been noted.

#### Comparison: 8.

The MEDOS HILITE 2800 & 2400 LT Oxygenators have the same device characteristics as the predicate device.

## 9. Test Data:

In accordance with FDA Guidance for Cardiopulmonary Bypass Oxygenators 510(k) submissions, the MEDOS HILITE 2800 and 2400 LT Oxygenators have completed extensive safety, performance, and validation testing prior to release. Final testing for the systems includes various performance tests designed to ensure that the device meets all of its functional requirements and performance specifications. Safety tests have further been performed to ensure the device complies with applicable industry and safety standards.

# 10. Literature Review:

A review of literature pertaining to the safety and effectiveness has been conducted. Appropriate safeguards have been incorporated in the design of the MEDOS HILITE 2800 & 2400 LT Oxygenator.

### 11. Conclusions:

Based upon the testing and comparison to the predicate device, the MEDOS HILITE 2800 & 2400 LT Oxygenator have the same intended use, with similar technological characteristics. MEDOS Medizintechnik AG therefore posits that its device is equivalent in safety and effectiveness to the predicate device.

Confidential Summary 005.doc

# Section 5 510(k) Submission Page #

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Gish Biomedical, Inc. c/o Ms. Janet Peets Regulatory & Clinical Affairs Specialist 22942 Arroyo Vista Rancho Santa Margarita, CA 92688

#### K090450 Re:

Medos Hilite Pediatric Oxygenator, Models 2800, 2400 LT Regulation Number: 21 CFR 870.4350 Regulation Name: Oxygenator, cardiopulmonary bypass (with heat exchanger) Regulatory Class: Class II Product Code: DTZ Dated: August 25, 2009 Received: August 26, 2009

# Dear Ms. Peets:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

SFP 1 8 2009

{3}------------------------------------------------

# Page 2 - Ms. Janet Peets

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

una E. Vachner

Image /page/3/Picture/7 description: The image shows a black and white drawing of a symbol. The symbol appears to be a stylized letter, possibly a lowercase 'h' or a similar character. The lines are thick and slightly uneven, giving it a hand-drawn appearance. The symbol is simple and abstract, with a curved shape at the top and a straight line at the bottom.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# MEDOS Medizintechnik AG

510(k) Number K090450

1

Device Name: MEDOS HILITE 2800 & 2400 LT Oxygenator

## Indications for use:

The MEDOS HILITE 2800 & 2400 LT Hollow Fiber Oxygenators are indicated for use in procedures requiring the extracorporeal oxygenation of and carbon dioxide removal from human blood. The 2800 & 2400 LT are pediatric oxygenators intended for use at blood flow rates of 0.5 to 2.8 liters per minute for the HILITE 2800 and 0.5 to 2.4 liters per minute for the HILTE 2400 LT for periods of up to six (6.0) hours.

## Prescription Device:

Federal Law (US) restricts this device to sale by or on the order of a physician.

Prescription Use : Yes OR

Over-The-Counter Use: No

PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Duma D. Laines

(Division Sign-Off) Division it Cardiovascular Devices

510(K) Number_K690450

Confidential Indications forUse 004.doc Section 4 510(k) Submission

---

**Source:** [https://fda.innolitics.com/device/K090450](https://fda.innolitics.com/device/K090450)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
