XPI ANGLED ABUTMENTS

K090397 · Keystone Dental, Inc. · NHA · Jun 5, 2009 · Dental

Device Facts

Record IDK090397
Device NameXPI ANGLED ABUTMENTS
ApplicantKeystone Dental, Inc.
Product CodeNHA · Dental
Decision DateJun 5, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Intended Use

The XP1 Angled Abutment is intended for use in conjunction with the XP1 Dental Implant System in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit cement retained restorations. The angled abutment is indicated in cases where angle correction is required.

Device Story

XP1 Angled Abutment serves as interface between dental implant and prosthetic restoration; corrects prosthetic angulation for implants placed off-axis to occlusal load. Device features hex and taper design for engagement with internal implant geometry; secured via abutment screw. Used by dental professionals in clinical settings to support single or multiple-unit cement-retained crowns in edentulous mandibles and maxillae. Provides structural support for dental prosthetics; enables functional restoration in cases of non-ideal implant placement.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Constructed from Ti-6AL-4V ELI titanium alloy per ASTM F136. Features hex and taper distal interface for implant engagement and screw channel for fixation. Non-sterile, provided in medical grade packaging.

Indications for Use

Indicated for partially or fully edentulous patients requiring dental implant restorations where angle correction is necessary for single or multiple-unit cement-retained crowns.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Additional Information: K090397 XP1 Angled Abutments JUN - 5 2009 # 510(k) Summary # Keystone Dental. Inc. XP1® Angled Abutments ### ADMINISTRATIVE INFORMATION Manufacturer Name: Keystone Dental, Inc. 144 Middlesex Turnpike Burlington, MA 01803 Telephone: 1 (781) 272-9272 Fax: 1 (781) 272-9972 Official Contact: Carolyn Bitetti SVP, Quality/Regulatory Keystone Dental. Inc. Direct Line: 781-328-3305 Fax: 781-272-9972 Email: cbitetti@keystonedental.com #### DEVICE NAME AND CLASSIFICATION Trade/Proprietary Name: Common Name: Classification Regulations: XP1 Angled Abutments Abutment for dental implant Endosseous dental implant abutment (21 CFR 872.3630), Class II NHA Product Codes ## DEVICE CLASSIFICATION PANEL The Classification Panel for these devices is the Dental Products Panel: the devices are reviewed by the Dental Devices Branch. #### INTENDED USE The XP1 Angled Abutment is intended for use in conjunction with the XP1 Dental Implant System in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit cement retained restorations. The angled abutment is indicated in cases where angle correction is required. {1}------------------------------------------------ ## DEVICE DESCRIPTION The XP1 15° Angled Abutment is intended to be placed into a dental implant to support prosthetic restorations in partially or fully edentulous mandibles and maxillae. The angled abutment can be used for single or multiple-unit restorations and is intended for cement-retained crowns. The angled abutment is used to correct the prosthetic anqulation of implants that are placed off-axis to the occlusal load. The distal end of the angled abutment has a hex and taper design that engages the mating internal hex and taper of the implant. In addition, the angled abutment has a screw channel with a screw thread at the distal end through which an abutment screw is used to screw the abutment into the implant. The XP1 Angled Abutment and the abutment screws are machined from titanium allov (Ti-6AL-4V ELI) conforming to ASTM F136 Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surqical Implant Applications. The device is provided non-sterile to the user in medical grade packaging. ### EQUIVALENCE TO ORIGINAL CLEARED PRODUCT Keystone Dental, Inc. has demonstrated that, for the purposes of FDA's regulation of medical devices, the XP1 Angled Abutment is substantially equivalent in indications and design principles to predicate devices, each of which has been determined by FDA to be substantially equivalent to preamendment devices. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three wing-like shapes, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle. The text is in all capital letters and is evenly spaced around the circle. #### Public Health Service JUN - 5 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Carolyn Bitetti Senior Vice President, Quality/Regulatory Keystone Dental, Incorporated 144 Middlesex Turnpike Burlington, Masachussets, 01803 Re: K090397 Trade/Device Name: XP1 Angled Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: May 14, 2009 Received: May 15, 2009 Dear Ms. Bitetti: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 – Ms. Bitetti Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/cdrh/comp/ for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Susan Funner Susan Runner, D.D.S., MA Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K040347 XP1 Angled Abutments ## Indications for Use 510(k) Number (if known): K090397 Device Name: XP1® Angled Abutments Indications for Use: The XP1 Angled Abutment is intended for use in conjunction with the XP1 Dental Implant System in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit cement retained restorations. The angled abutment is indicated in cases where angle correction is required. Kevin Nale fr MSR (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices K090397 510(k) Number: Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)