← Product Code [EBF](/productcode/EBF) · K090391

# PULPDENT MICRO-HYBRID COMPOSITE (K090391)

_Pulpdent Corporation · EBF · May 11, 2009 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K090391

## Device Facts

- **Applicant:** Pulpdent Corporation
- **Product Code:** [EBF](/productcode/EBF.md)
- **Decision Date:** May 11, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3690
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K090391 · May 11, 2009 | PULPDENT MICRO-HYBRID COMPOSITE | Pulpdent Corporation | NIH Technology Assessment Conference report on dental restorative materials | The sponsor utilized findings from an NIH Technology Assessment Conference to support the safety and effectiveness of the device by referencing long-term, real-world clinical usage of similar dental composite materials. | NIH Technology Assessment; Long-term clinical usage; Dental restorative materials |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| NIH Technology Assessment Conference on Effects and Side-Effects of Dental Restorative Materials; Technology assessment / Literature review of clinical usage; Follow-up/Duration: Approximately 20 years | General population undergoing dental restorative procedures; Sample Size: Billions of procedures | Not applicable for this study | Safety, benefit-to-risk ratio, and incidence of short-term or long-term risks |

## Indications for Use

Pulpdent Micro-Hybrid Composite is a radiopaque, light-cured, fine-particle hybrid dental composite used for esthetic, direct resin restorations including all cavity classifications (I-VI) in anterior and posterior teeth, Class V restorations, and for direct veneering.

## Device Story

Pulpdent Micro-Hybrid Composite is a radiopaque, light-cured, fine-particle hybrid dental composite. It is used by dental professionals in a clinical setting for direct resin restorations of anterior and posterior teeth, covering all cavity classifications (I-VI) and direct veneering. The material is applied directly to the tooth structure and cured using a light source to harden the resin. This restorative process allows for the repair of tooth decay or structural defects, providing an esthetic and functional restoration for the patient.

## Clinical Evidence

No clinical data. The safety and effectiveness are supported by the substantial equivalence to legally marketed predicate devices and the established clinical history of similar dental restorative materials as cited in the NIH Technology Assessment Conference on Effects and Side-Effects of Dental Restorative Materials.

## Technological Characteristics

Radiopaque, light-cured, fine-particle hybrid dental composite. Formulated for direct resin restorations. No specific ASTM standards or software algorithms are described.

## Regulatory Identification

Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.

## Predicate Devices

- Ivoclar Vivadent 4 Seasons
- Dentsply/Caulk Esthet*X
- GC Gradia Direct

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# MAY 1 1 2009

### EXHIBIT 2

# SUMMARY OF SAFETY AND 🇪두ΕΣΤΙVENESS DATA

Kenneth J. Berk 80 Oakland Street PO Box 780 Watertown, MA 02472 USA

Telephone: 617-926-6666 Fax: 617-926-6262 ken@pulpdent.com

DEVICE:

Trade Name: PULPDENT MICRO-HYBRID COMPC¡SITE Classification Name: Tooth shade resin material FDA Product Code: 76 EBF, 21 CFR Part 872.3690

#### PREDICATE DEVICES:

Ivoclar Vivadent 4 Seasoris Dentsply/Caulk Esthet*X GC Gradia Direct

## DESCRIPTION AND INTENDED USE:

Pulpdent Micro-Hybrid Composite is a radiopaque, light-cured, fine-particle hybrid dental composite for esthetic, direct resin restorations.

Pulpdent Micro-Hybrid Composite is used by the dental professional for the restoration of all cavity classifications (I-VI) in anterior and posterior teeth, Class V restorations, and for direct veneering,

## COMPARISON WITH PREDICATE PRODUCTS:

Pulpdent Micro-Hybrid Composite is substantially equivalent in design, composition, performance, intended use, safety and effectiveness to the predicate products listed above.

#### SAFETY AND EFFECTIVENESS:

Pulpdent Micro-Hybrid Composite is substantially equivalent in design, composition, performance, intended use, safety and effectiveness to the predically products listed above. The products have been found substantially equivalent under the 510(k) Premarket Notification process as Class II Dental Devices under CFR 872.3690.

According to the NH Technology Assessment Conlerence on Effects and Side-Effects of Dental Restorative Materials: "General usage of these materials over about 20 years indicates a high benefit to-risk ration..both composites and glass ionomers are relatively trouble-free. There is no evidence of short-term of long-term risk...There is no suspicion of any problems after virtually billions of procedures in the United States."

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three overlapping wings, symbolizing health, services, and humanism. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle emblem.

Public Health Service

MAY 1 1 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Kenneth J. Berk Director Pulpdent Corporation 80 Oakland Street Watertown, Massachusetts 02472

Re: K090391

Trade/Device Name: Pulpdent Micro-Hybrid Composite Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Code: EBF Dated: February 12, 2009 Received: February 18, 2009

Dear Mr. Berk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 Mr. Berk

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other l'ederal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Parl 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Susan Lumoe

Susan Runner, D.D.S., MA Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Ko90391 510(k) Number (if known):

Device Name: Pulpdent Micro-Hybrid Composite

Indications For Use:

Pulpdent Micro-Hybrid Composite is a radiopaque, light-cured, fine-particle hybrid dental composite used for esthetic, direct resin restorations including all cavity classifications (1-V),) anterior and posterior restorations, Class V restorations and direct veneering.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

K-Mulu for WSR

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of 1

510(k) Number: K090821

---

**Source:** [https://fda.innolitics.com/device/K090391](https://fda.innolitics.com/device/K090391)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com