K090341 · Danmar Products, Inc. · MVA · Jan 6, 2010 · Neurology
Device Facts
Record ID
K090341
Device Name
MICHIGAN CRANIAL RESHAPING ORTHOSIS
Applicant
Danmar Products, Inc.
Product Code
MVA · Neurology
Decision Date
Jan 6, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5970
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Danmar Products Michigan Cranial Reshaping Orthosis is intended for prescription use to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape. A mold is made of the baby's head to create the orthotic. A plaster cast or a laser-scanning accessory may be used to create a scan of the infant's head to create the mold. The device is indicated for infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic and brachycephalic shaped heads.
Device Story
Cranial orthosis (helmet) provides passive pressure to prominent cranial regions to improve symmetry/shape. Input: 3D head scan (via Omega Scanner) or traditional manual plaster cast. Transformation: Orthotist modifies scan file; Danmar uses Omega carver to create positive foam mold; orthosis fabricated with inner hypoallergenic foam and semi-rigid plastic outer shell. Output: Custom-fitted cranial helmet. Used in clinical settings by orthotists; patient wears device to treat plagiocephaly. Benefits: Non-invasive correction of cranial deformities. Manufacturing process updated to include digital scanning/carving, replacing manual plaster mold steps for improved efficiency while maintaining identical clinical function to predicate.
Clinical Evidence
Bench testing only. Biocompatibility data confirms inner lining is non-irritating and nontoxic. Mechanical testing demonstrates device does not break or shatter under impact. No clinical trials performed; equivalence established via technological comparison and performance testing.
Technological Characteristics
Cranial orthosis; inner hypoallergenic foam (1/4" to 1/2" thick), semi-rigid plastic outer shell. 3D shape capture via Class 1 Laser (Omega Scanner). Automated fabrication via Omega Carver. Connectivity: electronic file transfer of scan data. Sterilization: not specified.
Indications for Use
Indicated for infants 3 to 18 months old with moderate to severe nonsynostotic positional plagiocephaly, including plagiocephalic and brachycephalic head shapes. Contraindicated for patients with craniosynostosis, scaphocephalic-shaped heads, or hydrocephalus.
Regulatory Classification
Identification
A cranial orthosis is a device that is intended for medical purposes to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
Special Controls
*Classification.* Class II (special controls) (prescription use in accordance with § 801.109 of this chapter, biocompatibility testing, and labeling (contraindications, warnings, precautions, adverse events, instructions for physicians and parents)).
In addition to the general controls of the act, the Dynamic Orthotic Cranioplasty - DOC™ Band is subject to the following special controls in order to provide reasonable assurance of the safety and effectiveness: (1) The sale, distribution and use of this device are restricted to prescription use in accordance with 21 CFR 801.109. (2) The labeling must include (a) contraindications for the use of the device on infants with synostosis or with hydrocephalus; (b) warnings indicating the need: (i) to evaluate head circumference measurements and neurological status at intervals appropriate to the infant’s age and rate of head growth, and to describe steps that should be taken in order to reduce the potential for restriction of cranial growth and possible impairment of brain growth and development; (ii) to evaluate the skin at frequent intervals, e.g., every three to four hours, and to describe steps that should be taken if skin irritation or breakdown occurs; (c) precautions indicating the need: (i) to additionally treat torticollis, if the positional plagiocephaly is associated with torticollis; (ii) to evaluate device fit and to describe the steps that should be taken in order to reduce the potential for restriction of cranial growth, possible impairment of brain growth and development and skin irritation and/or breakdown; (iii) to evaluate the structural integrity of the device and to describe the steps that should be taken to reduce the potential for the device to slip out of place and cause asphyxiation or trauma to the eyes or skin; (d) adverse events, i.e., skin irritation and breakdown that have occurred with the use of this device; (e) clinician’s instructions for casting the infant, for fitting the device, and for care and use of the device; and (f) parents’ instructions for care and use of the device. (3) The materials must be assessed for biocompatibility with testing appropriate for long term direct skin contact.
{0}------------------------------------------------
## Danmar Products, Inc. Michigan Cranial Reshaping Orthosis
October 13, 2009
JAN - 6 2010
The following summary is provided pursuant to Section 513 (I) (3) (A) of the Federal Food Drug and Cosmetic Act:
Submitter Information:
Kay Fuller, RAC VCI, LLC on behalf of Danmar Products, Inc. 221 Jackson Industrial Drive Ann Arbor, MI 48103
| <b>Contact Information:</b> | Kay Fuller, RAC |
|-----------------------------|-----------------|
| | VCI, LLC |
| | 734-274-4680 |
Device Name:
Proprietary Name: Michigan Cranial Reshaping Orthosis
Cranial Reshaping Orthosis
Common Name: Cranial Helmet
Classification Name: Cranial Orthosis
Classification Code:
OAN, Cranial Orthosis, Laser-Scan MVA. Cranial Orthosis 21 CFR §882.5970
Predicate Device Equivalence:
The Danmar Products Michigan Cranial Reshaping Orthosis is substantially equivalent to the Hanger Cranial Band, cleared for US commercialization via K072566 on January 9, 2008, the Boston Band Cranial Remolding Orthosis, cleared for US commercialization via K072862 on January 22, 2008 and the Michigan Cranial Helmet, cleared for US commercialization via K003630, May 29, 2001.
Device Description:
The Danmar Products Michigan Cranial Reshaping Orthosis a cranial orthosis that provides passive pressure to prominent regions of an infant's cranium to improve cranial symmetry and/or shape. It is fabricated from a positive mold, obtained from traditional manual casting methods or a 3-D scan of the infant's head, with front and rear sections that are comprised of
{1}------------------------------------------------
an inner, hypoallergenic, soft foam that is ½ " to ¼" thick, and an outer shell made of a, semi-rigid plastic. The Michigan Cranial Reshaping Orthosis is essentially identical to the predicate device known as the Danmar Products Michigan Cranial Helmet, (K003630) with exceptions only to the validated manufacturing improvements noted herein. The new manufacturing methods vield a substantially equivalent device as compared with the Danmar predicate device and manufacturing method.
The Danmar Products Michigan Cranial Reshaping Orthosis is intended for prescription use to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape. A mold is made of the baby's head to create the orthotic. A plaster cast or a laser-scanning accessory may be used to create a scan of the infant's head to create the mold. The device is indicated for infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic and brachycephalic shaped heads.
Comparison of Technological Characteristics:
The Danmar Products Michigan Cranial Reshaping Orthosis device incorporates the same technological characteristics as the legally marketed predicate devices noted in this Premarket Notification submission. The proposed fabrication changes noted in this submission include the addition of a 3-D shape capture method (Omega Scanner) and the Omega Carver. The Omega Scanner is labeled as a Class 1 Laser and therefore it may be utilized by a qualified user without eye protection when utilized under normal operating conditions. The finished subject device technological characteristics and principals of operations remain identical to the Danmar predicate device, noted herein. The following Tables illustrate the finished subject device remains identical to the Danmar predicate device (K003630).
## Intended Use:
{2}------------------------------------------------
| steps | Predicate Device<br>Manual Process Steps | Subject Device<br>Omega Process Steps | steps |
|-------|------------------------------------------------------------------------------|-----------------------------------------------------------------|-------|
| 1 | Orthotist makes negative mold<br>of baby's head | Orthotist scans baby's head using<br>Orthotist's Omega scanner | 1 |
| | | Scan file modified by Orthotist, as<br>needed for symmetry | 2 |
| 2 | Orthotist sends negative mold to Danmar | Orthotist sends scan file to Danmar<br>electronically | 3 |
| 3 | Danmar makes positive plaster mold | Danmar carves positive foam plastic<br>mold with Omega carver | 4 |
| 4 | Danmar modifies positive plaster mold<br>according to Orthotist instructions | | |
| 5 | Danmar wraps nylon stockinette over<br>plaster mold | Danmar wraps nylon stockinette over<br>foam plastic mold | 5 |
| 6 | Danmar fabricates orthosis according to<br>conventional process | Danmar fabricates orthosis according<br>to conventional process | 6 |
and the comments of the comments of the comments of
{3}------------------------------------------------
| Comparison<br>Criteria | Subject<br>Device | Predicate<br>Device<br>K003630 | Predicate Device<br>K072862 | Predicate Device<br>K072566 |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code<br>Prescription Device | OAN, MVA<br>Yes | OAN<br>Yes | OAN, MVA<br>Yes | OAN, MVA<br>Yes |
| Indications for Use | "The Danmar Products<br>Michigan Cranial<br>Reshaping Orthosis is<br>intended for<br>prescription use to<br>apply pressure to<br>prominent regions of<br>an infant's cranium in<br>order to improve<br>cranial symmetry<br>and/or shape. A mold is<br>made of the baby's<br>head to create the<br>orthotic. A plaster cast<br>or a laser-scanning<br>accessory may be used<br>to create a 3-D scan of<br>the infant's head to<br>create the mold. The<br>device is indicated for<br>infants from 3 to 18<br>months of age, with<br>moderate to severe<br>nonsynostotic<br>positional<br>plagiocephaly,<br>including infants with<br>plagiocephalic and<br>brachycephalic shaped<br>heads." | "The Danmar Products<br>Michigan Cranial<br>Helmet is intended for<br>prescription use to be<br>used to apply pressure<br>to prominent regions of<br>an infant's cranium in<br>order to improve<br>cranial symmetry<br>and/or shape in infants<br>from 3 to 18 months of<br>age, with moderate to<br>severe nonsynostotic<br>positional<br>plagiocephaly,<br>including infants with<br>plagiocephalic - and<br>brachycephalic-shaped<br>heads." | "The Boston Band Cranial Molding<br>Orthosis is intended for medical<br>purposes to passively hold prominent<br>cranial regions of an infant's skull in<br>order to improve cranial symmetry<br>and/or shape in infants from three to<br>eighteen months of age, with<br>nonsynostotic positional plagiocephaly,<br>including infants with plagiocephalic,<br>brachycephalic and scaphocephalic<br>patterned head shapes." | "Intended for medical purposes to apply<br>static or gentle pressure to prominent<br>regions of an infant's cranium to<br>improve cranial symmetry or shape. To<br>treat infants from three to eighteen<br>months of age with moderate to severe<br>non-synostotic positional plagiocephaly,<br>including plagiocephalic,<br>brachycephalic, scaphocephalic-shaped<br>heads." |
| Contraindications | Not for use in<br>patients with<br>craniosynostosis,<br>scaphocephalic-<br>shaped heads;<br>hydrocephalus | Not for use in<br>patients with<br>craniosynostosis,<br>hydrocephalus | Unknown | Unknown |
| Materials /<br>Biocompatibility | Same/similar as<br>Predicates | Hypo-allergenic,<br>non-irritating and<br>nontoxic inner<br>foam lining;<br>semi-rigid plastic<br>outer shell | Hypoallergenic polyethylene<br>foam; thermo-formable<br>plastic | Hypoallergenic<br>polyethylene foam;<br>polypropylene copolymer<br>outer shell |
Summary of Device Evaluation:
The literature on this and similar predicate devices demonstrates that the Danmar Products Michigan Cranial Reshaping Orthosis performs as intended. Biocompatibility data demonstrates that the device's inner lining is non-irritating and nontoxic. The subject device validated fabrication modifications noted herein will yield a finished subject device equivalent to the Danmar predicate device (K003630). Performance testing results demonstrate that the device will not break or shatter when subjected to impact.
{4}------------------------------------------------
Conclusions:
.
Based on the above information and the information contained in this Premarket Notification, we conclude the Danmar Products Michigan Cranial Reshaping Orthosis is substantially equivalent to the noted legally marketed predicate devices.
:
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird-like figure. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Validation & Compliance Institute c/o Ms. Kay Fuller, RAC Regulatory Consultant, VCI, LLC 537 Fort Dearborn St. Dearborn, MI 48124
JAN - 62010
Re: K090341
Trade/Device Name: Michigan Cranial Reshaping Orthosis Regulation Number: 21 CFR 882.5970 Regulation Name: Cranial Orthosis Regulatory Class: II Product Code: MVA, OAN Dated: December 12, 2009 Received: December 16, 2009
Dear Ms. Fuller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interence commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Dr to and Cosmetic Act (Act) that do not require approval of a premarket approval application (DMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your devin (1117) found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
. th, m
Malvina B. Eydelman, M.D Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known): K090341
Device Name: Michigan Cranial Reshaping Orthosis
Indications For Use:
The Danmar Products Michigan Cranial Reshaping Orthosis is intended for prescription use to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape. A mold is made of the baby's head to create the orthotic. A plaster cast or a laser-scanning accessory may be used to create a scan of the infant's head to create the mold. The device is indicated for infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic and brachycephalic shaped heads.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Jeffrey Toy
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
K090341 510(k) Number_
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.