DIMENSION VISTA ENZYMATIC CREATININE FLEX REAGENT CARTRIDGE, AND ENZYMATIC CREATININE CALIBRATOR, MODELS K1270, KC270
K090330 · Siemens Healthcare Diagnostics, Inc. · JFY · May 22, 2009 · Clinical Chemistry
Device Facts
Record ID
K090330
Device Name
DIMENSION VISTA ENZYMATIC CREATININE FLEX REAGENT CARTRIDGE, AND ENZYMATIC CREATININE CALIBRATOR, MODELS K1270, KC270
Applicant
Siemens Healthcare Diagnostics, Inc.
Product Code
JFY · Clinical Chemistry
Decision Date
May 22, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1225
Device Class
Class 2
Indications for Use
The ECREA method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma and urine on the Dimension Vista® System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for other urine analytes. The ECREA CAL is an in vitro diagnostic product for the calibration of the Enzymatic Creatinine (ECREA) method on the Dimension Vista® System.
Device Story
Dimension Vista® ECREA Flex® reagent cartridge is an in vitro diagnostic test for quantitative creatinine measurement in human serum, plasma, and urine. The system uses a bichromatic end-point enzymatic method; reagents include TAPS buffer, creatinase, sarcosine oxidase, HTIB, creatininase, horseradish peroxidase, 4-aminophenazone, and potassium hexacyanoferrate (II). The device is used on the Dimension Vista® System in clinical laboratory settings by trained technicians. The system processes samples automatically; results are provided to clinicians to assist in diagnosing and treating renal disease or monitoring dialysis. The ECREA CAL is a liquid, bovine serum albumin-based calibrator used to calibrate the ECREA method on the system.
Clinical Evidence
Bench testing only. Precision evaluated via 20-day ANOVA (CLSI EP5-A); total CVs ranged 1.7-4.4% for serum and 2.8-3.97% for urine. Linearity assessed (EP6-A) with serum range 0.14-20 mg/dL and urine 2.8-400 mg/dL. Method comparison (EP9-A2) against predicate (n=130 serum, n=126 urine) showed high correlation (r=1.00 and 0.999). Commutability evaluated against ID-GC/MS reference method (r=0.9985 serum, r=0.998 urine). Interference testing (EP7-A) defined thresholds for bilirubin, hemolysis, and lipemia.
Indicated for quantitative measurement of creatinine in human serum, plasma, and urine to aid in diagnosis and treatment of renal diseases, monitoring renal dialysis, and as a calculation basis for other urine analytes. Intended for use on the Dimension Vista® System.
Regulatory Classification
Identification
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Predicate Devices
Roche Creatinine Plus Reagent (k003261)
Dimension Chem I Cal (k860021)
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k090330
B. Purpose for Submission:
New device
C. Measurand:
Creatinine
D. Type of Test:
Enzymatic, quantitative
E. Applicant:
Siemens Healthcare Diagnostics, Inc.
F. Proprietary and Established Names:
Dimension Vista Enzymatic Creatinine Flex Reagent Cartridge
Dimension Vista Enzymatic Creatinine Calibrator
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1225 Creatinine Test System
21 CFR 862.1150 Calibrator
2. Classification:
Class II
3. Product code:
JFY and JIT, respectively
{1}
4. Panel:
Clinical Chemistry
H. Intended Use:
1. Intended use(s):
See indications for use statements below.
2. Indication(s) for use:
The ECREA method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma and urine on the Dimension Vista® System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for other urine analytes.
The ECREA CAL is an in vitro diagnostic product for the calibration of the Enzymatic Creatinine (ECREA) method on the Dimension Vista® System.
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
Performance characteristics were provided for the Dimension Vista System 1500
I. Device Description:
The Dimension Vista® ECREA Flex® reagent cartridge is a prepackaged in-vitro diagnostic test method. The reagents contained in the Dimension Vista® ECREA Flex® reagent cartridge are: Reagent 1 – TAPS buffer, creatinase, sarcosine oxidase, HTIB; Reagent 2 – TAPS buffer, creatininase, horseradish peroxidase, 4-aminophenazone, and potassium hexacyanoferrate (II).
The Dimension Vista® ECREA CAL is an ECREA CAL is a liquid, bovine serum albumin, based product containing creatinine. The kit consists of six vials, three vials per level (A, B), 2.5 mL per vial. The approximate values are 0.95 mg/dL (Cal A) and 20 mg/dL (Cal B).
J. Substantial Equivalence Information:
1. Predicate device name(s):
Roche Creatinine Plus Reagent and Dimension Chem I Cal
{2}
2. Predicate K number(s):
k003261 and k860021, respectively.
3. Comparison with predicate:
| | Dimension Vista® ECREA Flex® reagent cartridge | Creatinine Plus Reagent (k003261) |
| --- | --- | --- |
| Similarities | | |
| Measurement | Bichromatic end point | Bichromatic end point |
| Sample Type | Plasma, serum, and urine | Plasma, serum and urine |
| Differences | | |
| Intended Use | The ECREA method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma, and urine on the Dimension Vista® System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for other urine analytes. | Enzymatic in vitro assay for the direct quantitative determination of creatinine in human serum, plasma and urine using Roche clinical chemistry analyzers. |
| Measuring Range | Serum, Plasma - 0.14 -20.0 mg/dL Urine - 2.80 - 400 mg/dL | Serum, Plasma -0.03 - 30 mg/dL Urine - 0.3 - 400 mg/dL |
| Sample Size | 2.7 μL | 6 μL |
| Feature | ECREA CAL | Dimension® Chem I CAL (DC18B) (k860021) |
| --- | --- | --- |
| Similarities | | |
| Matrix | Bovine Albumin | Bovine Albumin |
| Differences | | |
| Intended Use | The ECREA CAL is an in vitro diagnostic product for the calibration of the Enzymatic Creatinine (ECREA) method on the Dimension Vista® System. | The Dimension® Chemistry I Calibrator is an in vitro diagnostic product to be used to calibrate the Dimension® clinical chemistry system for the Calcium (CA), Creatinine (CREA/ECRE), Glucose (GLU/GLUC), Urea Nitrogen (BUN), and Uric Acid (URCA) methods. |
| Analyte | Creatinine | Calcium, Creatinine, Glucose, Urea |
{3}
| | | Nitrogen, Uric Acid |
| --- | --- | --- |
| Matrix | Bovine Albumin | Bovine Albumin |
| Form | Liquid | Lyophilized |
| Volume | Six vials, three vials per level (A, B), 2.5 mL per vial | Six vials, two vials per level (1, 2, 3), 2.0 mL per vial |
| Levels | Three levels: 0, 0.95, 20.0 mg/dL. | Three levels: 0.0, 11.1, 22.4 mg/dL |
K. Standard/Guidance Document Referenced (if applicable):
Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline (EP5-A)
Interference Testing in Clinical Chemistry; Approved Guideline (EP7-A)
Protocols for Determination of Limits of Detection and Limits of Quantitation (EP17-A)
Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline (EP9-A 1995)
Evaluation of the Linearity of Quantitative Measurement (EP6-A)
L. Test Principle:
Creatininase (in the first reagent) hydrolyzes creatinine in a sample to creatine. Creatine is hydrolyzed by creatinase to sarcosine and urea. Sarcosine from this reaction is oxidized by sarcosine oxidase to glycine and formaldehyde and hydrogen peroxide. The hydrogen peroxide reacts with 4-aminoantipyrine and an acid in the presence of peroxidase to yield a quinine imine chromogen. The resulting change in absorbance at 540 and 770 nm is proportional to the creatinine concentration in the sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision studies for serum and urine samples were evaluated using CLSI EP5-A as a guideline. The sponsor conducted a 20 day ANOVA study with the Dimension Vista 1500 instrument. The sponsor conducted within run precision (repeatability), between run precision, between day precision and within lab precision (total) with two serum pools, two serum controls, two urine pools and two urine controls. For within-run precision, samples were run in singlet for twenty days (n=20). For total precision, samples were tested twice daily, in duplicate over 20 days (n=80). Results of the precision studies
{4}
for serum and urine samples are shown below.
| | | Serum Pool | | Serum Control | | Urine Pool | | Urine Control | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | 1 | 2 | 1 | 2 | 1 | 2 | 1 | 2 |
| Within-run (repeatability) | Mean (mg/dL) | 0.61 | 1.57 | 0.79 | 5.81 | 112.19 | 282.19 | 64.10 | 149.84 |
| | SD (mg/dL) | 0.02 | 0.02 | 0.02 | 0.05 | 1.8 | 3.51 | 1.57 | 2.74 |
| | % CV | 2.9 | 1.4 | 2.4 | 0.84 | 1.63 | 1.2 | 2.45 | 1.83 |
| Between-run | Mean (mg/dL) | 0.61 | 1.57 | 0.79 | 5.81 | 112.19 | 282.19 | 64.10 | 149.84 |
| | SD (mg/dL) | 0.01 | 0.02 | 0.01 | 0.09 | 2.6 | 6.46 | 1.74 | 4.01 |
| | % CV | 2.0 | 1.4 | 1.9 | 1.5 | 2.3 | 2.29 | 2.7 | 2.67 |
| Between-day | Mean (mg/dL) | 0.61 | 1.57 | 0.79 | 5.81 | 112.19 | 282.19 | 64.10 | 149.84 |
| | SD (mg/dL) | 0.02 | 0.02 | 0.02 | 0.00 | 0.00 | 4.53 | 1.0 | 0.94 |
| | % CV | 2.7 | 1.3 | 2.3 | 0.00 | 0.00 | 1.6 | 1.55 | 0.61 |
| Within-lab (total) | Mean (mg/dL) | 0.61 | 1.57 | 0.79 | 5.81 | 112.19 | 282.19 | 64.10 | 149.84 |
| | SD (mg/dL) | 0.03 | 0.04 | 0.03 | 0.10 | 3.15 | 8.64 | 2.55 | 4.94 |
| | % CV | 4.4 | 2.4 | 3.8 | 1.7 | 2.8 | 3.06 | 3.97 | 3.30 |
## b. Linearity/assay reportable range:
Serum and urine linearity were assessed via recovery studies using EP6-A as a guideline. Studies are conducted on the Dimension Vista 1500 instrument
Serum
Serum samples were assayed in replicates of five and linear regression equations were determined. Serum samples were spiked with anhydrous creatinine and nine lower concentrations were prepared using Dimension Vista Calibrator level 1 as a diluent. Serum samples concentrations ranged from 0.0 to $25.46\mathrm{mg / dL}$. The serum linear equation is $y = 0.993x - 0.14$ and the recoveries ranged from 87.9 to $100\%$. The linear equation and recovery results and limit of detection (LoD) in d. below support the sponsor's claimed serum linear range of 0.14 to $20\mathrm{mg / dL}$.
Urine
Urine samples were assayed in replicates of five and linear regression equations were determined. Urine samples were spiked with anhydrous creatinine directly into urine to make a high sample and nine lower
{5}
concentrations were prepared using deionized water as a diluent. Urine concentrations ranged from 0.00 to 464.80 mg/dL. The urine linear equation is y=0.991x-2.96 and the recoveries ranged from 93.7 to 106.9%. The linear equation and recovery results and LoD in d. below support the sponsor's claimed urine linear range of 2.8 to 400 mg/dL.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability
The Dimension Vista Enzymatic Creatinine Calibrator is traceable and prepared gravimetrically from the NIST standard reference material SRM914.
Value assignment
Six levels of the in house master calibrators are prepared from the stock solution that is traceability to NIST 914 SRM Creatinine standard reference material. Level 1 does not contain analyte and is assigned a value of 0 mg/dL. The creatinine standards used in these calibrators is prepared using an in-house gravimetric preparation. Standards were tested in replicates of 5 per level assayed with multiple lots on multiple instruments. The sponsors acceptance criteria for the obtained values for calibrator A and B are 0.8 to 1.0 mg/dL and 20 to 22 mg/dL respectively. Working calibrators are traceable to the in-house master calibrators with similar protocols.
Stability
Stability characteristics of the Dimension Vista ECREA Calibrator were determined stressed and accelerated studies. Real-time studies are on-going. Four different stressed conditions were test at five time periods. The results from the studies support the sponsor's claimed shelf-life of 12 months at 4 °C, closed (sealed) stability of 30 days on the instrument and open vial (punctured) stability of 5 days at 2-8 °C.
d. Detection limit:
The sponsor assessed the Limit of the Blank (LoB) and Limit of Detection (LoD) according to the CLSI EP17-A guideline. The LoB was determined by assaying five blank samples (deionized water) in replicates of four (n=20) and adding 2 standard deviations of the mean. The LoD was determined by assaying four replicates of 5 samples (with concentrations close to 4 times the LoB) per day for three days. The samples were assayed on four instruments by one operator. The results for LoB are 0.07 mg/dL and for LoD are 0.14 mg/dL for serum. The LoD for urine was calculated to be 2.8 mg/dL.
6
{6}
7
# Analytical specificity:
Interference studies to determine the effects of unconjugated bilirubin, hemolysis and lipemia were performed according to CLSI EP-7A. The sponsor states that interferences are considered to be significant if the bias between the test and control samples are greater than 10%.
Assay performance claims have been established on the Dimension Vista System 1500 by testing a serum pool containing approximately 1.0 mg/dL creatinine to the following concentrations of each interferent:
Unconjugated and conjugated Bilirubin – 30 and 40 mg/dL;
Results:
Bilirubin: No significant interference at 30 mg/dL.
Concentrations greater than 30 mg/dL bilirubin may cause interference.
Hemolysis – 500 and 600 mg/dL;
Results:
Hemolysis: No significant interference at 500 mg/dL.
Concentrations greater than 500 mg/dL may cause interference. The sponsor states in the package insert the following “moderately hemolyzed specimens should not be used with the ECREA method”.
Lipemia (intralipid) – 1000 and 4000 mg/dL.
Results:
Lipemia (Intralipid): No significant interference at 1000 mg/dL.
Concentrations greater than 1000 mg/dL may cause interference.
The sponsor has the following interference section in the package insert:
Hemoglobin at 600 mg/dL [0.37 mmol/L] decreases ECREA results by 10% at 1.02 mg/dL [90.2 μmol/L] creatinine.
Bilirubin (unconjugated) at 40 mg/dL [707 μmol/L] decreases ECREA results by 19% at 1.01 mg/dL [89.3 μmol/L] creatinine.
Bilirubin (conjugated) at 40 /dL [707 μmol/L] decreases ECREA results by 16% at 1.01 mg/dL [89.3 μmol/L] creatinine.
Lipemia at 3000 mg/dL [33.9 mmol/L] increases ECREA by 25% at 1.01 mg/dL [89.3 μmol/L] creatinine.
In addition, the sponsor has an extensive list of substances tested in the package insert and found no significant interferences.
{7}
f. Assay cut-off:
Not applicable.
## 2. Comparison studies:
a. Method comparison with predicate device:
A comparison study was performed using CLSI EP9-A2 as a guideline with this device and a commercially available enzymatic method.
One-hundred and thirty serum samples that ranged from 0.57 - 19.9 mg/dL were assayed in singlicate with the current method using the Dimension Vista 1500 (y-axis) and a creatinine assay using the Hitachi 917 (x-axis). The resulting linear regression equation is y=1.029x-0.177 mg/dL with a correlation coefficient of 1.00.
One-hundred and twenty-six urine samples that ranged from 8.2 to 391 mg/dL were assayed in singlicate with the current method using the Dimension Vista 1500 (y-axis) and a creatinine assay using the Hitachi 917 (x-axis). The resulting linear regression equation is y=1.057x-6.11 mg/dL with a correlation coefficient of 0.999.
The sponsor has conducted a method comparison to a reference procedure called Isotope Dilution-Gas Chromatography/ Mass Spectrometry (ID-GC/MS) to evaluate commutatability. Thirty-two serum samples ranging from 0.5 to 40.9 mg/dL were evaluated against the ID-GC/MS method. The resulting linear regression equation was y=1.027x-0.015 with a correlation coefficient of 0.9985. Twenty urine samples ranging from 25.7 to 404.0 mg/dL were also evaluated against the ID-GC/MS method. The resulting linear regression equation was y=1.019x-3.089 with a correlation coefficient of 0.998.
b. Matrix comparison:
Fifty-six matched sets of serum(x)/ lithium heparin plasma (y)/ sodium heparin (y) samples ranging from 0.53 to 17.25 were assayed with the Dimension Vista 1500 using EP9-A2 as a guideline. Sixteen of the fifty-six samples were spiked to obtain higher creatinine concentrations. The following tables show the linear regression results from the comparison of serum to the plasmas tested.
| Serum | Li-Heparin | Na-Heparin |
| --- | --- | --- |
| Linear Regression Equation | Y=0.9745x + 0.0634 | Y=0.978x + 0.0745 |
| Correlation Coefficient | 0.9981 | 0.9965 |
The sponsor concluded that lithium heparin and sodium heparin are acceptable
{8}
anticoagulants.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
The sponsor references literature for the expected values as follows:
Expected Values:
Serum/plasma¹
Males: 0.67 – 1.17 mg/dL [59.2 – 104 μmol/L]
Females: 0.51 – 0.95 mg/dL [45.1 – 84 μmol/L]
Random Urine¹
Males: 40.0 – 278 mg/dL [3.54 – 24.6 mmol/L]
Females: 29.0 – 226 mg/dL [2.56 – 20.0 mmol/L]
24-Hour Urine Excretion²
Males: 0.87 – 2.41 g/day [51.2– 142 mmol/day]
Females: 0.67 – 1.59 g/day [5.92 – 141 mmol/day]
¹ Mazzachi BC, Peake MJ, Ehrhardt V. Reference range and method comparison studyies for enzymatic and Jaffe creatinine assays in plasma and serum and early morning urine. Clin. Lab. 2000; 46:53-55.
² Junge W, Wilke B, Halabi A., et al. Determination of reference intervals for serum creatinine, creatinine excretion and creatinine clearance with an enzymatic and a modified Jaffe method. Clin Chim Acta 2004; 344; 137-148.
9
{9}
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.