← Product Code [GEX](/productcode/GEX) · K090312

# TRIA ACNE TREATMENT SYSTEM, MODEL ATS-1 OTC (K090312)

_Tria Beauty, Inc. · GEX · Jan 5, 2010 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K090312

## Device Facts

- **Applicant:** Tria Beauty, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Jan 5, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The ATS-1 OTC is generally indicated to treat dermatological conditions and specifically indicated to treat mild to moderate inflammatory acne vulgaris. It is intended for over-the-counter use.

## Device Story

Handheld, electrically powered LED therapy system; emits therapeutic blue light for treatment of mild to moderate inflammatory acne vulgaris. Designed for over-the-counter, patient self-use. Device applies light directly to affected skin areas to achieve therapeutic effect. Mechanism of action relies on light-based dermatological treatment consistent with predicate devices. Intended to provide accessible acne management outside clinical settings.

## Clinical Evidence

Performance and consumer usability data submitted. No specific clinical trial results or metrics (e.g., sensitivity, specificity) provided in the summary document.

## Technological Characteristics

Handheld, electrically powered device; utilizes light emitting diodes (LEDs) to generate therapeutic blue light. Class II device under 21 CFR 878.4810.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- ATS-1 by TRIA Beauty, Inc. ([K060567](/device/K060567.md))
- ThermaClear by Therative, Inc. ([K060563](/device/K060563.md))
- Zeno by Tyrell, Inc. ([K043377](/device/K043377.md))
- Tanda by Pharos Life Corporation ([K080591](/device/K080591.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KC090312

#### 510(k) SUMMARY

# ATS-1 OTC

JAN - 5 2010

# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

TRIA Beauty, Inc. 5880 W. Las Positas Blvd., Suite 52 Pleasanton, CA 94588-8552 Phone: 925-701-2549 Facsimile: 925-701-2598 Contact Person: Tobin C. Island, Ph.D. Date Prepared: February 5, 2009

#### Name of Device and Name/Address of Sponsor

TRIA Acne Treatment System (ATS-1 OTC) TRIA Beauty, Inc. 5880 W. Las Positas Blvd., Suite 52 -Pleasanton, CA 94588-8552

### Common or Usual Name

Light Emitting Diode Therapy System

#### Classification Name

Laser Instrument, Surgical, Powered Regulation Number: 21 C.F.R.§ 878.4810 Product Code: GEX

#### Predicate Devices

ATS-1 by TRIA Beauty, Inc. (formerly SpectraGenics, Inc.) (K060567) ThermaClear by Therative, Inc. (K060653) Zeno by Tyrell, Inc. (K043377) Tanda by Pharos Life Corporation (K080591)

{1}------------------------------------------------

#### Intended Use / Indications for Use

- The ATS-1 OTC is generally indicated to treat dermatological conditions and specifically indicated to treat mild to moderate inflammatory acne vulgaris. It is intended for over-the-counter use.
#### Technological Characteristics

The ATS-1 OTC is a handheld, electrically powered device that uses LEDs (light emitting diodes) to produce therapeutic blue light.

# Performance Data

न्न

TRIA Beauty claims that the ATS-1 OTC performance data is substantially equivalent to the predicate devices, including and especially the company's previously cleared ATS-1 (K060567). Performance and consumer usability data is submitted with this 510(k) notification.

### Substantial Equivalence

The ATS-1 OTC has substantially the same intended use, indications for use, technological parameters, and mechanism of action as the predicate device(s). Furthermore, the performance and consumer usability data demonstrates that any technological differences or changes in labeling do not raise new questions of safety or efficacy or alter the device's intended therapeutic effect in comparison to the predicates. Therefore, the ATS-1 OTC is substantially equivalent to the predicate devices.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird with outstretched wings, facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

JAN - 5 2010

TRIA Beauty, Inc. % Hogan and Hartson LLP Mr. Jonathan S. Kahan Columbia Square 555 13th Street, N.W. Washington, District of Columbia 20004

Re: K090312

Trade/Device Name: TRIA Acne Treatment System (ATS-1 OTC) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: December 09, 2009 Received: December 09, 2009

Dear Mr. Kahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 – Mr. Jonathan S. Kahan

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson

Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications for Use Statement

510(k) Number (if known):

Device Name: ATS-1 OTC

Indications for Use:

The TRIA Beauty ATS-1 OTC is generally indicated to treat dermatological conditions and specifically indicated to treat mild to moderate inflammatory acne vulgaris.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE------------------------------------------------------------------------------------------------------------------------------------------NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Q

Division Sign-Off) Division Sign-Off)
Division of Surgical ( rthopedic,
Division of Surgical ( res ിഴിടിന്ന് of Suid Restorative Dev 510(k) Number -

19

Page of of

---

**Source:** [https://fda.innolitics.com/device/K090312](https://fda.innolitics.com/device/K090312)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
