POWER STANDING DANI

K090256 · Davismade, Inc. · IPL · Jul 6, 2009 · Physical Medicine

Device Facts

Record IDK090256
Device NamePOWER STANDING DANI
ApplicantDavismade, Inc.
Product CodeIPL · Physical Medicine
Decision DateJul 6, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3900
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Power Standing Dani stand-up wheelchair is a sit to stand mobility device equipped with power driven rear mounted wheels intended for the pediatric population with disabilities which limits their ability to ambulate without assistance. The product is designed for indoor/outdoor use on normal pedestrian terrain such as asphalt, grass, and gravel and not indicated for slopes exceeding 6° in a downhill, uphill or lateral orientation. The device is not intended for use while being transported in a moving vehicle.

Device Story

Power Standing Dani is a pediatric sit-to-stand mobility device; combines prone board/standing platform with rear-mounted power drive system. Input: user joystick commands. Transformation: joystick signals control differential power to rear-mounted 110W motors (e-fix system) for propulsion and steering. Output: motorized mobility in standing position. Used in indoor/outdoor pedestrian environments; operated by pediatric patient. Benefits: provides psychological/physiological advantages of standing to patients with restrictive strength/physical attributes preventing manual self-propulsion. Healthcare provider uses device to facilitate independent mobility for patients with neurological conditions.

Clinical Evidence

No clinical data. Substantial equivalence is based on the use of an existing, cleared manual standing platform design combined with an FDA-cleared power add-on conversion kit.

Technological Characteristics

Steel base frame; prone board with adjustable lateral, anterior, and posterior support components. Power system: Ulrich Alber e-fix rear-mounted drive with two 110W motors, brakes, and gearboxes. Battery: 2x12V (12Ah or 17Ah). Steering: differential power via joystick controller. Wheel sizes: 12" or 24".

Indications for Use

Indicated for pediatric patients with disabilities limiting independent ambulation, including spina bifida, cerebral palsy, muscular dystrophy, spinal muscular atrophy, spinal cord injury, and other neurological disorders. Not indicated for slopes exceeding 6 degrees or use in moving vehicles.

Regulatory Classification

Identification

A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K040256 JUL - 6 2009 DavisMade, Inc. 510(k) Submission: Power Standing Dani ## 510(K) SUMMARY Submitter: DavisMade, Inc. 2511 Davison Rd. Flint, MI 48506 810-233-9706 (Phone) 810-233-9716 (Fax) info@standingdani.com (e-mail) FDA Registration #: 1833039 Daniel Davis Contact Person: 27 January 09 Preparation Date: Proprietary name: Power Standing Dani Electric Power Wheelchair, Stand-up Classification Name: Regulatory Class: Class II Product Code: IPL Identification of Predicate Devices: Lifestand - LSC Lifestand (K041535) DavisMade Standing Dani (K890602) Ulrich Alber e-fix ® power add-on (K041535)/ Intended Use: The Power Standing Dani stand-up wheelchair is a sit to stand mobility device equipped with power driven rear mounted wheels intended for the pediatric population with disabilities which limits their ability to ambulate without assistance. The product is designed for indoor/outdoor use on normal pedestrian terrain such as asphalt, grass, and gravel and not indicated for slopes exceeding 6° in a downhill, uphill or lateral orientation. The device is not intended for use while being transported in a moving vehicle. Device Description: The Power Standing Dani stand-up wheelchair (wheelstand) is equipped with a vertical positioning prone board with an attached standing platform, inclined slightly forward, to which the pediatric is supported with lateral, anterior and posterior components in the upright position. The prone board and standing platform are comprised of positioning components that are interchangeable and whose location can be adjusted to accommodate the precise needs for each individual. The prone board and attached standing platform are securely mounted to an adjustable steel base {1}------------------------------------------------ 510(k) Submission: Power Standing Dani DavisMade, Inc. frame, to which the power driven rear wheels and front suspension caster wheels are mounted. The drive power is provided through a rear mounted wheel system, branded as the e-fix® is supplied by Ulrich Alber. Gmbh. Each of the two drive wheels contains a 110 Watt motor, brake and gear box in its hub. The wheels sizes are available in 12" or 24" diameter. The wheel motors are powered with 2x12V (12Ah or 17Ah) batteries, which are secured in a solid mounting rack. Front suspension castor wheels support the front of the frame and allow indirect steering by applying differential power to the rear drive wheels through the joystick controller. Safety and Effectiveness: The Power Standing Dani was developed based on extensive knowledge and experience gained by providing standing mobility for pediatrics requiring assistive technology since the introduction of the Manual Standing Dani® in 1990. This knowledge and experience has shown prone standing with adjustable support systems, allows the pediatrics optimum psychological and physiological benefits. > The manually propelled Standing Dani, however, is not intended for those pediatrics with restrictive strength and physical attributes that limit their ability to be independently mobile. The addition of a safe and effective joy stick controlled power drive add-on provides that pediatric population the same psychological and physiological benefits as those capable of manually selfpropelling. > The Power Standing Dani utilizes the identical prone board and standing platform as the Manual Standing Dani and incorporates an FDA cleared wheelchair power add-on conversion kit. > > ii {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is depicted in black, and the text is also in black. ## Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DavisMade, Incorporated % Mr. Daniel W. Davis Chief Executive Officer & Founder 2511 Davison Road Flint, Michigan 48506 Re: K090256 Trade/Device Name: The Power Standing Dani Regulation Number: 21 CFR 890.3900 Regulation Name: Standup Wheelchair Regulatory Class: II Product Code: IPL Dated: April 17, 2009 Received: June 16, 2009 Dear Mr. Davis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. JUL - 6 2009 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing {3}------------------------------------------------ Page 2-Mr. Daniel W. Davis practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Submission: Power Standing Dani DavisMade, Inc. ## STATEMENT OF INDICATIONS FOR USE 510(k) Number: (If known) Indications for use: The Power Standing Dani stand-up wheelchair (wheelstand) is a sit to stand mobility device equipped a removable prone board and standing platform with a joystick controlled power driven rear mounted wheels, providing the pediatric population with disabilities the providing the pocking to independently ambulate in the standing position. Types of disability diagnosis include spina bifida, cerebral palsy, muscular dystrophy, spinal muscular atrophy, spinal cord injury, and various other neurological disorders. Proprietary name: Power Standing Dani Prescription Use Over-The Counter Use (Per 21 CFR 801 Subpart D) (21 CFT 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) OR Concurrence of CDRH, Office of Device Evaluation (ODE) XX Lawrence D. Taylor (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K090256
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...