← Product Code [FJK](/productcode/FJK) · K090255

# BLOOD TUBING SETS (STERILE FLUID PATH) (K090255)

_Nxstage Medical, Inc. · FJK · Feb 10, 2009 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K090255

## Device Facts

- **Applicant:** Nxstage Medical, Inc.
- **Product Code:** [FJK](/productcode/FJK.md)
- **Decision Date:** Feb 10, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5820
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The blood tubing sets are indicated for use with a medically prescribed hemodialyzer. The suitability of a particular configuration is the responsibility of the physician in charge.

## Device Story

Blood tubing sets connect blood access devices to hemodialysis or hemofiltration machines and hemodialyzers/hemofilters. Components include air trap chambers, filters, injection sites, pump segments, heparin/saline lines, pressure pillows, and pressure monitoring lines (with/without transducer protectors). Used in clinical settings during extracorporeal procedures; operated by trained healthcare professionals. Facilitates blood flow and monitoring during dialysis; ensures compatibility between various dialysis machines and clinical configurations.

## Clinical Evidence

Bench testing only. Performance, verification, and validation testing were conducted to characterize the device and demonstrate substantial equivalence to the predicate.

## Technological Characteristics

Extracorporeal blood tubing set; includes air trap chambers, filters, injection sites, pump segments, and pressure monitoring lines. Design and materials are similar to the predicate device. Sterile fluid path. No electronic or software components.

## Regulatory Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

## Predicate Devices

- Arterial-Venous Blood Tubing Sets ([K080807](/device/K080807.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K090255
Page 1 of 3

EB 1 0 2009

### NxStage Medical, Inc. Blood Tubing Sets 510(k) Device Modification

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of SMDA 1990.

A. Submitter's Information:

Name:

NxStage Medical Inc.

Address:

439 South Union Street, 5th Floor Lawrence, MA 01843 United States

FDA Establishment Owner/Operator Number:

Contact Person:

Phone: Fax:

Manufacturing Site:

FDA Establishment Registration Number:

Manufacturing/Sterilization Site:

> FDA Establishment Registration Number:

Manufacturing/Sterilization Site:

> FDA Establishment Registration Number:

> > Sterilization Site:

9045797

Michael Doyle Manager, Regulatory Affairs

(978) 687-4746 (978) 687-4750

MEDIMEXICO, S. DE R.L. DE C.V. Av. Valle imperial No. 10523 Parque industrial Valle Sur Tijuana, B.C., Mexico 22180

9616074

Kawasumi Laboratories Co. Ltd. 55/26-27 m-13 Phaholyothin Rd. Km-46 Klong Nueng, Klong Luang Pratumtanee, Thailand 12120

9680437

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Kawasumi Laboratories Co. Ltd. 48 Moo 8 Ratchasima-Chokchai Rd., Tambon Tha-ang, Amphur Chokchai, Nakhorn Rachasima, Thailand 30190

9615908

Steris Corporation Isomedix Services, Inc. 1000 S. Sarah Place Ontario, CA 91761

{1}------------------------------------------------

K090255
Page 2 of 3

## NxStage Medical, Inc. Blood Tubing Sets 510{k} Device Modification

| FDA Establishment Registration Number:     | Contract sterilizer                                                                          |
|--------------------------------------------|----------------------------------------------------------------------------------------------|
| Sterilization Site:                        | STERIS Corporation<br>Isomedix Service, Inc.<br>7685 St. Andrews Ave.<br>San Diego, CA 92154 |
| FDA Establishment Registration Number:     | Contract sterilizer                                                                          |
| Trade/Proprietary Name:                    | Blood Tubing Sets                                                                            |
| Common/Usual Name:                         | Tubing sets                                                                                  |
| Classification Name:                       | Sets, Tubing, Blood, With and Without Anti-<br>Regurgitation Valve                           |
| Regulation Number:                         | 876.5820                                                                                     |
| Product Code:                              | FJK                                                                                          |
| Device Classification:                     | Class II                                                                                     |
| Device Panel:                              | Gastroenterology/Urology                                                                     |
| Substantial Equivalence/Predicate Devices: |                                                                                              |

#### C. Substar

The subject blood tubing set is substantially equivalent to the following legally marketed predicate device, previously cleared by the FDA;

- · Arterial-Venous Blood Tubing Sets, K080807, July 18, 2008
#### D. Device Description/Indications for Use:

The blood tubing sets consist of a family of products which are used during extracorporeal procedures to provide a means to connect blood access devices to a hemodialysis or hemofiltration machine and a hemodialyzer or hemofilter. To facilitate a hemodialysis procedure, for example, the arterial and/or venous blood tubing sets may contain features such as air trap chambers, filters, injection sites, pump segments, heparin infusion lines, saline administration lines, pressure pillows, priming sets, and pressure monitoring lines with or without transducer protectors. Numerous product codes are produced to accommodate various manufacturer's dialysis machines, as well as differing clinical preferences and clinical procedures.

#### Indications for use:

B. Device

The blood tubing sets are indicated for use with a medically prescribed hemodialyzer. The suitability of a particular configuration is the responsibility of the physician in charge.

#### E. Technological Characteristics:

The subject device has the same technological characteristics and is similar in design and configuration as compared to the predicate device. The subject device is designed with similar components and features that are also used in the predicate device.

#### F. Summary of Non-Clinical Test/Performance Testing - Bench

NxStage Medical, Inc. believes that the information and data provided in this submission clearly describes the subject device and demonstrates that the device is adequately

{2}------------------------------------------------

K09c255
Page 3 of 3

## NxStage Medical, Inc. Blood Tubing Sets 510(k) Device Modification

designed for the labeled indications for use. Performance, verification and validation testing was conducted to characterize performance of the subject blood tubing set to provide a basis of comparison to the predicate device as all features are not identical. Results of this testing have documented that the subject blood tubing set is substantially equivalent to the predicate device and is suitable for the labeled indications for use.

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three wing-like shapes, positioned to the right of a circular text element. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" and is arranged around the circumference of the circle. The bird-like figure is abstract and appears to be in motion.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## FEB 1 0 2009

Mr. Michael Doyle Manager, Regulatory Affairs NxStage Medical, Inc. 439 South Union Street, 5th Floor LAWRENCE MA 01843

Re: K090255

Trade/Device Name: Blood Tubing Sets Regulation Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: FJK, KOC Dated: January 30, 2009 Received: February 2, 2009

Dear Mr. Doyle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device. to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology)                      | (240) 276-0120 |
| Other          |                                  | (240) 276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppor/index.html.

Sincerely yours,

Laura B. Morris

anine M. M. irris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): 〈○90255

Device Name: Blood Tubing Sets (Sterile Fluid Path)

Indications for Use:

The blood tubing sets are indicated for use with a medically prescribed hemodialyzer. The suitability of a particular configuration is the responsibility of the physician in charge.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K090255

Page 1 of 1

Page No. 19

---

**Source:** [https://fda.innolitics.com/device/K090255](https://fda.innolitics.com/device/K090255)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
