K090164 · Biolitec, Inc. · GEX · Jun 26, 2009 · General, Plastic Surgery
Device Facts
Record ID
K090164
Device Name
CERALAS MULTIWAVELENGTH 980/1470NM DIODE LASER
Applicant
Biolitec, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 26, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is intended for delivery of laser light to soft tissue in the contact and non contact mode during surgical procedures including via endoscopes. The Ceralas Multiwavelength 980/1470 Diode Laser is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, cardiothoracic surgery, dental applications, and endovenous occlusion of the greater saphenous vein. The Multiwavelength laser is further indicated for laser assisted lipolysis.
Device Story
Diode laser system delivers 980nm and 1470nm laser energy to soft tissue; used in contact or non-contact modes, including via endoscopes. Operated by physicians in clinical/surgical settings (OR, clinic). Device transforms electrical energy into laser light for surgical applications: incision, excision, vaporization, ablation, hemostasis, coagulation. Output is laser light delivered to target tissue; healthcare provider uses output to perform surgical procedures, potentially reducing bleeding and tissue trauma compared to traditional methods. Used with ELVes Procedure Kit for endovenous occlusion.
Clinical Evidence
Bench testing only; no clinical data presented. Performance testing demonstrated no significant difference compared to cleared predicate devices.
Technological Characteristics
Diode laser system emitting 980nm and 1470nm wavelengths. Operates in contact and non-contact modes. Intended for soft tissue surgical applications. No specific materials, software architecture, or connectivity details provided.
Indications for Use
Indicated for patients requiring soft tissue surgery (incision, excision, vaporization, ablation, hemostasis, coagulation) across various specialties including ENT, arthroscopy, GI, general/plastic/podiatric surgery, urology, gynecology, neurosurgery, pulmonary, cardiothoracic, and dental. Also indicated for endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux and laser-assisted lipolysis.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
090164 Page (1) of (5)
# 510(k) SUMMARY
# Biolitec Inc.'s Ceralas Multiwavelength 980/1470nm Diode Laser System
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Hogan & Hartson 555 13th Street NW Washington DC 20004
JUN 26 2009
(202) 637-5794 Phone: Facsimile: (202) 637-5910
Jonathan S. Kahan Contact Person:
January 21, 2009 Date Prepared:
## Name of Device and Name/Address of Sponsor
Ceralas Multiwavelength 980/1470nm Diode Laser
Biolitec, Inc. 515 Shaker Road East Longmeadow, MA 01028
# Common or Usual Name
Diode Laser
## Classification Name
Laser, Surgical Diode Laser System, 21 C.F.R. 878.4810, Product Code GEX
## Predicate Devices
Cynosure Smart Lipo Multiwavelength Laser (K080121) Biolitec Ceralas 980nm Diode Laser (K072106) Biolitec Ceralas 1470nm Diode Laser (K073063)
\ \ \ \ DC - 061213/00001 - 2831829 v2
{1}------------------------------------------------
090164 page (2) of (5)
#### Intended Use / Indications for Use
The device is intended for delivery of laser light to soft tissue in the contact and non contact mode during surgical procedures including via endoscopes. The Ceralas Multiwavelength 980/1470nm Diode Laser is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, cardiothoracic surgery, dental applications, and endovenous occlusion of the greater saphenous vein. The Multiwavelength laser is further indicated for laser assisted lipolysis.
The device is specifically indicated for use as follows:
# Ear, Nose and Throat and Oral Surgery (Otolaryngology)
Hemostasis, incision, excision, ablation, coagulation, and vaporization of tissue from the ear, nose, throat and adjacent areas including soft tissue in the oral cavity. Examples include:
- . Removal of benign lesions from the ear, nose and throat
- Excision and vaporization of vocal cord nodules and polyps .
- . Incision and excision of carcinoma in situ
- Ablation and vaporization of hyperkeratosis .
- Excision of carcinoma of the larynx .
- . Laryngeal papillomectomy
- Excision and vaporization of herpes simplex I and II ●
- Neck dissection .
## Arthroscopy
Hemostasis, incision, excision, coagulation, vaporization and ablation of joint tissues during arthroscopic surgery. Examples include:
- Menisectomy
- . Synovectorny
- Chondromalacia ●
## Gastroenterology
Hemostasis, incision, excision, ablation, coagulation and vaporization of tissue in the upper and lower gastrointestinal tracts and also with endoscopic procedures. Examples include:
- . Hemostasis of upper and lower GI bleeding
- . Excision and vaporization of colorectal carcinoma
- � Excision of polyps
{2}------------------------------------------------
K090164 Page 3 of (5)
# General Surgery, Dermatology, Plastic Surgery and Podiatry
Excision, ablation, vaporization and photocoagulation of skin lesions, hemostasis, incision, excision, vaporization, ablation and debulking of soft tissue, abdominal, rectal, skin, fat or muscle tissue and dermabrasion. Examples include:
- Matrixectomy .
- Excision of neuromas .
- Excision of periungual and subungual warts ●
- Excision of plantar warts �
- Excision of keloids �
- . Liver resection
- . Excision of cutaneous lesions
- . Hemorrhoidectomy
- . Appendectomy
- . Debridement of decubitus ulcers
- . Hepatobiliary tumors
- . Mastectomy
- . Dermabrasion
- . Vaporization and hemostasis of capillary hemangioma
- Excision, vaporization and hemostasis of abdominal tumors .
- . Excision, vaporization and hemostasis of rectal pathology
- Pilonidal cystectomy .
- . Herniorapphy
. Adhesiolysis
- . Parathyroidectomy
- . Laparoscopic cholecystectomy
- . Thyroidectomy
- . Resection of organs
- . Debridement of wounds
- Photocoagulation of teleangectasia of the legs and face .
- Photocoagulation of vascular lesions of the face and extremities .
- Endovascular coagulation of the greater saphenous vein of the thigh in ● patients with superficial vein reflux.
- . Treatment of reticular veins and branch varicosities
#### Urology
Excision, vaporization, incision, coagulation, ablation and hemostasis of urological tissues. Examples include:
- Vaporization of urethral tumors .
- Release of urethral stricture .
- . Removal of bladder neck obstruction
- Excision and vaporization of condyloma .
- Lesions of external genitalia ●
- Vaporization of the prostate to treat benign prostatic hyperplasia (BPH) .
{3}------------------------------------------------
090164 Page (4) of (5)
## Gynecology
Ablation, excision, incision, coagulation, hemostasis and vaporization of gynecological tissue. Examples include:
- Endometrial ablation .
- Excision or vaporization of condylomata acuminate .
- Vaporization of cervical intraepithelial neoplasia .
- Cervical conization .
- Menorrhagia �
## Neurosurgery
Vaporization, coagulation, excision, incision, ablation and hemostasis of soft tissue. Examples include: hemostasis in conjunction with menigiomas
#### Cardiac Surgery
Hemostasis and coagulation of soft tissue, including cardiac tissue.
## Pulmonary Surgery
Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system. Examples include:
- . Tracheobronchial malignancy or stricture
- Benign and malignant pulmonary obstruction .
- Endoscopic pulmonary applications �
# Dental Applications
Indicated for the following applications on intraoral and extraoral soft tissue (including marginal and interdental gingival and epithelial lining of free gingival): frenectomy, frenotomy, biopsy, operculectomy, implant recovery, gingivectomy, gingivoplasty, gingival troughing, crown lengthening, hemostasis of donor site, removal of granulation tissue, laser assisted flap surgery, debridement of diseased epithelial lining, incisions and draining of abscesses, tissue retraction for impressions, papillectomy, vestibuloplasy, excision of lesions, exposure of unerupted/partially erupted teeth, leukoplakia, removal of hyperplastic tissues, treatment of aphthous ulcers and sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket), pulpotomy, pulpotomy as an adjunct to root canal therapy and light activation of bleaching materials for teeth whitening.
# Endovenous Occlusion of the Greater Saphenous Vein in Patients with Superficial Vein Reflux
Indicated for use with the ELVes Procedure Kit in the endovascular coagulation of the Greater Saphenous Vein (GSV) of the thigh in patients with Superficial Vein Reflux.
{4}------------------------------------------------
091164 Page (5) of (5)
# Technological Characteristics
The Ceralas Multiwavelength 980/1470 Diode Laser emits laser energy at the 980nm and 1470nm wavelengths in a manner substantially similar to the Ceralas 980 and Ceralas 1470.
## Performance Data
Performance testing of the Ceralas Multiwavelength 980/1470 Diode Laser demonstrates no significant difference as compared to the cleared predicate devices.
## Substantial Equivalence
The Ceralas Multiwavelength 980/1470nm Diode Laser is substantially equivalent to the Cynosure Smart Lipo Multiwavelength Laser, the Biolitec Ceralas 980mm Diode Laser and the Ceralas 1470nm Diode Laser. The Ceralas Multiwavelength 980/1470mm Diode Laser has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the Ceralas Multiwavelength 980/1470mm Diode Laser and its predicate devices raises no new issues of safety or effectiveness. Thus, the Ceralas Multiwavelength 980/1470nm Diode Laser is substantially equivalent.
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Public Health Service
JUN 26 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Biolitec, Inc. % Hogan & Hartson LLP Mr. Jonathan S. Kahan Columbia Square 555 13th Street Northwest Washington, District of Columbia 20004
Re: K090164
Trade/Device Name: Ceralas Multiwavelength 980/147nm Diode Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: June 2, 2009
Dear Mr. Kahan:
Received: June 3, 2009
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
# Page 2 - Mr. Jonathan S. Kahan
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark McMillan
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use Statement
| 510(k) Number (if known):_ | KC |
|----------------------------|----|
|----------------------------|----|
Device Name: Ceralas Multiwavelength 980/1470nm Diode Laser
90164
#### Indications for Use:
The device is intended for delivery of laser light to soft tissue in the contact and non contact mode during surgical procedures including via endoscopes. The Ceralas Multiwavelength 980/1470 Diode Laser is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, cardiothoracic surgery, dental applications, and endovenous occlusion of the greater saphenous vein. Multiwavelength laser is further indicated for laser assisted lipolysis
The device is specifically indicated for use as follows:
# Ear, Nose and Throat and Oral Surgery (Otolaryngology)
Hemostasis, incision, excision, ablation, coagulation, and vaporization of tissue from the ear, nose, throat and adjacent areas including soft tissue in the oral cavity. Examples include:
- Removal of benign lesions from the ear, nose and throat .
- Excision and vaporization of vocal cord nodules and polyps
- Incision and excision of carcinoma in situ
- Ablation and vaporization of hyperkeratosis
- Excision of carcinoma of the larynx
- Laryngeal papillomectomy
- Excision and vaporization of herpes simplex I and II
- Neck dissection
#### Arthroscopy
Hemostasis, incision, excision, coagulation, vaporization and ablation of joint tissues during arthroscopic surgery. Examples include:
- Menisectomy
- Synovectomy
- Chondromalacia
# Gastroenterology
Hemostasis, incision, excision, ablation, coagulation and vaporization of tissue in the upper and lower gastrointestinal tracts and also with endoscopic procedures. Examples include: منها
15
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
\\DĊ. 061218/000001 - 2831829 v2
510(k) Number K090164
{8}------------------------------------------------
K090164
Page ② of ④
- Hemostasis of upper and lower GI bleeding
- Excision and vaporization of colorectal carcinoma
- Excision of polyps
# General Surgery, Dermatology, Plastic Surgery and Podiatry
Excision, ablation, vaporization and photocoagulation of skin lesions, hemostasis, incision, excision, vaporization, ablation and debulking of soft tissue, abdominal, rectal, skin, fat or muscle tissue and dermabrasion. Examples include:
- Matrixectomy
- Excision of neuromas
- Excision of periungual and subungual warts
- Excision of plantar warts
- Excision of keloids
- Liver resection
- Excision of cutaneous lesions
- Hemorrhoidectomy
- Appendectomy
- Debridement of decubitus ulcers
- Hepatobiliary tumors
- Mastectomy
- Dermabrasion
- Vaporization and hemostasis of capillary hemangioma
- Excision, vaporization and hemostasis of abdominal turnors
- Excision, vaporization and hemostasis of rectal pathology
- Pilonidal cystectomy
- Herniorapphy
- Adhesiolysis
- Parathyroidectomy
- Laparoscopic cholecystectomy
- Thyroidectomy
- Resection of organs
- Debridement of wounds
- Photocoagulation of teleangectasia of the legs and face
- Photocoagulation of vascular lesions of the face and extremities
- Endovascular coagulation of the greater saphenous vein of the thigh in patients with superficial vein reflux.
- Treatment of reticular veins and branch varicosities ()
#### Urology
Excision, vaporization, incision, coagulation, ablation and hemostasis of urological tissues. Examples include:
16
Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K090164
\\\DC - 061213/000001 - 2831829 v2
{9}------------------------------------------------
KOGJ164
ge of (4)
- Vaporization of urethral tumors
- Release of urethral stricture
- Removal of bladder neck obstruction
- Excision and vaporization of condyloma
- Lesions of external genitalia
- Vaporization of the prostate to treat benign prostatic hyperplasia (BPH)
# Gynecology
Ablation, excision, incision, coagulation, hemostasis and vaporization of gynecological tissue. Examples include:
- Endometrial ablation .
- Excision or vaporization of condylomata acuminate .
- Vaporization of cervical intraepithelial neoplasia
- Cervical conization
- Menorrhagia
# Neurosurgerv
Vaporization, coagulation, excision, incision, ablation and hemostasis of soft tissue. Examples include: hemostasis in conjunction with menigiomas
## Cardiac Surgery
Hemostasis and coagulation of soft tissue, including cardiac tissue.
#### Pulmonary Surgery
Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system. Examples include:
- Tracheobronchial malignancy or stricture .
- Benign and malignant pulmonary obstruction .
- Endoscopic pulmonary applications
# Dental Applications
Indicated for the following applications on intraoral and extraoral soft tissue (including marginal and interdental gingival and epithelial lining of free gingival): frenectomy, frenotomy, biopsy, operculectomy, implant recovery, gingivectomy, gingivoplasty, gingival troughing, crown lengthening, hemostasis of donor site, removal of granulation tissue, laser assisted flap surgery, debridement of diseased epithelial lining, incisions and draining of abscesses, tissue retraction for impressions, papillectomy, vestibuloplasy, excision of lesions, exposure of unerupted/partially erupted teeth, leukoplakia, removal of hyperplastic tissues, treatment of aphthous ulcers and sulcular debridement (removal of diseased or inflamed soft tissue in the
17
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
\\\DC - 061213/000001 - 2831829 v2
510(k) Number_
er K090164
{10}------------------------------------------------
periodontal pocket), pulpotomy, pulpotomy as an adjunct to root canal therapy and light activation of bleaching materials for teeth whitening.
Endovenous Occlusion of the Greater Saphenous Vein in Patients with Superficial Vein Reflux
Indicated for use with the ELVes Procedure Kit in the endovascular coagulation of the Greater Saphenous Vein (GSV) of the thigh in patients with Superficial Vein Reflux.
Prescription Use _ X (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
k090164
# (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R. Oslen Jensen
Page 4 of 4
(Division Sign-Oll) (Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K090164
II. II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.