K090136 · STERIS Corporation · FQO · Mar 20, 2009 · General, Plastic Surgery
Device Facts
Record ID
K090136
Device Name
STERIS CMAX XLT SURGICAL TABLE
Applicant
STERIS Corporation
Product Code
FQO · General, Plastic Surgery
Decision Date
Mar 20, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4960
Device Class
Class 1
Indications for Use
The STERIS® 5085 SRT is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation. The STERIS® 5085 SRT accommodates all general surgical procedures including but not limited to, cardiac and vascular, endoscopic, gynecology, urology, nephrectomy, neurology, ophthalmologic, orthopedics and other procedures requiring intraoperative fluoroscopic C-arm imaging and also supports laparoscopic surgical technique for the largest surgical patients. The STERIS® 5085 SRT enables patient transport on hard level surfaces within the surgical suite (from pre-operative areas to the operating room and again from the operating room to post operative recovery).
Device Story
STERIS 5085 SRT is a mobile, electro-hydraulically operated surgical table; supports general surgical procedures and intraoperative fluoroscopic C-arm imaging. Features include powered lateral tilt, Trendelenburg/reverse Trendelenburg, Zip-Slide movable tabletop, and adjustable height. Device enables patient transport on hard level surfaces within surgical suites. Operated by clinical staff in operating rooms; provides stable platform for surgery and patient positioning. Benefits include high weight capacity, extended width, and low table elevation for improved access to large patients.
Clinical Evidence
Bench testing only. Compliance with UL and IEC standards for medical electrical equipment, including safety, electromagnetic compatibility, and usability, was demonstrated.
Technological Characteristics
Electro-hydraulic surgical table; mobile form factor. Features powered lateral tilt, Trendelenburg, and height adjustment. Complies with UL 60601-1, EN/IEC 60601-1, 60601-1-2, 60601-1-4, 60601-1-6, 60601-2-46, and CAN/CSA C22.2 No. 601.1-M90. Programmable electrical medical system.
Indications for Use
Indicated for patients undergoing general surgical procedures, including cardiac, vascular, endoscopic, gynecological, urological, nephrectomy, neurological, ophthalmological, and orthopedic surgeries, as well as procedures requiring intraoperative fluoroscopic C-arm imaging and laparoscopic techniques. Supports patient transport on hard level surfaces within the surgical suite.
Regulatory Classification
Identification
Operating tables and accessories and operating chairs and accessories are AC-powered or air-powered devices, usually with movable components, intended for use during diagnostic examinations or surgical procedures to support and position a patient.
{1}------------------------------------------------
page 1 of 3
STERIS®
MAR 2 0 2009
# 510(k) Summary For - . STERIS®5085 SRT
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (216) 354-2600 Fax No: (216) 639-4459
Contact:
John R. Scoville. Fellow Regulatory Affairs Telephone: (440) 392-7330 Fax No: (440) 357-9198
Summary Date: Submission:
March 05, 2009 K090136
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{2}------------------------------------------------
### Page 2 of 3 K090136
page 2 of 3
## STERIS Response to 021809 Request for Additional Information K090136 - STERIS 5085 SRT
### 1. Device Name
Trade Name:
Common/Usual Name:
Classification Name:
STERIS® 5085 SRT
STERIS® 5085 SRT Surgical Table
Operating Table classified as Class I device (Product Code [FQO] per 21 CFR 878.4960
#### Predicate Devices 2.
- K041924 Hausted POWERTRAN Series Stretcher ●
- K971307 Amsco 3085SP Surgical Table .
- K022309 Stryker ZOOM motorized Stretcher ●
### 3. Description of Device
The STERIS® 5085 SRT Surgical Table is a mobile, electro-hydraulically operated surgical table designed to support all general surgical procedures including cardiac and vascular, endoscopic, gynecology, urology, nephrectomy, neurology, ophthalmology and orthopedics with the addition of STERIS table accessories. The STERIS® 5085 SRT Surgical Table features powered lateral tilt, Trendelenburg / reverse Trendelenburg, Zip-Slide™ movable tabletop, and adjustable height functions. The STERIS® 5085 SRT has a patient transport feature that allows the user to transport patients to and from the surgical suite on hard levelsurfaces.
### 4. Intended Use
The STERIS® 5085 SRT is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation. The STERIS® 5085 SRT accommodates all general surgical procedures including but not limited to, cardiac and vascular, endoscopic, gynecology, urology, nephrectomy, neurology, ophthalmologic, orthopedics and other procedures requiring intraoperative fluoroscopic C-arm imaging and also supports laparoscopic surgical technique for the largest surgical patients.
The STERIS® 5085 SRT enables patient transport on hard level surfaces within the surgical suite (from pre-operative areas to the operating room and again from the operating room to post operative recovery).
{3}------------------------------------------------
STERIS Response to 021809 Request for Additional Information K090136 - STERIS 5085 SRT
#### Description of Safety and Substantial Equivalence 5.
STERIS Corporation believes that the STERIS® 5085 SRT Surgical Table is Substantially Equivalent to the predicate devices: K041924 Hausted POWERTRAN Series Stretcher, K971307 Amsco 3085SP Surgical Table and K022309 Stryker ZOOM motorized Stretcher and the differences between the proposed and predicate devices do not raise any new issues of safety and efficacy.
The STERIS® 5085 SRT Surgical Table as well as the predicate devices have been designed to meet UL and IEC requirements. Significant safety and performance characteristics are tested to ensure compliance with specifications and that the proposed device is as safe, as effective, and performs as well as or better than the predicate devices. No performance standards or special controls applicable to powered table patient transport devices have been established under sections 513 or 514 of the FD& C Act.
The proposed device also complies with the following voluntary standards:
UL 60601-1 (Amended by UL 60601-1 Rev 1, UL 60601-1): Medical Electrical Equipment, Part 1: General Requirements for Safety
EN/IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance for Medical Electrical Equipment, Part 1: General Requirements for Safety
EN/IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
EN/IEC 60601-1-4: Medical electrical equipment - Part 1-4: General requirements for safety - Collateral Standard: Programmable electrical medical systems.
EN/IEC 60601-1-6: Medical electrical equipment - Part 1-6: General requirements for safety - Collateral standard: Usability .
EN/IEC 60601-2-46: Medical electrical equipment - Part 2-46: Particular requirements for the safety of operating tables.
CAN/CSA C22.2 No. 601.1-M90: Medical Electrical Equipment - General Requirements for Safety.
A-4
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract eagle design with three stylized wing feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES.USA" are arranged in a circular pattern around the eagle.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Steris Corporation % Mr. John R. Scoville, Jr. Fellow, Regulatory Affairs 5960 Heisley Road Mentor, Ohio 44060-1834
MAR 2 0 2009
Re: K090136
Trade/Device Name: STERIS® 5085 SRT Regulation Number: 21 CFR 878.4960 Regulation Name: Operating tables and accessories and operating chairs and accessories Regulatory Class: I Product Code: FOO Dated: March 5, 2009 Received: March 6, 2009
Dear Mr. Scoville:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
# Page 2 - Mr. John R. Scoville, Jr.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Sincerely yours,
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# K090136
## STERIS Response to 021809 Request for Additional Information K090136 - STERIS 5085 SRT
# Indications for Use
510(k) Number (if known): K090136
Device Name:
STERIS® 5085 SRT
Indications For Use:
The STERIS® 5085 SRT is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation. The STERIS® 5085 SRT accommodates all general surgical procedures including but not limited to, cardiac and vascular, endoscopic, gynecology, urology, nephrectomy, neurology, ophthalmologic, orthopedics and other procedures requiring intraoperative fluoroscopic C-arm imaging and also supports laparoscopic surgical technique for the largest surgical patients.
The STERIS® 5085 SRT enables patient transport on hard level surfaces within the surgical suite (from pre-operative areas to the operating room and again from the operating room to post operative recovery).
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nix M. Olen for mxm
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
510(k) Number K090136
Page 1 of 1
CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.