The On-Call® Plus Blood Glucose Monitoring System is an electrochemical enzymatic assay for the quantitative detection of glucose in capillary whole blood from the fingertip, forearm, and/or palm by people with diabetes at home and by healthcare professionals as an aid in the monitoring the effectiveness of diabetes control programs. The On-Call® Plus Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus, nor intended for use on neonates. The On-Call® Plus Blood Glucose control solution is for use with the On-Call® Plus Blood Glucose meter and strips as a quality control check to verify the accuracy of blood glucose test results.
Device Story
System uses electrochemical enzymatic assay to measure glucose in capillary whole blood; input sample (1 µL) applied to test strip containing glucose oxidase and mediator; meter measures electrical current generated by reaction; calculates glucose concentration; displays plasma-equivalent result. Used by patients at home or healthcare professionals in clinics. Provides quantitative blood glucose readings to monitor diabetes control programs. Includes memory for 300 records with time/date stamps. Data port allows transfer of stored records. System includes meter, test strips, and control solution for quality verification. Benefits include portable, rapid monitoring of glycemic status.
Clinical Evidence
Clinical study conducted with lay persons and trained technicians. Compared On-Call® Plus system accuracy against YSI Model 2300 STAT PLUS reference. Results demonstrated lay persons obtained readings comparable to trained technicians. Users reported satisfaction with ease of operation and performance. Bench testing included repeatability, intermediate precision, linearity, interference, hematocrit (30-55%), temperature, altitude, and sample volume effects. System met ISO 15197 accuracy requirements.
Technological Characteristics
Electrochemical enzymatic assay (glucose oxidase biosensor). Powered by 3.0V CR 2032 coin cell battery. Dimensions: 85mm x 54mm x 20.5mm. Weight: 49.5g. Data port: 9600 baud, 8 data bits, 1 stop bit, no parity. Operating range: 1.1-33.3 mmol/L (20-600 mg/dL). Hematocrit range: 30-55%. Plasma-equivalent calibration.
Indications for Use
Indicated for people with diabetes and healthcare professionals for quantitative glucose monitoring in capillary whole blood (fingertip, forearm, palm). Not for neonates, diabetes diagnosis, or screening. Contraindicated for patients who are dehydrated, in shock, critically ill, or in a hyperosmolar state.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
One Touch Ultra Blood Glucose Monitoring System (K002134)
Submission Summary (Full Text)
{1}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k090057
B. Purpose for Submission:
New device
C. Measurand:
Whole blood glucose
D. Type of Test:
Whole blood glucose concentration through a quantitative amperometric assay
(Glucose Oxidase)
E. Applicant:
Acon Laboratories Co.
F. Proprietary and Established Names:
On-Call Plus Blood Glucose Monitoring System
G. Regulatory Information:
1. Regulation section:
21CFR Sec.-862.1345-Glucose test system.
21CFR Sec.-862.1660 Quality control material (assayed and unassayed).
2. Classification:
2, 2, 1 (reserved)
3. Product code:
NBW - System, Test, Blood Glucose, Over the Counter
CGA - Glucose Oxidase, Glucose
JJX - Quality Control Material (Assayed and Unassayed)
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
See indication(s) for use below
{2}
2. Indication(s) for use:
The On-Call® Plus Blood Glucose Monitoring System is an electrochemical enzymatic assay for the quantitative detection of glucose in capillary whole blood from the fingertip, forearm, and/or palm by people with diabetes at home and by healthcare professionals as an aid in the monitoring the effectiveness of diabetes control programs.
The On-Call® Plus Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus, nor intended for use on neonates.
The On-Call® Plus Blood Glucose control solution is for use with the On-Call® Plus Blood Glucose meter and strips as a quality control check to verify the accuracy of blood glucose test results.
3. Special conditions for use statement(s):
- Not for neonatal use
- Not for screening or diagnosis of diabetes mellitus
- Alternative site testing is for use at times of steady state only
- For Over-the-Counter use
- Not for use in patients who are dehydrated, in shock, critically ill, or in hyperosmolar state
4. Special instrument requirements:
On-Call® Plus Blood Glucose Meter
I. Device Description:
The On-Call® Plus Blood Glucose Monitoring System consists of glucose meter, blood glucose test strips and two levels of control solution.
J. Substantial Equivalence Information:
1. Predicate device name(s):
One Touch Ultra Blood Glucose Monitoring System
2. Predicate 510(k) number(s):
k002134
{3}
3. Comparison with predicate:
| Features | On-Call® Plus Blood Glucose Monitoring System | One Touch Ultra Blood Glucose Monitoring System (K002134) |
| --- | --- | --- |
| Similarities | | |
| Measurement Range | 20 to 600 mg/dL (1.1-33.3 mmol/L) | Same |
| Result Calibration | Plasma-equivalent | Same |
| Sample | Fresh capillary whole blood | Same |
| Minimum Sample Size | 1 μL | Same |
| Assay Method | Glucose oxidase biosensor | Same |
| Power Source | One (1) CR 2032 3.0V coin cell battery | Same |
| Battery Life | 12 months or approximately 1,000 tests | Same |
| Glucose Units of Measure | mg/dL | Same |
| Hematocrit Range | 30-55% | Same |
| Automatic Shutoff | Two minutes after last user action | Same |
| Differences | | |
| Test Time | 10 seconds | 5 seconds |
| Memory | Up to 300 records with time and date | 150 blood glucose and control solution tests |
| Meter Size | 85 mm x 54 mm x 20.5 mm | 3.12” x 2.25” x 0.85” |
| Weight | Approximately 49.5 g (with battery installed) | 1.5 ounces with battery (Approximately 42 g) |
| Operating Temperature | 5-45°C (41 - 113°F) | 6-44°C (43 - 111°F) |
| Operating Relative Humidity | 20-90% (non-condensing) | 10-90% |
| Sample Site | Fingertip, palm and forearm | Fingertip and forearm |
K. Standard/Guidance Document Referenced (if applicable):
- EN 61010-2-101 - Safety requirements for electrical equipment for measurement, control and laboratory use. Particular requirements for in vitro diagnostic (IVD) medical equipment. 12/31/2002
- IEC 60068-2-64 - Environmental Testing. Test Methods. Test Fh: Vibration, broad-band random (digital control) and guidance 04//30/2009
{4}
- CLSI EP7-A2 - Interference testing in Clinical Chemistry; Approved Guideline - second edition 11/23/2005
- CLSI EP6-A - Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. 04/01/2003
- ISO 15197: In vitro diagnostic test systems - Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus - 2003
- Draft Guidance for Industry and FDA Staff - Total Product Life Cycle for Portable Invasive Blood Glucose Monitoring Systems - Issue Date: 10/24/2006
## L. Test Principle:
The On-Call® Plus Blood Glucose Test Strips are thin strips with a chemical reagent system using glucose oxidase. They work with the On-Call® Plus Blood Glucose Meter to measure the glucose concentration in capillary whole blood. Blood is applied to the end tip of the test strip. The blood is then automatically absorbed into the reaction cell. The reaction takes place in the reaction cell. A transient electrical current is formed during the reaction which is detected by the meter. The blood glucose concentration is then calculated based on the electrical current. The result is then shown on the meter display. The meters are calibrated to display plasma equivalent results.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The repeatability precision of the On-Call® Plus Blood Glucose Monitoring System was evaluated according to EN ISO 15197:2003, Section 7.2.2 Repeatability evaluation, with three test strip lots using blood samples at five glucose concentrations measured 10 times on 10 glucose meters each.
| Avg | SD | CV |
| --- | --- | --- |
| Level 1 - 30 to 50 mg/dL | | |
| 43.4 | 2.47 | 5.70% |
| Level 2 - 51 to 110 mg/dL | | |
| 75.7 | 2.63 | 3.47% |
| Level 3 - 111 to 150 mg/dL | | |
| 143.3 | 4.39 | 3.07% |
| Level 4 - 151 to 250 mg/dL | | |
| 238.0 | 7.05 | 2.97% |
| Level 5 - 251 to 500 mg/dL | | |
| 377.9 | 12.03 | 3.17% |
The intermediate precision of the On-Call® Plus Blood Glucose Monitoring System was evaluated according to EN ISO 15197, Section 7.2.3 Intermediate
{5}
precision evaluation, with three test strip lots using control solution at three glucose concentrations over 10 days on 10 glucose meters each.
| Avg | SD | CV |
| --- | --- | --- |
| Level 1: 30 to 50 mg/dL | | |
| 39.5 | 1.98 | 5.0% |
| Level 2: 96 to 144 mg/dL | | |
| 132.1 | 5.12 | 3.9% |
| Level 3: 280 to 420 mg/dL | | |
| 360.6 | 11.92 | 3.3% |
b. Linearity/assay reportable range:
The linearity of the On-Call® Plus Blood Glucose Monitoring System was evaluated with three test strip lots using blood samples at 11 glucose concentrations. The 11 levels of glucose concentrations were prepared from venous blood samples containing heparin anticoagulant at approximately $42\%$ hematocrit level. The range of glucose concentrations used (2 to $667~\mathrm{mg / dL}$ ) is representative of the measuring range of the meter. The results of meter readings vs. YSI Plasma Values were compared using linear regression analysis.
| Strip Lot | Slope | Intercept | R | R² |
| --- | --- | --- | --- | --- |
| 1 | 0.9966 | 0.9355 | 0.9983 | 0.9966 |
| 2 | 0.9971 | 0.7032 | 0.9985 | 0.9970 |
| 3 | 0.9959 | 1.1810 | 0.9982 | 0.9965 |
The claimed range of measurement is 20 to $600\mathrm{mg / dL}$
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
System traceability and validation of calibration and control materials are traceable to NIST standard reference material (SRM) 917b, D-glucose (dextrose) which was used for calibration.
A real time stability study of the control solution with cap closed was conducted and found to support the 24 month shelf life of the control solution.
In addition, an open and closed stability study of the control solution was also done to mimic the actual use condition. The study results confirm the 3-month use life of the control solution.
d. Detection limit:
20 mg/dL based on the above linearity study.
{6}
# e. Analytical specificity:
The effect of interference on the On-Call® Plus Blood Glucose Monitoring System was evaluated with two test strip lots using venous blood samples containing heparin anticoagulant at approximately $42\%$ hematocrit level at three glucose concentrations (approximately $< 60 \, \mathrm{mg/dL}$ , $100 - 150 \, \mathrm{mg/dL}$ and $350 - 450 \, \mathrm{mg/dL}$ ). The blood samples were then used to prepare the test pool at low and high concentrations of the interfering substances and control pool.
The acceptance criteria were as follows:
Glucose concentration $< 75\mathrm{mg / dL}$
Average absolute bias for each interfering substance compared with control pool must be $\leq \pm 15\mathrm{mg / dL}$
Glucose concentration $\geq 75\mathrm{mg / dL}$
Average % Bias for each interfering substance compared with control pool must be $\leq \pm 15\%$
| Interfering Substances | Therapeutic/ Physiological Levels | Test Concentration – Low | Test Concentration – High | On-Call® Plus |
| --- | --- | --- | --- | --- |
| Acetaminophen | 1.0-3.0 mg/dL | 4 mg/dL | 20 mg/dL | No interference at therapeutic levels and levels ≤4 mg/dL; interference only at super therapeutic levels |
| Ascorbic Acid | 0.4-2.0 mg/dL | 3 mg/dL | 6 mg/dL | No interference at therapeutic levels and levels ≤3 mg/dL; interference only at super therapeutic levels |
| Bilirubin | 0.3-1.2 mg/dL | 20 mg/dL | 40 mg/dL | No interference at |
| | | | | therapeutic levels |
| Cholesterol | 0.2-0.5 mg/dL | 10 mg/dL | 20 mg/dL | No interference at therapeutic levels |
| Cholesterol + Cholesterol | 0.1-0.2 mg/dL | 10 mg/dL | 20 mg/dL | No |
{7}
| Interfering Substances | Therapeutic/ Physiological Levels | Test Concentration - Low | Test Concentration - High | On-Call® Plus |
| --- | --- | --- | --- | --- |
| | | | | physiological levels up to high test concentration |
| Creatinine | 0.6-1.3 mg/dL | 1.5 mg/dL | 5 mg/dL | No interference at physiological levels up to high test concentration |
| Dopamine | 0.03 mg/dL | 0.03 mg/dL | 0.09 mg/dL | No interference at therapeutic levels up to high test concentration |
| L-Dopa | 0.02-0.3 mg/dL | 0.3 mg/dL | 3 mg/dL | No interference at therapeutic levels up to high test concentration |
| Methyl-Dopa | 0.1-0.75 mg/dL | 0.75 mg/dL | 1.5 mg/dL | No interference at therapeutic levels up to high test concentration |
| Ibuprofen | 1.0-7.0 mg/dL | 7 mg/dL | 50 mg/dL | No interference at therapeutic levels up to high test concentration |
| Salicylate Acid | 10-30 mg/dL | 30 mg/dL | 60 mg/dL | No interference at therapeutic levels up to high test concentration |
{8}
| Interfering Substances | Therapeutic/ Physiological Levels | Test Concentration – Low | Test Concentration – High | On-Call® Plus |
| --- | --- | --- | --- | --- |
| Tetracycline | 0.2-0.5 mg/dL | 0.5 mg/dL | 1.5 mg/dL | No interference at therapeutic levels up to high test concentration |
| Tolazamide | 2.0-2.5 mg/dL | 5.0 mg/dL | 10 mg/dL | No interference at therapeutic levels up to high test concentration |
## Hematocrit study
The effect of different hematocrit levels was evaluated on the On-Call® Plus Blood Glucose Monitoring System with three test strip lots using blood samples at four glucose concentrations (40-60, 80-120, 250-300, and 450-550 mg/dL). The four levels of glucose concentrations were prepared from venous blood samples containing heparin anticoagulant at six different hematocrit levels ranging from 30-55%. Glucose results for each concentration and hematocrit level were compared to the same plasma glucose concentration on the YSI at 40% Hematocrit using the acceptance criteria below.
- Glucose concentration ≥75 mg/dL:
- Average % Bias (Average strip reading vs. Plasma YSI Value at 40% Hct): within ±15% bias
- Individual %Bias (Individual strip reading vs. Plasma YSI Value at 40% Hct): within ±15% bias
- Glucose concentration < 75 mg/dL:
- Average Bias (Average strip reading vs. Plasma YSI Value at 40% Hct): within ±15 mg/dL bias
- Individual Bias (Individual strip reading vs. Plasma YSI Value at 40% Hct): within ±15 mg/dL bias
Results indicate that the On-Call® Plus Blood Glucose Monitoring System can provide accurate results when measuring blood samples having 30–55% hematocrit ranges covering measurement range of the system.
{9}
# Altitude study
An altitude effect study was performed by testing each of 15 different samples at sea level and at elevated altitude of 8,516 ft. The On-Call® Plus Blood Glucose Monitoring System meets acceptance criteria of having more than 95% of the data points for all three strip lots within ±20% bias when glucose concentration is ≥75 mg/dL, and ±15 mg/dL when glucose concentration is <75 mg/dL when compared to plasma YSI values. The results indicate that the On-Call® Plus Blood Glucose Monitoring System meets the accuracy acceptance criteria for testing at both sea level and elevated altitude of 8,516 ft.
Temperature and humidity studies were performed per ISO 15197:2003 and showed that the meter can be used at temperatures from 5-45°C and at a relative humidity from 20-90%.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The system accuracy evaluation for the On-Call® Plus Blood Glucose Monitoring System has been performed according to EN ISO 15197:2003, Section 7.3 System accuracy evaluation.
A trained technician performed testing with each study subject’s capillary blood samples from fingertip, palm and forearm sites on the studied system. The testing was performed using the same method provided in the instructions for use for the studied system. Exceptions are allowed for very low (<50 mg/dL) and very high (>400 mg/dL) glucose concentration samples. Per EN ISO 15197:2003, samples with glucose concentration of <50 mg/L and >400 mg/L can be prepared by incubating the samples or by supplementing additional glucose to obtain the samples in these very low and very high glucose concentration levels. For each subject, an additional fingerstick capillary blood sample was also collected into microtainer tubes containing heparin anticoagulant for the YSI reference instrument testing and hematocrit level testing.
Study Site:
The tested glucose concentration range was from 43.6 to 473 mg/dL for fingertip capillary blood testing. The tested glucose concentration range was from 47.5 to 399 mg/dL for palm and forearm capillary blood testing. The subject blood hematocrit range is from 32% to 49%.
{10}
| Linear Regression Summary Table: Meter Reading (y) vs. Plasma YSI Value (x) (YSI using fingertip sample) | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Sample Site | Lot | Slope | Intercept | R | R² | N |
| Fingertip | 1 | 0.9972 | -4.6130 | 0.9924 | 0.9849 | 244 |
| Fingertip | 2 | 0.9649 | -0.5640 | 0.9917 | 0.9834 | 244 |
| Fingertip | 3 | 0.9848 | 3.8831 | 0.9898 | 0.9797 | 244 |
| Palm | 1 | 0.9702 | 2.8354 | 0.9821 | 0.9645 | 214 |
| Palm | 2 | 0.9614 | 4.4561 | 0.9820 | 0.9642 | 214 |
| Palm | 3 | 0.9747 | 9.8885 | 0.9819 | 0.9641 | 214 |
| Forearm | 1 | 0.9419 | 5.5952 | 0.9778 | 0.9560 | 214 |
| Forearm | 2 | 0.9226 | 9.0375 | 0.9814 | 0.9632 | 214 |
| Forearm | 3 | 0.9410 | 12.656 | 0.9778 | 0.9561 | 214 |
| Fingertip Sample On-Call® Plus Result (vs. Plasma YSI with Fingertip sample) | | | | |
| --- | --- | --- | --- | --- |
| Strip Lot: 390346 | | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% | |
| 97/208 (46.6%) | 159/208 (76.4%) | 194/208 (93.3%) | 208/208 (100%) | |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | | |
| 11/36 (30.6%) | 29/36 (80.6%) | 36/36 (100%) | | |
| Strip Lot: 390350 | | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% | |
| 94/208 (45.2%) | 159/208 (76.4%) | 196/208 (94.2%) | 208/208 (100%) | |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | | |
| 16/36 (44.4%) | 30/36 (83.3%) | 36/36 (100%) | | |
| Strip Lot: 390357 | | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% | |
| 100/208 (48.1%) | 164/208 (78.8%) | 193/208 (92.8%) | 208/208 (100%) | |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | | |
| 16/36 (44.4%) | 25/36 (69.4%) | 36/36 (100%) | | |
| Palm Sample On-Call® Plus Result (vs. Plasma YSI with Fingertip sample) | | | | |
| Strip Lot: 390346 | | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% | |
| 88/198 (44.4%) | 144/198 (72.7%) | 187/198 (94.4%) | 197/198 (99.5%) | |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | | |
| 10/16 (62.5%) | 16/16 (100%) | 16/16 (100%) | | |
{11}
| Strip Lot: 390350 | | | |
| --- | --- | --- | --- |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 85/198 (42.9%) | 143/198 (72.2%) | 186/198 (93.9%) | 198/198 (100%) |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | |
| 7/16 (43.8%) | 16/16 (100%) | 16/16 (100%) | |
| Strip Lot: 390357 | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 72/198 (36.4%) | 127/198 (64.1%) | 167/198 (84.3%) | 197/198 (99.5%) |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | |
| Within ±5 mg/dL | Within ±5 mg/dL | Within ±15 mg/dL | |
| 8/16 (50.0%) | 15/16 (93.8%) | 16/16 (100%) | |
| Forearm Sample On-Call® Plus Result (vs. Plasma YSI with Fingertip sample) | | | |
| Strip Lot: 390346 | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 73/198 (36.9%) | 131/198 (66.2%) | 172/198 (86.9%) | 197/198 (99.5%) |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | |
| 14/16 (87.5%) | 16/16 (100%) | 16/16 (100%) | |
| Strip Lot: 390350 | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 77/198 (38.9%) | 154/198 (77.8%) | 185/198 (93.4%) | 198/198 (100%) |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | |
| 10/16 (62.5%) | 15/16 (93.8%) | 16/16 (100%) | |
| Strip Lot: 390357 | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 71/198 (35.9%) | 134/198 (67.7%) | 169/198 (85.4%) | 197/198 (99.5%) |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | |
| 12/16 (75.0%) | 15/16 (93.8%) | 16/16 (100%) | |
b. Matrix comparison:
Not Applicable - capillary whole blood is the only indicated matrix
3. Clinical studies:
a. Clinical Sensitivity: Not applicable
{12}
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
The objective of the user performance evaluation study was to demonstrate that the intended users are able to operate the On-Call® Plus Blood Glucose Monitoring System (studied system) as well as a trained technician. The user study was performed for fingertip, palm, and forearm sample sites to demonstrate that acceptable user performance can be obtained at palm and forearm alternative sample sites in addition to the fingertip sample site.
The Flesch-Kincaid Grade Level Score of 6.0 to 7.8 was obtained with the User's Manual and the test strip and control Package Inserts of the On-Call® Plus Blood Glucose Monitoring System. In addition, a questionnaire administered to the participating lay persons indicated user satisfaction with the ease of operation by following the Instructions for Use in the User's Manual and the overall performance of the On-Call® Plus Blood Glucose Monitoring System.
Device labeling was given to each study subject to allow the subject to learn and understand how to run testing for the studied system. After reading the instructions, each study subject performed fingerstick testing on the studied system by themselves. After each study subject tested themselves, a trained technician also ran the same testing with the same subject's fingerstick blood. The same procedure for fingerstick testing was then repeated for palm and forearm alternative sites. In the study for each subject, additional fingerstick capillary blood was also collected into microtainer tubes containing heparin anticoagulant for the YSI reference instrument testing and hematocrit level testing.
A user performance study was performed by testing 107 different subjects over 13 days. Both the layperson and technician performed the testing for each subject sample for the user study. The tested glucose concentration range was from 47.5 to 399 mg/dL.
The user performance on the On-Call® Plus Blood Glucose Monitoring System was evaluated according to EN ISO 15197:2003, Section 8 User performance evaluation. The user study was performed for fingertip, palm, and forearm sample sites to show that acceptable user performance can be obtained at palm and forearm alternative sample sites in addition to the fingertip sample site.
{13}
Layperson compared to YSI:
| Fingertip Sample Site Tested by Layperson | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Strip Lot: 390346 | | | | | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | | | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% | | | | |
| 89/198 (44.9%) | 161/198 (81.3%) | 185/198 (93.4%) | 198/198 (100%) | | | | |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | | | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | | | | | |
| 11/16 (68.8%) | 16/16 (100%) | 16/16 (100%) | | | | | |
| Strip Lot: 390350 | | | | | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | | | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% | | | | |
| 86/198 (43.4%) | 156/198 (78.8%) | 190/198 (96.0%) | 198/198 (100%) | | | | |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | | | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | | | | | |
| 11/16 (68.8%) | 16/16 (100%) | 16/16 (100%) | | | | | |
{14}
| Strip Lot: 390357 | | | |
| --- | --- | --- | --- |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 89/198 (44.9%) | 155/198 (78.3%) | 183/198 (92.4%) | 197/198 (99.5%) |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | |
| 12/16 (75.0%) | 15/16 (93.8%) | 16/16 (100%) | |
| Palm Sample Site Tested by Layperson | | | |
| Strip Lot: 390346 | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 97/198 (49.0%) | 155/198 (78.3%) | 185/198 (93.4%) | 196/198 (99.0%) |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | |
| 12/16 (75.0%) | 15/16 (93.8%) | 16/16 (100%) | |
| Strip Lot: 390350 | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 89/198 (44.9%) | 163/198 (82.3%) | 185/198 (93.4%) | 198/198 (100%) |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | |
| 10/16 (62.5%) | 15/16 (93.8%) | 16/16 (100%) | |
| Strip Lot: 390357 | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 83/198 (41.9%) | 135/198 (68.2%) | 170/198 (85.9%) | 196/198 (99.0%) |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | |
| 9/16 (56.3%) | 15/16 (93.8%) | 16/16 (100%) | |
| Forearm Sample Site Tested by Layperson | | | |
| Strip Lot: 390346 | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 88/198 (44.4%) | 149/198 (75.3%) | 179/198 (90.4%) | 198/198 (100%) |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | |
| 8/16 (50.0%) | 16/16 (100%) | 16/16 (100%) | |
| Strip Lot: 390350 | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 92/198 (46.5%) | 147/198 (74.2%) | 180/198 (90.9%) | 197/198 (99.5%) |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | |
{15}
| 7/16 (43.8%) | 16/16 (100%) | 16/16 (100%) | |
| --- | --- | --- | --- |
| Strip Lot: 390357 | | | |
| System Accuracy Results for Glucose Concentration ≥75 mg/dL | | | |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 70/198 (35.4%) | 130/198 (65.7%) | 167/198 (84.3%) | 195/198 (98.5%) |
| System Accuracy Results for Glucose Concentration <75 mg/dL | | | |
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | |
| 9/16 (56.3%) | 14/16 (87.5%) | 16/16 (100%) | |
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The sponsor included the following expected values for non-diabetic normal glucose levels in their strip labeling*:
| Time | Range, mg/dL |
| --- | --- |
| Fasting and Before Meals | 70 – 100 |
| 2 hours after Meals | Less than 140 |
*ADA Clinical Practice Recommendations, 2003
N. Instrument Name:
On-Call® Plus Blood Glucose Meter
O. System Descriptions:
1. Modes of Operation:
Each test strip is single use and must be replaced with a new strip for additional readings.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
The applicant has provided documentation that indicates the device was designed and developed under good software life-cycle processes.
{16}
3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected
4. Specimen Sampling and Handling:
This device is intended to be used with capillary whole blood from the finger, palm, or forearm only. Since the whole blood sample is applied directly to the test strip there are no special handling or storage issues.
5. Calibration:
A calibration code chip is provided with each vial of test strips.
6. Quality Control:
The sponsor is providing a one level of glucose control solution with this device. Two levels are available for purchase separately, as stated in the labeling. When a test strip is inserted into the meter, a control can be run. An acceptable range for each control level is printed on the test strip vial label. The user is referred to the user manual and customer support for problems.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
None
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.