ASFORA BULLET CAGE SYSTEM

K090048 · Medical Designs, LLC · MAX · Aug 14, 2009 · Orthopedic

Device Facts

Record IDK090048
Device NameASFORA BULLET CAGE SYSTEM
ApplicantMedical Designs, LLC
Product CodeMAX · Orthopedic
Decision DateAug 14, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Asfora Bullet Cage™ (ABC) is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) and instability in the lumbar spine at one or two levels from L2 to S1. The DDD patients may also have up to Grade I spondylolisthesis at the involved level. The ABC cage devices are used with autogenous bone graft and are implanted via an open posterior approach. Patients should have had at least six (6) months of non-operative treatment prior to implant.

Device Story

Asfora Bullet Cage is a titanium intervertebral fusion device; hollow, threaded, self-tapping design with closed tapered medial ends. Features six slit apertures for autogenous bone graft placement; distal end includes threaded cap. Available in 5 diameters (10-18mm) and 2 lengths (21-25mm). Implanted by surgeons via open posterior approach to immobilize adjacent vertebrae and promote arthrodesis in lumbar spine. Device provides structural support during fusion process; benefits patients by stabilizing spinal segments affected by degenerative disc disease.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and bench testing.

Technological Characteristics

Material: Titanium. Design: Hollow, threaded, self-tapping cage with closed tapered medial ends and six longitudinal slit apertures. Dimensions: 5 diameters (10mm, 12mm, 14mm, 16mm, 18mm) and 2 lengths (21mm, 25mm). Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) and lumbar instability (L2-S1, 1-2 levels), including those with up to Grade I spondylolisthesis. Requires 6 months prior non-operative treatment. Used with autogenous bone graft via open posterior approach.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # AUG 1 4 2009 ## 2. 510(k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with 21 CFR 807.92 ### General Information: | A. | Submitted By: | Medical Designs, LLC.<br>1210 W. 18th Street, Suite 104<br>Sioux Falls, South Dakota 57104<br>Tel: 605-275-1032<br>Fax: 605-335-3734 | |----|----------------------|--------------------------------------------------------------------------------------------------------------------------------------| | | Contact Person: | Kristi Vondra, Vice President of Operations | | | Date Prepared: | August 12, 2009 | | B. | Device Trade Name: | Asfora Bullet Cage™ | | | Classification Name: | Orthopedic Devices, Prosthetic Devices<br>Intervertebral Fusion Device with Bone Graft, Lumbar<br>21 CFR 888.3080 (MAX) | | C. | Predicate Devices: | INTER FIX™ Threaded Fusion Device (P970013;<br>Downclassified to Class II)<br>Ray TFC™ Device (P950019; Downclassified to Class II) | #### Device Description: D. The Asfora Bullet Cage™ is comprised of two hollow, threaded, self-tapping titanium fusion devices (cages) with closed tapered medial ends. The distal end has a small orifice that can be closed with a threaded cap. There are six equidistant slit apertures along the longitudinal axis of the device for placement of autogenous bone graft material. The cage has large cutting threads, angled toward the proximal end. The Asfora Bullet Cage™ is available in ten sizes: 5 diameters (10mm, 12mm, 14mm, 16mm, and 18mm) and 2 lengths (21mm or 25mm). This device, with associated instrumentation is intended for use in posterior lumbar interbody fusion to immobilize adjacent vertebrae and promote arthrodesis (fusion) across the disc space. #### E. Indications for Use: The Asfora Bullet Cage™ (ABC) is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) and instability in the lumbar spine at {1}------------------------------------------------ one or two levels from L2 to S1. The DDD patients may also have up to Grade I spondylolisthesis at the involved level. The ABC cage devices are used with autogenous bone graft and are implanted via an open posterior approach. Patients should have had at least six (6) months of non-operative treatment prior to implant. ### Substantial Equivalence Comparison of Technical Characteristics to Prodicate Device(s): F. The clinical and performance data provide adequate information to demonstrate that the Asfora Bullet Cage™ and its predicates, the INTER FIX™ Threaded Fusion Device (P970013; Downclassified to Class II) and the Ray RFC™ Device (P950019; Downclassified to Class II) are substantially equivalent. H. Summary: > Testing and comparison of technological characteristics and intended uses found that all components of the Asfora Bullet Cage™ are equivalent to the predicates. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three parallel lines that curve and converge, resembling a human figure or abstract representation of health and well-being. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 Medical Designs, LLC % Kristi Vondra 1210 W 18th St. North Center, Suite 104 Sioux Falls, South Dakota 57104 Re: K090048 AUG 1 4 2009 Trade/Device Name: Asfora Bullet Cage Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: August 12, 2009 Received: August 13, 2009 Dear Ms. Vondra: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 – Ms. Kristi Vondra If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Sincerely yours, Oaubay buelus Mark N. Mettler Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Indications for Use 510(k) Number (if known): K090048 The Asfora Bullet Cage™ Device Name: Indications for Use: The Asfora Bullet Cage™ (ABC) is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) and instability in the lumbar spine at one or two levels from L2 to S1. The DDD patients may also have up to Grade I spondylolisthesis at the involved level. The ABC cage devices are used with autogenous bone graft and are implanted via an open posterior approach. Patients should have had at least six (6) months of non-operative treatment prior to implant. X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) 00 (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices Page 1 of 1 1090048 510(k) Number_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...