MEDICALGORITHMICS REAL-TIME ECG MONITOR AND ARRHYTHMIA DETECTOR, MODEL POCKETECG
Applicant
Medicalgorithmics SP Z.O.O.
Product Code
DSI · Cardiovascular
Decision Date
May 22, 2009
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
1. Patients who have a demonstrated need for cardiac monitoring. These may include but are not limited to patients who require monitoring for: a) non-life threatening arrhythmias such as supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, PACs, PSVT) and ventricular ectopy; b) evaluation of bradyarrhythmias and intermittent bundle branch block, including after cardiovascular surgery and myocardial infarction; and c) arrhythmias associated with co-morbid conditions such as hyperthyroidism or chronic lung disease 2. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath) 3. Patients with palpitations with or without known arrhythmias to obtain correlation of rhythm with symptoms. 4. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation) 5. Patients recovering from cardiac surgery who are indicated for outpatient arrhythmia monitoring 6. Patients with diagnosed sleep disordered breathing including sleep apnea (obstructive, central) to evaluate possible nocturnal arrhythmias 7. Patients requiring arrhythmia evaluation of etiology of stroke or transient cerebral ischemia, possibly secondary to atrial fibrillation or atrial flutter. 8. Data from the device may be used by another device to analyze, measure or report QT interval. The device is not intended to sound any alarms for QT interval changes.
Device Story
Ambulatory ECG monitor; captures electrographic signals via patient-worn sensor; transmits data via Bluetooth to Windows Mobile PDA; PDA runs proprietary software for real-time arrhythmia detection and beat classification; transmits ECG data and annotations via cellular network to remote server; Monitoring Center staff review data; enables correlation of rhythm with symptoms; aids in drug efficacy monitoring and post-surgical/stroke evaluation; benefits patients through remote, continuous cardiac monitoring.
Clinical Evidence
Bench testing only. Device performance validated against AAMI/ANSI EC57:1998/(R)2003 standards for cardiac rhythm and ST-segment measurement algorithms.
Technological Characteristics
System comprises ECG sensor/transmitter, Bluetooth link, and Windows Mobile 5.x/6.x PDA with GSM/CDMA modem. Standards: IEC 60601-1, IEC 60601-1-2, AAMI/ANSI EC38, AAMI/ANSI EC57. Connectivity: Bluetooth (sensor-to-PDA), cellular (PDA-to-server).
Indications for Use
Indicated for patients requiring cardiac arrhythmia monitoring, including those with supraventricular/ventricular arrhythmias, bradyarrhythmias, bundle branch block, symptoms of syncope/dizziness/dyspnea/palpitations, post-cardiac surgery recovery, sleep-disordered breathing, or stroke/TIA etiology evaluation. Contraindicated for patients requiring inpatient monitoring for life-threatening arrhythmias or those requiring hospitalization.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Card Guard's CG-6108 Continuous ECG Monitor and Arrhythmia Detector (K071995)
Submission Summary (Full Text)
{0}------------------------------------------------
| 510(k) Summary | December 1, 2008 K090037 P1/3 |
|----------------|-------------------------------|
|----------------|-------------------------------|
| 1. Submitter Name and Address | Medicalgorithmics LLC<br>245 West 107th St., Suite 11A<br>New York, NY 10025, USA |
|-------------------------------|-----------------------------------------------------------------------------------|
| Contact Person | Martin Jasinski, phone (917) 9419581,<br>fax (817) 5829527 |
| 2. Device | |
| Trade name: | PocketECG - Medicalgorithmics Real-Time ECG<br>Monitor and Arrhythmia Detector |
| Classification name: | Arrhythmia Detector and Alarm |
| Product code: | DSI |
| Regulation no: | 870.1025 |
| Class: | Class II, Special Controls |
#### 3. Substantial Equivalence
The selected predicate devices are:
- 1. CardioNet's Ambulatory ECG Monitor, K072558 (Reg. Number 870.1025 Product Code DSI)
- Card Guard's CG-6108 Continuous ECG. Monitor and Arrhythmia Detector, 2. K071995 (Reg. Number 870.1025, Product Code DS1)
#### 4. Device Description
PocketECG - Medicalgorithmics Real-Time ECG Monitor and Arrhythmia Detector is an ambulatory ECG monitor which analyzes electrographic signal, classifies all detected heart beats and recognizes rhythm abnormalities. All detection results, including annotations for every detected heart beat and ECG signal are transmitted via cellular telephony network to a remote server accessible by a Monitoring Center for reviewing by trained medical staff. The data transmission is automatically triggered when abnormalities are detected, or periodically in case of normal ECG.
The patient worn transmitter streams via Bluetooth link the ECG signal to a Windows Mobile operated PDA (Personal Digital Assistant) device with mobile phone capabilities. The PDA runs Medicalgorithmics proprietary software which detects the ECG annotations and manages the data transmission. The PDA device stores entire ECG on its storage card.
#### 510(k) Summary
Medicalgorithmics 510(k) Premarket Notification
{1}------------------------------------------------
510(k) Summary
K090037
p2/3
# 5. Indications for Use and contradictions
The indications for use for the PocketECG monitor are as follows:
- · 1. Patients who have a demonstrated need for cardiac monitoring. These may include but are not limited to patients who require monitoring for: a) non-life threatening arrhythmias such as supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, PACs, PSVT) and ventricular ectopy; b) evaluation of bradyarrhythmias and intermittent bundle branch block, including after cardiovascular surgery and myocardial infarction; and c) arrhythmias associated with co-morbid conditions such as hyperthyroidism or chronic lung disease
- 2. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath)
- 3. Patients with palpitations with or without known arrhythmias to obtain correlation of rhythm with symptoms.
- 4. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation)
- 5. Patients recovering from cardiac surgery who are indicated for outpatient arrhythmia monitoring
- 6. Patients with diagnosed sleep disordered breathing including sleep apnea (obstructive, central) to evaluate possible nocturnal arrhythmias
- 7. Patients requiring arrhythmia evaluation of etiology of stroke or transient cerebral ischemia, possibly secondary to atrial fibrillation or atrial flutter.
- 8. Data from the device may be used by another device to analyze, measure or report QT interval. The device is not intended to sound any alarms for QT interval changes.
## Contradictions:
- 1. Patients with potentially life-threatening arrhythmias who require inpatient monitoring.
- 2. Patients who the attending physician thinks should be hospitalized.
## 6. Technological comparison to predicate devices
The first technology difference between the subject device is that the predicate devices use customized PDA size monitors, while the subject device uses of-the-shelf PDA with the following minimum requirements:
- 1. Windows Mobile 5.x or 6.x Operating System,
- 2. built in GSM/CDMA modem,
- 3. built in Bluetooth module for communication with the ECG Transmitter,
- 4. replaceable Storage Card slot for cards of minimum 1 GB capacity,
- 5. USB port
Example PDAs meeting the above criteria are: HTC Touch: http://www.htc.com/www/product.aspx?id=362
{2}------------------------------------------------
510(k) Summary
Medicalgorithmics 510(k) Premarket Notification
Image /page/2/Picture/2 description: The image shows handwritten text. The text at the top reads "K09 0037". Below that, the text reads "P 3/3". The handwriting appears to be on a white background.
Hewlett Packard iPAQ hw6940
http://h10010.www1.hp.com/wwpc/us/en/sm/WF06a/215348-64929-314903-215381-1822489.html
The second technological difference between the subject device and the predicate devices is that the subject device uses an arrhythmia analysis algorithm developed by Medicalgorithmics while the predicate devices use arrhythmia analysis algorithm licensed from Mortara (K072558) or their proprietary algorithms (K071995).
The third technological difference between the subject device and the predicate devices is that the subject device uses its own ECG sensor and transmitter, while the predicate devices use their own manufactured ECG sensors.
#### 7. Referenced standards
The Medicalgorithmics ECG Monitor and Arrhythmia Detector, PocketECG meets the requirements of following performance standards in accordance with FDA Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm.
- IEC 60601-1:1999 "Medical Electrical Equipment Part 1: General . Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995"
- IEC 60601-1-2:2001/A1:2004 "Medical Electrical Equipment Part 1-2: . Requirements for Safety; Electromagnetic Compatibility --General Requirements and Tests" Class B
- AMI/ANSI EC38:2007 Medical electrical equipment Part 2-47: Particular . requirements for the safety, including essential performance, of ambulatory electrocardiographic systems
- AAMI / ANSI EC57:1998/(R)2003 Testing and Reporting Performance . Results of Cardiac Rhythm and ST-Segment Measurement Algorithms
Quality management system - Medical devices is in conformance with the standards: PN-EN ISO 9001:2001 and PN-EN ISO 13485:2005.
### 8. Substantial Equivalence Conclusion
Medicalgorithmics ECG Monitor and Arrhythmia Detector, PocketECG is safe, effective and substantially equivalent to the predicate devices as supported by the descriptive information and the performance testing. The subject device is composed of off-the-shelf, certified devices and components fully complying with the US safety and EMC standards.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the eagle. The eagle is depicted in a simple, line-art style, and the text is in a sans-serif font.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAY 2 2 2009
MedicAlgorithmics, Sp. Z O.O. c/o Mr. Martin Jasinski MedicAlgorithmics, LLC 245 West 107th St, Suite 11A New York, NY 10025
Re: K090037
Trade/Device Name: MedicAlgorithmics Real-Time ECG Monitor and Arrhythmia Detector, Model PocketECG
Regulation Number: 21 CFR 870.1025
Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm)
Regulatory Class: Class II (special controls)
Product Code: DSI, MLO
Dated: March 31, 2009 Received: May 8, 2009
Dear Mr. Jasinski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA) it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
R. Vachner
A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):_K090037
Device Name: PocketECG-Medicalgorithmics Real-Time ECG Monitor and Arrhythmia Detector
Indications For Use:
The indications for use for the PocketECG monitor are as follows:
ﺎ . Patients who have a demonstrated need for cardiac monitoring. These may include but are not limited to patients who require monitoring for: a) non-life threatening arrhythmias such as supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, PACs, PSVT) and ventricular ectopy; b) evaluation of bradyarrhythmias and intermittent bundle branch block, including after cardiovascular surgery and myocardial infarction; and c) arrhythmias associated with co-morbid conditions such as hyperthyroidism or chronic lung disease
2. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath)
- Patients with palpitations with or without known arrhythmias to obtain correlation of 3. rhythm with symptoms.
AND/OR Over-The-Counter Use × Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nuna R. Valuner
(Division Sign-Off) Division of Cardiovascular Devices 510(k) Number Ko90037
Page 5-1
{6}------------------------------------------------
- 4. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation)
- న. Patients recovering from cardiac surgery who are indicated for outpatient arrhythmja monitoring
- 6. Patients with diagnosed sleep disordered breathing including sleep apnea (obstructive, central) to evaluate possible nocturnal arrhythmias
- 7. Patients requiring arrhythmia evaluation of etiology of stroke or transient cerebral ischemia, possibly secondary to atrial fibrillation or atrial flutter.
- 8. Data from the device may be used by another device to analyze, measure or report QT interval. The device is not intended to sound any alarms for QT interval changes.
#### Contradictions:
- 1 . Patients with potentially life-threatening arrhythmias who require inpatient monitoring.
- 2. Patients who the attending physician thinks should be hospitalized.
# (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.