Clinical samples from patients with suspected or acute hemostasis disorders (peri- or postoperative or ICU)
Clinical samples from patients in routine clinical settings were used to perform a method comparison study between the ROTEM delta system and the predicate TEG 5000 series analyzer.
Patients with suspected or acute hemostasis disorders in the peri- or postoperative phase or at the ICU; Sample Size: Varies by assay (e.g., n=36 for In-TEM, n=67 for Hep-TEM, n=74 for NATEM); Number of Sites: 3
Haemoscope Corporation Thrombelastograph® Coagulation Analyzer (TEG®) 5000 Series
The ROTEM® delta Thromboelastometry System is designed for invitro diagnostic use by professionals in a laboratory environment. The ROTEM® system is intended to be used to provide a qualitative and quantitative indication of the coagulation state of a blood sample. For this purpose the ROTEM® system records the clot firmness changes in a sample of citrated whole blood as the sample clots, retracts and Ivses in real time. The analyzer output consists of a qualitative graphical representation (mirrored coagulation curve - clot firmness over time) and several defined numerical parameters describing the curve quantitatively. The ROTEM® system provides specific blood modifiers (so-called reagents) intended to be used with the system, as additive to the blood sample. The results of the ROTEM® analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history (anamnesis), the clinical picture and, if necessary, further coagulations tests.
Device Story
ROTEM® delta is a 4-channel thromboelastometry system for in-vitro coagulation analysis. Input: 300 μl citrated whole blood sample in a cylindrical cup. Operation: A pin oscillates within the blood sample; optical detection system (4 CCD chips) monitors pin movement. As blood clots, clot firmness increases, obstructing pin oscillation; rotational movement is converted into an amplitude (0-100mm). System includes integrated computer, electronic pipette, and reagents (in-TEM®, hep-TEM®, star-TEM®). Output: Qualitative graphical mirrored coagulation curve and quantitative parameters (CT, CFT, alpha angle, MCF, lysis). Used in laboratory environments by professionals. Results assist clinicians in assessing hemostasis, hemorrhage, and thrombosis risks alongside patient history and clinical picture.
Clinical Evidence
Bench testing only. Performance compared to TEG® 5000 using NATEM, INTEM, and HEPTEM assays. Primary endpoint (clot firmness) showed equality (MCF vs MA). Secondary kinetic parameters (CT, CFT, Alpha Angle) showed linear correlation (r>0.8). Reference ranges established via CSL1 protocols on three US clinical sample groups. Evaluated interference from aprotinin, tranexamic acid, EACA, heparin, and urokinase. ROTEM® demonstrated higher heparin sensitivity than TEG®.
Technological Characteristics
Four-channel thromboelastometry system. Sensing: optical detection of oscillating pin movement. Energy: electrical. Connectivity: standalone instrument with integrated computer. Software: Linux-based, locked. Reagents: ellagic acid, heparinase, calcium chloride. Sterilization: N/A (reagents/consumables). Standards: IEC 61010-1, EN 61326-1, IEC 62304, ISO 13485, ISO 14971.
Indications for Use
Indicated for adult patients where evaluation of blood coagulation properties is desired, including assessment of peri-operative hemostasis.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
Thrombelastograph® Coagulation Analyzer (TEG®) - 5000 Series (K002177)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k083842
B. Purpose for Submission:
Bundled submission for clearance of new instrument and associated assays
C. Measurand:
Coagulation Factors (intrinsic pathway)
D. Type of Test:
Clotting
E. Applicant:
Pentapharm GmbH
F. Proprietary and Established Names:
ROTEM® delta Thromboelastometry System:
in-TEM® Assay
hep-TEM® Assay
NATEM® Assay
star-TEM®
ROTEM® delta instrument
ROTROL N
ROTROL P
G. Regulatory Information:
1. Regulation section:
21 CFR 864.5425 - Multipurpose system for in vitro coagulation studies
2. Classification:
Class II
3. Product code:
JPA - System, Multipurpose for in vitro coagulation studies
4. Panel:
81 Hematology
H. Intended Use:
1. Intended use(s):
The ROTEM® delta Thromboelastometry System is designed for in vitro diagnostic use by professionals in a laboratory environment. The ROTEM® delta is intended to provide a qualitative and quantitative indication of the coagulation state of a blood sample. For this purpose the ROTEM® delta records the clot firmness changes in a sample of citrated whole blood as the sample clots, retracts and lyses in real time. The analyzer output consists of a qualitative graphical representation (mirrored coagulation curve – clot firmness over time) and several defined numerical parameters describing the curve quantitatively.
The in-TEM® assay is a semi-quantitative in vitro diagnostic assay used to monitor the coagulation process via the intrinsic pathway in citrated whole blood specimens. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of
{1}
Clot formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)). The assay is intended for professional use in the clinical laboratory on the ROTEM® delta Instrument.
The hep-TEM® assay is a semi-quantitative in vitro diagnostic assay used to monitor the coagulation process via the intrinsic pathway in the presence of heparin, in citrated whole blood specimens. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)). The assay is intended for professional use in the clinical laboratory on the ROTEM® delta Instrument.
The NATEM® assay is a semi-quantitative in vitro diagnostic assay used to monitor the coagulation process contact activated by the surface of the measurement cell, in citrated whole blood specimens. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)). The assay is intended for professional use in the clinical laboratory on the ROTEM® delta Instrument.
The star-TEM® reagent is intended for use as recalcification reagent in the NATEM and in-TEM on the ROTEM® delta Thromboelastometry System.
2. Indication(s) for use:
The ROTEM® delta Thromboelastometry System is a non-invasive diagnostic instrument designed to monitor and analyze the coagulation state of a blood sample in order to assist in the assessment of patient clinical hemostasis conditions. The indication for ROTEM® delta is with adult patients where an evaluation of their blood coagulation properties is desired. Coagulation evaluations with the ROTEM® delta System are commonly used to assess clinical conditions such as peri-operative hemostasis.
3. Special conditions for use statement(s):
Prescription Use Only
4. Special instrument requirements:
ROTEM® delta instrument
I. Device Description:
The ROTEM® delta Thromboelastometry System consists of a four-column instrument (with integrated computer module, computer controlled electronic pipette, software), system assays (in-TEM®, hep-TEM®, NATEM®, and star-TEM®), quality controls (ROTROL N, ROTROL P) and measurement cells (cup and pin pro).
The in-TEM® assay contains ellagic acid, the hep-TEM® includes heparinase and calcium chloride, and star-TEM® calcium chloride.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Haemoscope Corporation Thrombelastograph® Coagulation Analyzer (TEG®) 5000 Series
2. Predicate 510(k) number(s):
k002177
3. Comparison with predicate:
{2}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | The ROTEM® delta is intended to provide a qualitative and quantitative indication of the coagulation state of a blood sample. For this purpose the ROTEM® delta records the clot firmness changes in a sample of citrated whole blood as the sample clots, retracts and lyses in real time. | The TEG - 5000 Series Analyzer is intended to be used to provide a quantitative and qualitative indication of the coagulation state of a blood sample by monitoring, measuring, analyzing and reporting coagulation parameter information. The Thrombelastograph (TEG) Coagulation Analyzer TEG - 5000 Series records the kinetic changes in a sample of whole blood, plasma or platelet rich-plasma as the sample clots, retracts and/or lyses (breaks apart). |
| Measuring Technique | Shear elasticity | Same |
| Reagents | Heparinase I, CaCl2 | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Signal Generation | Oscillating pin in stationary cup | Oscillating cup around a stationary pin |
| Pipetting | Electronic pipette | Manual Pipette |
| Measuring Channels | 4 | 2 |
| Reagents | Ellagic acid activator | Kaolin activator |
# K. Standard/Guidance Document Referenced (if applicable):
Guidance for Industry and FDA Staff: 510(k) Submissions for Coagulation Instruments (June 19, 2003)
Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)
DIN EN 61010-1:2002 Safety requirements for electrical equipment for measurement, control and laboratory use - Part 1: General requirements (IEC 61010-1:2001); German version EN 61010-1:2001
EN 61326-1:1997 + A1, A2, A3 Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements (IEC 61326-1:2005); German Version EN 61326-1:2006 (includes EN 61000-4-2:2001, EN 61000-4-3:2003, EN 61000-4-4:2005, EN 61000-4-5:2001, EN 61000-4-6:2001, EN 61000-4-11:2005 and EN 55011/CISPR 11 cl. B)
EN 55022 cl.B:1998 + A1, A2 Information technology equipment - Radio disturbance characteristics - Limits and methods of measurement (IEC/CISPR 22: 1997 + A1:2000 + A2:2002); German Version EN 55022:1998 + Corrigendum July 2003 + A1:2000+ Corrigendum April 2003 +A2:2003 FCC part 15 allows the application of CISPR 22 class B as an alternative to FCC cl. B; ANSI C.63.4 is taken into account.
EN 61000-3-2:2000 Electromagnetic compatibility (EMC) - Part 3-2: Limits - Limits from harmonic currents emissions (equipment input current $\leq 16A$ per phase) (IEC 61000-3-2:2000); German Version EN 61000-3-2:2000
{3}
EN 61000-3-3:1995 + A1 Electromagnetic compatibility (EMC) – Part 3-3:Limits – Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems, for equipment with rated current ≤16A per phase and not subjected to conditional connection (IEC 61000-3-3:1994 + A1:2001); German Version EN 61000-3-3:1995 + A1:2001
IEC 62304 Ed. 1.0, Medical device software – Software life cycle processes. (Software / Informatics) Date of Standard: 2006
CEN 13640 Stability testing of in vitro diagnostic reagents. (In Vitro Diagnostics) Date of Standard: 2002
EN ISO 13485:2003 Medical devices – Quality management systems – Requirements for regulatory purposes (ISO 13485:2003) German Version EN ISO 13485:2003
EN ISO 14971:2001 + A1 Medical devices – Application of risk management to medical devices (ISO 14971:2000 + A1:2003) German Version EN ISO 14971:2001 + A1:2003
CLSI EP09-A2, Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline. (In Vitro Diagnostics) Date of Standard: 2002
CLSI EP05-A, Vol.19, No. 2 Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline Date of Standard: 1999
CLSI C28-A2, How to Define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline – Second Edition. (In Vitro Diagnostics) Date of Standard: 2000
CLSI EP07-A, Vol. 22, No. 27 Interference Testing in Clinical Chemistry; Approved Guideline
L. Test Principle:
Blood sample is added to a cylindrical blood holding cup. An oscillating pin is immersed into the blood holding cup. Movement is obstructed when a clot forms and attaches to the pin and cup surfaces. As the clot becomes firmer, the rotational movement of the pin is reduced. The motion of the pin is detected by an optical detection system. The rotational movement of the pin is converted into amplitude. An amplitude of 0 mm means unobstructed rotation, and an amplitude of 100 mm can be regarded as blocking of the pin by the clot. The TEM amplitude is a measure of the clot firmness.
The following parameters are used to describe the clot firmness:
MCF (mm) – Maximum clot firmness (maximum amplitude) of the developed clot during the test
A10, A20 – Clot firmness (amplitude) at the time points 10 and 20 minutes after CT
CT (s) – Clotting time The time from test start until first significant level of clot firmness (2mm) is reached
CFT (s) – Clot formation time. The time from the CT/R until a clot firmness of 20 mm is reached
α – Alpha angle. Angle between the baseline and a tangent to the clotting curve through the 2 mm (CT/R) point.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
{4}
Dose-response curves for heparin, for dilution and for urokinase on the INTEM assay, were presented to demonstrate thromboelastograph principles: coagulation kinetics, clot firmness, and hyperfibrinolysis.
Dose response to UF heparin was presented to demonstrate coagulation kinetics. It is the most widely used anticoagulant in clinical practice, and it inhibits plasma coagulation (mainly FIIa). Heparin was spiked in the concentrations 0.1, 0.2, 0.4, 0.8, and 1.6 U/ml. A dose-dependent effect was seen on all INTEM parameters.
Dose response to dilution affect was presented to demonstrate clot firmness. Blood samples depleted of platelets or fibrinogen decreases clot firmness. To simulate the depletion of platelets and fibrinogen due to plasma expander substitution in a patient, normal blood was diluted in saline to 90, 80, 70, 60 and 50%. A pronounced dose-dependent effect was seen on the INTEM A20 (Figure 10), the CFT and the Alpha Angle. The effect on the CT parameter was very weak. Only at the 50% dilution the CT results began to prolong in comparison to the 100% whole blood control, but the values were never elevated above the upper reference limit.
Hyperfibrinolysis dose response was demonstrated in-vitro, through spiking studies with the physiological plasmin activator urokinase was spiked in the concentrations 10, 20, 30, 40, and 50 U/ml. In most cases, there was a clot lysis within 1 hour from 20 U/ml on. A decrease of the median Lysis Onset Time (LOT) and A20 parameter was seen from 20 U/ml on. The other parameters CT, CFT, and Alpha were not affected as within the urokinase concentration range tested the clots fully before the clot lysed.
## a. Precision/Reproducibility:
Within-run (channel-to-channel), total, and operator-to-operator precision were evaluated. The sample pool consisted of the following sample types:
- Normal blood samples in the reference ranges
- Level 1 blood samples (at the medical decision limit between normal and hypocoagulable): normal blood samples diluted with physiological saline solution and spiked with a direct thrombin inhibitor
- Level 2 blood samples outside the reference ranges (at the medical decision limit between normal and hypercoagulable): normal sample spiked with fibrinogen (up to approximately 8 g/L)
- ROTROL N (level 1 - mimicking 'normal')
- ROTROL P (level 2 - mimicking 'pathological')
Within-run precision was assessed by performing five consecutive runs of each sample/control. Each run was performed using four different channels of ROTEM® in parallel which resulted in 20 replicates per test and sample.
Total precision was determined by analyzing two levels of control in duplicate on two separate runs, over 20 working days. The two test runs were separated by at least two hours.
Between-operator precision was investigated by five different operators analyzing the two control samples during one working day, using two different channels per test and sample. Results were deemed acceptable based on the following:
{5}
in-TEM®/ hep-TEM®
| | CT
CV (%) | CFT
CV (%) | A-angle
CV (%) | A20
CV (%) |
| --- | --- | --- | --- | --- |
| Within-run | <10 | <20 | <5 | <5 |
| Between Operator | <10 | <30 | <5 | <5 |
NATEM®
| | CT
CV (%) | CFT
CV (%) | A-angle
CV (%) | A20
CV (%) |
| --- | --- | --- | --- | --- |
| Within-run | <15 | <20 | <8 | <8 |
| Between Operator | <15 | <30 | <8 | <8 |
Results are summarized below:
in-TEM®
| | CT
CV (%) | CFT
CV (%) | A-angle
CV (%) | A20
CV (%) |
| --- | --- | --- | --- | --- |
| Within-run | 5.0 | 11.3 | 2.0 | 1.7 |
| Between Operator | 2.7 | 13.7 | 0.6 | 5.4 |
hep-TEM®
| | CT
CV (%) | CFT
CV (%) | A-angle
CV (%) | A20
CV (%) |
| --- | --- | --- | --- | --- |
| Within-run | 6.6 | 12.8 | 2.6 | 2.47 |
NATEM®
| | CT
CV (%) | CFT
CV (%) | A-angle
CV (%) | A20
CV (%) |
| --- | --- | --- | --- | --- |
| Within-run | 5.2 | 10.8 | 7.4 | 3.9 |
| Between Operator | 10.1 | 4.4 | 1.1 | 5.0 |
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Bi-level lyophilized, plasma-based controls to monitor precision and accuracy.
Control ranges are assigned as combined ranges from three centers. Each center determines the ranges over 3 days, performing two determinations twice daily.
Stability
ROTROL N is stable for eight hours after reconstitution, and ROTROL P is stable for four hours after reconstitution. Data was submitted that supported the 8-day in-
{6}
TEM® and NATEM® reagent stability claim, and the 30-day hep-TEM® stability claim.
d. Detection limit:
Not Applicable
e. Analytical specificity:
in-TEM® assay
Aprotinin, tranexamic acid and epsilon-aminocaproic acid (EACA) interference were evaluated per CLSI EP7A. The three potential interferents were spiked in-vitro with two concentrations of the antifibrinolytic substances and compared to the non-spiked interferent free control sample.
| Interferent | Maximum Concentration | Lower Concentration |
| --- | --- | --- |
| Aprotinin | 400 KIU/ml | 200 KIU/ml |
| Tranexamic Acid | 60 ug/ml | 30 ug/ml |
| EACA | 600 ug/ml | 300 ug/ml |
hep-TEM® assay
Inactivation of low molecular weight and unfractionated heparin was demonstrated for the he-TEM® assay;
| Unfractionated Heparin (UFH) (Anti-FXa U/ml) | CT in-TEM (sec) | CT hep-TEM (sec) |
| --- | --- | --- |
| 0 | 166 | - |
| 1 | >600 | 146 |
| 10 | >>1000 | 161 |
| Low molecular weight Heparin (Anti-FXa U/ml) | CT in-TEM (sec) | CT hep-TEM (sec) |
| --- | --- | --- |
| 0 | 170 | 172 |
| 1 | 279 | 181 |
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Clinical samples from patients with suspected or acute hemostasis disorders in the peri- or postoperative phase or at the ICU were included in the study. The study included patients with normal coagulation, hypocoagulable and hypercoagulable states in order to obtain results over the whole range of the individual parameter results. A unique sample identifier, the reason for the surgery, and a brief qualitative description of the clinical finding at the time of blood draw were obtained for each patient included in the study. Patients were recruited from three (3) US sites. The samples for the activated tests were analyzed within two hours from blood draw, the non-activated test (NATEM) was analyzed within one hour.
In-TEM
{7}
| | n | Min ROT EM®/ TEG ® | Max ROT EM®/ TEG ® | Slope Deming | Intercept Deming | Slope OLS | Intercept OLS | R OLS |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| CT vs. R | 3
6 | 2/3 | 10/23 | 0.38 | 0.56 | 0.37 | 0.60 | 0.9392 |
| CFT vs. K | 3
5 | 1/1 | 5/4 | 1.00 | -0.13 | 0.85 | 0.13 | 0.8484 |
| α vs. Angle | 3
5 | 50/43 | 84/79 | 0.65 | 31.14 | 0.54 | 38.24 | 0.7561 |
| MCF vs. MA | 3
5 | 41/41 | 82/84 | 1.08 | -8.40 | 1.03 | -5.35 | 0.9589 |
8
{8}
Hep-TEM
| | n | Min ROT EM®/ TEG ® | Max ROT EM®/ TEG ® | Slope Demi ng | Interc ept Demi ng | Slo pe OL S | Interc ept OLS | R OLS |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| CT vs. R | 6 7 | 2/2 | 9/24 | 0.25 | 1.33 | 0.2 5 | 1.37 | 0.845 3 |
| CFT vs. K | 6 7 | 1/1 | 6/6 | 1.26 | -0.59 | 0.9 9 | -0.09 | 0.819 3 |
| α vs. Angle | 6 7 | 44/38 | 84/81 | 0.87 | 15.71 | 0.6 7 | 28.86 | 0.740 8 |
| MCF vs. MA | 6 6 | 35/36 | 80/81 | 1.06 | -5.77 | 0.9 5 | 0.69 | 0.904 7 |
NATEM
| | n | Min ROT EM® / TEG ® | Max ROT EM® / TEG ® | Slope Demi ng | Intercept Demi ng | Slope OL S | Intercept OLS | R OLS |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| CT vs. R | 7 4 | 3/2 | 58/91 | 0.39 | 4.94 | 0.3 7 | 5.27 | 0.823 2 |
| CFT vs. K | 6 4 | 1/1 | 18/26 | 0.72 | 0.96 | 0.6 4 | 1.25 | 0.852 4 |
| α vs. Angle | 6 1 | 9/1 | 78/77 | 0.84 | 11.02 | 0.7 6 | 15.42 | 0.889 3 |
| MCF vs. MA | 6 1 | 34/39 | 84/87 | 1.01 | -6.79 | 0.9 2 | -0.89 | 0.905 9 |
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The reference ranges were determined according the recommendations from guideline CLSI C28-A2. 149 patients were recruited according to pre-determined inclusion/exclusion criteria.
{9}
In-TEM®/hep-TEM®
| CT (sec) | CFT (sec) | A-angle (°) | A20 (mm) | MCF (mm) |
| --- | --- | --- | --- | --- |
| 122-208 | 45-110 | 70-81 | 51-72 | 51-72 |
NATEM®
| CT (sec) | CFT (sec) | A-angle(°) | A20 (mm) | MCF (mm) |
| --- | --- | --- | --- | --- |
| 254-837 | 72-357 | 39-75 | 40-67 | 46-69 |
Each laboratory is recommended to establish a site specific reference range.
N. Instrument Name:
ROTEM® delta
O. System Descriptions:
1. Modes of Operation:
Automatic
2. Software:
Instrument uses Linux operating system. No unauthorized software can be installed by the user. No user access at the operating system level is allowed. Software on the measurement path is used for measurement result calculation, controlling functions, monitoring the measurement process and data analysis. The measurement is represented in a graphical picture and as numeric results.
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
3. Specimen Identification:
Sample identification information is manually entered into the instrument.
4. Specimen Sampling and Handling:
Samples are collected in 3.2% sodium citrate, and pre-warmed to 37°C in a sample holding cup. The instrument automatically pipettes samples and reagents. Samples are stored at room temperature, and must be used within four hours of collection.
5. Calibration:
The ROTEM® delta device uses a two tier calibration. A factory set calibration that involves the elasticity of the spring connector, and a zero-signal auto-calibration that is monitored as long as the instrument is turned on.
6. Quality Control:
The controls lyophilized plasma based reagents that have to be reconstituted fresh for each test run. The controls are recommended to be run with each series of assays and with any change of personnel operating the tests. The package inserts of ROTROL N and ROTROL P show batch-specific target reference ranges for each parameter of the ROTEM® tests. The material can therefore not only be used to monitor precision, but also the accuracy of the system (device, reagents, and test operation).
P. Other Supportive Instrument Performance Characteristics Data Not Covered In the "Performance Characteristics" Section above:
None.
{10}
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
11
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.