WEDGELOC SUTURE ANCHOR WITH OPTI-FIBER SUTURES, MODEL 1000-01-055
Applicant
Medshape Solutions
Product Code
HWC · Orthopedic
Decision Date
Mar 25, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MedShape Solutions, Inc., WedgeLoc™ Suture Anchor with Opti-Fiber™ Sutures are intended for fixation of suture to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow and pelvis in the following procedures: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Shoulder: Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Foot/Ankle: Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Knee: Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotial Band Tenodesis Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial collateral Ligament Reconstruction
Device Story
WedgeLoc™ Suture Anchor is a sterile, single-use, two-piece 'push-in' bone anchor; consists of crosslinked Methyl methacrylate (MMA) polymer body and PEEK-OPTIMA® eyelet. Pre-loaded with two non-absorbable USP No. 2 Opti-Fiber™ polyblend sutures. Device utilizes compression to secure eyelet and sutures into pre-drilled bone hole. Used by surgeons in orthopedic procedures (shoulder, foot/ankle, knee, hand/wrist, elbow, pelvis) to approximate soft tissue to bone. Output is mechanical fixation of suture to bone; facilitates healing by maintaining tissue position. Benefits include secure soft tissue attachment during surgical repair.
Clinical Evidence
Bench testing only. In-vitro functional performance testing conducted in cancellous and cortical bone analogues. Biocompatibility screening performed in compliance with ISO 10993:2003.
Indicated for fixation of suture to bone in shoulder, foot/ankle, knee, hand/wrist, elbow, and pelvis for soft tissue repair procedures including rotator cuff, Bankart, SLAP, biceps tenodesis, ligament reconstructions, and tendon repairs.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for Medshape Solutions. The word "MEDSHAPE" is in large, bold, black letters. Underneath the word "MEDSHAPE" is a line, and below that is the word "SOLUTIONS" in smaller, black letters.
510(k) Page __________________________________________________________________________________________________________________________________________________________________ of ___________________________________________________________________________________________________________________________________________________________________________
## 510(k) Summary
K083792 510(k) Number:
March 18th, 2008 Date Prepared:
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
- Submitter: A. MedShape Solutions, Inc. 1575 Northside Drive, Suite 440 Atlanta, Georgia 30318
- B. Company Contact: Jack Griffis Vice President, Research & Development (404) 249-9156 x11 (direct) (404) 249-9158 (fax) Jack Griffis@MedShapeSolutions.com
- ். Preparer: Max Sherman Sherman Consulting Services, Inc. 339 East Wildwood Drive Warsaw, Indiana 46580 (574) 269-1928 (direct) (574) 267-3845 (fax)
- Device Information: D. WedgeLoc™ Suture Anchor with Opti-Fiber™ Sutures Trade Name: Suture Anchor Common Name:
- Fastener, Fixation, Non-degradable, Soft Tissue E. Classification Name: HWC/MBI and GAT 21 CFR 888.3040
- . F. Predicate Device(s): Arthrex PushLock™ suture anchor, K051219 Arthrex FiberWIRE™ polyblend suture, K021434
{1}------------------------------------------------
- G. Physical Description:
The proposed WedgeLoc™ Suture Anchor is a sterile, single use, two piece "push-in" bone anchor. The suture anchor is designed to use the principles of compression to force the eyelet, pre-loaded with non-absorbable suture, into a hole created in the bone. The suture anchor body is comprised of a crosslinked Methyl methacrylate (MMA) polymer (comprised of Memori™ 7111. refer to MAF #1533) and PEEK eyelet (comprised of PEEK-OPTIMA® by Invibio®, refer to MAF #1209).
Each suture anchor is preloaded with two non-absorbable USP No. 2 sutures. The polyblend surgical suture provided with the proposed WedgeLoc™ Suture Anchor is provided by CP Medical, and has previously been cleared for market under K041894. In addition, these sutures meet USP requirements as described in the USP monograph for non-absorbable surgical sutures. Opti-Fiber™ non-absorbable polyblend surgical sutures are made with a braided, twisted core of polyethylene terephthalate and a polyethylene cover and are identical to the CP-Fiber™ USP No. 2 sutures described in K041894. In addition, the predicate Arthrex FiberWIRE™ polyblend suture was listed as the predicate for the CP-Fiber™ in the premarket notification K041894.
- H. Indications for Use:
The WedgeLoc™ Suture Anchor and Opti-Fiber™ sutures are indicated for fixation of suture to bone in the shoulder, foot/ankle, knee, hand/wrist and elbow in the following procedures:
- Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Shoulder: Tenodesis, Acromio-Clavicular Separation Repair. Biceps Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
- Lateral Stabilization, Medial Stabilization, Achilles Tendon Foot/Ankle: Repair. Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair
- Medial Collateral Ligament Repair, Lateral Collateral Ligament Knee: Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
- Collateral Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Ligament Reconstruction
- Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Elbow: Radial collateral Ligament Reconstruction
Image /page/1/Picture/12 description: The image shows the logo for MEDSHAPE. The word "MEDSHAPE" is in large, bold, black letters. Underneath "MEDSHAPE" is the word "SOLUTIONS" in smaller, thinner, black letters.
{2}------------------------------------------------
K083792(pg 3/3)
510(k) Page ______________ of_
l. Comparison of Technological Characteristics: The WedgeLoc™ Suture Anchor and Opti-Fiber™ sutures are substantially equivalent in design, function and intended use to the following predicate devices:
Arthrex PushLock™ suture anchor, K051219 Arthrex FiberWIRE™ polyblend suture, K021434
Both suture anchors are comprised of a two part assembly; main body and evelet. The evelets of both suture anchors are made from PEEK-OPTIMA® material. The main body of the proposed WedgeLoc™ Suture Anchor differs from the predicate device as it is comprised of a cross-linked Methyl methacrylate (MMA) (Memori™ 7111) instead of PEEK-OPTIMA® material. The Memori™ 7111 has successfully passed biocompatibility screening in compliance with ISO 10993:2003 Biological evaluation of medical devices (refer to MAF #1533). Both anchors are deployed using a "push in" or "tap in" process for securing their polyblend sutures into bone. Both suture anchors have the same indications for use.
Both predicate and proposed Opti-Fiber™ sutures are indicated for use in soft tissue approximation; both meet USP for non-absorbable sutures, may be silicone coated or uncoated, and may be dyed or undyed with an appropriate FDA listed color additive. The proposed Opti-Fiber™ suture has previously been cleared for market as per K041894.
Functional performance testing was conducted in-vitro. This testing was conducted in both cancellous and cortical bone analogues. Analysis of the results substantiates the statement that the proposed device performs as well as or better than the predicate device.
Jack Griffis Vice President, Research & Development
Image /page/2/Picture/9 description: The image shows the logo for MEDSHAPE SOLUTIONS. The word "MEDSHAPE" is in large, bold, black letters. Below it, the word "SOLUTIONS" is in smaller, thinner, black letters with a line above it.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image contains the text 'Public Health Service'. The text is in a simple, sans-serif font and is horizontally aligned. The text appears to be part of a document or sign.
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms.
MAR 2 5 2009
ood and Drug Administration 200 Corporate Boulevard Rockville MD 20850
MedShape Solutions % Mr. Jack Griffis Vice President, Research & Development 1575 Northside Drive, Suite 440 Atlanta, Georgia 30318
Re: K083792
Trade/Device Name: WedgeLoc™ Suture Anchor with Opti-Fiber™ Sutures Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic fixation fastener Regulatory Class: II Product Code: HWC, MBI, GAT Dated: March 18, 2009 Received: April 12, 2009
Dear Mr. Griffis:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becared the device is substantially equivalent (for the indications felerenced above and nave acterificated predicate devices marketed in interstate 101 use stated in the encreate) to regary che enactment date of the Medical Device Amendments, or to commence provide to May 20, 1978, the encordance with the provisions of the Federal Food, Drug, devices that have been require approval of a premarket approval application (PMA). and Cosment Acr (Act) that to not required to the general controls provisions of the Act. The r ou may, merciole, market the device, boojes resuirements for annual registration, listing of general Controls provisions of the rist labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it II your device is classified (soc above) into Existing major regulations affecting your device can may be subject to such additional oblines - Lifeting and to 898. In addition, FDA may be found in the Gode acements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issuation of a backed complies with other requirements of the Act
that FDA has made a determination that your device complies with other must that TDA has made a determination administered by other Federal agencies. You must or any Federal statutes and regulations daminities or registration and listing (21 comply with an the Act 3 requirements, n.v., good manufacturing practice requirements as set CFK Part 807); labeling (21 CFR Part 820); and if applicable, the electronic a forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
## Page 2 - Mr. Jack Griffis
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
For Peter D. Rummell
Mark N. Melkerson ) 0 ( Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Page ___ of _
## Indications for Use
510(k) Number: K083792
Device Name: WedgeLoc™ Suture Anchor and Opti-Fiber™ Sutures
Indications for Use:
The MedShape Solutions, Inc., WedgeLoc™ Suture Anchor with Opti-Fiber™ Sutures are intended for fixation of suture to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow and pelvis in the following procedures:
- Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Shoulder: Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
- Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Foot/Ankle: Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair
- Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Knee: Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
- Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction
- Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial collateral Ligament Reconstruction
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
|-------------------------------------------------|---|--------|------------------------------------------------|--|
|-------------------------------------------------|---|--------|------------------------------------------------|--|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| (k) Number | 14083792 |
|------------|----------|
|------------|----------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.