BIOLOX DELTA CERAMIC FEMORAL HEADS

K083762 · Smith & Nephew, Inc. · LZO · Mar 11, 2009 · Orthopedic

Device Facts

Record IDK083762
Device NameBIOLOX DELTA CERAMIC FEMORAL HEADS
ApplicantSmith & Nephew, Inc.
Product CodeLZO · Orthopedic
Decision DateMar 11, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis; or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthrifis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. The Biolox Delta Ceramic femoral heads are for single use only.

Device Story

Biolox Delta Ceramic femoral head; 12/14 taper junction; mechanical lock to Smith & Nephew femoral hip stems; articulates against polyethylene acetabular liner. Used in OR by orthopedic surgeons during primary or revision hip arthroplasty. Provides load-bearing surface for hip joint replacement; restores joint function; alleviates pain from degenerative or traumatic hip conditions. Single-use device.

Clinical Evidence

Bench testing only. No clinical data provided. Mechanical testing data indicates the device is capable of withstanding expected in vivo loading without failure.

Technological Characteristics

Biolox Delta ceramic material; 12/14 taper connection; femoral head form factor; intended for use with Smith & Nephew femoral hip stems and polyethylene acetabular liners. Single-use.

Indications for Use

Indicated for individuals undergoing primary or revision hip surgery due to trauma, inflammatory joint disease (e.g., rheumatoid arthritis), or noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, pelvic fracture, diastrophic variant, remote osteomyelitis, nonunion, femoral neck/trochanteric fractures, femoral osteotomy, Girdlestone resection, fracture dislocation, or deformity correction).

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Summary of Safety and Effectiveness for Summary of Salety and Enectiveness for Smith & Nephew Biolox Delta Ceramic Femoral Heads #### Contact Person and Address Date of Summary: August 27, 2008 Mandy L. Coe Regulatory Affairs Specialist Smith & Nephew, Inc. Orthopaedic Division 1450 E. Brooks Road Memphis, Tennessee 38116 (901) 399-6277 Name of Device: Smith & Nephew Biolox Delta Ceramic Femoral Heads Common Name: Femoral Head Device Classification: 21 CFR 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis - Class II Device Product Code: LZO #### Device Description The Biolox Delta Ceramic femoral heads feature a 12/14 taper and are intended to be used with existing Smith & Nephew femoral hip stems. The ceramic femoral head mechanically locks with the femoral hip stem via a taper junction and articulates against a Smith & Nephew polyethylene acetabular liner. The subject devices are similar in design and function to the Biolox Forte Alumina Ceramic femoral heads cleared via 510(k) premarket notifications K981847, K991162, and K022958 respectively. ## Mechanical Testing A review of the mechanical data indicated that the Smith & Nephew Biolox Delta Ceramic femoral heads are equivalent to devices currently cleared for market and are capable of withstanding expected in vivo loading without failure. #### Intended Use Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis; or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthrifis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. The Biolox Delta Ceramic femoral heads are for single use only. ## Substantial Equivalence Information The Smith & Nephew Biolox Delta Ceramic femoral heads are similar in overall design, material and indications to the Smith & Nephew Biolox Forte Alumina Ceramic femoral heads cleared via 510(k) premarket notifications K981847, K991162, and K022958 as well as the Zimmer Biolox Delta Ceramic femoral heads cleared via K071535. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle. MAR 1 1 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Smith & Nephew, Inc. % Ms. Mandy Coe Regulatory Affairs Specialist 1450 E. Brooks Road -Memphis, Tennessee 38116 Re: K083762 Trade/Device Name: Biolox Delta Ceramic Femoral Heads Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulation Class: Class II Product Code: LZO Dated: February 27, 2009 Received: March 2, 2009 Dear Ms. Coe: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can . be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {2}------------------------------------------------ Page 2 -- Ms. Mandy Coe forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. L. Arimid. for Mark N. Melkerson Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known): K083762 Device Name: Biolox Delta Ceramic Femoral Heads Indications for Use: Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomvelitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques: femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Smith & Nephew Biolox Delta Ceramic femoral heads are for single use only. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) SIVISION SIDU-1 · Vivision of General, Restorative nd Neurological Devices ber Page 1 of Indications
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...