← Product Code [GGW](/productcode/GGW) · K083729

# HEMOCLOT QUANTI- V-L, FACTOR V-L CALIBRATOR, BIOPHEN V-L CAL (UNDILUTED), BIOPHEN ACT. PC-R CONTROL PLASMA AND BIOPHEN (K083729)

_Aniara Diagnostica, LLC · GGW · May 8, 2009 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K083729

## Device Facts

- **Applicant:** Aniara Diagnostica, LLC
- **Product Code:** [GGW](/productcode/GGW.md)
- **Decision Date:** May 8, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.7925
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

HEMOCLOT Quanti VL is an in vitro diagnostic test for the quantitative determination of Factor V Leiden (FVL) activity in human citrated plasma, using automated or manual methods, by its resistance to the action of Activated Protein C ( APC). Patients with the Factor V-L (mutalion R506Q), are exposed to an increased thrombotic risk. Biophen V-L CAL (undilated), at defined Factor V-Leiden (FV-L) concentrations, for the callibration of Factor V-Leiden aclivity on human plasma, using the Hemoclot Quanti V-L kit. Factor V-L Cafibrator, pre diluted (1:20) human plasmas. at defined Factor V-Leiden (FV-L) concentrations, for the calibration of Factor V-L activity on human plasma using Hemoclot Quanti V-L Kit. Biophen Act PC-r Control Plasma contains human plasma, presenting an Activated Protein C Resistance (APC-R) is used as quality control plasma for the testing of Activated Protein C resistance (APC-R) on human plasma using Hemoclot Quarti V-L Kit. (Updated) BIOPHEN Normal Control Plasma is a set of 12 vials of normal citrated human plasma for the quality control of some coagulation factors. The following table shows parameters, which are measured using assays from HYPHEN BioMed or from other manufacturers, and according to the package inserts: Assays Reagents Manufacturers Reference ATIII Biophen ATIII Hyphen Biomed 221102/221105 Protein C Biophen Protein C Hyphen Biomed 221202/221205 aPC resistance (FV Leiden) Hemoclot Quanti V-L Hyphen Biomed CK065K Lupus Anticoagulant DVVtest®/DVVconfirm® American Diagnostica 810/815/815L DVVtest, DVVconfirm are registered trademarks from American Diagnostica, Inc. The BIOPHEN Normal Control Plasma is tested for the absence of Lupus Anticoagulant and can be used as a negative control for this investigation. This control plasma is also tested for the absence of Activated Protein C resistance (Act PC-7). When the APTT is performed with or without Activated Protein C (APC) the ratio obtained (APTT + APC/APTT) is ≥2.00.

## Device Story

HEMOCLOT Quanti V-L is an in vitro diagnostic clotting assay measuring Factor V Leiden (FVL) activity in human citrated plasma. Input: patient citrated plasma. Process: plasma mixed with clotting factors (prothrombin, fibrinogen, protein S, APC) and heparin-neutralizing substance; purified Factor Xa and phospholipids added; clotting initiated by calcium. Principle: clotting time is inversely proportional to FVL concentration; resistance to APC indicates FVL mutation. Output: quantitative FVL activity level. Used in clinical laboratories; operated by technicians/lab personnel. Results interpreted by clinicians to identify APC resistance associated with Factor V R506Q mutation. Benefits: aids in diagnosis of thrombophilic conditions.

## Clinical Evidence

Bench testing only. Four-site method comparison study (n=189) compared Hemoclot Quanti V-L to predicate (Coatest APCr). Results: 94.18% total agreement, 89.74% positive agreement, 97.30% negative agreement. Comparison with molecular biology diagnosis (n=63) showed 97% agreement. Precision studies (intra- and inter-assay) yielded CV < 6%. Linearity validated between 10% and 100% FVL (r² > 0.98). Specificity confirmed with factor-deficient plasmas and samples with heparin or AVK therapy.

## Technological Characteristics

Lyophilized reagents (human fibrinogen, prothrombin, Protein S, APC, Factor Xa, rabbit brain phospholipids). Clotting-based assay. Quantitative measurement via calibration curves. Compatible with automated coagulation analyzers (e.g., STAR, KC10) or manual water bath. Reagents stable at 2-8°C.

## Regulatory Identification

A partial thromboplastin time test is a device used for primary screening for coagulation abnormalities, for evaluation of the effect of therapy on procoagulant disorders, and as an assay for coagulation factor deficiencies of the intrinsic coagulation pathway.

## Predicate Devices

- COATEST APC Resistance V ([K963111](/device/K963111.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k083729

B. Purpose for Submission:
Bundled submission for clearance of a new Factor V-Leiden (FVL) assay and its associated calibrators and control, and modification for a currently cleared coagulation control (k043451).

C. Measurand:
Factor V-Leiden

D. Type of Test:
Clotting Assay

E. Applicant:
ANIARA DIAGNOSTICA, LLC

F. Proprietary and Established Names:
Hemoclot Quanti V-L
Factor V-L Calibrator
Biophen V-L Cal (Undiluted)
Biophen ACT PC-r Control Plasma
Biophen Normal Control

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  GGW, Test Partial | Class II | 21 CFR 864.7925 | 81 Hematology  |
|  Thrombin Time |  | 21 CFR 864.542521 |   |
|  GGN, Plasma, |  | CFR 862.1150 |   |
|  Coagulation Control |  |  |   |
|  JIT, Calibrator, |  |  |   |
|  Secondary |  |  |   |

H. Intended Use:

1. Intended use(s):
HEMOCLOT Quanti V-L is a clotting method for measuring the Factor V Leiden (FVL) activity in human citrated plasma, by its resistance to the action of Activated Protein C (APC).

Factor V-L Calibrator is lyophilized, pre-diluted (1:20) human plasmas, at defined Factor V-Leiden (FV-L) concentrations, for the calibration of Factor V-L activity quantitative clotting assay on human citrated plasma, using the HEMOCLOT Quanti. V-L Kit (ref CK065K).

BIOPHEN V-L CAL (Undiluted) is lyophilized, undiluted human plasmas, at defined Factor V-Leiden (FV-L) concentrations for the calibration curve of Factor

{1}

V-L activity quantitative clotting assay on human citrated plasma, using HEMOCLOT Quanti V-L Kit (ref CK065K).

**BIOPHEN Act PC-r Control Plasma** kit contains human plasma, presenting an activated Protein C Resistance (APC-R), usually correlated with the genetic mutation of Factor V R506Q. This plasma is used as quality control plasma for the testing of Activated Protein C Resistance (APC-R).

**BIOPHEN Normal Control Plasma** is a set of 12 vials of normal citrated human plasma for the quality control of some coagulation factors (ATIII, Protein C, aPC resistance FV Leiden, Lupus Anticoagulant).

2. **Indication(s) for use:**
Same as Intended Use.

3. **Special conditions for use statement(s):**
For Prescription use only.

4. **Special instrument requirements:**
Sigma Diagnostics KC 1A (k955724)
Sigma Diagnostics AMAX 400 (k972260)
American Bioproducts STA (k942117)
American Bioproducts STA-R (k082675)
Manual (Water Bath)

I. **Device Description:**

**HEMOCLOT Quanti V-L** is an in vitro diagnostic kit containing two reagents to measure the Factor V Leiden concentration in citrated plasma, by its resistance to the Activated Protein C. Reagent 1 is a clotting mixture containing human Fibrinogen, human Prothrombin, and Protein S at a constant concentration, optimized for the assay, and human Activated Protein C, lyophilized. It also contains a heparin neutralizing substance. Reagent 2 is Purified Human Factor Xa, containing rabbit brain phospholipids (cephalin), lyophilized. Calibrators and controls are required and sold separately.

**FVL Calibrator** kit: 12 vials (3 sets of 4 vials) of 1 mL of prediluted human plasma (1:20) at different concentrations of FV-L, ranging from about 10% to 100%.

**Biophen V-L Cal (Undiluted)** kit: 9 vials (3 sets of 3 vials) of 0.5 mL of undiluted human plasma at different concentrations of FV-L, to cover the assay range, from about 10% to 100%. (3 vials for each concentration, 4 points for the calibration curve).

**Act. PC-r Control Plasma** kit: 12 vials of 0.5 mL of human plasma, presenting an APC-R, citrated and lyophilized.

**BIOPHEN Normal Control Plasma**: 12 vials of 1 mL of normal citrated human plasma, lyophilized.

2

{2}

J. Substantial Equivalence Information:

|   | New Device | Predicate Device  |
| --- | --- | --- |
|   | HEMOCLOT Quanti V-L, Factor V-L Calibrator, Biophen V-L CAL (undiluted), Biophen Act PC-r Control Plasma, Biophen Normal Control Plasma | COATEST APC Resistance V (K963111). Control Plasmas Level 1 and Level 2 are included in this device kit.  |
|   | Similarities |   |
|  Intended Use | Measurement of Factor V Leiden activity in human citrated plasma by its resistance to the action of Activated Protein C. | Same  |
|  Assay Type | Clotting Method | Same  |
|  Test Sample | Human Citrated Plasma | Same  |
|  Assay Principle | Measurement of Factor V Leiden coagulation activity based on insensitivity of Factor V Leiden to the action of Activated Protein C. | Same  |
|  Stability of Unopened Reagents | Stable at 2-8°C until expiry date | Same  |
|  Controls, Normal | Citrated human plasma, lyophilized, to provide quality control in the normal range. APC V Ratio > 2. | Same  |
|  Controls, Abnormal (APC Resistant) | Citrated human plasma, lyophilized, to provide quality control for APC resistant range of the assay. APC V Ratio < 1.80 | Same  |
|   | Differences |   |
|  Assay Principle | Assay is performed in the presence of Activated Protein C and Protein S (one single test for each patient). In the presence of APC, prolongation of clotting time is inversely related to the amount of Factor V Leiden. Normal Factor V is not measured. APC resistance due to Factor V Leiden is indicated when the %FVL value is above or equal to the cut-off value. | Assay is performed in the presence and absence of Activated Protein C (two tests for each patient). In the presence of APC, prolongation of clotting time is directly related to the concentration of Normal Factor V, and inversely related to the amount of Factor V Leiden. APC resistance due to Factor V Leiden is indicated when the APC-V ratio is below or equal to the cut-off value.  |
|  Assay Calibration | Assay of calibrator plasma at defined Factor V Leiden concentrations. Calibrator plasmas available in prediluted (FVL Calibrator) and undiluted (BIOPHEN V-L Cal) form, lyophilized. | Not Applicable.  |

{3}

4

# K. Standard/Guidance Document Referenced (if applicable):

|  STANDARDS  |
| --- |
|  Title and Reference Number  |
|  |

Other Standards

|  GUIDANCE  |   |   |   |
| --- | --- | --- | --- |
|  Document Title | Office | Division | Web Page  |
|  |   |   |   |

# L. Test Principle:

The Hemoclot Quanti V-L kit is a clotting method triggered by purified Factor Xa, containing Phospholipids, and Calcium, and is performed on diluted test plasma, in presence of Activated Protein C and Protein S. In the first step, diluted plasma is mixed with purified clotting Factors [clotting mixture containing Prothrombin, Fibrinogen, Protein S and Activated Protein C (APC)], in a constant and optimized concentration. Then, the purified Factor Xa in the presence of Phospholipids (PLP) is added. Clotting is initiated by the addition of Calcium (Ca2+). The clotting time is then recorded. Clotting time measured is inversely proportional to the concentration of factor V-L. There is an inverse linear relationship, on a lin-log graph paper, between the factor V-L concentration and the corresponding clotting time.

FVL Calibrator and Biophen V-L Cal (Undiluted) are two calibrators that may be used with the HEMOCLOT Quanti.V-L kit. Biophen V-L CAL (Undiluted) is lyophilized undiluted human plasma containing different concentrations of FV-L, to cover the assay range, from about 10% to 100%. FVL Calibrator is lyophilized prediluted human plasma (1:20) at different concentrations of FV-L, to cover the assay range, from about 10% to 100%.

Biophen Normal Control Plasma and Biophen Act. PC-r Control Plasma are quality control plasmas which may be used with the kit.

# M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Two lots of Hemoclot Quanti V-L were analyzed for inter-assay precision. In one study the assay was performed with the waterbath, over four days using the internal reference calibrator at 4 different concentrations (10%, 25%, 50%, 100%) of FVL. Normal Control and the Biophen APCr control were used as controls. In another study using the waterbath, 9 vials of reagent was assayed at the 100% FVL level by several technicians. Inter-assay precision was also demonstrated using the KC10 analyzer by performing 10 assay runs over

{4}

several days using 2 technicians. The calibration curve was run, which represents the entire range of the assay, 10%-100% FVL.

Intra-assay precision was demonstrated on the waterbath by assaying 10 reagent vials from the same lot at the 100% FVL Level, in the single run.

All analyses yielded CV&lt;6%.

Additional precision testing was performed at the medical decision points (10%, 25%) on the KC10 (10 runs), STA-R (10 runs), and waterbath (5 runs). Two samples were assayed over 3 days using 3 technicians. For each run a reference calibration curve and Normal Control and Biophen APCr was performed.

b. Linearity/assay reportable range:

The linearity of the Hemoclot Quanti V-L assay was evaluated by use of FVL Calibrator plasmas at 4 different concentrations of FVL, 10%, 25%, 50%, and 100%. The linearity of the assay was validated on the waterbath and several commercial analyzers (KC10, STA, STAR and Amax 400). The assay is linear between 10% and 100% FVL, r²&gt;0.98. The linearity of the assay below 10% is not determined. Any value below 10% is considered normal. Values between 25% and 75% are expected for heterozygous plasmas. Values &gt;75% are expected for homozygous plasmas. Values between 10 -25% are considered inconclusive.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The Internal Reference Standard used for assigning values to the calibrators and controls is traceable to pooled normal and APCr human plasmas, controlled for normal Factor V content against commercially available reference plasma pools, and verified against an NIBSC standard for normal Factor V (SSC/ISTH Secondary Coagulation Standard).

The calibration curve for the Hemoclot Quanti V-L kit is established with either Biophen V-L cal (undiluted) or Factor V-L Calibrator kits. By definition, normal plasma contains 0% FVL and heterozygous plasma, from patients carrying the R506Q mutation, contains 50% FVL. For the Biophen V-L cal (undiluted) kit, calibrators are made from mixtures of pooled heterozygous plasmas and pooled normal plasmas to yield calibration plasmas at concentrations of 10%, 25% and 50%. For the Factor V-L Calibrator kit, calibrators are made from mixtures of pooled heterozygous plasmas and pooled normal plasmas and diluted appropriately to yield calibration plasmas at concentrations of 10%, 25%, 50% and 100% FVL, ready to use with the Hemoclot Quanti V-L kit. The concentration of each calibrator level is accurately defined, against an internal reference standard, as the mean value from at least 10 vials tested in the same series by manual and/or automated methods, using the Hemoclot Quanti V-L Assay.

5

{5}

Biophen Act. PC-r Control Plasma and Biophen Normal Control Plasma (510k #043451) may be used for quality control with the Hemoclot Quanti V-L kit. Biophen Act PC-r Control Plasma is derived from pooled heterozygous plasma, from patients carrying the R506Q mutation, and contains about 50% FVL. Biophen Normal Control Plasma is derived from pooled normal plasmas and contains about 0% FVL. The FVL concentration of each control is accurately defined, against an internal reference standard, as the mean value of at least 3 vials tested in at least 3 independent series (i.e., at least 9 vials tested) by manual and/or automated methods, using the Hemoclot Quanti V-L Assay.

Stability: Reconstituted reagents are stable for 24 hrs at 2-8°C or 8 hrs at room temperature (18-25°C). Overheating studies support 30 month expiration dating for lyophilized reagents stored at 2-8°C.

d. Detection limit:

Not applicable

e. Analytical specificity:

The specificity of the assay was demonstrated by analyzing factor deficient plasmas. Results from Factor deficient plasmas (V, X, II, PS, PC) showed no interference. Since the assay can be performed in patients with heparin therapy (up to 1 IU/mL), plasma samples spiked with heparin were tested and showed no significant interference as the result of a heparin neutralizing substance in Reagent 1. To determine whether the assay can be performed in patients with Anti-Vitamin K (AVK) therapy, plasma samples from APCr patients with AVK therapy were tested and yielded expected values.

f. Assay cut-off:

The normal cut-off range was determined and validated by analyzing normal patient plasmas and aPCr patient plasmas, carrying R506Q mutation (FVL) identified with molecular technology as Heterozygous (HTZ) or Homozygous (HMZ). Analyses were conducted on water bath, KC10 and STAR analyzers. Normal plasma is expected to measure &lt;10% FVL. Plasmas from patients with Factor V Leiden are expected to measure between 25% and 75% FVL (heterozygous) or &gt;75% (homozygous).

Using the Hemoclot Quanti V-L method, the normal plasma range was determined by analyzing 20 normal plasma samples from a French blood bank on three different instruments with normal and APCr Controls and a selection of ActPCr plasma samples. To verify the cut-off, 42 plasma samples from FVL/HTZ patients out of therapy, 20 plasma samples from FVL/HTZ patients under Anti Vitamin K (AVK) therapy and 18 plasma samples from FVL/HMZ patients were analyzed. Results were acceptable with normal samples measuring &lt;10% FVL, 25% and 75% FVL for FVL/HTZ and &gt;75% FVL for FVL/HMZ samples.

2. Comparison studies:

a. Method comparison with predicate device:

A four site (two in the US and two in Europe) comparison study was

{6}

conducted. 189 plasma samples were analyzed using the test (Hemoclot Quanti V-L) and predicate (Coatest APCr) method. Samples were selected from patients who were previously tested for coagulation disorders and represented both normal and APC-Resistant populations (about 60% of the samples were from normal patients and 40% from APCr patients). Patient plasma samples were selected to cover the range of the assay. Overall, there was 94% agreement between the two methods for determining APC resistance due to the Factor V Leiden mutation. In one study, samples were identified as heterozygous (25), homozygous (11), or normal (27) with regard to FVL mutation using molecular biology methods. There was 97% agreement between the results of test method and the molecular biology diagnosis with regard to APCr resistance.

|  All Sites | Predicate  |   |   |
| --- | --- | --- | --- |
|   |   |  Normal | Abnormal  |
|  Hemoclot Quanti V-L | Normal | 108 | 7  |
|   |  Abnormal | 1 | 70  |
|   |  Inconclusive* | 2 | 1  |

*Outside the range of the Hemoclot Quanti V-L

Percent Total Agreement 94.18% (178/189)

Percent Positive Agreement 89.74% (70/78)

Percent Negative Agreement 97.30% (108/111)

b. Matrix comparison:
Not applicable.

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.

4. Clinical cut-off:
Not applicable.

5. Expected values/Reference range:
Normal plasma is expected to measure &lt;10% FVL. Plasmas from patients with Factor V Leiden are expected to measure between 25% and 75% FVL (heterozygous) or &gt;75% (homozygous).

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K083729](https://fda.innolitics.com/device/K083729)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
