← Product Code [FRN](/productcode/FRN) · K083689

# INFUSOMAT SPACE VOLUMETRIC INFUSION PUMP SYSTEM (K083689)

_B.Braun Medical, Inc. · FRN · Jul 1, 2009 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K083689

## Device Facts

- **Applicant:** B.Braun Medical, Inc.
- **Product Code:** [FRN](/productcode/FRN.md)
- **Decision Date:** Jul 1, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Infusomat® Space Volumetric Infusion Pump System is an electrical, external, volumetric infusion pump system indicated for use with adults, pediatrics and neonates and is intended to provide infusions of parenteral fluids/medications, blood and blood products indicated for infusion through FDA approved routes of administration.

## Device Story

Infusomat® Space is a transportable, linear peristaltic volumetric infusion pump. It delivers parenteral fluids, medications, blood, and blood products. Operated by trained healthcare professionals in hospitals, home care, outpatient, and transport settings. System integrates with SpaceStation docking system featuring SpaceCom communication software. Inputs include infusion parameters entered by clinicians and barcode-scanned data. Features include PCA functionality, outbound wireless data transmission, and barcode-based medication error reduction. SpaceCom supports Ethernet, USB, and serial interfaces; uses CAN bus for internal communication and proprietary protocols for hospital information system integration. Output consists of controlled fluid delivery and data transmission to hospital management systems. Benefits include standardized infusion protocols, medication error reduction, and centralized asset management.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing, design comparison, and technological similarities to predicate devices.

## Technological Characteristics

Linear peristaltic pump; injection-molded thermoplastic components; 12V DC or battery power; CAN bus communication; PCMCIA network card; barcode reader interface (PS2/USB); supports Ethernet, serial, and USB connectivity. Software-controlled infusion parameters and data transmission.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- Infusomat® Space Volumetric Infusion Pump System ([K062700](/device/K062700.md))
- Outlook® Safety Infusion System ([K011975](/device/K011975.md))
- MEDLEY™ PCA Module ([K032233](/device/K032233.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

B. Braun Medical Inc. 510(k) Premarket Notification

K083689

510(k) Summary

JUL - 1 2009

# SUBMITTER:

B. Braun Medical Inc. 1601 Wallace Drive Ste. 150 Carrollton, Texas 75006-6690

Contact: Linda Morgan, R.N., B.S.N. Regulatory Affairs Manager Phone: 972.466.5070

#### DEVICE MANUFACTURED:

B. Braun Melsungen AG Carl-Braun-Strasse 1 34212 Melsungen GERMANY

DEVICE NAME:

Infusomat® Space Volumetric Infusion Pump System

COMMON OR USUAL NAME:

Infusion Pump

## DEVICE CLASSIFICATION:

Class II per Code of Federal Regulation Title 21 §880.5725: Infusion Pump, product code FRN and PCA Infusion Pump, product code MEA

CLASSIFICATION PANEL:

PREDICATE

General Hospital The Infusomat® Space Volumetric Infusion Pump System is of comparable type and is substantially equivalent to the original submission of the Infusomat® Space Volumetric Infusion Pump System marketed under cleared 510(k) K062700. The predicate device for barcode and wireless

data transmission functionalities is B. Braun Medical Inc. Outlook® Safety Infusion System, marketed under cleared The predicate device for PCA 510(k) K011975. functionality is the MEDLEY'IM PCA Module marketed by ALARIS Medical Systems Inc. under cleared 510(k) K032233.

{1}------------------------------------------------

#### DESCRIPTION:

The Infusomat® Space Volumetric Infusion Pump System includes an external transportable electronic infusion pump and pump accessories. B. Braun Medical Inc. has taken great efforts in consulting with groups of actual users of infusion systems as well as hospital administrations from pharmacists to information technology professionals to develop barcode, out-bound wireless data transmission and PCA features that not only enhance the administration of infusions but are also tools in the development of medication error reduction programs while ensuring the validity of transferred data. The Infusomat® Space Volumetric Infusion Pump System can be customized to meet the needs of a particular institution by allowing facilities to enable or disable particular aspects of the bar coding and wireless features, such as nurse ID matching, or by assigning dose or rate limits to particular medications.

** Although the only additions to the Infusomat® Space Volumetric Infusion Pump System are barcode, outbound wireless data transmission and PCA functionality, for convenience of review, the device description may include previously cleared features **

The mechanics of the Infusomat Space have not changed since the original submission.

The Infusomat® Space utilizes a linear peristaltic pumping mechanism and is intended to provide infusions of parenteral fluids. The Infusomat® Space Volumetric Infusion Pump System is intended to be used by trained healthcare professionals in healthcare facilities, home care, outpatient, and medical transport environments. A trained Biomedical Technician must perform a complete set-up of the pump prior to use in a clinical setting.

The system is intended to provide intermittent or continuous flow of parenteral fluids to the patient. Parenteral fluids may include standard fluids and/or medications indicated for infusion as well as blood and blood products.

{2}------------------------------------------------

#### B. Braun Medical Inc. 510(k) Premarket Notification

## SpaceStation with SpaceCom

The B. Braun Space Station is a 12V DC or battery powered flexible docking and communication system for the medical workplace, in particular the intensive medical care. It serves the perspicuous accommodation of the infusion and infusion-syringe pumps Infusomat® Space and Perfusor® Space.

SpaceCom is a communication software that has been integrated into the SpaceStation. SpaceCom supports different interfaces like Ethernet, PS2-Keyboard, Serial, USB ports and PCMCIA network card. For bar-coding, a barcode image reader can be connected to the PS2-Keyboard or USB interface. The pumps are connected with connectors on the inner backside to the SpaceStation. These connectors provide the voltage supply, distribute the addresses in the Space system via a serial interface, transfer data via a bus system (CAN bus) and transmit a staff call, which may be pending.

The Infusomat® Space Volumetric Infusion Pump System must be housed in a SpaceStation that has SpaceCom software installed in order to accomplish the advanced features of bar coding and out-bound wireless data transmission. (SpaceStations are available with or without SpaceCom).

The software communication (CAN bus) allows out-bound flow of information from the infusion pump to an existing hospital information management system. The PCMCIA network card resides in the SpaceStation with SpaceCom. All data transmitted must use the B. Braun information protocol for proprietary data exchange.

The implementation of the PCA feature furthers B. Braun's vision that one pump system can be used throughout an entire facility. The intent of B. Braun is pain therapy management, not only to add a new "pain pump" feature to the existing platform, but to facilitate the hospital-wide management of post-operative pain treatment. Having one system that may be used to administer a wide variety of infusion therapies allow healthcare professionals the benefit of only having to learn one device hospital-wide. The use of one device house-wide helps to ensure consistency in infusion protocols. Medication delivery will be enhanced

{3}------------------------------------------------

by understanding and using the advanced features of the Infusomat® Space and will optimize asset utilization within a healthcare institution. Increased functionality and data acquisition of the Infusomat® Space Volumetric Infusion Pump System support medication error reducing programs and can be used to audit the effectiveness of those programs.

The Infusomat® Space Volumetric Infusion Pump System INTENDED USE: is an electrical, external, volumetric infusion pump system indicated for use with adults, pediatrics and neonates and is intended to provide infusions of of parenteral fluids/medications, blood and blood products indicated for infusion through FDA approved routes of administration.

> The addition of bar code, out-bound wireless data transmission and PCA technologies offer clinicians greater flexibility in administering and managing infusion therapies while using one pump system throughout the healthcare facility. All device programming, and data entry of infusion parameters is performed by trained healthcare professionals according to a physician's order.

The Infusomat®Space Volumetric Infusion Pump System is of comparable type and is substantially equivalent to the original submission of the Infusomat® Space Volumetric Infusion Pump System marketed under cleared 510(k) K062700. The predicate device for barcode and out-bound wireless data transmission functionalities is B. Braun Medical Inc. Outlook Safety Infusion System, marketed under cleared 510(k) K011975. The predicate device for PCA functionality is the MEDLEYTM PCA Module marketed by ALARIS Medical Systems Inc. under cleared 510(k) K032233.

The subject device is a computer- controlled electrical, external, lineal peristaltic infusion pump composed of injection molded thermoplastic components. The components of the subject and all non-solution contact parts. The subject and predicates have implemented the advanced features of infusion therapies in a similar manner

## SUBSTANTIAL EQUIVALENCE:

K083689. PAGE 4 of 5

{4}------------------------------------------------

B. Braun Medical Inc. 510(k) Premarket Notification

> by developing communication protocols that send and receive infusion data in a barcode or wireless (out-bound) format. The software communication (CAN bus) allows flow of information from the infusion pump to an existing hospital information management system. The PCMCIA network card resides in the SpaceStation with SpaceCom. All data transmitted must be use the B. Braun proprietary information protocol, ensuring the integrity of data received or sent.

> The included accessories of the Infusomat® Space Volumetric Infusion Pump System are similar to the predicate devices in that they all offer the clinician maximum flexibility in providing infusion therapy. The subject and predicate systems provide possibilities to clean up and organize the patient bedside by storage of pumps in a filing and communication system. This filing and communication system provides central power supply to the single pumps and tubing retainers to prevent IV line confusion. Also the subject and the predicate provide a means for easy transportation of the modules as well as of the system. Additionally the subject and predicate devices are used to customize the menu items and drug library parameters for each facility. The barcode readers and PCA button accessories are similar in that they both allow for ease of patient use and safeguard the controlled medications being administered.

{5}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and service. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", indicating the department's name and national affiliation. The seal is simple and monochromatic.

Public Health Service

JUL - 1 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Linda Morgan, R.N., B.S.N. Regulatory Affairs Manager B. Braun Medical Incorporated 1601 Wallace Drive, Suite 150 Carrollton, Texas 75006-6690

Re: K083689

Trade/Device Name: Infusomat® Space Volumetric Infusion Pump System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: June 15, 2009 Received: June 17, 2009

Dear Ms. Morgan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

#### Page 2 - Ms. Morgan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Anthony D. Wntar for

Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

B. Braun Medical Inc.

510(k) Premarket Notification

# Indications for Use Statement

Page of of

510(k) Number (if known): K083689

Device Name:

Infusomat® Space Volumetric Infusion Pump System

Indications For Use:

The Infusomat® Space Volumetric Infusion Pump System is an electrical, external, volumetric infusion pump system indicated for use with adults, pediatrics and neonates and is intended to provide infusions of parenteral fluids/medications, blood and blood products indicated for infusion through FDA approved routes of administration.

Prescription Use X (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K083689

---

**Source:** [https://fda.innolitics.com/device/K083689](https://fda.innolitics.com/device/K083689)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
