K083687 · Anhui Kangda Medical Products Co., Ltd. · FPA · Jun 11, 2009 · General Hospital
Device Facts
Record ID
K083687
Device Name
BAIXIN INTRAVASCULAR ADMINISTRATION SET
Applicant
Anhui Kangda Medical Products Co., Ltd.
Product Code
FPA · General Hospital
Decision Date
Jun 11, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the BAIXINTM Intravascular Administration Set is used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein.
Device Story
BAIXIN™ Intravascular Administration Set facilitates fluid delivery from a container to a patient's venous system. The device consists of a closure-piercing device, drip chamber, fluid filter, tubing, flow regulator, and male conical fitting. It operates via gravity feed. The device is sterile, single-use, and intended for clinical use. Healthcare providers use the flow regulator to adjust fluid delivery rates. The device ensures controlled fluid administration, benefiting patients requiring intravenous therapy.
Clinical Evidence
Bench testing only. Testing included sterility, pyrogenicity, haemolysis, toxicity, chemical extractables (metal ions, reducing matter), EO residue, particulate contamination, leakage, tensile strength, flow rate, and mechanical performance of components (closure-piercing device, air-inlet, tubing, filter, drip chamber, flow regulator, and fittings) per ISO 8536-4-2007.
Technological Characteristics
Gravity-fed intravascular administration set. Components include closure-piercing device, drip chamber, fluid filter, tubing (min 1500mm), flow regulator, and male conical fitting (ISO 594-1/2 compliant). Materials evaluated for biocompatibility (toxicity, haemolysis, pyrogenicity). Sterilized for single use. Conforms to ISO 8536-4-2007.
Indications for Use
Indicated for the administration of fluids from a container to a patient's vascular system via a needle or catheter inserted into a vein.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Submission Summary (Full Text)
{0}------------------------------------------------
K083687
## 510(K) SUMMARY
#### Submitted by:
Anhui Kangda Medical Products Co., Ltd
88 Zhenxing Road,
Yangcun Town,
Tianchang City,
Anhui Province
P. R. China
Phone: +86 0550 7093893
+86 0550 7093800 Fax:
Preparation Date: Jun. 10. 2008
### Contact Person/Prepared by:
Official Correspondent: bian wei giang SHANGHAI CARELIFE INTERNATIONAL TRADING CO., LTD. 1707 vingiao bldg 58 jinxin road jinqiao, pudong, shanghai, CHINA 201206 Phone: 86-21-50324423
US Agent: ROBERTJ. EDWARDS BOCA MEDICAL PRODUCTS, INC. 3550 nw 126th ave coral springs, FL 33065 Phone: 954 346 Fax: 954 346 Email: bob@bocapharmacal.com
510(K) Summary of Safety and Effectiveness for the:
Trade/Proprietary Name: BAIXIN™ Intravascular Administration Set
Common Name: INTRAVASCULAR ADMINISTRATION SET
Classification Name: set, administration, intravascular
Class: Il
JUN 1 1 2009
{1}------------------------------------------------
### Procodes: FPA - set, administration, intravascular
## Device Description:
BAIXIN™ Intravascular Administration Set is used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. This device is sterile and for single use only, so the following precautions shall be observed during use:
- (1) Don't use when visual damage on its packaging is detected or the
- protective cap is found not in position.
- (2) Only for single use, discard after use.
- (3) Use before expiration date.
- (4) It shall be used immediately after the individual package is opened.
#### Intended Use:
The intended use of the BAIXINTM Intravascular Administration Set is used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein.
## Statement to conform to ISO 8536-4-2007:
Anhui Kangda Medical Products Co., Ltd. has established that its family of syringes corm to the FDA recognized consensus standard, ISO 8536-4-2007, Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed. Data supporting conformance with the standard is available from Anhui Kangda Medical Products Co., Ltd.
## Summary of Testing:
All materials used in the fabrication of the Anhui Kangda piston syringe and hypodermic needles were evaluated for:
| Test items | Requirements | |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterility | No evidence of growth of BI | |
| Pyrogenicity | Free from pyrogenicity (Bacterial endotoxin test) | |
| Haemolysis | The infusion set shall be free from haemolytic reactions | |
| Toxicity | Free from toxicity | |
| Reducing<br>(oxidizable) matter | The total amount of potassium permanganate solution used<br>[c(KMnO4)=0.002mol/l]shall not exceed 2.0ml; | |
| Metal ions | The extract shall not contain in total more than 1µg/ml of cadmium, chromium, copper, lead and tin. | |
| | Not more than 0.1µg/ml of cadmium, when determined by atomic<br>absorption spectroscopy (AAS) or equivalent method. | |
| Titration acidity or<br>alkalinity | Not more than 1ml of either | |
| | standard volumetric solution | |
| | shall be required for the | |
| | | indicator to change to the<br>colour grey |
| Residue on<br>evaporation | | The total amount of dry<br>residue shall not exceed 5 mg. |
| UV absorption of<br>extract solution | | The extract solution S, shall<br>not show absorption greater<br>than 0.1. |
| EO residue | | Less than 0.5mg/set |
| Particulate<br>contamination | | Contamination index<90 |
| Leakage | | Immerse the infusion set, with one end blocked, in water at 20°C to<br>30°C and apply an internal air pressure of 50kPa above atmospheric<br>pressure for 15s. Examine the infusion set for air leakage. Apply an<br>internal excess pressure of -20kPa at (23±1) °C and (40±1) °C for 15s.<br>inspect whether air enters the infusion set. |
| Tensile strength | | Withstand a static tensile force of not less than 15N for 15s;<br>Dimensions |
| Closure-piercing<br>device | | The closure-piercing device<br>shall be capable of piercing<br>and penetrating the closure<br>without prepiercing. No coring should occur during this<br>procedure. |
| Air-inlet device | | The air-inlet device shall be<br>provided with an air filter to<br>prevent the ingress of<br>microorganisms into the<br>container<br>Air went trough rate≥90%<br>The air-inlet device shall be separate from, or integral with, the<br>closure-piercing device.<br>The air admitted into the<br>container shall not become<br>entrained in the liquid outflow.<br>The flow rate of fluid is not<br>reduced by more than 20% of<br>that from a freely ventilated<br>container. |
| Tubing | | The tubing shall be transparent or sufficiently translucent so that the<br>interface of air and water during the passage of air bubbles can be<br>observed with normal or corrected vision.<br>The tubing length distal to the drip chamber shall be not less than<br>1500mm in length. |
| Fluid filter | | The retention of latex particles<br>on the filter shall be not less<br>than 80% |
| Dimensions of the<br>drip chamber | | There shall be a distance of<br>not less than 40mm between<br>the drip tube and fluid filter<br>The wall of the drip chamber<br>shall not be closer than 5mm<br>to the end of the drip tube<br>The drip tube shall be such<br>that 20 drops of distilled water |
| | or 60 drops of distilled water at:<br>23±2°C and at a flow rate of 60<br>drops/min±10drops/min<br>deliver a volume of<br>$1ml±0.1ml(1g±0.1g)$ | |
| Flow regulator | The flow regulator shall adjust<br>the flow of the infusion solution<br>between zero and maximum.<br>The regulation length should be no less than 30mm. | |
| Flow rate | 20 drops, 10min>1000ml | |
| Injection port | There shall be no leakage of<br>more than one falling drop of<br>water. | |
| Male conical fitting | In accordance with ISO594-1 or ISO594-2 | |
| Protective cap | The protective caps shall<br>maintain the sterility of the<br>closure-piercing device, the<br>male conical fitting and the<br>interior of the infusion set. | |
| Individual package | The infusion set and/or the air-inlet device shall be individually packed<br>so that they remain sterile during storage. The unit container shall be<br>sealed in a tamper-evident manner. | |
| no flattened portions<br>and kinks | The infusion set and/or the air-inlet devices shall be packed and<br>sterilized in such a way that there are no flattened portions or kinks<br>when they are ready for use | |
{2}------------------------------------------------
.
:
: 1
: 上
: 上
.
이 지원 기준
{3}------------------------------------------------
K08368
# Conclusion:
The materials, performance, and operational features of both the submitted device and the predicate device are substantially equivalent and are safe and effective for their intended use. :
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter.
JUN 1 1 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Anhui Kangda Medical Products Company, Limited C/O Mr. Bian Wei Qiang Shanghai Carelife International Trading Company, Limited 1707 Yingiao Building 58 Jinxin Road, Jinqiao Pudong, Shanghai CHINA 201206
K083687 Re:
> Trade/Device Name: BAIXINTM Intravascular Administration Set (BAIXINTM) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: June 3, 2009 Received: June 3, 2009
Dear Mr. Qiang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Page 2- Mr. Oiang
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/Centers_Offices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Anthony, is mistaken for
Susan Dunner, D.D.S., M.S.
Susan Runner, D.D.S., M. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K083687
# Indications for Use
510(k) Number (if known): _ K083687
Device Name: BAIXIN™ Intravascular Administration Set (BAIXIN™)
Indications for Use: Intravascular Administration Set is used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein.
Over-The-Counter Use AND/OR Prescription Use (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
sta all
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: k083687
VIII
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.