PACS SCAN

K083618 · Pacsgear, Inc. · LLZ · Jan 28, 2009 · Radiology

Device Facts

Record IDK083618
Device NamePACS SCAN
ApplicantPacsgear, Inc.
Product CodeLLZ · Radiology
Decision DateJan 28, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The PacsSCAN Medical Imaging Software with QC Module is intended to be used by authorized staff to perform various quality control operations on PACSGEAR imaging studies before they are made available to other locations in the network. These operations include scanning film images, capturing video images, confirming or editing patient demographics, reviewing the history of the study, adding or removing images. combining with another study, renumbering images, editing patient orientation information, and setting or editing routing information.

Device Story

PacsSCAN Medical Imaging Software with QC Module is a software tool for clinical staff to manage and perform quality control on medical imaging studies. Inputs include digitized film images and captured video images. The software allows users to confirm/edit patient demographics, review study history, manipulate images (add/remove, renumber, flip, rotate, pan, zoom), combine studies, and configure routing information. It supports JPEG and JPEG2000 lossy/lossless compression and provides automatic LUT compensation. Used in clinical environments by authorized staff to ensure image quality and data integrity before studies are transmitted across a network. The output is a processed, verified imaging study ready for clinical use, which facilitates accurate clinical decision-making by ensuring clinicians receive correctly oriented and identified images.

Clinical Evidence

No clinical data provided. Software validation was performed to establish the device's ability to meet its intended use and specifications.

Technological Characteristics

Software-based image processing system. Features include film digitization, video capture, JPEG/JPEG2000 compression, automatic LUT compensation, and image processing (segmentation/smoothing). Connectivity includes DICOM Print support. Operates as a networked imaging software module.

Indications for Use

Indicated for use by authorized clinical staff to perform quality control operations on medical imaging studies prior to network distribution. No specific patient population, age, or disease state contraindications are defined.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K083618 ### 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92. ## Submitter Information: PACSGEAR, Inc. 7020 Koll Center Pkwy. Suite 100 Pleasanton, CA 94566 JAN 2 8 2009 ## Date Summary Prepared: July 2, 2008 ### Contact Person: Ian P. Gordon Emergo Group, Inc. Telephone: 512-327-9997 Fax: 512-327-9998 #### Device Name: Trade Name(s): PacsSCAN Medical Imaging Software with QC Module Classification Name: Image Processing System Classification Regulation: 892.2050 Panel: Radiology Product Code: LLZ ### Predicate Device Information: Device Name: RadWorks Medical Imaging Software with Quality Control Module Manufacturer: Applicare Medical Imaging, B.V. Reference: K982862 ### Device Description: The PacsSCAN Medical Imaging Software with OC Module is intended to be used by authorized staff to perform various quality control operations on PacsSCAN imaging studies before they are made available to other locations in the network. These operations include digitizing film images, capturing video images, confirming or editing patient demographics, reviewing the history of the study, adding or removing images, combining with another study, renumbering images, editing patient orientation information, and setting or editing routing information. ## Intended Use: The PacsSCAN Medical Imaging Software with QC Module is intended to be used by authorized staff to perform various quality control operations on PACSGEAR imaging studies before they are made available to other locations in the network. These operations include scanning film images, capturing video images, confirming or editing patient demographics, reviewing the history of the study, adding or removing images. combining with another study, renumbering images, editing patient orientation information, and setting or editing routing information. {2}------------------------------------------------ | Specification | PACSGEAR Medical<br>Imaging Software | RadWorks Medical<br>Imaging Software | |--------------------------------------------------------------|------------------------------------------------|--------------------------------------| | Digitize film | Yes | Yes | | Capture video | Yes | Yes | | Edit patient demographics | Yes | Yes | | Review reports/study<br>history | Yes | Yes | | Add/remove images | Yes | Yes | | Combine studies | Yes | Yes | | Renumber Images | Yes | Yes | | Edit patient orientation<br>information | Yes | Yes | | Set/edit routing information | Yes | Yes | | Image review | Flip/Rotate/Pan/Zoom | Flip/Rotate/Pan/Zoom | | JPEG lossy/lossless<br>compression | Yes and JPEG2000<br>lossy/lossless compression | Yes | | LUT compensation | Yes (automatic) | Yes (manual) | | Image processing (image<br>segmentation, image<br>smoothing) | Yes | Yes | | DICOM Print | Yes | Yes | ## Comparison to Predicate Device: ## Testing and Conclusions: Software validation has established the device's ability to meet its intended use and established specifications. {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 PACSGEAR, Inc. JAN 2 8 2009 c/o Mr. Ian Gordon Senior Vice President Emergo Group, Inc. 1705 S. Capitol of Texas Highway, Suite 500 AUSTIN TX 78746 Re: K083618 Trade/Device Name: PacsSCAN Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 20, 2008 Received: December 9, 2008 ## Dear Mr. Gordon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-011 | |----------------|----------------------------------|---------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-011 | | 21 CFR 892.xxx | (Radiology) | (240) 276-012 | | Other | | (240) 276-010 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.support/ind/subml/ Sincerely yours. Janine M. Morris Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Indications for Use 510(k) Number (if known): 1083618 Device Name: PacsSCAN Medical Imaging Software with QC Module Indications for Use: The PacsSCAN Medical Imaging Software with QC Module is intended to be used by authorized staff to perform various quality control operations on PACSGEAR imaging studies before they are made available to other locations in the network. These operations include scanning film images, capturing video images, confirming or editing patient demographics, reviewing the history of the study, adding or removing images, combining with another study, renumbering images, editing patient orientation information, and setting or editing routing information. Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![image](image) (Division Sign-off) Division of Reproductive, Abdominal a Radiological Devices 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...