PATHOLASE FAMILY OF PINPOINTE AND PINPOINTE FOOTLASER ND: YAG LASERS
K083616 · Patholase, Inc. · GEX · Jun 12, 2009 · General, Plastic Surgery
Device Facts
Record ID
K083616
Device Name
PATHOLASE FAMILY OF PINPOINTE AND PINPOINTE FOOTLASER ND: YAG LASERS
Applicant
Patholase, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 12, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd:YAG Lasers and the delivery accessories that are used with them are intended for use in surgical procedures involving open, laparoscopic and endoscopic ablation, vaporization, incision, and coagulation of soft tissue in the medical specialties of general and cosmetic dentistry, otolaryngology/ENT surgery, and dermatology & plastic surgery including: Oropharangeal / Dental Surgery... General Surgery... Endonasal Surgery... Podiatry... Dermatology and Plastic Surgery...
Device Story
PathoLase PinPointe/FootLaser is a 1064 nm Nd:YAG surgical laser system. Main console includes control panel, laser rod, flashlamp, and 630-680 nm red aiming beam. Delivery accessories include reusable/sterilizable optical fibers, handpieces, and disposable tips for contact or non-contact use. Operated by physicians in clinical settings (OR, clinic). System uses footswitch for activation. Laser energy enables ablation, vaporization, incision, and coagulation of soft tissue. Aiming beam assists in precise placement. Benefits include controlled tissue interaction for various surgical specialties.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
1064 nm Nd:YAG solid-state laser; 630-680 nm diode aiming beam. Components: main console, footswitch, optical fibers (200-1000 um), handpieces, and tips. Reusable components are steam autoclave sterilizable. Connectivity: remote interlock connector. Energy source: electrical power cord.
Indications for Use
Indicated for patients requiring soft tissue ablation, vaporization, incision, excision, or coagulation in dentistry, ENT, general surgery, podiatry, and dermatology. Specific procedures include abscess drainage, lesion/tumor removal, tonsillectomy, matrixectomy, wart removal, tattoo removal, and various laparoscopic/endoscopic surgeries. No specific age or gender contraindications provided.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Modified Lumenis VersaPulse PowerSuite Holmium (Ho:YAG) and Dual Wavelength (Ho:YAG/Nd:YAG) Surgical Lasers and Delivery Devices with Accessories - 2100 nm & 1064 nm (K011703)
{0}------------------------------------------------
KO83616
## JUN 1 2 2009
## Section 5 - 510(k) Summary
#### I. General Information
Submitter:
PathoLase, Inc. 275 Airpark Boulevard Suite 500 Chico, CA 95973 Tel: 230-809-3800
Contact Person:
John Strisower Chairman & Chief Executive, PathoLase, Inc.
Summary Preparation Date: May 6, 2009
#### II. Names
Device Names:
PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd: Y AG Lasers (and delivery device accessories)
Primary Classification Names: . Laser Powered Surgical Instruments (and Accessories)
#### III. Predicate Devices
- Modified Lumenis VersaPulse PowerSuite Holmium (Ho:YAG) and Dual . Wavelength (Ho:YAG) Nd:YAG) Surgical Lasers and Delivery Devices with Accessories - 2100 nm & 1064 nm (K011703)
- Incisive InPulse Dental Laser 1064 nm (K011423). �
#### IV. Product Description
The PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd:YAG Lasers are comprised of the following main components:
- . Main console containing the major electrical components, including:
- A Control/ Display Panel with the:
- Keyswitch (that controls authorized access to the laser system); .
- emergency Laser Stop button; .
- Displays (laser emission indicator, average power, pulse energy, . repetition rate)
- . Standby button (default mode when laser system turned on - places system into the Standby mode preventing laser emission).
- Ready button (places system into the Ready mode allowing laser . emission when the footswitch is depressed and a fiber optic is properly attached);
- A 1064 nm treatment laser (solid state Nd: YAG laser rod) with flashlamp and associated light regulation components and electronics;
- A 630 -680 nm (red) aiming beam diode laser;
- A Delivery device fiber-optic connector port;
- A Remote interlock connector (External door interlock connector);
{1}------------------------------------------------
# KOS 3616
- A Connector ports for the footswitch and power cord;
- A Accessory holder (attached to the rear of the main console);
- . Footswitch:
A
- Medical grade power cord;
- Delivery Devices for Non-Contact and Contact with Intact Skin/Tissue:
- Guide Tip -No Standoff: Reusable, cleanable, tip is provided for noncontact use to direct and control the placement of the laser beam (free beam) at the treatment location. The Guide tip attaches to the end of the handpiece. The optical fiber is threaded through the handpiece and fits securely into the bore of the Guide tip;
- With Standoff: Reusable, cleanable, tip is provided for Guide Tip minimal-contact with intact skin/ tissue to direct and control the placement of the laser beam at the treatment location. The Guide tip attaches to the end of the handpiece. The optical fiber is threaded through the handpiece and fits securely into the bore of the Guide tip;
- Delivery Devices for Contact with Breached Surfaces (Previously Cleared in K011423):
- A Optical Fibers -Reusable, cleanable, sterilizable optical fibers (range of 200 - 1000 um diameter) provided non-sterile, clean and ready for sterilization (steam autoclave).
- Handpieces -Reusable, cleanable, sterilizable handpieces (large and small diameter shafts) provided non-sterile, clean and ready for sterilization (steam autoclave). The optical fiber is threaded through the handpiece and secured and held in place with the handpiece locking cap:
- A Handpiece Tips - Disposable single-use tips are provided in straight and curved configurations and are used to direct and control the placement of the optical fiber tip at the treatment The handpiece tips attach to the end of the location. handpiece. The optical fiber is threaded through both the handpiece and the handpiece tip;
- Accessories: ●
A
- A Safety Glasses
- Tools: .
- A Optical Fiber Striper;
- A Optical Fiber Cleaver (carbide wedge, ceramic, or equivalent scribe for cleaving the optical fibers).
#### V. Rationale for Substantial Equivalence
The PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd:YAG Lasers share the same or similar indications for use, device operation, overall technical and functional capabilities, and therefore is substantially equivalent to the predicate devices.
#### VI. Safety and Effectiveness Information
The review of the indications for use and technical characteristics provided demonstrates that the PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd:YAG Lasers are substantially equivalent to the predicate devices.
{2}------------------------------------------------
083616
#### VII. Conclusion
The PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd:YAG Lasers were found to be substantially equivalent to the predicate devices.
The PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd:YAG Lasers share identical indications for use, similar design features, and functional features with, and thus are substantially equivalent to, the predicate devices.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
PathoLase, Incorporated A L Voss Associates % Ms. Anne Worden 3637 Bernal Avenue Pleasanton, California 94566
JUN 1 2 2003
Re: K083616
Trade/Device Name: PathoLase Family of PinPointe™ and PinPointe™
FootLaser™ Nd:YAG Lasers
Regulation Number: 21 CFR 878.4810
Regulation Name: Lascr Surgical Instrument for Use In General And Plastic Surgery and In Dermatology
Regulatory Class: II Product Code: GEX Dated: May 6, 2009 Received: May 8, 2009
Dear Ms. Worden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration
{4}------------------------------------------------
Page 2-Ms. Worden
and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours
for
Mark N. Melkerson
Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K083616
Device Name: PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd: YAG Lasers
Indications for Use:
The PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd:YAG Lasers and the delivery accessories that are used with them are intended for use in surgical procedures involving open, laparoscopic and endoscopic ablation, vaporization, incision, and coagulation of soft tissue in the medical specialties of general and cosmetic dentistry, otolaryngology/ENT surgery, and dermatology & plastic surgery including:
## Oropharangeal / Dental Surgery
Indicated for:
- � Abscess incision and drainage
- Aphthous ulcers treatment .
- Biopsies, excisional and incisional .
- Crown lengthening .
- Exposure of unerupted / partially erupted teeth
- Fibroma removal
- Frenectomy
- Frenotomy
- Gingival incision and excision
- Gingivectomy
- Gingivoplasty
- Hemostatis
- Implant recovery
- Lesion (tumor) removal
- Leukoplakia
- Operculectomy
- Oral papillectomy
- Pulpotomy
Prescription Use (Part 21 CFR 801 Subpart D) ** Page 1 of 4 * AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nulke Odes Formen
(Division S (Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number
Page 1 of 4
Premarket Notification, 510(k) Submission for: PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd: YAG Lasers Section 4: Indications for Use
{6}------------------------------------------------
510(k) Number (if known): K083616
Device Name: PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd: YAG Lasers
Indications for Use - Continued:
## Oropharangeal / Dental Surgery - Continued
- . Pulpotomy as adjunct to root canal therapy
- Removal of filling material such as gutta percha or resin as adjunct treatment during root . canal re-treatment
- Selective ablation of enamel (first degree) caries removal ●
- Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal . pocket) to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility
- Tissue retraction for impressions .
- . Vestibuloplasty
### General Surgery
Indicated for:
.
- Open, laparoscopic, and endoscopic general surgery (ablation, vaporization, incision,
- excision, and-coagulation-of-soft-tissue) including: -
- Cholecystectomy
- Lymphadenectomy
- Mastectomy
- Partial nephrectomy
- Hepatectomy
- Pilonidal cystectomy
- Pancreatectomy
- Resection of lipoma
- Splenectomy
- Pelvic adhesiolysis
- Hemorrhoidectomy
Removal of lesions
Prescription Use V
AND/OR (Part 21 CFR 801 Subpart D)
*** Page 2 of 4 *** Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil K. Oslen Forman
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K083616
Page 2 of 4
Premarket Notification, 510(k) Submission for: PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd: YAG Lasers
Section 4: Indications for Use
{7}------------------------------------------------
510{k) Number (if known): K083616
Device Name: PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd:YAG Lasers
Indications for Use - Continued:
#### General Surgery - Continued
- Thyroidectomy -
- Removal of polyps
- Parathyroidectomy
- Removal of tumors
- Herniorrhaphy
- Tumor biopsy
- Tonsillectomy
- Debridement of decubitus ulcers
- Appendectomy
#### Endonasal Surgery
Endonasal surgery (ablation, vaporization, incision, and coagulation of soft tissue) including:
- Lesions or tumors of the oral, nasal, glossal, pharyngeal & laryngeal tissues ●
- Tonsillectomy .
- Adenoidectomy . ●
#### Podiatry
Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue) including:
- . Matrixectomy
- Periungual and subungual warts .
- Plantar warts
- Radical nail excision
- Neuromas
## ** Page 3 of 4 **
> Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Nul R.P. O'sada Firman
(Division Sign-C Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K083616
Page 3 of 4
Premarket Notification, 510(k) Submission for: PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd: YAG Lasers Section 4: Indications for Use
{8}------------------------------------------------
510(k) Number (if known): _ K083616
Device Name: PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd:YAG Lasers
Indications for Use - Continued:
## Dermatology and Plastic Surgery
Dermatology and plastic surgery (ablation, vaporization, excision, and coagulation of soft tissue) including:
- Lesions of skin and subcutaneous tissue �
- Telangiectasia
- Port wine lesions
- Spider veins
- Hemangiomas
- Plantar warts
- Periungual and subungual warts
- Removal of tattoos
- Debridement of decubitus ulcer
- Treatment of keloids
*** Page 4 of 4 ***
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Net Rf Ogden for mktm
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number
Page 4 of 4
Premarket Notification, 510(k) Submission for: PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd: YAG Lasers Section 4: Indications for Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.