REEKROSS FAMILY OR OTW PTA CATHETERS

K083596 · ClearStream Technologies , Ltd. · DQY · Mar 2, 2009 · Cardiovascular

Device Facts

Record IDK083596
Device NameREEKROSS FAMILY OR OTW PTA CATHETERS
ApplicantClearStream Technologies , Ltd.
Product CodeDQY · Cardiovascular
Decision DateMar 2, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

Balloon dilatation of the femoral, popliteal and infra popliteal arteries. These catheters are not designed to be used in the coronary arteries.

Device Story

ReeKross OTW PTA catheter is an over-the-wire balloon dilation catheter for angioplasty; used for femoral, popliteal, and infra-popliteal arteries. Device consists of stainless steel guidewire shaft with laser-cut spiral at distal end; spiral sealed with polymer bodybond material to isolate inflation lumen from guidewire lumen. Copolymer inflation lumen extends from proximal hub to balloon shoulder. Balloon material is Nylon/Pebax blend. Platinum radiopaque bands swaged onto bodybond section for fluoroscopic visualization. Operated by physicians in clinical setting; device facilitates vessel dilation via inflation/deflation of balloon using contrast solution. Benefits patient by restoring blood flow in peripheral arteries.

Clinical Evidence

Bench testing only.

Technological Characteristics

Stainless steel guidewire shaft with laser-cut spiral; polymer bodybond sealant; copolymer inflation lumen; Nylon/Pebax blend balloon; platinum radiopaque bands. Over-the-wire configuration. Sterilization method not specified.

Indications for Use

Indicated for balloon dilatation of femoral, popliteal, and infra-popliteal arteries. Contraindicated for use in coronary arteries.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Premarket Notification - ClearStream Technologies PTA Catheters MAR = 2 2009 # 510(k) SUMMARY K083596 11:13:13 The summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. Submitter Information: ClearStream Technologies Ltd, Moyne Upper, Enniscorthy, Co Wexford, Ireland. Phone: 00 353 53 9237111 Fax: 00 353 53 9237100 Date Summary Prepared: March 19, 2008 Contact Person: Fiona Ní Mhullain Device Name: Trade Name(s): ReeKross™, ReeKross™14 and ReeKross™18 over the wire Peripheral Transluminal Angioplasty (PTA) Catheters Classification Name: Percutaneous Catheter Classification Regulation: 21 CFR 870.1250 Panel: Cardiovascular Product Code: LIT ## Predicate Device Information: Device Name: Savvy Long Manufacturer: ClearStream Technologies Ltd Reference: K072947 ## Device Description: # CONFIDEN {1}------------------------------------------------ #### Premarket Notification - ClearStream Technologies PTA Catheters The ReeKross family of over the wire PTA catheters are a balloon dilation catheter for angioplasty. They consist of a stainless steel guidewire shaft that extends from the proximal hub to the distal tip of the catheter. This stainless steel shaft has a spiral cut positioned at its distal end. The spiral cut is laser cut and passivated following completion. A polymer bodybond material is used to seal the spiral cut and ensure that the contrast solution remains in the inflation lumen only and does not leak into the guidewire lumen. This bodybond material is fused onto the outside diameter of the stainless steel shaft. A copolymer inflation lumen extends from the proximal hub of the catheter to the proximal shoulder of the balloon. The lumen of the shaft is used for the purpose of inflating and deflating the balloon. This outer is thermally fused to the balloon's proximal shoulder. The balloon material is a Nylon/Pebax blend. Low Profile Platinum Radiopaque bands are used to locate the balloon under fluoroscopy. These bands are swaged onto the bodybond material sealing the spiral cut section of the guidewire lumen. ## Intended Use: Balloon dilatation of the femoral, popliteal and infra popliteal arteries. These catheters are not designed to be used in the coronary arteries. #### Comparison to Predicate Device: This device is equivalent to the predicate device in intended use, design characteristics and general physical characteristics. #### Testing and Conclusion: Performance testing and compliance with recognized standards have established substantial equivalence with regard to safety and effectiveness. {2}------------------------------------------------ Page 3 - Mr. Harry van Vugt (Please include 510(k) number here: K083596) | HFZ # | Last Name | Date | HFZ # | Last Name | Date | HFZ # | Last Name | Date | |-------|-----------|---------|-------|-----------|------|-------|-----------|------| | 450 | Courtney | 22209 | | | | | | | | 450 | Comage | 2/2/01 | | | | | | | | 2-450 | Vchuller | 7/27/09 | | | | | | | HFZ-401 DMC cc: HFZ-404 510(k) Staff HFZ-450 Division D.O. Prepared by:TDCourtney:myb:02/27/09 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services, USA. The seal features a stylized eagle with outstretched wings, symbolizing the department's mission to protect the health of all Americans. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ClearStream Technologies. Inc. c/o Mr. Robert A. van Boxtel KEMA Quality B.V. Utrechtseweg 310 Arnhem Netherlands NL-6812 AR MAR - 2 2009 Re: K083596 Trade Name: Reekross OTW PTA Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: February 19, 2009 Received: February 20, 2009 Dear Mr. van Boxtel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Or to and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your deview can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in {4}------------------------------------------------ Page 2 - Mr. Robert van Boxtel the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. una R. Vulin Image /page/4/Picture/6 description: The image shows a handwritten signature or symbol. It appears to be a stylized combination of letters, possibly an 'S' and an 'm' or 'n', connected in a cursive style. The stroke begins with a curve, forms a loop, and ends with a downward stroke. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Premarket Notification - ClearStream Technologies PTA Catheters Indications for Use **510(k) Number (if known):** Device Name: ClearStream Technologies PTA Catheters #### Indications for Use: Balloon dilatation of the femoral, popliteal and infra-popliteal arteries. These catheters are not designed to be used in the coronary arteries. Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Duna R. de Lunes Division Sign-Off) Division of Cardiovascular Devices 10(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...