ANTI-WRINKLE LIGHT, MODEL 1302 AND PAIN THERAPY LIGHT, MODEL 1301

K083580 · Led Intellectual Properties, LLC · ILY · Jun 22, 2009 · Physical Medicine

Device Facts

Record IDK083580
Device NameANTI-WRINKLE LIGHT, MODEL 1302 AND PAIN THERAPY LIGHT, MODEL 1301
ApplicantLed Intellectual Properties, LLC
Product CodeILY · Physical Medicine
Decision DateJun 22, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Pain Therapy Light, model: PTL, is an Over-The-Counter hand held device intended to emit energy in the visible and near infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain or muscle spasm, the temporary increase in local and blood circulation, and the temporary relaxation of muscles.

Device Story

Pain Therapy Light Model 1301 is a handheld LED device emitting infrared light (630, 660, 855, and 940nm). Operated by the user in a home or clinical setting, the device uses an LED array to provide topical heating to the skin. The device features a simple integrated chip, manual on/off switch, and a 9V AC-to-DC power supply. By elevating local tissue temperature, the device aims to provide temporary relief of minor muscle and joint pain, stiffness, and muscle spasms, while promoting local circulation and muscle relaxation. The user applies the wand directly to the treatment area as needed.

Clinical Evidence

No clinical data. The submission relies on the device's substantial equivalence to legally marketed predicate devices and a history of safe and effective use in the United States for over 7 years prior to submission.

Technological Characteristics

Handheld wand configuration; LED array emitting 630, 660, 855, and 940nm; 9V AC-to-DC power supply; simple integrated chip with manual on/off switch and electronic resistor. Non-sterile, non-implantable. Complies with Section 878 performance standards for infrared lamps; OSHA/NRTL listed.

Indications for Use

Indicated for OTC use by adults for temporary relief of minor muscle/joint pain, stiffness, minor arthritis pain, muscle spasms, temporary increase in local blood circulation, and temporary muscle relaxation via topical heating.

Regulatory Classification

Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K083580 # 510(k) Application Summary K083580 - Revision IV, Submission Date: December 19th, 2010 JUN 2 2 2009 1. Submitter Information: AEGIS Regulatory, Inc. - Robert T. Wagner 1131 Anthem View Lane Knoxville. TN 37922 Tel.: 865-982-5552 Email: bob@bobwagner.net > For Manufacturer: LED Intellectual Properties, LLC Attn: Steve Marquese 9832 Chesterfield Circle Santa Ana. CA 92705 Tel .: (949) 394-2427 #### General Information 2. - 2.1 Classification Name: ILY - Infrared Lamp, therapeutic heating (Dual Use) - 2.2 Common/Usual Name: LightStim - 2.3 Proprietary Names: Pain Therapy Light, Model 1301 - 2.4 Classification: Class II - 2.5 Classification Number: 1.) 878.5500 #### 3. Description: Pain Therapy Light Model 1301 is a hand-held device consisting of low intensity light emitting diodes (LED's) that provide infra-red light to the skin. Physical components include an LED array emitting 630, 660, 855 and 940nm respectively in a handheld "wand" configuration containing a simple "in/out" integrated chip with external, manual on/off switch, a single electronic resistor, and a removable 9 volt AC-to-DC power supply. Treatment times are as required by the user, and specified by the User's Manuals for product code ILY indications. #### 4. Intended Use: #### LightStim Pain Relief Light (handheld wand) Product Code - ILY Indications for Use: The Pain Therapy Light, model: PTL, is an Over-The-Counter hand held device intended to emit energy in the visible and near infrared spectrum to provide topical heating for the purpose of elevating {1}------------------------------------------------ tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain or muscle spasm, the temporary increase in local and blood circulation, and the temporary relaxation of muscles. ## 5. Substantial Equivalence to Predicate Device(s): Pain Therapy Light: This device is substantially equivalent to the following predicate devices, which are currently in safe and effective commerce: - 1. Quantum Warp 10 Light Delivery System K032229 - 2. Pain Relief Bio-Beam 660+940 & Med Light 1630+2630 K042813 - 3. Terra Quant MQ2000 V.5 & handy Rx MQ2007 K061614 - 4. Acubeam, Super Nova, Dio K022888 ### 6. Performance Standards: These devices have been tested under and are in compliance with performance standards that have been established for such devices under Section 878 of the Federal Food, Drug, and Cosmetics Act. All electrical and radiological products made by the applicant have been OSHA/NRTL listed, and have received constituent marks. ### 7. Labeling: As attached and in User's Manuals (Attached Version 2.0) - Numerous technical warnings and advisements to insure proper use and maintenance. #### 8. Statement of Safety and Effectiveness: These products have been used safely and effectively in clinical testing applications within the United States, with no negative reports or claims submitted to FDA / CDRH, for over 7 years prior to the date of this submission. ### 9. Over-The-Counter Variance Request: As identical predicates of this device have been in safe and effective applications by layperson users, Over-The-Counter Variance is requested. ## 10. Design and Use of the Device(s): | Is the device intended for prescription use? | NO | |--------------------------------------------------------------------|-----| | Is the device intended for over-the-counter use? | YES | | Are its components derived from a tissue or other biologic source? | NO | | Is the device provided sterile? | NO | | Is the device intended for single use? | NO | {2}------------------------------------------------ | Is the device a reprocessed single use device? | NO | |--------------------------------------------------|----| | Does the device contain a drug? | NO | | Does the device contain a biologic? | NO | | Does the device use software? | NO | | Does the submission include clinical information | NO | | Is the device implanted? | NO | · {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service . Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines representing wings and a stylized body. JUN 22 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 LED Intellectual Properties, LLC % AEGIS Regulatory, Inc. Mr. Roberty Wagner 1131 Anthem View Lane Knoxville, Tennessee 37922 Re: K083580 Trade/Device Name: LED Intellectual Properties, LLC - Pain Therapy Light model: PTL Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY Dated: June 5, 2009 Received; June 17, 2009 Dear Mr. Wagner: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Mr. Roberty Wagner forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N. Millman Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K083580 p. 1 of 1 Indications for Use 510(k) Number: K083580 Device Name: LED Intellectual Properties, LLC - Pain Therapy Light model: PTL Indications For Use: The Pain Therapy Light, model: PTL, is an Over-The-Counter hand held device intended to emit energy in the visible and near infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain or muscle spasm, the temporary increase in local and blood circulation, and the temporary relaxation of muscles. Prescription Use AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Nolbert Ogden forman Division Sign-Off Concurrence of CDRH, Office of Device Evaluation (0) Division of Surgical, Orthopedic, and Restorative Devices Page 1 of 510(k) Number.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...