K083578 · Noras Mri Products GmbH · MOS · Jan 16, 2009 · Radiology
Device Facts
Record ID
K083578
Device Name
NORAS CPC MULTIFUNCTIONAL COIL
Applicant
Noras Mri Products GmbH
Product Code
MOS · Radiology
Decision Date
Jan 16, 2009
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The intended use of the Noras CPC is, in conjunction with a magnetic resonance scanner, body-imaging for diagnostic with a MRI system. It is indicated for use as a diagnostic imaging device transverse, sagittal, coronal and oblique images of the internal structures of special regions of the human body. When interpreted by a trained physician, these unages provide information that can be useful in determining diagnosis. The application area shouldn't be bigger than 20 cm because of the depth of penetration of 10 cm per coll. The outcome of this is diagnostic images of toes, feet, ankle, knee, finger, hand, wrist, elbow, shoulder, carotid artery, inner ear, baby heart and other parts of the body with sımılar structure can be taken.
Device Story
Multifunctional MRI receive-only coil; 2x4-channel phased array design; spherical configuration. Used with 1.5T and 3T MRI systems; compatible with Siemens MRI tables. Pre-tuned fixed load; no user tuning required. Captures MR signals from specific body regions (extremities, carotid, inner ear, heart); produces transverse, sagittal, coronal, and oblique images. Operated by MRI technicians/radiologists in clinical imaging settings. Output interpreted by physicians to assist in diagnosis. Benefits include improved access to regions of interest and high-quality diagnostic imaging.
Clinical Evidence
Bench testing only. Performance and safety parameters verified against NEMA standards and IEC 60601-2-33. Testing conducted on Siemens Avanto 1.5T system to validate safety and performance for the 1.5T series.
Technological Characteristics
2x4-channel phased array receive-only coil; spherical geometry; 1.5T and 3T MRI compatible. Pre-tuned fixed load. Connectivity via system-specific plugs. Conforms to IEC 60601-1 (3rd ed) and IEC 60601-2-33. No active software algorithm; hardware-based signal reception.
Indications for Use
Indicated for diagnostic imaging of internal structures of human body regions (toes, feet, ankle, knee, finger, hand, wrist, elbow, shoulder, carotid artery, inner ear, baby heart) in patients requiring MRI. Application area limited to 20 cm diameter due to 10 cm penetration depth per coil. For use by trained physicians.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Model CG-WHC18-H150-AP Wrist Hand Coil 1.5T (K071882)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "KO83578" in a handwritten-like font. Below this, the text "Section 9 510(k) summary" is printed in a standard font. The text appears to be part of a document or label.
## CPC Multipurpose Coil
Date of Summary Preparation: November 20, 2008
JAN 1 6 2009
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.
1. General Information
### Importer/Distributor
Name and Address Noras MRI products GmbH Leibnizstrasse 4 97204 Höchberg Germany
Establishment Registration Number 3004929307
### Manufacturing Site
Name and Address
Noras MRI products GmbH
Leıbnızstrasse 4
97204 Höchberg
Germany
Establishment Registration Number 3004929307
Owner/Operator 9071737
{1}------------------------------------------------
# 2. Contact Person
Zahed Sedighiani Noras MRI products GmbH Leibnizstr 4 97204 Höchberg, Germany Telephone: (+49) 931/29927-17 Fax: (+49) 931/29927-20 Email Address: zahed sedightanı@noras de
## 3. Device Name and Classification
| Trade Name: | Noras CPC - Multifunctional Coil |
|-----------------------|--------------------------------------|
| Common Name: | CPC |
| Classification Name: | Magnetic Resonance Diagnostic Device |
| Classification Panel: | Radiology |
| CFR Number: | 21 CFR § 892 1000 |
| Device Class: | II |
| Product Code: | 90MOS and LNH |
## 4. Device Description
1
The CPC is a multifuntional coil for 1 5T and 3T MRI-systems, with 2x4-channel phased array receive only colls The coils are pre-tuned in the fixed load and no further tuning or matching is required for the user
The coll is divided into two spherical parts Each half contains a 4 channel receive only array The coils were used together with a parallel imaging procedure.
Each coll has its own plug The plug is specific to the MRI-system and can't be mixed up with a not designated MRI-system
The base plate is flat and is suitable on each Siemens MRI-system table
{2}------------------------------------------------
## 5. Intended Use
The intended use of the Noras CPC is, in conjunction with a magnetic resonance scanner, body-imaging for diagnostic with a MRI system
It is indicated for use as a diagnostic imaging device transverse, sagittal, coronal and oblique images of the internal structures of special regions of the human body When interpreted by a trained physician, these unages provide information that can be useful in determining diagnosis
The application area shouldn't be bigger than 20 cm because of the depth of penetration of 10 cm per coll The outcome of this is diagnostic images of toes, feet, ankle, knee, finger, hand, wrist, elbow, shoulder, carotid artery, inner ear, baby heart and other parts of the body with sımılar structure can be taken
# 6. Substantial Equivalence
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Noras believes that, within the meaning of the Safe Medical Devices Act of 1990, the CPC multifunctional coil is substantially equivalent to the following coils.
| Coil Name | Premarket Notification | Clearance Date |
|-------------------------------------------------|------------------------|-------------------|
| Machnet Carotids Coil<br>Array Assembly | K 012491 | October 24, 2001 |
| Siemens Flex Loop<br>Coil Set 3T | K 063313 | November 17, 2006 |
| Invivo Precision Eight<br>Wrist Array Coil 1 5T | K 993477 | January 11, 1999 |
| Phased Array Shoulder Coil | K 963356 | November 22, 1996 |
| Model 455GE, Phased Array<br>Wrist Coil | K 010074 | April 3, 1996 |
| Model CG-WHC18-H150-<br>AP Wrist Hand Coil 1 5T | K 071882 | July 18, 2007 |
{3}------------------------------------------------
7. Summary of Technological Characteristics of the Principal Device as Compared with the Predicate Devices
The MRI multifunctional CPC coil has been designed to examine the internal structures of special regions of the human body To get MR images it is mandatory to use the loop colls of the MRT System manufacturer The approach of this new product was, to give the physician a more comfortable access to the region of interest in the internal structures of special regions of the human body to produce transverse, sagittal, coronal and oblique images, which provide information that can be useful in determining diagnosis
# 8. General Safety and Effectiveness Concerns
The complete system CPC will conform the harmonized standard of IEC 60601-1 30 edition. Medical electrical equipment - General requirements for basic safety and essential performance
The CPC coil will conform to the FDA recognized NEMA standards for the measurement of performance and safety parameters and the IEC standards for safety issues with the Magnetic Resonance Imaging Devices, IEC 60601-2-33 2003 This will assure that the performance of this device can be considered safe and effective when used with the currently available MAGNETOM systems
All tests are performed on the Siemens Avanto This is the most critical engine for safety and performance of the Siemens 1 5T series Passing these tests gives the Siemens approval for the complete 1 5T series
# 9. Conclusion as to Substantial Equivalence
Noras believes that, within the definition of the Safe Medical Devices Act of 1990, the Noras multifunctional CPC colls are substantially equivalent to the predicate devices listed above
\$\frac{1}{2}\$
Hubert Noras
President
1 -
1
November 27, 2008
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo with a stylized eagle design. The eagle is depicted with three curved lines forming its body and wings. The logo is surrounded by text that reads 'U.S. Department of Health & Human Services, USA'. The text is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 6 2009
Mr Zahed Sedighianı Regulatory Affairs Manager Noras MRI Products GmbH Leibnizstr 4, Hochberg, 97204 GERMANY
Re K083578 Trade/Device Name CPC Regulation Number 21 CFR 892 1000 Regulation Name Magnetic resonance diagnostic device Regulatory Class II Product Code MOS Dated November 27, 2008 Received December 3, 2008
Dear Mr Sedighiani
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food., Orug and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA) You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual regustration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 In addition, FDA may publish further announcements concerning your device in the Federal Register
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies You must comply with all the Act's requirements, including but not limited to. registration and listing (21 CFR Part 807), labeling (21 CFR Part 801, good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter
| 21 CFR 876 xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884 xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892 xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807 97) For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveyllance at 240-276-3474 For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464 You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http //www fda gov/cdrh/industry suppot/index html
Sincerely yours,
Janina M. Norris
anıne M Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (1f known) _ KOB 35 28
Device Name CPC
Indications for Use
The intended use of the Noras CPC is, in conjunction with a Magnetic Resonance Scanner, Body-imaging for diagnostic with a MRI system
It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of special regions of the human body When ınterpreted by a trained physician, these images provide information that can be use in determining diagnosis
The application area shouldn't be bigger than 20 cm because of the depth of 10 om per coll The outcome of this is diagnostic images of toes, feet, ankle, knee, finger, hand, wrist, elbow, shoulder, carotid artery, inner ear and other parts of the body with sumilar structure can be taken
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Arnie Potte Wher
Division of Reproductive, Abdominal a Radiological Device 510(k) Number
(Posted November 13, 2003)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.