K083577 · Stellate Systems · GWQ · Aug 29, 2011 · Neurology
Device Facts
Record ID
K083577
Device Name
HARMONIE SYSTEM
Applicant
Stellate Systems
Product Code
GWQ · Neurology
Decision Date
Aug 29, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Harmonie System is indicated for the acquisition, display, review, analysis and archiving of physiological signals including patient video obtained during routine EEG examinations, Long-Term Monitoring in Epilepsy (LTM) and Sleep Studies (polysomnography or PSG) in patients of all ages. The seizure detection component of the software is intended for post-hoc review for marking sections of the EEG that may correspond to electrographic seizures in adults only. The Burst-Suppression component is intended to be used for the quantification and recognition (mark) of sections of the EEG and PSG that may contain patterns of interest. The seizure detection component and Burst-Suppression Module are for use in adults only. The Harmonie software is not intended to replace conventional devices or methods used for sleep monitoring in critical care or intra-operative settings. In no way are any of the system functions represented as being in and of themselves diagnostic. The system requires competent user input, and its output must be reviewed and interpreted by a physician or other trained health care professional who will exercise professional judgment in using this information.
Device Story
Data acquisition system for EEG/PSG signals; records up to 128 channels; sample rates 200-2000Hz. Amplifies, filters, digitizes signals; optional synchronized MPEG4 digital video recording. Used in clinics, hospitals, or ICU settings by physicians/technicians. Data stored on workstation for local/remote review. Software modules include EEG analysis, PSG scoring (sleep staging, event marking), and automated detection of seizure/burst-suppression patterns. Output displayed on monitors for physician interpretation; assists in clinical decision-making for epilepsy and sleep disorders. Benefits include streamlined long-term monitoring review and standardized data archiving.
Clinical Evidence
Bench testing only. Verification and validation activities performed to establish performance, reliability, and functionality against published safety standards and predicate specifications. No clinical trial data provided.
Technological Characteristics
Data acquisition system; 20-bit amplifiers (DUO, eAmp, EEG64); 128-channel capacity; programmable filters (0.1-500Hz); sampling rates up to 2000Hz. Components include cart-mounted workstations, panel PCs, and optional photic stimulators. Connectivity: local or remote station archiving. Software: Harmonie EEG/PSG modules. Sterilization: N/A (non-invasive).
Indications for Use
Indicated for acquisition, display, review, analysis, and archiving of EEG, PSG, and synchronized video signals in patients of all ages. Seizure detection and Burst-Suppression components indicated for adult use only. Not for use in critical care or intra-operative sleep monitoring.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for "xltek, a division of natus". The logo consists of a stylized graphic to the left of the text "xltek". The graphic appears to be a stylized representation of a person with arms raised, with a star above their head. A dashed line is below the graphic and the text. Below the dashed line is the text "a division of natus" in a smaller font size.
# 510(K) Summary
AUG 2 9 2011
Natus Medical Incorporated DBA Excel-Tech Ltd. 2568 Bristol Circle. Oakville, Ontario, L6H 5S1
Date: Aug 11, 2011
Name of the device - Harmonie System
Common Name - Electroencephalograph and Polysomnograph
Classification name - EEG/PSG Monitoring (21 CFR 882.1400 Product Code GWQ)
The Harmonie System is substantially equivalent to K010728 - Harmonie Schwarzer EEG System and K053606 - BE Plus / AURA-LTM64 Amplifier.
#### Device Description:
The Harmonie System is a data acquisition device that can record up to 128 channels of EEG and PSG signals at sample rates ranging from 200Hz to 2000Hz per channel according to the specifications of each model. EEG and PSG signals are amplified, filtered and digitized. A Photic Stimulator is optionally provided with the system. The system can also concurrently record synchronized digital patient video, through an optional MPEG4 encoder card and the DV software module. The digital data (EEG and digital video) is stored on the computer workstation's hard disk for review and analysis, either during or after a recording session.
Following is the functional description of various Harmonie models:
- 1. Harmonie System with DUO Recorder
This is a basic EEG recording system, typically used for routine EEG examination. The DUO recorder is a 20-bit, 44-channel amplifier with user programmable filters, automatic anti-aliasing and sampling rates of up to 2,000 Hz. All components are mounted on a robust enclosed cart, which can accommodate a Photic Stimulator and various models of CCTV cameras, as accessories. The system includes Harmonie EEG software that allows acquisition, display, review, analysis and archiving of EEG data locally or at a remote station.
- 2. Harmonie-S System with DUO Recorder
The Harmonie-S system uses same DUO recorder described above but uses PSG software to record sleep studies. It includes optionally digital video recording. The amplifier and video camera are placed in the room. The computer and monitor are placed in an observation room where the PSG technologist works. The software includes the Harmonie EEG and the PSG modules. The
{1}------------------------------------------------
PSG module is designed specifically for the visualization and scoring of variables commonly used in sleep studies, such as the EEG, EMG, EKG, oxygen saturation, body position , eye movements and respiration signals. The system allows performing the main functions of a sleep study: sleep staging, manual event marking, displaying graphs of sleep stages and events, and various sleep report generation.
#### MESA II System with DUO Recorder 3.
MESA-II is a portable system suitable for EEG examinations, also using the DUO EEG amplifier. It offers the full functionality of the standard Harmonie system, but with a laptop computer and other components mounted on a cart with a smaller foot-print, which enables EEG examinations to be performed in space-limited environments. The cart can accommodate a Photic Stimulator and various models of CCTV cameras, as accessories. The system includes Harmonie EEG software that allows acquisition, display, review, analysis and archiving of EEG data locally or at a remote station.
#### Vita ICU System with DUO Recorder 4.
The Vita ICU system is an EEG recording system designed to facilitate EEG recordings in the ICU, although it can also be used as a routine EEG system. It includes optionally digital video recording. The Vita ICU system is composed of a panel PC (computer and monitor in the same physical entity), including a touch-screen monitor, and is mounted on small footprint cart. It includes the Harmonie EEG software as well as software specific for this application (the Vita ICU software). The Vita ICU software includes a touch-screen interface to facilitate user-interaction with the device.
- 5. Harmonie-E System with eAmp Recorder
The Harmonie-E system with eAmp recorder is an EEG recording system designed for LTM particularly in the context of the evaluation of patients with epilepsy. The eAmp is an IP-enabled amplifier with the capability to acquire data for up to 128 channels (by combining two 64-channel units). It includes programmable filters from 0.1 to 500Hz and sampling rates up to 2,000Hz that allow measurement of higher frequency EEG. The system includes digital video recording. The amplifier and video camera are placed in the room and the recording computer and monitor and placed in a control room. The system includes Harmonie EEG software that allows acquisition, display, review, analysis and archiving of EEG data locally or at a remote station. The Harmonie-E module for the automatic marking of EEG events of interest, i.e. spike patterns and seizure patterns. This post-hock review marking facilitates the review of the long EEG recordings.
- Harmonie-E System with EEG64 Recorder 6.
The Harmonie-E system with Schwarzer EEG64 recorder is an EEG recording system designed for LTM particularly in the context of the evaluation of patients with epilepsy. The Schwarzer EEG64 amplifier is equivalent to eAmp with slightly different filtering characteristics than the eAMP amplifier. The system includes Harmonie EEG software that allows acquisition, display, review, analysis and archiving of EEG data locally or at a remote station.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for "xltek", which appears to be a company name. The logo consists of a stylized graphic to the left of the text "xltek". Below the text is the phrase "a division of natus", which is smaller in size. The logo is simple and uses a clean font.
### Intended Use:
The Harmonie System is indicated for the acquisition, display, review, analysis and archiving of physiological signals including patient video obtained during routine EEG examinations, Long-Term Monitoring in Epilepsy (LTM) and Sleep Studies (polysomnography or PSG) in patients of all ages.
The seizure detection component of the software is intended for post-hoc review for marking sections of the EEG that may correspond to electrographic seizures in adults only. The Burst-Suppression component is intended to be used for the quantification and recognition (mark) of sections of the EEG and PSG that may contain patterns of interest. The seizure detection component and Burst-Suppression Module are for use in adults only.
The Harmonie software is not intended to replace conventional devices or methods used for sleep monitoring in critical care or intra-operative settings.
In no way are any of the system functions represented as being in and of themselves diagnostic. The system requires competent user input, and its output must be reviewed and interpreted by a physician or other trained health care professional who will exercise professional judgment in using this information.
#### Performance Data
Bench testing and verification and validation activities were performed to establish the performance, reliability and the functionality of the Harmonie system. Pass/Fail criteria were established based on the published specifications of applicable safety standards, predicate and the current device.
#### Conclusion
The conclusion drawn from the test results and review of scientific literatures support is that the device and the indications for use are substantially equivalent to the predicate device and do not raise new questions of safety and effectiveness.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Excel-Tech, Ltd Division, Natus Medical Incorporated c/o Ms. Goldy Singh Director of Quality & Regulatory 2568 Bristol Circle, Oakville, Ontario Canada L6H 5S1
Re: K083577
Trade/Device Name: Harmonie System Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: GWQ Dated: April 8, 2011 Received: April 8, 2011
AUG 2 9 2011
Dear Ms. Singh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Of ice are of your may (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Goldy Singh
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely
f
Der
2.n
Mark N. Melkerso Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K083577
Device Name: Harmonie System
Indications For Use:
The Harmonie System is indicated for the acquisition, display, review, analysis and archiving of physiological signals including patient video obtained during routine EEG examinations, Long-Term Monitoring in Epilepsy (LTM) and Sleep Studies (polysomnography or PSG) in patients of all ages.
The seizure detection component of the software is intended for post-hoc review for marking sections of the EEG that may correspond to electrographic seizures in adults only. The Burst-Suppression component is intended to be used for the quantification and recognition (mark) of sections of the EEG and PSG that may contain patterns of interest. The seizure detection component and Burst-Suppression Module are for use in adults only.
The Harmonie software is not intended to replace conventional devices or methods used for sleep monitoring in critical care or intra-operative settings.
In no way are any of the system functions represented as being in and of themselves diagnostic. The system requires competent user input, and its output must be reviewed and interpreted by a physician or other trained health care professional who will exercise professional judgment in using this information.
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number K083577
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.