← Product Code [EOB](/productcode/EOB) · K083553

# NASO-LARYNGO-PHARYNGOSCOPE (K083553)

_Schoelly Imaging, Inc. · EOB · Jan 16, 2009 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K083553

## Device Facts

- **Applicant:** Schoelly Imaging, Inc.
- **Product Code:** [EOB](/productcode/EOB.md)
- **Decision Date:** Jan 16, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 874.4760
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat

## Indications for Use

The Schoelly Naso-Laryngo-Pharyngoscope is intended to be used by qualified surgeons and physicians to visualize and observe the pharynx and the larynx.

## Device Story

Flexible, steerable fiberoptic endoscope for diagnostic visualization of pharynx and larynx. Device captures optical images via fiber bundles; transmits to eyepiece or camera for physician observation. Operated by surgeons/physicians in clinical settings. Provides direct visualization to assist in diagnostic assessment of throat/larynx anatomy. Non-sterile, requires EtO sterilization before use. Benefits include improved diagnostic access to upper airway structures.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance validated through comparison of optical and mechanical specifications against predicate device.

## Technological Characteristics

Flexible fiberoptic endoscope; 3.2mm distal tip diameter; 300mm working length; 140-degree angulation; 70-degree field of view; 7,000 fiber count. Compliant with IEC 60601-1 and IEC 60601-2-18. Supplied non-sterile; validated for EtO sterilization (6% EtO/94% CO2, 131°F, 120 min exposure). Compatible with chemical disinfectants including Gigasept FF, Lysetol FF, Helipur HplusN, and Cidex.

## Regulatory Identification

A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.

## Predicate Devices

- Schoelly Naso-Laryngo-Pharyngoscope ([K991560](/device/K991560.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Special 510(k) Summary

|         | A) Submitted by Schoelly Imaging, Inc<br>100 Hartwell St<br>West Boylston, MA 01583<br>Registration 8043903 | JAN 1 6 2009 |
|---------|-------------------------------------------------------------------------------------------------------------|--------------|
| Contact | MEDIcept<br>200 Homer Ave<br>Ashland, MA 01721<br>F. David Rothkopf                                         |              |

Date of Application November 26, 2008

B) Device Name Naso-Laryngo-Pharyngoscope

508-231-8842 x20 508-231-8861 Fax

Common Name Nasopharyngoscope

Device Class 21 CFR 874 4760 CLASS II

- Product Code EOB
C) Predicate Naso-Laryngo-Pharyngoscope (K991560)

D) Device Description

The Schoelly Naso-Laryngo-Pharyngoscope is a flexible and steerable fiberoptic endoscope for diagnostic purposes

- E) Intended Use
The Schoelly Naso-Laryngo-Pharyngoscope is intended to be used by qualified surgeons and physicians to visualize and observe the pharynx and the larynx

- F) Comparison to Predicate Device(s)
The Schoelly Naso-Laryngoscope has the same intended use, target population, clinical setting, and technology as its predicate device Schoelly Naso-Laryngo-Pharyngoscone (K991560) There is no change to in compatibility with the environment or other devices, go differences in electrical, mechanical or chemical safety There are slight changes in the optical and mechanical specifications and patient contact materials, all of which have non-significant impact on device performance All changes have been made under design control, and validation was performed and conformance assured

Schoelly Imaging, Inc Special 510(k) Naso-Laryngo-Pharyngoscope

{1}------------------------------------------------

| Features                                                                                                                                                                                                                        | Schoelly Naso-Laryngoscope<br>(the Special 510K application<br>herein)                                                                                                                                                                                                                                                        | Schoelly-Naso-<br>Laryngoscope<br>(K991560)                                                                                                                                                                                                                                                                      |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                                                                                                                                                                                                                    | Visualize and observe the<br>pharynx and the larynx                                                                                                                                                                                                                                                                           | Visualize and observe<br>the pharynx and the<br>larynx                                                                                                                                                                                                                                                           |
| Design                                                                                                                                                                                                                          | Flexible and steerable fiberoptic<br>endoscope for diagnostic<br>purposes                                                                                                                                                                                                                                                     | Flexible and steerable<br>fiberoptic endoscope<br>for diagnostic<br>purposes                                                                                                                                                                                                                                     |
| Parts number                                                                                                                                                                                                                    | 31 1001s                                                                                                                                                                                                                                                                                                                      | 31 0001s                                                                                                                                                                                                                                                                                                         |
| Specifications                                                                                                                                                                                                                  | 31 0001s/31 1002s                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                  |
| Tip tubing diameter                                                                                                                                                                                                             | $3.2 mm$                                                                                                                                                                                                                                                                                                                      | $3.8 mm$                                                                                                                                                                                                                                                                                                         |
| Working Length                                                                                                                                                                                                                  | $300 mm$                                                                                                                                                                                                                                                                                                                      | $300 mm$                                                                                                                                                                                                                                                                                                         |
| Angulation Up                                                                                                                                                                                                                   | $140^0$                                                                                                                                                                                                                                                                                                                       | $150^0$                                                                                                                                                                                                                                                                                                          |
| Angulation Down                                                                                                                                                                                                                 | $140^0$                                                                                                                                                                                                                                                                                                                       | $150^0$                                                                                                                                                                                                                                                                                                          |
| Direction of view                                                                                                                                                                                                               | $0^0$                                                                                                                                                                                                                                                                                                                         | $0^0$                                                                                                                                                                                                                                                                                                            |
| Field of View                                                                                                                                                                                                                   | $70^0$                                                                                                                                                                                                                                                                                                                        | $85^0$                                                                                                                                                                                                                                                                                                           |
| Length of distal tip<br>deflection                                                                                                                                                                                              | $20 mm$                                                                                                                                                                                                                                                                                                                       | $25 mm$                                                                                                                                                                                                                                                                                                          |
| Depth of field                                                                                                                                                                                                                  | $6 - 1000 mm$                                                                                                                                                                                                                                                                                                                 | $3mm - 50mm$                                                                                                                                                                                                                                                                                                     |
| Total Fibers (pixels)                                                                                                                                                                                                           | $7,000$                                                                                                                                                                                                                                                                                                                       | $10,000$                                                                                                                                                                                                                                                                                                         |
| Fiber size                                                                                                                                                                                                                      | $8 3 \mu m$                                                                                                                                                                                                                                                                                                                   | $8 3 \mu m$                                                                                                                                                                                                                                                                                                      |
| Magnification³                                                                                                                                                                                                                  | $7X$                                                                                                                                                                                                                                                                                                                          | $2 5 X$                                                                                                                                                                                                                                                                                                          |
| Energy used                                                                                                                                                                                                                     | IEC601-1 and IEC 60601-2-18 2000 compliant                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                  |
| Labeling<br>Sterilization*/Cleaning                                                                                                                                                                                             | The Instruction Manual states<br>Sterilization                                                                                                                                                                                                                                                                                | Users Manual states                                                                                                                                                                                                                                                                                              |
| *device is supplied as<br>non-sterile                                                                                                                                                                                           | Section 7 5 Flexible Schoelly<br>endoscopes with air-exhaust valve can<br>also be gas sterilized using ethylene<br>oxide "<br>Specifications<br>Gas mixture 6% EtO, 94% CO₂<br>Temperature 131°F+/-5°F, 55°C +/-2°C<br>Relative air humidity 40-90%<br>Pressure (overpressure), 1 7 bar<br>(170kpa)<br>Exposure time 120 mins | "The Schoelly Naso-<br>Laryngo-Pharyngoscope is<br>provided non-sterile, and<br>must be sterilized using<br>EtO sterilization prior to<br>use<br>"<br>Specifications<br>EO concentration 600 +/-<br>25 mg/L<br>Temperature 54+/- 2°C<br>(58°C) max)<br>Relative humidity 70% +/-<br>5%<br>Pressure 14 +/- 1 PSIG |
| Cleaning                                                                                                                                                                                                                        | Gas exposure time 4 hours                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                  |
| Section 7 2 states                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                  |
| "Place the endoscope and the<br>unscrewed parts in a suitable<br>disinfection solution (can also double up<br>as a cleaning solution) The disinfection<br>solution used should be permitted by the<br>manufacturer for such use |                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                  |
| Material compatibility releases exist for<br>the following disinfectants<br>*Gigasept FF, Schulke & Mayr GmbH<br>*Lysetol FF<br>*Helipur HplusN, B Braun Medical AG<br>*Cidex, Johnson & Johnson             "                  | "To dissolve various<br>impurities, use mild<br>cleaning substances"                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                  |

{2}------------------------------------------------

## Naso-Laryngo-Pharyngoscope Special 510(k) Premarket-Notification Submission

" Manufactured in two resolutions, 12,000 and 16,000 pixels

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 1 6 2009

Schoelly Imaging, Inc c/o David McNally 100 Hartwell Street West Boylston, MA 01583

Re K083553

Trade/Device Name Naso-Pharyngoscope Regulation Number 21 CFR 874 4760 Regulation Name Nasopharyngoscope (flexible or rigid) and accessories Regulatory Class Class, II Product Code EOB Dated November 20, 2008 Received December 18, 2008

Dear Mr McNally

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA) You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, insting of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807), labeling (21 CFR Part 801), good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050

ﻧﺴﻴ

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807 97) For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Drysson of Surveillance Systems at 240-276-3464 You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely vours.

M.B. Egelston, mD

Malvina B Eydelman, M D Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Ko83553

## Indications for Use

510(k) Number (if known) Ko 8 3 53 3

Device Name Naso-Laryngo-Pharyngoscope

Indications for Use

The Schoelly Naso-Laryngo-Pharyngoscope is intended to be used by qualified surgeons and physicians to visualize and observe the pharynx and the larynx.

Prescription Use X 21CFR 801, Subpart D OR Over-the-Counter Use_21CFR 801 109

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NECESSARY)

> ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.. . ..... . ..... .. .. .. .. ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... .. 16 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

Concurrence of CDRH, Office of Device InVitro Diagnostic Device Evaluation and Safety (OIVD)

Karen H. Baker

Division Sign Off Office of Device InVitro Diagnostic Device Evaluation and Safety

510 (K) K083553

Schoelly Imaging, Inc Special 510(k) Naso-Laryngo-Pharyngoscope

---

**Source:** [https://fda.innolitics.com/device/K083553](https://fda.innolitics.com/device/K083553)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
