K083544 · Kat Implants, LLC · NHA · Feb 2, 2009 · Dental
Device Facts
Record ID
K083544
Device Name
KAT IMPLANT SYSTEM
Applicant
Kat Implants, LLC
Product Code
NHA · Dental
Decision Date
Feb 2, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
KAT Implant System is intended to restore edentulous areas of maxilla and mandible, to provide support for removable dentures, fixed bridges, or to be used as a single tooth replacement. Single or splinted implants can be immediately loaded if good primary stability and appropriate occlusal loading are achieved. The implant can be placed in extraction sites or healed alveolar ridges. Immediate loading may not be appropriate in Type IV bone due to difficulty in achieving primary stability. KAT 3.0mm implant is indicated for use in maxillary lateral or mandibular lateral and central incisors in single or multiple units to support prosthesis, such as artificial teeth. The implant can be placed in extraction sites or healed alveolar ridges and can be immediately loaded when good primary stability is achieved and the functional load is appropriate. Immediate loading may not be appropriate in Type IV bone due to difficulty in achieving primary stability. KAT 2.5mm is a self-tapping titanium alloy threaded screw indicated for transitional and long-term intra-bony applications, such as providing support for transitional or long term crowns, bridges and dentures. KAT 2.5mm may also be used for inter-radicular transitional application.
Device Story
KAT Implant System comprises dental implants, abutments, and associated surgical/restorative instruments. Implants feature external V-shaped threads for bone insertion; 3.5mm-5.0mm diameters include horizontal fins. All implants utilize a 3.1mm diameter 1.5-degree taper abutment receiving post; components connect via locking taper. Used by dentists/oral surgeons in clinical settings to restore edentulous sites. Implants placed in extraction sites or healed alveolar ridges; support removable dentures, fixed bridges, or single crowns. Immediate loading possible given sufficient primary stability and appropriate occlusal loading. System provides mechanical foundation for prosthetic teeth; improves patient masticatory function and aesthetics.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Materials: Titanium alloy Ti-6Al-4V ELI (ASTM F136). Surfaces: Machined, aluminum oxide grit blasted, and nitric acid passivated. Design: Threaded screw with 1.5-degree taper locking connection. Diameters: 2.5mm to 5.0mm. Lengths: 6mm to 14mm. Mechanical fixation via external threads and horizontal fins.
Indications for Use
Indicated for patients requiring restoration of edentulous areas in maxilla or mandible, including single tooth replacement, removable dentures, or fixed bridges. KAT 3.0mm indicated for maxillary lateral or mandibular lateral and central incisors. KAT 2.5mm indicated for transitional and long-term intra-bony applications. Contraindicated for immediate loading in Type IV bone due to primary stability concerns.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
LaminOss® Osteocompressive Dental System (K982925)
{0}------------------------------------------------
K083544
### 5. 510(k) Summary
## -2 2009
5
3
This summary of safety and effectiveness information is being provided in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is:
#### Submitter's Information:
Vitali Bondar, D.D.S. Owner and CEO KAT Implants LLC 15 Ryc Street, Suite 115 Portsmouth, NH 03801 Phone: (603) 427-0084 Fax: (603) 427-0045
## Date the Summary was Prepared: November 20, 2008
#### Device name:
- Common Name: Dental Implant and Abutment .
- 하 Trade Name: KAT Implant System
- Classification name: Endosseous Dental Implant (21 CFR 872.3640, # Product code DZE) and Endosseous Dental Implant Abutment (21 CFR 872.3630, Product Code NHA)
Device classification: Class II
#### Intended use
KAT Implant System is intended to restore edentulous areas of maxilla and mandible, to provide support for removable dentures, fixed bridges, or to be used as a single tooth replacement. Single or splinted implants can be immediately loaded if good primary stability and appropriate occlusal loading are achieved. The implants can be placed in extraction sites or healed alveolar ridges. Immediate loading may not be appropriate in Type IV bone due to difficulty in achieving primary stability.
KAT 3.0mm implant is indicated for use in maxillary lateral or mandibular lateral and central incisors in single or multiple units to support prosthesis, such as artifical teeth. The implant can be placed in extraction sites or healed alveolar ridges and can be immediately loaded when good primary stability is achieved and the functional load is appropriate. Immediate loading may not be appropriate in Type IV bone due to difficulty in achieving primary stability.
{1}------------------------------------------------
K 083544
ு
KAT 2.5mm is a self-tapping titanium alloy threaded screw indicated for transitional and long-term intra-bony applications, such as providing support for transitional and crowns, bridges and dentures. KAT 2.5mm may also be used for inter-rollul br transitional application.
# The legally marketed devices to which the equivalence is claimed [807.92(a)(3)]:
Predicate device: LaminOss® Osteocompressive Dental System Applicant: Impladent Inc. (USA) 510(k) number: K982925
Predicate device: Bicon Dental Implant 4.5 x 6.5mm Applicant: Bicon Inc. (USA) 510(k) number: K050712
Predicate device: IMTEC Sedax MDI 1.8mm Applicant: IMTEC Corporation (USA) 510(k) number; K031106
Predicate device: OsseoSpeed 4.0S-6mm Applicant: Astra Tech Inc. (USA) 510(k) number: K063779
Predicate device: OsseoSpeed Narrow Applicant: Astra Tech Inc. (USA) 510(k) number: K080396
> KAT Dental Implant System has the following similarities to the predicate devices:
- Has the same intended use; .
- Uses the same operating principles; .
- Incorporates the same basic design; 1
- Incorporates the same materials; and
- Is packaged and sterilized using the same or equivalent materials and processes.
{2}------------------------------------------------
K083544
3 of 3
#### Description of Device
KAT implant system consists of dental implants, implant abutments, healing abutments, temporary abutments, screw retained framework abutment, spacer, cylinder, healing collar and the instruments for placement and restoration of the implants. KAT implants are supplied in 2.5, 3.0, 3.5, 4.3 and 5.0mm diameters. KAT 2.5 and 3.0mm implants are supplied in 10, 12 and 14mm length. IAAT 3.5mm is supplied in 8, 10, 12 and 14mm length. KAT 4.3 and 5.0mm inplants are supplied in 6, 8, 10, 12 and 14mm length. External V-shaped thread is utilized to screw the implants into the bone. Horizontal fins are placed in between the thread and the abutment receiving portion of 3.5, 4.3 and 5.0mm implace All implants have 3.1mm diameter 1.5 degree taper abutment receiving post. KAT abutments are available in 4.2, 4.7, 5.5 and 6.5mm diameter. KAT abutments can be attached to any KAT implant via locking taper connection.
#### Material composition
Titanium alloy Ti-6Al-4V ELI is used to manufacture implants, abutments and other components of the KAT implant system conforming to ASTM F136. Surfaces of the implants and abutments include: machined surfaces and aluminum oxide grit blasted and nitric acid passivated surfaces.
## Summary of the technological characteristics:
The fundamental scientific technology of the device is identical or very similar to the referenced predicate devices. The components of KAT dental implant system are substantially equivalent to the listed predicate devices because of the similarities in design, material and intended use.
#### Conclusions Drawn:
KAT Dental Implant System has the same intended use as, and technological characteristics similar to the legally marketed predicate devices. Any differences in the technological characteristics did not raise new issues of safety and effectiveness. Therefore the new implant system is substantially equivalent to the predicate devices.
7
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three bars extending from its back, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## FEB -2 2009
Vitali Bondar, D.D.S. Owner and Chief Executive Officer KAT Implants L.L.C. 15 Rye Street, Suite 115 Portsmouth, New Hampshire 03801
Re: K083544
> Trade/Device Name: KAT Implant System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: January 27, 2009 Received: January 27, 2009
Dear Dr. Bondar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898, In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Dr. Bondar
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Anthony D. Watson for
Cindy V. Reichard, M.D.
Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
K083544
# 4. Indications for Use Statement
510(k) Number: KO835 YU
Device Name: KAT Implant System
#### Indications for Use: -
KAT Implant System is intended to restore edentulous areas of maxilla and mandible, to provide support for removable dentures, fixed bridges, or to be used as a single tooth replacement. Single or splinted implants can be immediately loaded if good primary stability and appropriate occlusal loading are achieved. The implant can be placed in extraction sites or healed alveolar ridges. Immediate loading may not be appropriate in Type IV bone due to difficulty in achieving primary stability.
KAT 3.0mm implant is indicated for use in maxillary lateral or mandibular lateral and central incisors in single or multiple units to support prosthesis, such as artificial teeth. The implant can be placed in extraction sites or healed alveolar ridges and can be immediately loaded when good primary stability is achieved and the functional load is appropriate. Immediate loading may not be appropriate in Type IV bone due to difficulty in achieving primary stability.
KAT 2.5mm is a self-tapping titanium alloy threaded screw indicated for transitional and long-lerm intra-bony applications, such as providing support for transitional or long term trowns, bridges and dentures. KAT 2.5mm may also be used for inter-radicular transitional application.
{6}------------------------------------------------
207:2
K083544
X Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runon
(Division Sign-Off)
appenenen Sign-Olf)
Division of Anesthesiology, General Hospital Infection Control Ancsthestology, General
Infection Control, Dental Devices
Page of
510(k) Number: K083544
## FEB - 2 2009
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.