← Product Code [JPA](/productcode/JPA) · K083518

# ACL ACUSTAR, HEMOSIL ACUSTAR D-DIMER, HEMOSIL ACUSTAR D-DIMER CONTROLS (K083518)

_Instrumentation Laboratory CO · JPA · Mar 13, 2009 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K083518

## Device Facts

- **Applicant:** Instrumentation Laboratory CO
- **Product Code:** [JPA](/productcode/JPA.md)
- **Decision Date:** Mar 13, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5425
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K083518 · Mar 13, 2009 | ACL ACUSTAR, HEMOSIL ACUSTAR D-DIMER, HEMOSIL ACUSTAR D-DIMER CONTROLS | Instrumentation Laboratory CO | Retrospective clinical plasma samples from patients with suspected VTE | The study evaluated the clinical performance (sensitivity, specificity, and negative predictive value) of the HemosIL AcuStar D-Dimer assay using 344 frozen clinical samples from patients presenting with suspected venous thromboembolism. | D-Dimer; VTE; Clinical performance; Retrospective samples |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective clinical performance study | Patients admitted to an emergency unit with suspected PE or DVT; Sample Size: 344 | Standard objective tests | Sensitivity, Specificity, NPV |

## Indications for Use

ACL AcuStar: Automated immunoassay analyzer designed specifically for in vitro diagnostic use in a clinical laboratory. The assay analysis is based on chemiluminescent technology. The system provides results for both direct measurements and calculated parameters. HemosIL AcuStar D-Dimer: Fully automated chemiluminescent immunoassay for the quantitative determination of D-Dimer in human citrated plasma on the ACL AcuStar as an aid in the diagnosis of venous thromboembolism (VTE) [deep vein thrombosis (DVT) and pulmonary embolism (PE)]. HemosIL AcuStar D-Dimer Controls: For the quality control of D-Dimer assay performed on the ACL AcuStar.

## Device Story

ACL AcuStar is an automated, software-driven, bench-top chemiluminescent immunoassay analyzer for clinical laboratories. It processes human citrated plasma samples using magnetic particles coated with anti-D-Dimer monoclonal antibodies and isoluminol-labeled tracer antibodies. The system performs a two-step incubation and wash cycle, measuring relative light units (RLUs) via an internal luminometer. RLUs are proportional to D-Dimer concentration. The device includes a control computer for user interface, data management, and automated sample handling/dilution. Results assist clinicians in diagnosing venous thromboembolism (VTE). The system uses barcode readers for specimen identification and supports automated on-board dilutions to expand the reportable range.

## Clinical Evidence

Clinical study evaluated 344 frozen citrated plasma samples from emergency unit patients suspected of VTE. 97 were VTE positive, 247 negative. Using a 500 ng/mL cut-off, the assay demonstrated 100% clinical sensitivity (95% CI 96.3-100.0%), 55.5% specificity (95% CI 49.0-61.8%), and 100% negative predictive value (95% CI 97.3-100.0%). Method comparison studies against VIDAS and ACL TOP predicates showed good correlation (r=0.89 to 0.98).

## Technological Characteristics

Bench-top automated chemiluminescent immunoassay analyzer. Uses magnetic particle separation and isoluminol-labeled monoclonal antibody tracers. Dimensions/form factor: bench-top. Connectivity: barcode reader for specimen/rack identification. Software: automated, 4-parameter logistic curve (4PLC) data reduction. Sterilization: N/A (reagents). Materials: mouse monoclonal antibodies, magnetic particles, isoluminol tracer.

## Regulatory Identification

A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

## Special Controls

*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

## Predicate Devices

- VIDAS Instrument ([K891385](/device/K891385.md))
- VIDAS D-Dimer Exclusion Assay ([K040882](/device/K040882.md))
- ACL TOP ([K073377](/device/K073377.md))
- HemosIL D-Dimer HS ([K070927](/device/K070927.md))
- HemosIL D-Dimer Controls ([K972696](/device/K972696.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
K083518

B. Purpose for Submission:
Clearance of a new instrument, assay, and controls

C. Measurand:
D-Dimer

D. Type of Test:
Quantitative, Chemiluminescent Immunoassay

E. Applicant:
Instrumentation Laboratories

F. Proprietary and Established Names:
ACL AcuStar
HemosIL AcuStar D-Dimer
HemosIL AcuStar D-dimer Controls

G. Regulatory Information:
1. Regulation section:
21 CFR 864.7320
2. Classification:
Class II
3. Product code:
JPA (Instrument)

{1}

GGN (Controls)

DAP (Assay)

4. Panel:
81 Hematology

H. Intended Use:

1. Intended use(s):
ACL AcuStar is an automated immunoassay analyzer designed specifically for in vitro diagnostic use in a clinical laboratory. The assay analysis is based on chemiluminescent technology. The system provides results for both direct measurements and calculated parameters.

HemosIL AcuStar D-dimer is a fully automated chemiluminescent immunoassay for the quantitative determination of D-Dimer in human citrated plasma on the ACL AcuStar as an aid in the diagnosis of venous thromboembolism (VTE).

HemosIL AcuStar D-Dimer controls are for the quality control of D-dimer assay performed on the ACL AcuStar.

2. Indication(s) for use:

3. Special conditions for use statement(s):

4. Special instrument requirements:
ACL AcuStar

I. Device Description:

The ACL AcuStar is automated, software driven, bench-top analyzer consisting of (1) a main unit that provides sample and reagent handling, sample testing and result measurement hardware, and (2) a control computer that provides the user interface, data and instrument management.

The AcuStar D-Dimer Assay is a 100 determination kit consisting of 1 vial of a magnetic particle suspension coated with mouse monoclonal antibody directed against D-Dimer, 1 vial of tracer consisting of anti-XDP mouse monoclonal antibody labeled with isoluminol,

2

{2}

diluent, buffers, perservative and stabilizers.

The AcuStar D-Dimer Controls are tri-level controls of partially purified D-Dimer obtained by digestion of Factor XIIIa cross-linked human fibrin with human plasmin.

## J. Substantial Equivalence Information:

1. Predicate device name(s):
- BioMerieux VIDAS Instrument
- BioMerieux VIDAS D-Dimer Exclusion Assay
- Instrumentation Laboratories ACL TOP
- Instrumentation Laboratories HemosIL D-Dimer HS
- Instrumentation laboratories HemosIL D-Dimer Controls

2. Predicate 510(k) number(s):
- K891385
- K040882
- K073377
- K070927
- K972696

3. Comparison with predicate:
- Predicate 1 – ACL TOP (K073377) with HemosIL D-Dimer HS (K070927) and HemosIL D-Dimer Controls (K972696)
- Predicate 2 – VIDAS Instrument (K8913850 with VIDAS Exclusion Assay (K040882)

Similarities
|  Item | Device | Predicate 1 | Predicate 2  |
| --- | --- | --- | --- |
|  Intended Use | Quantitative determination of D-Dimer | same | same  |
|  Sample | Citrate Plasma | Same | same  |

{3}

4

Differences

|  Item | Device | Predicate 1 | Predicate 2  |
| --- | --- | --- | --- |
|  Assay principle | Chemiluminescent | Immunoturbidometric | EIA with fluorescent detection  |
|  Detection Limit | 6.51 ng/ml | 21 ng/ml | 45 ng/mL  |
|  Cut-off | 500 ng/ml | 230 ng/ml | 500 ng/ml  |

K. Standard/Guidance Document Referenced (if applicable):

EP05-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline, 2nd Ed., 08/20/2004

C28-A2: How to Define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline, 2nd Ed., 06/01/2000

EP07-A2: Interference Testing in Clinical Chemistry; Approved Guideline, 2nd Ed., 11/23/2005

EP09-A2: Method Comparison and Bias Estimation, 2nd Ed., 09/20/2002

L. Test Principle:

The ACL AcuStar D-Dimer Chemiluminescent Immunoassay is a 2-step assay. In the first step, assay buffer, magnetic particles coated with a monoclonal antibody specific to D-Dimer is mixed with a sample. Following an incubation, during which any D-Dimer in the sample is captured by the antibody coated particles. In part two, after a washing step which removes any unbound materials, a tracer, consisting of a monoclonal antibody specific to D-Dimer conjugated with isoluminol is added to the sample cuvette. During a short incubation, the tracer binds to any D-Dimer captured by the antibody coated particle. After a washing step to remove any unbound tracer, the sample cuvette is sent to the luminometer, and the chemiluminescent reaction is measured as relative light units (RLU's). The RLU's are directly proportional to the D-Dimer concentration in the sample.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

2 replicates per run, 2 runs per day, for 20 days (N=80 for each sample level) on 2

{4}

different ACL AcuStar instruments.

|  ACL AcuStar No. 1  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Control Level | N | Mean ng/mL | CV% Within Run | CV% Between Run | CV% Total  |
|  Low D-dimer Control | 80 | 232 | 5.5 | 1.4 | 6.0  |
|  High D-dimer Control | 80 | 875 | 2.2 | 3.4 | 4.6  |
|  Very High D-Dimer Control | 80 | 8918 | 2.2 | 5.5 | 5.9  |
|  D-Dimer Calibrator Level 1 | 80 | 362 | 2.9 | 2.9 | 4.9  |
|  ACL AcuStar No. 2  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Control Level | N | Mean ng/mL | CV% Within Run | CV% Between Run | CV% Total  |
|  Low D-dimer Control | 80 | 234 | 4.0 | 2.8 | 6.8  |
|  High D-dimer Control | 80 | 841 | 2.3 | 2.9 | 4.9  |
|  Very High D-Dimer Control | 80 | 8467 | 2.5 | 3.7 | 5.6  |
|  D-Dimer Calibrator Level 1 | 80 | 358 | 2.7 | 3.1 | 5.4  |

# b. Linearity/assay reportable range:

Dilutions of a high concentration D-Dimer plasma sample ( $\sim 95000\mathrm{ng/mL}$ ) were prepared in AcuStar D-Dimer Assay Sample Diluent. Each level was tested in quadruplicate using two different lots of AcuStar D-dimer reagents. Results demonstrated linearity of  $54.3 - 74000\mathrm{ng/mL}$  with Auto Rerun Off. When the Auto Rerun is activated, the ACL AcuStar makes an on-board dilution and automatically corrects the final result for the dilution factor, which expands the test range to  $54.3 - 1110000\mathrm{ng/mL}$ . No inhibitory prozone was detected on the highest concentrations tested.

# c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The submission contained Cartridge on-board stability, Calibrator and Control on-board stability, and partial on-board Control stability (to simulate normal usage) data that supported the cartridge, control and calibrator stability claims.

{5}

d. Detection limit:

20 replicates of House Standards Level 1 and 2 were run on an ACL AcuStar using 2 different lots of AcuStar D-Dimer reagents. Detection Limit is calculated as:

$$
\left[ S _ {1} \right] \times \frac {3 \times S D _ {0}}{R L U _ {1} - R L U _ {0}}
$$

The submission contained data to support the detection limit claim of 6.51 ng/mL

e. Analytical specificity:

The highest concentration of hemoglobin (500 mg/dL), bilirubin (18 mg/dL), triglycerides (1250 mg/dL), heparin (Low molecular Weight and Unfractionated) (2 IU/mL, rheumatoid factor (450 IU/mL), and human anti-mouse antibody (1 µg/mL) were each spiked into two samples (low and high D-Dimer concentrations) and the results compared to the unspiked sample results. To demonstrate the lack of interference of fibrinogen to the AcuStar D-Dimer assay, a positive D-Dimer sample was diluted to multiple levels with sample diluent and D-Dimer depleted plasma, and the results compared. Data demonstrated no significant interference by;

- Hemoglobin up to 500 mg/dL
- Bilirubin up to 18 mg/dL
- Triglycerides up to 1250 mg/dL
- Heparin up to 2 IU/mL
- Rheumatoid factor up to 450 IU/mL)
- Human anti-mouse antibody up to 1 µg/mL

f. Assay cut-off:

Assay cut-off was validated through ROC analysis using 150 frozen samples from non-consecutive outpatients suspected of VTE. The samples were selected to include 30% positive VTE samples and were tested with the HemosIL AcuStar D-Dimer Assay on the ACL AcuStar. A 100% Sensitivity, 35.2% Specificity, and 100% Negative Predicative Value was obtained based on a 500 ng/mL cut-off.

2. Comparison studies:

a. Method comparison with predicate device:

{6}

7

In-House Study

150 frozen samples from non-consecutive outpatients suspected of VTE and 30 samples diagnosed with Disseminated Intravascular Coagulation (DIC) were tested in singlicate with HemosIL AcuStar D-Dimer on the ACL AcuStar versus the VIDAS D-Dimer Exclusion Assay. (y = 1.16 x -247, r=0.90)

Field Study #1

102 patient samples, 14 of which from patients diagnosed with VTE, were tested in singlicate with HemosIL AcuStar D-Dimer on the ACL AcuStar versus the HemosIL D-Dimer HS on an ACL TOP. Sample results ranged from 6 – 12629 ng/mL y=2.54 +383.39, r=0.89.

Field Study #2

100 patient samples, 24 of which obtained from patients diagnosed with VTE, were tested in single CATE with HemosIL AcuStar D-Dimer on the ACL AcuStar versus the BioMerieux VIDAS D-Dimer Exclusion Assay. Sample results ranged from 77 – 63131 ng/mL. y = 1.09 -51.74, r= 0.98

b. Matrix comparison:

50 fresh plasma samples were obtained from an emergency room and immediately analyzed in duplicate with the HemosIL AcuStar D-dimer reagents on an ACL AcuStar instrument and then split into two aliquots. One aliquot was stored at 2-8° C, and the other at -70° C. After one week, each aliquot was tested in duplicate, and compared to the fresh sample results. Results demonstrated good correlation ;

Fresh vs 2-8 ° C : y = 1.04 x -323, r=0.99

Fresh vs -70° C : y = 1.03x -290, r=0.99

3. Clinical studies:

344 frozen citrated plasma samples were collected from patients admitted to an emergency unit with suspected VTE. 97 were confirmed as VTE positive by standard objective tests and the remaining 247 were confirmed as negative

a. Clinical Sensitivity:

Based on the cut-off of 500 ng/mL, 100% (95% CI 96.3 to 100.0%) sensitivity and 100% (95% CI 97.3 to 100.0%) negative predictive value was obtained.

b. Clinical specificity:

{7}

Based on the cut-off of 500 ng/mL, the study yielded a specificity of 55.5% (95% CI 49.0 to 61.8%)

c. Other clinical supportive data (when a. and b. are not applicable):

4. Clinical cut-off:

5. Expected values/Reference range:

189 citrated plasma samples obtained from healthy blood bank donors were tested in singlicate on an ACL AcuStar using one lot of HemosIL AcuStar D-Dimer reagents. 9 samples were deemed far outliers and removed from the analysis. The D-Dimer concentration distribution was not normal; therefore, non parametric statistics were applied, resulting in the following:

|  Non-Parametric Limits | 95% Limit | 90% CI  |
| --- | --- | --- |
|  Lower | 65.5 | 53.1 -73.1  |
|  Upper | 630 | 564 - 802  |

The device's package insert recommends that each laboratory establish its own normal range.

N. Instrument Name:

ACL AcuStar

O. System Descriptions:

1. Modes of Operation:

Automatic

2. Software:

FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes ☐ X or No ☐

3. Specimen Identification:

{8}

Barcode reader. for both the specimen and rack. If there is no sample in a rack position, the barcode reader reads the barcode on the back of the rack position that indicates an empty location.

4. Specimen Sampling and Handling:

Specimens are sampled from an open tube and automatically diluted

5. Calibration:

The assay kit contains lot dependent, bi-level lyophilized calibrators. The assay uses a 4 Parameter Logistic Curve (4PLC) fit data reduction method to generate a Master Curve which is stored in the instrument through the cartridge barcode. With the measurement of calibrators, the predefined Master Curve is transformed to a new, instrument specific 4PLC Working Curve.

6. Quality Control:

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Q. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

{9}

10

---

**Source:** [https://fda.innolitics.com/device/K083518](https://fda.innolitics.com/device/K083518)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
