← Product Code [NHA](/productcode/NHA) · K083474

# ZIMMER ANGLED CONTOUR ZIRCONIA ABUTMENT (K083474)

_Zimmer Dental, Inc. · NHA · Dec 23, 2008 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K083474

## Device Facts

- **Applicant:** Zimmer Dental, Inc.
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Dec 23, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The Zimmer® Angled Contour Zirconia Abutment is used as a terminal or intermediate abutment for a cemented prosthesis. The abutment can be used for a single or multiple-unit restoration in anterior and pre-molar regions.

## Device Story

Zimmer Angled Contour Zirconia Abutment is a dental implant component used as a terminal or intermediate abutment for cemented prostheses. It features a pre-defined offset margin—lower on the buccal aspect, higher on the lingual aspect—to reduce the need for chairside preparation. The abutment cone is fixed at a 17-degree angle. It is intended for use in anterior and pre-molar regions for single or multiple-unit restorations. The device is placed by a dental professional during restorative procedures to support a dental crown or bridge.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Zirconia abutment with a 17-degree angled cone and pre-defined offset margin. Designed for endosseous dental implant systems. No specific ASTM standards or software/connectivity features described.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- Zimmer Patient Specific Abutment, Internal Hex, Ceramic

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

<083474

| / | 07 |
|---|----|
| i |    |

Image /page/0/Picture/2 description: The image shows the logo for Zimmer Dental. The logo consists of a circle with a stylized letter Z inside it. Below the circle, the word "zimmer" is written in a serif font, with a line underneath it. Below the line, the word "dental" is written in a smaller, sans-serif font.

1900 Aston Avenue Carlsbad, CA 92008 760.929.4300 (ph) 760.431.7811 (fax)

Zimmer Denta

| 510k No.: |      |
|-----------|------|
| Page No.: | 12.9 |

DEC 2 3 2008

## Special 510(k): Device Modification PRE-MARKET NOTIFICATION 510(k) 510(k) SUMMARY (21CFR807.92(a))

1. Submitter's Information:

| Name:          | Zimmer Dental Inc. |
|----------------|--------------------|
| Address:       | 1900 Aston Ave.    |
|                | Carlsbad, CA 92008 |
| Phone:         | 760-929-4300       |
| Contact:       | William Fisher     |
| Date Prepared: | November 19, 2008  |

- 2. Device Name: Zimmer Angled Contour Zirconia Abutment (cat no. ZRA351A, ZRA352A, ZRA461A, ZRA462A)
Device Classification Name: Endosseous Dental Implant aBUTMENT

- റ്റ് Predicate Device(s): Zimmer Patient Specific Abutment, Internal Hex, Ceramic
- র্বা Device Description:

The new Abutments have sa prepared margin and cone. The Abutments feature a pre-defined offset margin that is lower on the buccal aspect and higher on the lingual aspect to minimize the need to further prepare the Abutment by user. The cone portion of the Abutment is set at a 17 degree andle.

## ട. Intended Use:

The Zimmer® Angled Contour Zirconia Abutment is used as a terminal or intermediate abutment for a cemented prosthesis. The abutment can be used for a single or multiple-unit restoration in anterior and pre-molar regions.

- 6. Device Comparison:
The new device is equivalent in design with Predicate. The new device is a dimensional modification to the Predicate. It differs from the Predicate in that the cone portion of the new device is at a 17 degree angle, while the Patient Specific is angled per customer requirements. The materials, general structure, and function in the endosseous implant system remains the same as the Predicate Device.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular fashion around the bird symbol.

## Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. William Fisher Regulatory Affairs Associate Zimmer Dental, Incorporated 1900 Aston Avenue Carlsbad, California 92008

DEC 2 3 2008

Re: K083474

Trade/Device Name: Zimmer® Angled Contour Zirconia Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: November 20, 2008 Received: November 24, 2008

Dear Mr. Fisher:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

## Page 2 - Mr. Fisher

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric' s (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Chu S. Lin, Ph.D.

hiu S. Lin. Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Zimmer Dental

510(k) No. K083474

Attachment 12.13 1 of 1

Image /page/3/Picture/3 description: The image shows a logo for "ZURBRUGG dental". The logo consists of a circle with some text and a symbol inside, above the word "ZURBRUGG" in a stylized font, and the word "dental" in a smaller, simpler font below. The text within the circle is difficult to read due to the image quality.

Indications for Use

510(k) Number (if known): __

Device Name: Zimmer® Angled Contour Zirconia Abutment

Indications For Use:

The Zimmer® Angled Contour Zirconia Abutment is used as a terminal or intermediate abutment for a cemented prosthesis. The abutment can be used for a single or multiple-unit restoration in anterior and pre-molar regions.

X Prescription Use (Part 21 CFR 801 Subpart D)

i

AND/OR

Over-The-Counter Use_ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Qve

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K083474

---

**Source:** [https://fda.innolitics.com/device/K083474](https://fda.innolitics.com/device/K083474)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
