COLONCAD API 3.1
K083423 · Medicsight Plc · NWE · May 17, 2011 · Radiology
Device Facts
| Record ID | K083423 |
| Device Name | COLONCAD API 3.1 |
| Applicant | Medicsight Plc |
| Product Code | NWE · Radiology |
| Decision Date | May 17, 2011 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.2050 |
| Device Class | Class 2 |
| Attributes | Software as a Medical Device |
Intended Use
Medicsight ColonCAD™ API is a non-invasive computer aided detection (CAD) image analysis software tool designed to assist radiologists in the detection of colorectal polyps during their review of digital images derived from CT colonography (CTC). This CAD software post-processes the CTC image data obtained from multi-detector computed tomography (MDCT) scanners. The device is intended to be used on patients referred for a CT Colonography examination, as an overlay tool to prompt the radiologist to colonic findings that have been identified by the device. The CAD can assist radiologists after they have made an initial review of all the CTC image data, supporting their evaluation ("second read").
Device Story
ColonCAD API is a software-based computer-aided detection (CAD) tool for radiologists; integrates into existing medical imaging systems via API. Inputs: digital image data from multi-detector CT (MDCT) scanners. Operation: uses mathematical image processing to identify potential polyps; highlights findings as 'CAD marks' on 2D/3D images. Workflow: radiologist performs initial review of CTC data; CAD provides second-read support by prompting for potential findings; radiologist makes final determination. Benefit: assists in polyp detection; improves diagnostic accuracy. Not intended for independent patient management decisions.
Clinical Evidence
Clinical validation performed via a Multi-Reader Multi-Case (MRMC) study. Results demonstrated that radiologists' accuracy for detecting colorectal polyps of any size was significantly higher with CAD assistance compared to unassisted reading, as measured by the segment-level area under the ROC curve (AUC).
Technological Characteristics
Medical imaging software; DICOM 3.0 compliant. Post-processing analysis of CTC image data using mathematical image processing techniques. Deployed as an API for integration with third-party visualization systems. Software-based detection algorithm.
Indications for Use
Indicated for patients referred for CT colonography (CTC) examination to assist radiologists in detecting colorectal polyps. Used as a second-read tool to highlight potential findings in MDCT-derived digital images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
- Colon CAR 1.2, Medicsight PLC (K042674)
- syngo Colonography package with extended functionality, Siemens Medical Solutions, Inc. (K042605)
Related Devices
- K091529 — ICAD VERALOOK CTC CAD SOFTWARE · Icad, Inc. · Aug 4, 2010
- K111758 — CADENS (TM) COLON · Dog Microsystems, Inc. · Feb 14, 2012
- K033102 — MEDIC COLON MODEL, RELEASE 1 · Medicsight · Nov 5, 2003
- K252586 — CADDIE · Odin Medical Limited · Sep 12, 2025
- K123566 — SYNAPSE 3D COLON ANALYSIS · Fujifilm Medical Systems U.S.A, Inc. · Jan 22, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
KC83423
Page 1 of 4
#### 5 510(k) Summary of Safety and Effectiveness MAY 1 7 2011
This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in accordance with the requirements of 21 CFR Part 807, Subpart E and Section 807.92(c).
| Owner Name | Medicsight PLC |
|--------------|---------------------------------------|
| Address | Kensington Centre |
| | 66 Hammersmith Road<br>London W14 8UD |
| Phone | +44 207 605 7950 |
| Fax | +44 207 605 7951 |
| Contact | Ms. Amarjit Gill |
| Title | VP Regulatory and Quality Assurance |
| Phone | +44 207 605 7954 |
| Fax | +44 207 605 7951 |
| Summary Date | November 18th, 2008 |
# 5.1 Identification of Owner
#### Identification of Product 5.2
| Device Name | ColonCAD API |
|---------------------|---------------------------------------------------------------|
| Device Common Name | Medical imaging software for CT scanners |
| Classification Name | Colon Computed Tomography System,<br>Computer Aided Detection |
| Regulation Number | 21 CFR 892.2050, Class II, NWE |
| Manufacturer | Medicsight PLC |
P10563 Medicsight ColonCAD API 3.1 510(k) Premarket Notification
{1}------------------------------------------------
K083423
Page 2 of 4
## 5.3 Predicate Device
The predicate devices are:
- . Colon CAR 1.2. Medicsight PLC (K042674)
- . syngo Colonography package with extended functionality, Siemens Medical Solutions, Inc. (K042605).
## 5.4 Device Description
Medicsight ColonCAD API is a medical imaging software tool designed to assist radiologists in the detection of polyps in CT scans of the product is packaged as an Application Programming Interface (API) which allows it to be integrated into existing medical imaging solutions.
The ColonCAD API assists the radiologist in detecting colorectal polyps using mathematical image processing techniques. The CAD assists the radiologist by highlighting potential polyps in 2D and 3D image views. The results are displayed in the form of "CAD marks" on or near the potential polyps. The radiologist must assess every CT scan image to search for polyps and review the CAD marked images to determine if the indicated findings are polyps.
Patient management decisions should not be made solely on the results of ColonCAD analysis.
#### 5.5 Indications for Use
Medicsight ColonCAD™ API is a non-invasive computer aided detection (CAD) image analysis software tool designed to assist radiologists in the detection of colorectal polyps during their review of digital images derived from CT colonography (CTC). This CAD software post-processes the CTC image data obtained from multi-detector computed tomography (MDCT) scanners.
The device is intended to be used on patients referred for a CT Colonography examination, as an overlay tool to prompt the radiologist to colonic findings that have P10563 Medicsight ColonCAD API 3.1 510(k) Premarket Notification Page 16 of 151
{2}------------------------------------------------
KC83423
Page 3 of 4
been identified by the device. The CAD can assist radiologists after they have made an initial review of all the CTC image data, supporting their evaluation ("second read").
#### 5.6 Predicate Device Comparison
The ColonCAD API system and its predicate devices provide post-acquisition image analysis of CT colonography images which is used to assist the radiologist in the detection of colorectal polyps.
ColonCAD API is a modified version of its predicate Colon CAR Release 1.2, Medicsight PLC (K042674). The primary difference is the method of deployment. ColonCAD API is designed as an Application Programming Interface (API) thus enabling integration with other medical image visualization systems while Colon CAR Release 1.2 is designed as a stand-alone desktop application.
ColonCAD API is also substantially equivalent to "syngo - Colonography Package with Extended Functionality", Siemens (K042605).
Both predicate devices include tools which allow the enhancement of features in CT scans in order to assist the radiologist in the detection of colorectal polyps (Polyp Enhanced Viewing). ColonCAD API provides functionality which is substantially equivalent to these tools.
ColonCAD API has the same technological characteristics as the ColonCAR 1.2 predicate device. The ColonCAD API device uses the same underlying image processing technology as the ColonCAR 1.2 device to perform the post-processing analysis of CTC image data (i.e. it uses the same mathematical rationale, software architecture design, and code-base). The computer aided detection algorithm used in the ColonCAD API device is the same as the algorithm that was used in the ColonCAR 1.2 device. No additional features or functions were added to the ColonCAD API device except for the programming interface required to facilitate the display of the CAD marks on third party visualization systems.
P10563 Medicsight ColonCAD API 3.1 510(k) Premarket Notification
Page 17 of 151
{3}------------------------------------------------
Therefore, ColonCAD API is substantially equivalent to its predicate devices and does not introduce any new safety risks.
## 5.7 Performance Standards
No applicable FDA performance standards have been issued under the authority of Section 514.
ColonCAD API conforms to the DICOM Standard 3.0.
# 5.8 Summary of Studies
Medicsight has conducted non-clinical and clinical studies to verify, validate, and assess the performance of the ColonCAD API. Non-clinical studies include: functional testing, installation testing and internal clinical evaluations.
The intended use of ColonCAD API was validated in a clinical study. The results of the MRMC study demonstrated that radiologists' accuracy for detecting colorectal polyps of any size was significantly higher with CAD than in the unassisted read, as measured by the segment-level area under the ROC curve (AUC).
## 5.9 Conclusions
The intended use and technological characteristics of the ColonCAD API 3.1 device are similar to the predicate devices. Minor technological differences do not raise any new questions regarding the safety and effectiveness of the device. Thus, the ColonCAD API system is substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a stylized symbol on the left, followed by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in a serif font. A horizontal line is present below the text. The symbol appears to be an abstract representation of a human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Amarjit Gill VP Regulatory and Quality Assurance Medicsight PLC Kensington Centre 66 Hammersmith Road, W14 8UD LONDON
MAY 1 7 2011
Re: K083423
Trade/Device Name: ColonCAD API Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: NWE Dated: March 4, 2011 Received: March 7, 2011
Dear Ms. Gill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 5.10(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Patel
Mary S. Pastel, Sc.D. . Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Indication(s) for Use Statement ব
510(k) Number (if known): K083423 Device Name: ColonCAD API Indications for Use:
Medicsight ColonCAD™ API is a non-invasive computer aided detection (CAD) image analysis software tool designed to assist radiologists in the detection of colorectal polyps during their review of digital images derived from CT colonography (CTC). This CAD software post-processes the CTC image data obtained from multi-detector computed tomography (MDCT) scanners.
The device is intended to be used on patients referred for a CT Colonography examination, as an overlay tool to prompt the radiologist to colonic findings that have been identified by the device. The CAD can assist radiologists after they have made an initial review of all the CTC image data, supporting their evaluation ("second read").
Prescription Use _____________________________________________________________________________________________________________________________________________________________ Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Many S. Pratt
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K 683423
P10563 Medicsight ColonCAD API 3.1 510(k) Premarket Notification
্রে
Page 14 of 151