← Product Code [GEI](/productcode/GEI) · K083281

# EL LIGHTWAVE SUCTION ABLATOR, MODEL EL-2000-S, LIGHTWAVE SUCTION ABLATOR, MODEL IA-2000-S, LIGHTWAVE ABLATOR, MODEL IA- (K083281)

_Linvatec Corp. · GEI · Feb 25, 2009 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K083281

## Device Facts

- **Applicant:** Linvatec Corp.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Feb 25, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The ConMed Linvatec Lightwave® Ablators are intended to be used for electrosurgical cutting and coagulation in arthroscopic procedures, using a conductive fluid environment. Additionally, the suction/aspiration versions of these devices have the capability for suctioning irrigation fluids. Electrosurgical cutting and coagulation may be performed in various joints such as the shoulder, ankle, wrist, elbow, knee, and hip.

## Device Story

ConMed Linvatec LightWave® Ablators are electrosurgical devices for arthroscopic procedures. Input: electrical energy from external electrosurgical generators. Operation: electrical current delivered via insulation-coated electrode and ceramic insulator to tissue in conductive fluid environment; suction versions include aspiration capability for irrigation fluids. Used in OR by surgeons. Output: tissue cutting and coagulation; fluid removal. Benefits: enables precise tissue management in joint spaces. Hand or foot control operation.

## Clinical Evidence

Bench testing only. Devices demonstrated conformance to ANSI/AAMI HF18:2001 and IEC 60601-2-2:2006 standards.

## Technological Characteristics

Materials: insulation-coated electrode, ceramic insulator, integrated handle, cord set. Energy: electrosurgical (requires external generator). Connectivity: connects to standard electrosurgical generators. Form factor: hand or foot-controlled probe. Standards: ANSI/AAMI HF18:2001, IEC 60601-2-2:2006.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- LightWave® Integrated Electrode Ablator and LightWave® Integrated Electrode Suction Ablator ([K050923](/device/K050923.md))
- ArthroCare® ArthroWands® ([K082323](/device/K082323.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K083281/

FEB 2 5 2009

## 510(k) SUMMARY ConMed Linvatec EL LightWave® Suction Ablator

In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, ConMed Linvatec is hereby submitting the 510(k) Summary of Safety and Effectiveness for the 510(k) Number

## A. Submitter

ConMed Linvatec 11311 Concept Boulevard Largo, Florida 33773-4908 Registration Number: 1017294

#### B. Company Contact

Lisa Anderson Regulatory Affairs Specialist (727) 399-5501 Telephone (727) 399-5264 FAX

#### C. Device Name

| Trade Name:            | ConMed Linvatec LightWave® Ablators |
|------------------------|-------------------------------------|
| Common Name:           | Electrode                           |
| Classification Name:   | General & Plastic Surgery, 878.4400 |
| Proposed Class/Device: | Class II                            |
| Product Code:          | GEI                                 |

### D. Predicate/Legally Marketed Devices

ConMed Linvatec LightWave® Integrated Electrode Ablator and LightWave® Integrated Electrode Suction Ablator

510(k) # K050923

ArthroCare® ArthroWands®

## 510(k) # K082323

## E. Device Description

The ConMed Linvatec LightWave® Ablators are comprised of a range of nonsuction and suction devices, operable by hand or foot control. The devices

p. 1 of 2

{1}------------------------------------------------

consist of an electrical, insulation-coated electrode and ceramic insulator attached to an integrated handle and cord set, allowing attachment to commonly available electrosurgical generators.

## F. Intended Use

The ConMed Linvatec Lightwave® Ablators are intended to be used for electrosurqical cutting and coagulation in arthroscopic procedures, using a conductive fluid environment. Additionally, the suction/aspiration versions of these devices have the capability for suctioning irrigation fluids.

Electrosurgical cutting and coagulation may be performed in various joints such as the shoulder, ankle, wrist, elbow, knee, and hip.

#### G. Substantial Equivalence

The ConMed Linvatec LightWave® Ablators are substantially equivalent in intended use and design to the legally marketed predicate devices LightWave® Integrated Electrode Ablator and LightWave® Integrated Electrode Suction Ablator (ConMed Linvatec) and the ArthroWands (ArthroCare). Performance testing has demonstrated the devices conform to applicable sections of ANSI/AAMI HF18:2001 and IEC 60601-2-2:2006 and do not raise any concerns with safety or efficacy.

\$\rho. 2/2\$

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white and appears to be a scanned image.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

ConMed Linvatec % Ms. Lisa Anderson Regulatory Affairs Specialist 11311 Concept Boulevard Largo, Florida 33773

# FEB 2 5 2009

Re: K083281

Trade/Device Name: ConMed Linvatec LightWave® Ablators Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: January 6, 2009 Received: January 8, 2009

Dear Ms. Anderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

#### Page 2 - Ms. Lisa Anderson

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):

V

ConMed Linvatec LightWave® Ablators Device Name:

Indications for Use:

The ConMed Linvatec Lightwave® Ablators are intended to be used for electrosurgical cutting and coagulation in arthroscopic procedures, using a conductive fluid environment. Additionally, the suction/aspiration versions of these devices have the capability for suctioning irrigation fluids.

Electrosurgical cutting and coagulation may be performed in various joints such as the shoulder, ankle, wrist, elbow, knee, and hip.

y

f

Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)

510(k) Number

ﻛﺮ

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE If NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

7832

Neil Hodden Forren

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

---

**Source:** [https://fda.innolitics.com/device/K083281](https://fda.innolitics.com/device/K083281)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
