The NuTorque Programmable Electric System is used by trained dental professionals to perform general dental procedures including crown preparation, cavity preparation, crown finishing, inlay and the filling, polishing, prophylaxis and endodontic treatment. The device is a control unit which drives a low voltage dc electric micromotor via a handpiece hose. Power is supplied to the control unit by an ac power supply. The speed of the motor is controlled by the foot control of the dental unit. ISO Etype attachments are used to perform the various procedures. The maximum free run speed of the micromotor is 40,000 rpm's.
Device Story
NuTorque Programmable Electric System is an electric dental handpiece system. Components: control unit, power supply, cable, brushless micromotor. Input: user-selected settings via color touch screen; speed control via dental unit foot pedal. Operation: control unit drives low-voltage DC micromotor; supports 100-40,000 rpm; clockwise/counterclockwise rotation; variable light intensity. Used by dental professionals in clinical settings for grinding, trimming, drilling, polishing, and endodontic procedures. Output: mechanical rotation of ISO E-type attachments. Benefits: programmable custom settings (5 endodontic, 3 preparation) and demo-mode for verifying illumination, fiber optic delay, display contrast, and auto-reverse/forward settings prior to patient use.
Clinical Evidence
No clinical data. Performance testing conducted to validate safety and effectiveness, including electrical safety, electromagnetic compatibility, and software validation/verification in accordance with consensus standards.
Technological Characteristics
System includes control unit, AC power supply, cable, and brushless micromotor. Features color touch screen interface. Supports 5 custom endodontic settings and 3 preparation settings with 27 gear/torque ratio combinations. Includes demo-mode for setting verification. ISO E-type attachment compatible. Electrical safety and EMC testing performed.
Indications for Use
Indicated for trained dental professionals performing general dental procedures including crown/cavity preparation, crown finishing, inlay/filling, polishing, prophylaxis, and endodontic treatment.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
K083252
Image /page/0/Picture/1 description: The image shows the logo for DentalEZ Group. The text "DentalEZ" is in a bold, sans-serif font, with "Dental" on the top line and "EZ" on the same line. Below the "EZ" is the word "GROUP" in a smaller font. A curved line starts below the "G" in "GROUP" and extends up and around the right side of the logo, ending above the "Z" in "EZ".
StarDental Products 1816 Colonial Village Lane Lancaster, PA 17601-5864 717/291-1161 Fax 717/391-2757 www.dentalez.com
### 510(k) Summary NuTorque Programmable Electric System October 14, 2008
FEB 1 8 2009
1988 1993 3
### Company:
DentalEZ Inc., StarDental Division Owner/operator number 2520265
#### Contact Person:
Dale Braas Quality Manager DentalEZ Inc., StarDental Division 1816 Colonial Village Lane Lancaster, PA 17601 Phone: (717) 291-1161, ext. 4308 Fax: (717) 391-2757
### Proprietary/Trade Name:
NuTorque Programmable Electric System
#### Common/Usual Name:
Dental Handpiece
Classification: ... ..........................................................................................................................................................
Controller, Foot, Handpiece and Cord (per 21 CFR 872.4200)
#### Predicate Device:
Ti-Max NL400 manufactured by Nakanishi, Inc. Japan. (K011926)
Q. Berner
Image /page/0/Picture/18 description: The image shows the logo for StarDental. The text "StarDental" is written in a bold, sans-serif font. To the right of the text is a stylized star shape, with a portion of the star cut out. The logo is simple and modern.
{1}------------------------------------------------
#### Device Description:
The NuTorque Programmable Electric System is an electric handpiece system for use in performing various dental procedures including grinding and trimming. post and pin drilling, pin setting, finishing, polishing, endodontic, caries removal and prophylaxis.
The NuTorque Programmable Electric System is composed of a power supply. control unit, cable and brushless micromotor. The control unit controls the torque, speed and directional rotation of the motor. The system is programmed through the use of a color touch screen on the control unit. The motor has a rotational speed of 100- 40,000 rpm's and can be operated in a clockwise or counterclockwise direction. The system also provides variable light intensity to the handpiece.
#### Intended Use:
The NuTorque Programmable Electric System is used by trained dental professionals to perform general dental procedures including crown preparation, cavity preparation, crown finishing, inlay and the filling, polishing, prophylaxis and endodontic treatment.
The device is a control unit which drives a low voltage de electric micromotor via a handpiece hose. Power is supplied to the control unit by an ac power supply. The speed of the motor is controlled by the foot control of the dental unit. ISO Etype attachments are used to perform the various procedures. The maximum free run speed of the micromotor is 40,000 rpm's.
#### Technological Characteristics:
The NuTorque Programmable Electric System is composed of a control unit, power supply, cable and brushless micromotor. The system can be custom programmed with 5 custom settings in the endodontic mode and 3 custom settings in the preparation mode. Within these custom settings there are 27 available gear and torque ratio combinations available. The system also has a demo-mode which allows the operator to verify the selected settings for illumination, fiber optic delay, display contrast, autoreverse-forward and speaker prior to working on the patient. Performance testing was conducted to validate the safety and effectiveness of the NuTorque Programmable Electric System. This testing included the electrical safety, electromagnetic compatibility and validation and verification of the software. Testing was completed in accordance with recognized consensus standards.
**StarDental**
{2}------------------------------------------------
## Substantial Equivalence:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------. .
The determination of substantial equivalence is based on the premise that the proposed devices and the predicate devices have the same intended use, and similar technology and design. Both devices have the same means of operation and are used for the same procedures.
Image /page/2/Picture/2 description: The image contains the logo for StarDental. The text "StarDental" is written in a bold, sans-serif font, with the word "Star" slightly larger than "Dental". To the right of the text is a stylized star symbol, which is composed of several geometric shapes. The logo is black and white.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the three branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
# FEB 1 8 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DentalEZ Incorporated, StarDental Division C/o J.A. Van Vugt KEMA Quality B.V. Utrechtseweg 310 Amhem Netherlands NL-6812 AR
Re: K083252
> Trade/Device Name: NuTorque Programmable Electric System Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EBW Dated: February 2, 2009 Received: February 3, 2009
Dear Mr. Braas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Braas
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification ... The FDA finding of substantial.equivalence.of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
## Indications for Use
# 510(k) Number (if known): _ Ko 8325 2
Device Name: NuTorque Programmable Electric System
Indications for Use:
The NuTorque Programmable Electric System is used by trained dental professionals to perform general dental procedures including crown preparation, cavity preparation, crown finishing, inlay and the filling, polishing, prophylaxis and endodontic treatment.
The device is a control unit which drives a low voltage dc electric micromotor via a handpiece hose. Power is supplied to the control unit by an ac power supply. The speed of the motor is controlled by the for control of the dental unit. ISO E-type attachments are used to perform the various procedures. The maximum freerunspeed of the micromotor is 40,000 rpm's.
Prescription I Ise (Sait 2) CFR 801 Subpart D)
Over-The-Counter Use AND/OR . . . (21 CFR: 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susie Kane
(Division Sign-Off) Division of Anesthesiolory, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.