K083207 · Quanta System Spa · GEX · Nov 14, 2008 · General, Plastic Surgery
Device Facts
Record ID
K083207
Device Name
ULTRAWAVE III EX 1320
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 14, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Intended for general surgical applications; dermatology/plastic surgery; endoscopic/laparoscopic surgery; general surgery; gynecology; ENT; hemostasis; neurosurgery; oculoplastics; pulmonary surgery; thoracic surgery; urology; and orthopedics. The intended use of the integrated cooling system in the laser hand piece is to provide cooling of the skin prior to laser treatment, for the reduction of pain during laser treatment, to allow for the use of higher fluencies for laser treatments such as hair removal and vascular lesion, and to reduce the potential side effects of laser treatments.
Device Story
Multi-source laser system (Nd:YAG 1064/1320/532nm, Alexandrite 755nm, and IPL) for surgical and dermatological applications. Device utilizes microprocessor-based controller to manage laser/IPL emission, optical delivery, and integrated skin cooling. Operated by clinicians in clinical/surgical settings. Integrated skin cooler reduces pain and thermal side effects, enabling higher fluencies. Output consists of controlled light energy for tissue interaction (ablation, coagulation, photothermolysis). Clinical decision-making supported by laser-tissue interaction for specific pathologies (e.g., vascular lesions, hair follicles, soft tissue). Benefits include precise tissue management, reduced bleeding, and cosmetic improvement.
Clinical Evidence
Bench testing only. Performance data demonstrates capability for fractional photothermolysis (creation of microscopic islets of damage). No clinical trial data provided.
Technological Characteristics
Multi-wavelength laser system (Nd:YAG 1064/1320/532nm, Alexandrite 755nm) and IPL (400-1200nm range). Includes microprocessor-based controller, power supply, optical delivery system, and integrated skin cooling handpiece. Class II laser surgical instrument.
Indications for Use
Indicated for soft tissue incision, excision, coagulation, hemostasis, vaporization, and ablation across general, endoscopic, laparoscopic, ENT, gynecological, neurosurgical, orthopedic, pulmonary, thoracic, and urological procedures. Dermatology/plastic surgery indications include treatment of benign vascular lesions (e.g., port wine stains, hemangiomas, telangiectasia), benign pigmented lesions (e.g., lentigos, tattoos, nevi), facial wrinkles, acne, and permanent hair reduction for Fitzpatrick skin types I-VI. Contraindications not specified.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Quanta System S.p.A ULTRAWAVE II EX 1320 (K070805)
Adept Medical Concept 1064/532/755 Laser (K032218)
Submission Summary (Full Text)
{0}------------------------------------------------
# (083207
# 510(k) SUMMARY Quanta System Ultrawave III EX 1320
Title:
NOV 1 4 2008
| Submitter: | Quanta System SpA<br>via IV Novembre,116<br>21058 Solbiate<br>Olona VA / Italy |
|------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ไฟ | The The Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Chil |
Dr. Isabella Carrer Contact: Medical Division Manager
| Date Prepared: | July 11, 2008 |
|-------------------------|----------------------------------------------------------------------------------------|
| Device Trade Name: | Quanta System Ultrawave III EX 1320 |
| Common Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology |
| Classification<br>Name: | Instrument, surgical, powered, laser |
| Predicate Devices: | Quanta System S.p.A ULTRAWAVE II EX<br>1320 (K070805) Adept Medical Concept 1064/532/755 Laser |
|--------------------|------------------------------------------------------------------------------------------------|
|--------------------|------------------------------------------------------------------------------------------------|
- (K032218)
| Intended Use/<br>Indications for Use: | Device Name: | Ultrawave III EX 1320 |
|---------------------------------------|--------------|-----------------------|
| | | <i>Nd:YAG 1064nm</i> |
Intended for general surgical applications; dermatology/plastic surgery; endoscopic/laparoscopic surgery; general surgery; gynecology; ENT; hemostasis; neurosurgery; oculoplastics; pulmonary surgery; thoracic surgery; urology; and orthopedics.
# General Surgical Applications:
Incision, excision, coagulation, hemostasis, vaporization, and/or ablation of soft tissue in dermatology/plastic surgery, endoscopic/laparoscopic general surgery, gastrocnterology, general surgery, gynecology, head and neck/-otorhinolaryngology (ENT),
{1}------------------------------------------------
neurosurgery, oculoplastics, orthopedics, pulmonary surgery, thoracic surgery and urology.
## Dermatology/Plastic Surgery:
Coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lake, leg and spider veins and poikiloderma of Civatte and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis . It addition, the laser is intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of blue and/or black tattoos), and plaques.
The laser is also indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
The laser is also indicated for the treatment of facial wrinkles.
It is indicated for the removal of unwanted hair, for the stable long-term, or permanent, hair reduction through selective targeting of melanin in hair follicles, and for the treatment of pseudofolliculitis barbae (PFB).
It is indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
It is indicated for use on all skin types (Fitzpatrick I-VI) including tanned skin, and the removal and permanent reduction of unwanted hair in Fitzpatrick I-VI, including suntanned skin types.
#### Orthopedics:
Cutting, ablation, and/or hemostasis of intra-articular tissue in orthopedic surgical and arthroscopic applications.
# Pulmonary Surgery:
Palliative treatment of benign and malignant
{2}------------------------------------------------
pulmonary airway obtructions, including squamous cell carcinoma, adenocarcinoma, carcinoid, benign tumors, granulomas, and benign strictures.
## Thoracic Surgery:
Incision, excision, coagulating and vaporization of soft tissue. Thoracic applications, including but not limited to, isolation of vessels for endarterectomy and/or bypass grafts, wedge resections, thoractomy, formation of pockets; vaporization, coagulation, pacemaker incision/excision, debulking, and ablation of lung tissue (thoracoscopy).
## Urology:
All applications including superficial urinary bladder tumors, invasive bladder carcinoma, urethral strictures and lesions of the external genitalia (including condyloma acuminate).
## Nd:YAG 1320nm
Indicated for use in general surgery and dermatology for the incision, excision, ablation, vaporization, coagulation and haemostasis of soft tissue. It is also indicated for the treatment of periorbital and perioral wrinkles, fine lines and wrinkles, and the treatment of back acne and atrophic acne scars.
## Nd:YAG 532nm
For the coagulation and hemostasis of vascular lesions.
## Dermatology/Plastic Surgery:
For photocoagulation and hemostasis of vascular and cutaneous lesions in dermatology including but not limited to, the following general categories: vascular lesions [angiomas, hemangiomas (port wine), telangiectasia (facial or ex-tremitics telangiectasias, venous anomalies, leg veins] : benign pigmented lesions (nevi, lentigines, chloasma, cafe au- lait, tattoos (red and green ink), verrucae, skin tags, keratoses, plaques, cutaneous lesion treatment (hemostasis, color lightening, blanching, flattening, reduction of lesion
{3}------------------------------------------------
size.
## General Surgery:
Vaporizing, Coagulating, Incising , Excising, Debulking, and Ablating of Soft Tissue as well as in Endoscopic (e.g., laparoscopic) or open surgeries.
## Gastroenterology:
Tissue ablation and hemostasis in the gastrointestinal tract; Esophageal neoplastic obstructions, including squamous cell carcinoma and adenocarcinoma; Gastrointestinal hemostasis (including Varices, Esophagitis, Esophageal Ulcer, Mallory-Woiss tear, Gastric Ulcer, Angiodysplasia, Stomal Ulcers, Nonbleeding Ulcers, Gastric Erosions); Gastrointestinal Tissue ablation (Benign and Malignant neoplasm, Angiodysplasia, Polyps, Ulcer, Colitis, Hemorrhoids).
### Head and NecMOtorhinolaryngology (ENT):
Tissue incision, excision, ablation, and vessel hemostasis
# Hemostasis during Surgery:
Adjunctive coagulation and hemostasis (bleeding control) during surgery in endoscopic (e.g., laparoscopic) and open surgery.
## Neurosurgery:
Hemostasis for: Pituitary Tumor, Meningioma; hemagioblastoma: AVMs: Glioma; Glioblastoma; Astrocytoma; Oligodendroglioma.
### Ophthalmology:
Post-vitrectomy endophotocoagulation of the retina.
## Pulmonary Surgery:
Palliative treatment of benign and malignant pulmonary airway obstructions, including Squamous Cell Carcinoma, Adenocarcinoma; Carcinoid; Benign Tumors; Granulomas; Benign Strictures.
{4}------------------------------------------------
#### Thoracic Surgery:
Cutting (incision and excision), coagulating, and vaporizing of soft tissue Thoracic applications including, but not limited to: Isolation of vessels for endartecetomy and/or by-pass grafts, Wedge Resections, Thoractoniy, Formation of Pacemaker pockets. Vaporization, coagulation, incision and excision, debulking, and ablation oflung tissue (Thoracoscopy )
## Urology:
All applications including: Superficial urinary bladder tumors, invasive bladder carcinoma; Urethral Strictures; Lesions of the external genitalia (including condyloma acuminata).
# Alexandrite 755nm
Intended for coagulation and hemostatis of vascular lesions and the removal and permanent reduction of unwanted hair in Fitzpatrick skin types I-VI, including suntanned skin types. Also indicated for pigmented lesions and wrinkles.
#### IPL
## 590-1200nm : 625-1200nm; 650-1200nm
Indicated for permanent hair reduction.
#### 550-1200nm : 570-1200nm
Indicated for photocoagulation of dermatological vascular lesion ( i.e., face telangiectasia), photothermolysis of blood vessels (treatment of facial and leg veins), and treatment of benign pigmented lesions.
## 400-1200nm
Indicated for inflammatory acne (acne vulgaris).
### Integrated Skin Cooler
The intended use of the integrated cooling system in the
{5}------------------------------------------------
laser hand piece is to provide cooling of the skin prior to laser treatment, for the reduction of pain during laser treatment, to allow for the use of higher fluencies for laser treatments such as hair removal and vascular lesion, and to reduce the potential side effects of laser treatments.
Any other different use is considered incorrect.
The Ultrawave III EX 1320 consists of two laser sources and an IPL. The laser sources emit wavelengths of 1064nm (long pulse and short pulse),1320nm and 532nm for the Nd:YAG source and 755nm for the Alexandrite source. The IPI. has several different spectral ranges to be utilized for different applications: 400-1200nm, 550-1200nm, 570-1200nm, 590-1200nm, 625-1200nm and 650-1200nm. In addition, the Ultrawave III EX 1320 includes a power supply; a cooling system; an optical delivery system; a microprocessor based controller; an integral skin cooler; and safety features to ensure use of the appropriate laser, wavelength and hand piece.
Performance data was provided showing the Ultrawave III EX 1320 is capable of performing fractional photothermolysis, i.e., creation of pattern (lattice) of Performance Data microscopic islets of damage at superficial skin layers.
> The Quanta System Ultrawave III EX 1320 is as safe and effective as the predicate devices. The Ultrawave III EX 1320 has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Ultrawave III EX 1320 and its predicate devices raise no new issues of safety or effectiveness. Thus, the Ultrawave III EX 1320 is substantially equivalent.
# Technological Characteristics:
Substantial Equivalence:
{6}------------------------------------------------
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 4 2008
Quanta System S.P.A. % Regulatory Technology Services, LLC Mr. Mark Job 1394 25"1 Street Northwest Buffalo, Minnesota 55313
Re: K083207
Trade/Device Name: Ultrawave III EX 1320 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: October 30, 2008 Received: October 31, 2008
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{7}------------------------------------------------
Page 2 - Mr. Mark Job
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark M. Milliman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): 1 083207
Ultrawave III EX 1320 Device Name:
# Nd:YAG 1064nm
Intended for general surgical applications; dermatology/plastic surgery; endoscopic/laparoscopic surgery; general surgery; gynecology; ENT; hemostasis; neurosurgery; oculoplastics; pulmonary surgery; thoracic surgery; urology; and orthopedics.
# General Surgical Applications:
Incision, excision, coagulation, hemostasis, vaporization, and/or ablation of soft tissue in dermatology/plastic surgery, endoscopic/laparoscopic general surgery, gastroenterology, general surgery, gynecology, head and neck/-otorhinolaryngology (ENT), neurosurgery, oculoplastics, orthopedics, pulmonary surgery, thoracic surgery and urology.
# Dermatology/Plastic Surgery:
Coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lake, leg and spider veins and poikiloderma of Civatte and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis . It addition, the laser is intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of blue and/or black tattoos), and plaques.
The laser is also indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
The laser is also indicated for the treatment of facial wrinkles.
It is indicated for the removal of unwanted hair, for the stable long-term, or permanent, hair reduction through selective targeting of melanin in hair follicles, and for the treatment of pseudofolliculitis barbae (PFB).
It is indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
> (Division Sign-Division of General, Restorative, and Neurological Devices
**510(k) Number** K083207
\\\DC - 028794/00001 - 2439941 vI
{9}------------------------------------------------
It is indicated for use on all skin types (Fitzpatrick I-VI ) including tanned skin, and the removal and permanent reduction of unwanted hair in Fitzpatrick I-VI, including suntanned skin types.
## Orthopedics:
Cutting, ablation, and/or hemostasis of intra-articular tissue in orthopedic surgical and arthroscopic applications.
## Pulmonary Surgery:
Palliative treatment of benign and malignant pulmonary airway obtructions. including squamous cell carcinoma, adenocarcinoma, carcinoid, benign tumors, granulomas, and benign strictures.
## Thoracic Surgery:
Incision, excision, coagulating and vaporization of soft tissue. Thoracic applications, including but not limited to, isolation of vessels for endarterectorny and/or by-pass grafts, wedge resections, thoractomy, formation of pacemaker pockets; vaporization, coagulation, incision/excision, debulking, and ablation of lung tissue (thoracoscopy).
## Urology:
All applications including superficial urinary bladder tumors, invasive bladder carcinoma, urethral strictures and lesions of the external genitalia (including condyloma acuminate),
# Nd:YAG 1320nm
Indicated for use in general surgery and dermatology for the incision, excision, ablation, vaporization, coagulation and haemostasis of soft tissue. It is also indicated for the treatment of periorbital and perioral wrinkles, fine lines and wrinkles, and the treatment of back acne and atrophic acne scars.
## Nd:YAG 532nm
For the coagulation and hemostasis of vascular lesions.
Dermatology/Plastic Surgery:
Neil RP. Ogden for mxn
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** K083207
2
{10}------------------------------------------------
For photocoagulation and hemostasis of vascular and cutaneous lesions in dermatology including but not limited to, the following general categories: vascular lesions [angiomas, hemangiomas (port wine), telangiectasia (facial or ex-tremities telangiectasias, venous anomalies, leg veins] : benign pigmented lesions (nevi, lentigines, chloasma, cafe au- lait, tattoos (red and green ink), verrucae, skin tags, keratoses, plaques, cutaneous lesion treatment (hemostasis, color lightening, blanching, flattening, reduction of lesion size.
# General Surgery:
Vaporizing, Coagulating, Incising, Excising, Debulking, and Ablating of Soft Tissue as well as in Endoscopic (e.g., laparoscopic) or open surgeries.
# Gastroenterology:
Tissue ablation and hemostasis in the gastrointestinal tract; Esophageal neoplastic obstructions, including squamous cell carcinoma and adenocarcinoma; Gastrointestinal homostasis (including Varices, Esophagitis, Esophageal Ulcer, Mallory-Weiss tear, Gastric Ulcer, Angiodysplasia, Stomal Ulccrs, Non-bleeding Ulcers, Gastric Erosions); Gastrointestinal Tissue ablation (Benign and Malignant neoplasm, Angiodysplasia, Polyps, Ulcer, Colitis, Hemorrhoids).
# Head and NecMOtorhinolaryngology (ENT):
Tissue incision, excision, ablation, and vessel hemostasis
## Hemostasis during Surgery:
Adiunctive coagulation and hemostasis (bleeding control) during surgery in endoscopic (e.g., laparoscopic) and open surgery.
#### Neurosurgery:
Hemostasis for: Pituitary Tumor, Meningioma; hemagioblastoma; AVMs; Glioma; Glioblastoma; Astrocytoma; Oligodendroglioma.
## Ophthalmology:
Post-vitrectomy endophotocoagulation of the retina.
## Pulmonary Surgery:
Neil RR Hyde for nxm
(Division Sign-C Division of General. Restorative, and Neurological Devices
**510(k) Number** K083207
3
{11}------------------------------------------------
Palliative treatment of benign and malignant pulmonary airway obstructions, including Squamous Cell Carcinoma, Adenocarcinoma; Carcinoid; Benign Tumors; Granulomas; Benign Strictures.
# Thoracic Surgery:
Cutting (incision and excision), coagulating, and vaporizing of soft tissue Thoracic applications including, but not limited to: Isolation of vessels for endartecetomy and/or by-pass grafts, Wedge Resections, Thoractoniy, Formation of Pacemaker pockets. Vaporization, coagulation, incision and excision, debulking, and ablation oflung tissue (Thoracoscopy )
# Urology:
All applications including: Superficial urinary bladder tumors, invasive bladder carcinoma; Urethral Strictures; Lesions of the external genitalia (including condyloma acuminata).
# Alexandrite 755nm
Intended for coagulation and hemostatis of vascular lesions and the removal and permanent reduction of unwanted hair in Fitzpatrick skin types I-VI, including suntanned skin types. Also indicated for pigmented lesions and wrinkles.
# IPL
# 590-1200nm ; 625-1200nm; 650-1200nm
Indicated for permanent hair reduction.
## 550-1200nm : 570-1200nm
Indicated for photocoagulation of dermatological vascular lesion ( i.e., face telangiectasia), photothermolysis of blood vessels (treatment of facial and leg veins), and treatment of benign pigmented lesions.
## 400-1200nm
Indicated for inflammatory acne (acne vulgaris).
Nilre Ogle for nxm
(Division Sign-Off) Division of General. Restorative. and Neurological Devices
510(k) Number 085207
# Integrated Skin Cooler
4
{12}------------------------------------------------
The intended use of the integrated cooling system in the laser hand piece is to provide cooling of the skin prior to laser treatment, for the reduction of pain during laser treatment, to allow for the use of higher fluencies for laser treatments such as hair removal and vascular lesion, and to reduce the potential side effects of laser treatments.
Any other different use is considered incorrect.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ Use (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter
(21 C,F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Niel R.P. Ogle for mxm
(Division Sign-Ol Division of General, Restorative, and Neurological Devices
**510(k) Number** K083207
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.