THIRD EYE RETROSCOPE AUXILIARY ENDOSCOPY SYSTEM

K083180 · Avantis Medical Systems, Inc. · FDF · Feb 12, 2009 · Gastroenterology, Urology

Device Facts

Record IDK083180
Device NameTHIRD EYE RETROSCOPE AUXILIARY ENDOSCOPY SYSTEM
ApplicantAvantis Medical Systems, Inc.
Product CodeFDF · Gastroenterology, Urology
Decision DateFeb 12, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Avantis Medical Systems, Inc. Third Eye is indicated for use in the instrument channel of a conventional colonoscope to provide retrograde illumination and visualization of the colon for diagnostic purposes.

Device Story

Third Eye Retroscope is an auxiliary endoscopy system used during colonoscopy. Device consists of a disposable Retroscope and Cap, and a reusable Video Processor. After a colonoscope reaches the cecum, the Retroscope is inserted through the instrument channel; it automatically bends 180 degrees into a 'J' shape upon exiting the distal tip. The system provides continuous retrograde (backward-looking) visualization of the colon during colonoscope withdrawal. The Video Processor interfaces with the Retroscope, processes video signals, and provides user controls for light intensity and picture settings. The physician views the retrograde image on a monitor alongside the standard colonoscope view. This dual-view approach improves the detection of polyps and adenomas compared to traditional colonoscopy alone. The device is intended for use in clinical settings by trained physicians.

Clinical Evidence

Prospective, multi-center post-market study. Results demonstrated that retrograde viewing with the Third Eye significantly improved the detection of polyps and adenomas compared to traditional colonoscopy alone. Findings were statistically significant. Risks to patients were comparable to those reported in literature for traditional screening colonoscopies.

Technological Characteristics

System includes a disposable Retroscope, Cap, and a Video Processor. Retroscope features a flexible design that assumes a 'J' shape upon deployment. Video Processor provides electrical interface, signal processing, and user-adjustable light/picture settings. Connectivity includes video output to a monitor. Sterilization is maintained via sterile barrier packaging. Software is used for image processing and user control adjustments.

Indications for Use

Indicated for patients undergoing colonoscopy for diagnostic purposes; used as an auxiliary device with a conventional colonoscope to provide retrograde illumination and visualization of the colon.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K 0831.80 Image /page/0/Picture/1 description: The image shows the logo for Avantis Medical Systems. The logo consists of the word "avantis" in lowercase letters, with the words "medical systems" in smaller letters below it. Above the word "avantis" is a stylized graphic that appears to be two curved shapes with a circle in the middle. 510(k) SUMMARY FEB 1 2 2009 ## Applicant Information: a) | Date Summary Prepared | 09 February 2009 | |-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------| | Sponsor/Submitter | Avantis Medical Systems, Inc.<br>263 Santa Ana Court<br>Sunnyvale, CA 94085-4511 | | Correspondent Contact Information | Amrita Sethi<br>Director of Regulatory and Quality Affairs<br>Phone: 408-733-1901 x227<br>Fax: 408-733-1847<br>E-mail: asethi@avantismedical.com | ## Device Information: b) | Device Common Name | Endoscope | |--------------------------------------|------------------------------------------------------------------| | Device Trade & Proprietary Name | Third Eye® Retroscope® Auxiliary<br>Endoscopy System (Third Eye) | | Device Classification Name | Colonoscope (per 21CFR 876.1500) | | Device Classification Regulation | 21CFR 876.1500 | | Device Classification | Class II (special controls) | | Device Classification & Product Code | FDF, FDS, KNW, KOG | #### Identification of Predicate Device: c) The Avantis Third Eye is substantially equivalent in operation and fundamental scientific technology to its previous version, cleared under K070330. ### d) Device Description Summary: The Third Eye System consists of both disposable portions and facility equipment. The disposable portions include the Third Eye Retroscope and Caps. The facility equipment portions of the device include the Third Eye Video Processor and accessories. The Third Eye is designed as an auxiliary device for use during a colonoscopy {2}------------------------------------------------ Ko 83i80 Image /page/1/Picture/1 description: The image shows the logo for Avantis Medical Systems. The logo features the word "avantis" in a lowercase sans-serif font, with the words "medical systems" in a smaller font below. Above the word "avantis" is a graphic of two curved shapes facing each other with a circle in the middle. procedure. After a colonoscope has been advanced to the cecurn, the Third Eye Retroscope is inserted through the instrument channel of the colonoscope. As the Third Eye Retroscope emerges from the distal tip of the instrument channel of the colonoscope, it automatically bends 180 degrees to form a "J" shape. The Third Eye then provides a continuous retrograde image of the colon throughout the process of withdrawal of the colonoscope. The Third Eve Video Processor is a piece of electrical equipment that acts as an interface for the Third Eye Retroscope, the video output signals, and user input controls. The output video signals from the Video Processor are displayed on a monitor for viewing by the physician. Application software that runs on the Video Processor allows the user to adjust the light intensity and various picture settings of the Third Eve Retroscope image. - Intended Use: e) The Third Eye Retroscope is intended for use in the instrument channel of a conventional colonoscope to provide retrograde illumination and visualization of the colon for diagnostic purposes. ### Discussion of Substantial Equivalence to Predicate Device: f) The Third Eye has been modified to a) accommodate additional conventional colonoscope models and brands, b) provide product enhancements, c) improve manufacturability, and d) update software. The modified device is substantially equivalent to the cleared predicate device in fundamental design materials, processes, sterile barrier packaging, and sterilization. All changes have been subjected to risk analysis and have been verified to meet current design specifications. The IFU has been updated to a) add clinical data, b) clarify procedural directions based on the clinical experiences and c) expand the indications for use based on clinical data demonstrating that the Third Eye used with a traditional colonoscope improves detection of polyps and adenomas during colonoscopy as compared to the use of a colonoscope alone. Differences between the modified Third Eye System and the predicate device have no impact on safety and effectiveness. #### Scientific Technology g) The Third Eye Retroscope® Auxiliary Endoscopy System includes the Retroscope, a Video Processor, Cap and accessories. Modifications made to the device and accessories did not change the fundamental scientific technology of the device. The testing described demonstrates that the differences in the device components and {3}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for Avantis Medical Systems. The logo features the word "avantis" in lowercase letters, with the words "medical systems" underneath. Above the word "avantis" is a graphic of two curved shapes that are connected by a circle in the middle. R08 31 80 yg 3 of 3 accessories do not raise any new or unresolved issues for safety and efficacy. ## Summary of Supporting Non-Clinical Performance Data h) Bench verification testing was conducted to verify that the modified device meets the design inputs and intended performance characteristics. Results demonstrate that modifications made to the Third Eve design performs as intended. ### Summary of Supporting Clinical Data i) Results from a prospective, multi-center post-market study demonstrate that retrograde viewing of the colon with the Third Eye improves the detection of polyps and adenomas compared to traditional colonoscopy alone. Results were statistically significant. ## Conclusions j) Avantis Medical Systems, Inc. has determined, based on clinical data, that the Third Eye improves the detection of polyps and adenomas in patients for whom colonoscopies are indicated. Risks to patients in terms of adverse events are comparable to those reported in the literature for traditional screening colonoscopies. Use of the Third Eye did not increase the risks to the patients but offered a statistically significant benefit of improved detection. Results of the clinical study are included in the Instructions for Use. Any statement regarding "substantial equivalence" made in this 510(k) submission and summary only relates to whether the product addressed in this submission may be lawfully marketed without premarket approval or reclassification, and is not intended to be interpreted as an admission or any other type of evidence in any patent proceeding, including patent infringement, litigation or proceeding before any Patent Office. The present submission and statements therefore should not be construed as affecting or relating to the scope of any patent application, or to whether the product addressed in this submission, or its use, may be considered indistinct, from a patentability perspective, from any of the other devices referenced in this filing > 263 Santa Ana Ct., Sunnyvale, CA 94085 TEL: (408) 733-1901, FAX: (408) 733-1847 3 of 3 {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract image of an eagle. # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 1 2 2009 Ms. Amrita Sethi Director of Regulatory and Quality Affairs Avantis Medical Systems, Inc. 263 Santa Ana Court SUNNYVALE CA 94085-4511. Re: K083180 Trade/Device Name: Third Eye® Retroscope® Auxiliary Endoscopy System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDF and ODA Dated: February 4, 2009 Received: February 5, 2009 Dear Ms. Sethi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 | |----------------|----------------------------------|----------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 | | 21 CFR 892.xxx | (Radiology) | (240) 276-0120 | | Other | | (240) 276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html. Sincerely yours, Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT 510(k) Number (if known): K083180 Device Name: Third Eye® Retroscope® Auxiliary Endoscopy System (Third Eye) Indications for Use: The Avantis Medical Systems, Inc. Third Eye is indicated for use in the instrument channel of a conventional colonoscope to provide retrograde illumination and visualization of the colon for diagnostic purposes. Prescription Use × Over-The-Counter Use - AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Qubuthemen (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...