K083164 · Bio-Medical Research, Ltd. · NGX · Dec 16, 2008 · Physical Medicine
Device Facts
Record ID
K083164
Device Name
SYSTEM-ARMS, MODELS 390, X60
Applicant
Bio-Medical Research, Ltd.
Product Code
NGX · Physical Medicine
Decision Date
Dec 16, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
System Arms, Type 390, Model X60 is intended for use by healthy adults to apply Neuromuscular Electrical Simulation (NMES) via adhesive electrodes to healthy muscles of the upper arms in order to improve or facilitate the muscle performance of the triceps and biceps muscles. Proposed indications for use: System Arms (Type 390, X60) is intended to stimulate healthy muscles in order to improve or facilitate muscle performance of the triceps and biceps muscles. System Arms, Type 390, Model X60 is intended for over-the-counter use.
Device Story
System Arms, Type 390, Model X60 is a two-channel, battery-powered neuromuscular electrical stimulation (NMES) system. Components include two arm garments, a hand-held rechargeable control unit, and adhesive gel-based electrodes. The device is intended for over-the-counter use by healthy adults. The user applies the garment to the upper arms; the control unit delivers electrical stimulation to the triceps and biceps muscles via the electrodes. The system offers three user-selectable programs. Power is provided by a 3.6V NiMH rechargeable battery. The device is designed to improve or facilitate muscle performance. The control unit is interchangeable with other Slendertone System models. Internal wiring is over-molded to prevent moisture ingress and prevent user access to current paths.
Clinical Evidence
No clinical data submitted. Device safety and performance supported by compliance with international standards including IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, and IEC 60601-1-6.
Technological Characteristics
Two-channel, battery-powered (3.6V NiMH) NMES stimulator. Includes arm garments and adhesive gel-based electrodes. Over-molded internal connections for moisture protection. Complies with IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, and IEC 60601-1-6. Battery charger complies with IEC 60950 and UL 1950.
Indications for Use
Indicated for healthy adults to stimulate healthy muscles to improve or facilitate performance of triceps and biceps muscles. Intended for over-the-counter use.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{1}------------------------------------------------
DEC 1 6 2008
Image /page/0/Picture/1 description: The image shows the logo for BMR. The logo consists of the letters "BMR" in a bold, sans-serif font. To the left of the letters is a curved shape that resembles a crescent or a stylized "C", which partially encloses the letters. The entire logo is in black and white.
# l Research Ltd.
usiness Park West, Galway, Ireland - Fax: +353 (0)91 774301 353 (0)91 774300
KOr3164
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
#### 1. Contact Details
| Name: | Anne-Marie Keenan |
|------------|-------------------------------------------------------------------------------|
| Address: | Bio-Medical Research Ltd.,<br>Parkmore Business Park, West<br>Galway, Ireland |
| Telephone: | +353 91 774300 |
| Fax: | +353 91 774301 |
| E-Mail: | akeenan@bmr.ie |
| Prepared: | 24th October 2008 |
| Revised: | 2nd December 2008 |
#### 2. Device Name
| Trade Name of Device: | System Arms, Type 390, Model X60 |
|-----------------------|------------------------------------------------------|
| Common Name: | Muscle Stimulator |
| Classification Name: | Stimulator, muscle, powered, for muscle conditioning |
| Product Code: | NGX |
### 3. Identification of Equivalent Legally Marketed Device
| Device Trade Name: | EnerVive, Type 561 | System-Arms Type 390, E60 |
|--------------------|--------------------------|---------------------------|
| Manufacturer: | Bio-Medical Research Ltd | Bio-Medical Research Ltd |
| 510(k) Nos: | K071666 | K072553 |
#### 4. Description of Device
System Arms, Type 390, Model X60 is a two-channel, battery powered, muscle stimulation system. It is supplied with two arms garments (left & right arm), a hand-held rechargeable control unit, a pack of 4 adhesive backed gel based electrodes, instructions for use and a carry pouch.
The control unit is interchangeable between all the cleared System models from the Slendertone® range of garments (i.e. System-Abs Type 390, Model E10/X10, System-Shorts
{2}------------------------------------------------
Type 390, Model E20, System-Mini, Type 390, E30 and also System Arms Type 390, Model E60).
There are three programs available to users of the System Arms, Type 390, X60. Power is derived from a 3.6V NiMH rechargeable battery pack pre-installed in the unit.
All the internal connections are over-molded to prevent moisture ingress. The user has no access to the wiring or connectors within the garment and is unable to alter the current path.
For purposes of hygiene, the garment may be cleaned and instructions for belt care are included in the user manual.
#### 5. Statement of Intended Use/Indications for Use
System Arms, Type 390, Model X60 is intended for use by healthy adults to apply Neuromuscular Electrical Simulation (NMES) via adhesive electrodes to healthy muscles of the upper arms in order to improve or facilitate the muscle performance of the triceps and biceps muscles.
Proposed indications for use: System Arms (Type 390, X60) is intended to stimulate healthy muscles in order to improve or facilitate muscle performance of the triceps and biceps muscles.
System Arms, Type 390, Model X60 is intended for over-the-counter use.
#### 6. Technological Characteristics
The System Arms (Type 390, X60) incorporates the control unit and garment technology of the existing predicate System Arms (K072553).
#### 7. Clinical and Non-Clinical Tests
No new clinical studies have been submitted as part of this premarket notification.
The System Arms, Type 390, X60 device has been CE marked under the Medical Device Directive 93/42/EEC (NB No. 0366) and also complies with the following international safety standards:
Page 2 of 3
{3}------------------------------------------------
- IEC 60601-1:1988 & A1:1991, A2:1995 Medical electrical equipment Part 1: l General requirements for safety
- IEC 60601-2-10:1987 & Al 2001 Medical electrical equipment Part 2-10: 사 Particular requirements for the safety of nerve and muscle stimulators
- IEC 60601-1-2:2001 (EN 60601-1-2:2001) Medical electrical equipment Part 1-2: " General requirements for safety - Collateral standard: Electromagnetic compatibility -Requirements and tests
- CISPR 22:2003/CFR 47 Part 15:2005 제
- IEC 60601-1-6:2004 (EN 60601-1-6:2001) Medical electrical equipment Part 1-6: . General requirements for safety - Collateral standard: Usability
The battery charger complies to safety standards IEC 60950 and UL 1950
Page 3 of 3
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the eagle. The eagle is depicted in a simple, bold line drawing style. The text is in all caps and appears to be in a sans-serif font.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Biomedical Research Limited % Ms. Anne-Marie Keenan Parkmore Business Park, West Galway Ireland
# DEC 1 6 2008
Re: K083164
Trade/Device Name: System Arms (Type 390, X60) Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: December 2, 2008 Received: December 4, 2008
Dear Ms. Keenan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
#### Page 2 - Ms. Anne-Marie Keenan
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
00 Mark N. Melkerson
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known):
Device Name:
System Arms (Type 390, X60)
Indications for Use:
System Arms (Type 390, X60) is intended to stimulate healthy muscles in order to improve or facilitate muscle performance of the triceps and biceps muscles.
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use X AND/OR (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
FOR M. MELKERSON
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** K083164
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.