← Product Code [LLZ](/productcode/LLZ) · K083149

# ARTERIAL HEALTH PACKAGE (AHP) SOFTWARE (K083149)

_Siemens Medical Solutions USA, Inc. · LLZ · Jan 14, 2009 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K083149

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jan 14, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".

## Device Story

Arterial Health Package (AHP) software; operates on diagnostic ultrasound systems. Inputs: ultrasound images of carotid artery. Transformation: software tools measure intima-media thickness (IMT); references validated normative tables. Output: quantitative IMT measurements; visual/tabular data for patient communication. Used in clinical settings by physicians. Purpose: assist in evaluating cardiovascular disease risk and communicating findings to patients. Adheres to ASE Consensus Statement guidelines for carotid ultrasound use.

## Clinical Evidence

Bench testing only. Software verified and validated according to company design control process. No clinical study data provided.

## Technological Characteristics

Diagnostic ultrasound imaging software; provides measurement of anatomical structures (IMT) and analysis packages. Operates as an integrated software feature on ultrasound systems. Verification and validation performed per design controls.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- [K061980](/device/K061980.md)
- [K071036](/device/K071036.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KOr3149

JAN 1 4 2009

# Section 5

## 510(k) Summary

| <b>Sponsor:</b>              | Siemens Medical Solutions USA, Inc.,<br>Ultrasound Division<br>1230 Shorebird Way<br>Mountain View, California 94043 |
|------------------------------|----------------------------------------------------------------------------------------------------------------------|
| <b>Contact Person:</b>       | Sheila.W. Pickering<br>Telephone: (650) 965 5398<br>Fax: (650) 943 7053                                              |
| <b>Submission Date:</b>      | September 29, 2008                                                                                                   |
| <b>Device Name:</b>          | Siemens Diagnostic Ultrasound Systems Arterial Health Package<br>Software                                            |
| <b>Common Name:</b>          | Diagnostic Ultrasound System Imaging Software                                                                        |
| <b>Classification:</b>       |                                                                                                                      |
| <b>Regulatory Class:</b>     | II                                                                                                                   |
| <b>Review Category:</b>      | Tier II                                                                                                              |
| <b>Classification Panel:</b> | Radiology<br>System. Imaging Processing : 21 CFR 892.2050: LLZ                                                       |

#### A. Legally Marketed Predicate Devices

The modified software is substantially equivalent to the software cleared in K061980 and K071036.

#### B. Device Description:

The modified software features provide for the measurement of anatomical structures and for analysis packages that provide information.

#### C. Intended Use

The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".

#### D. Substantial Equivalence

The submission device is substantially equivalent to the predicate with regard to both intended use and technological characteristics.

#### E. Performance Data

The Arterial Health Package software has been verified and validated according to the company's design control process.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus has a staff with a snake winding around it, and a pair of wings at the top.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Siemens Medical Solutions USA, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street NW BUFFALO MN 55313

JAN 1 4 2009

Re: K083149

Trade/Device Name; Arterial Health Package (AHP) software . Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 26, 2008 Received: December 29, 2008

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology)                      | (240) 276-0120 |
| Other          |                                  | (240) 276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.

Sincerely yours,

Janine M. Morris

anine M. M. nrri Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Section 4

## Indications for Use

510(k) Number (if known):

Device Name:

Arterial Health Package (AHP) Software

Indications For Use:

The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the

"ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices

510(k) Number

009

---

**Source:** [https://fda.innolitics.com/device/K083149](https://fda.innolitics.com/device/K083149)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
