VERITAS COLLAGEN MATRIX (DRY)

K083039 · Synovis Surgical Innovations · FTM · Nov 26, 2008 · General, Plastic Surgery

Device Facts

Record IDK083039
Device NameVERITAS COLLAGEN MATRIX (DRY)
ApplicantSynovis Surgical Innovations
Product CodeFTM · General, Plastic Surgery
Decision DateNov 26, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Veritas Collagen Matrix (Dry) is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies; using surgical staplers when staple line reinforcement is needed. Veritas Collagen Matrix (Dry) can be used for reinforcement of staple lines during lung and bronchus resection, during gastric, bariatric, small bowel, mesentery, colon and colorectal procedures, and cardiac surgery (for example: occlusion of the left atrial appendage during open chest procedures). Veritas Collagen Matrix (Dry) minimizes tissue attachment to the device in case of direct contact with viscera.

Device Story

Veritas Collagen Matrix (Dry) is an implantable surgical mesh composed of non-crosslinked bovine pericardium. It serves as a buttress for surgical staple lines to reinforce tissue during various surgical procedures, including lung, gastric, bariatric, and cardiac surgeries. The device facilitates neo-collagen formation, neo-vascularization, and tissue remodeling while minimizing visceral tissue attachment. It is used by surgeons in clinical settings. This specific submission introduces a new adhesive method for attaching the buttress to the stapler.

Clinical Evidence

Bench testing only. Testing focused on the biocompatibility of the new adhesive and its mechanical performance in allowing the device to function as a staple line buttress.

Technological Characteristics

Implantable surgical mesh; non-crosslinked bovine pericardium; sourced from cattle <30 months old. New attachment method utilizes an adhesive. Device is intended for use with linear and circular surgical staplers.

Indications for Use

Indicated for patients requiring soft tissue repair or staple line reinforcement during lung, bronchus, gastric, bariatric, small bowel, mesentery, colon, colorectal, or cardiac surgeries.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 2.0 510(K) SUMMARY Submitted by K083039 | | 2575 University Avenue West<br>St. Paul, MN 55114-1024<br>Tel: 651-796-7300<br>Fax: 651-642-9018 | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | NOV 2 6 2008 | | Contact Person | Jodi Jorgenson<br>At address above | | Date Prepared | October 13, 2008 | | Device Trade Name: | Veritas® Collagen Matrix (Dry)<br>Marketed as Peri-Strips Dry® with Veritas® Collagen Matrix. | | Common Name | Surgical Mesh | | Classification Name | Mesh, Surgical<br>21 CFR 878.3300<br>Product Code: FTM | | Predicate devices | Veritas® Collagen Matrix (Dry) K041669 and Veritas®<br>Collagen Matrix (Wet) K062915<br>(Device acting as its own predicate.) | | Device Description | Veritas Collagen Matrix is an implantable surgical mesh<br>comprised of non-crosslinked bovine pericardium. Veritas<br>Collagen Matrix bovine pericardium is procured in the United<br>States from cattle less than 30 months of age. | | | Veritas Collagen Matrix allows for neo-collagen formation and<br>neo-vascularization of the implanted device and permits<br>replacement of the device with host tissue, or remodeling.<br>Veritas Collagen Matrix also minimizes tissue attachment to the<br>device in case of direct contact with viscera. | : : . . . Synovis Surgical Innovations Veritas Collagen Matrix (Dry) is used for buttressing and reinforcing surgical staple lines and is currently available in configurations for multiple sizes and brands of both linear and circular staplers. . : - . . {1}------------------------------------------------ | Statement of Intended use | Intended Use (Unchanged from previous clearances) | |---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The device is intended to be used as a staple line buttress | | | Indications for Use (Unchanged from previous clearances) | | | Veritas Collagen Matrix (Dry) is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies; using surgical staplers when staple line reinforcement is needed. | | | Veritas Collagen Matrix (Dry) can be used for reinforcement of staple lines during lung and bronchus resection, during gastric, bariatric, small bowel, mesentery, colon and colorectal procedures, and cardiac surgery (for example: occlusion of the left atrial appendage during open chest procedures). | | | Veritas Collagen Matrix (Dry) minimizes tissue attachment to the device in case of direct contact with viscera. | | Technological Comparisons | Veritas Collagen Matrix (Dry) is acting as its own predicate and is therefore substantially equivalent, having the same technological characteristics and intended use with the exception of the method of attachment of the buttress to the stapler, which is the subject of this submission. | | Technology/Device Testing | Veritas Collagen Matrix (Dry) is unchanged and previous testing applies. The testing presented in this submission is specifically related to the use of an adhesive, which is the new method for attachment of the buttress to the stapler. The testing focuses on the biocompatibility of the adhesive and its ability to allow the device to function as intended. | . The same of the same of the same of the states of the many · : : {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Synovis Surgical Innovations % Ms. Jodi Jorgenson Sr. Regulatory Affairs Specialist 2575 University Avenue West St. Paul, Minnesota 55114-1024 NOV 2 6 2008 Re: K083039 Trade/Device Name: Veritas Collagen Matrix (Dry) Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTM Dated: October 13, 2008 Received: October 14, 2008 Dear Ms. Jorgenson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ ## Page 2 - Ms. Jodi Jorgenson This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Mark M. Mulhern - Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K083039 Device Name: Veritas Collagen Matrix (Dry) Indications For Use: Veritas Collagen Matrix (Dry) is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers when staple line reinforcement is needed. Veritas Collagen Matrix (Dry) can be used for reinforcement of staple lines during lung and bronchus resection, during gastric, bariatric, small bowel, mesentery, colon and colorectal procedures, and cardiac surgery (for example, occlusion of the left atrial appendage during open chest procedures. Veritas Collagen Matrix (Dry) minimizes tissue attachment to the device in case of direct contact with viscera. Mark N. Milliken -torative, Division of Ge and Neurologic 510(k) Number K0830 Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...