ZUMA, MODEL 55-XXXX/56-XXXX

K082926 · Seaspine, Inc. · OVD · Dec 22, 2008 · Orthopedic

Device Facts

Record IDK082926
Device NameZUMA, MODEL 55-XXXX/56-XXXX
ApplicantSeaspine, Inc.
Product CodeOVD · Orthopedic
Decision DateDec 22, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

When used as an intervertebral body fusion device, the Zuma System is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autograft. Zuma is a stand alone system intended to be used with the bone screws provided and requires no additional supplementary fixation systems. When used as a Vertebral Body Replacement Device, the Zuma System is intended for use in the thoracolumbar spine (T1 to L5) to replace a collapsed, diseased, damaged or unstable complete or partial vertebral body due to tumor or trauma/fracture, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Zuma System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column, even in the absence of fusion for a prolonged period. Additionally, Zuma is intended for use with bone graft.

Device Story

Zuma is an implantable spinal device; constructed from PEEK and titanium; includes radiographic markers. Device features open central area for autograft bone graft material; secured to vertebral bodies via provided bone screws. Used in spinal fusion or vertebral body replacement procedures; intended for stand-alone use without supplementary fixation. Implanted by surgeons in clinical settings to restore spinal height, achieve decompression, and maintain biomechanical integrity of the spinal column. Benefits include stabilization of diseased or damaged vertebral segments and promotion of fusion.

Clinical Evidence

Bench testing only; no clinical data required.

Technological Characteristics

Implantable spinal device; materials: PEEK and titanium; includes radiographic markers; secured with bone screws; open central geometry for bone graft; stand-alone system.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1), including those with Grade 1 spondylolisthesis/retrolisthesis or prior non-fusion surgery, following six months of failed non-operative treatment. Also indicated for thoracolumbar (T1-L5) vertebral body replacement in patients with collapsed, diseased, damaged, or unstable vertebrae due to tumor or trauma/fracture.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary | Company Name: | SeaSpine, Inc.<br>2302 La Mirada Drive<br>Vista, CA 92081 | |-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Jeff Brittan<br>Senior Project Engineer<br>E-mail: jbrittan@seaspine.com<br>Phone: (760) 727-8399, Fax: (760) 727-8809 | | Date Prepared: | September 30, 2008 | | Trade Name: | Zuma™ | | Common Name: | Interbody Fusion Device<br>Vertebral Body Replacement Device | | Classification Name: | Intervertebral Fusion Device with Bone Graft, Lumbar<br>21 CFR 888.3080, Product Code MAX, Class II<br>Spinal Intervertebral Body Fixation Orthosis<br>21 CFR 888.3060, Product Code MQP, Class II | | Review Panel: | Orthopedic | | Device Description: | Zuma is an implantable spinal device made from polyetheretherketone<br>(PEEK) and titanium with markers for radiographic visualization; it is<br>secured to vertebral bodies with bone screws. The device has an open<br>central area for receiving bone graft material and is offered in a variety<br>of sizes and geometries to accommodate variations in pathology and<br>patient anatomy. | | Intended Use: | When used as an intervertebral body fusion device, the Zuma System is<br>intended for spinal fusion procedures at one or two contiguous levels<br>(L2-S1) in skeletally mature patients with degenerative disc disease<br>(DDD). DDD is defined as back pain of discogenic origin with<br>degeneration of the disc confirmed by history and radiographic studies.<br>DDD patients may also have up to Grade 1 spondylolisthesis or<br>retrolisthesis at the involved level(s). These patients may have had a<br>previous non-fusion spinal surgery at the involved spinal level(s). These<br>patients should have had six months of non-operative treatment. The<br>device is intended to be used with autograft. Zuma is a stand alone<br>system intended to be used with the bone screws provided and requires<br>no additional supplementary fixation systems. | | | collapsed, diseased, damaged or unstable complete or partial vertebral<br>body due to tumor or trauma/fracture, to achieve anterior<br>decompression of the spinal cord and neural tissues, and to restore the<br>height of a collapsed vertebral body. The Zuma System is designed to<br>restore the biomechanical integrity of the anterior, middle and posterior<br>spinal column, even in the absence of fusion for a prolonged period.<br>Additionally, Zuma is intended for use with bone graft. | | Substantial<br>Equivalence: | Zuma was shown to be substantially equivalent to predicate<br>devices through comparison in areas including intended use, design,<br>materials, and function. | | Performance Data: | Mechanical testing results indicated that Zuma possessed appropriate<br>properties for its intended use and is substantially equivalent to<br>predicate devices. Clinical data was not required for this device. | When used as a Vertebral Body Replacement Device, the Zuma System is intended for use in the thoracolumbar spine (T1 to L5) to replace a page 1 of 2 {1}------------------------------------------------ {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the bird symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 SFP 1.2 201 SeaSpine, Inc. % Mr. Jeff Brittan Senior Project Engineer 2302 La Mirada Drive Vista, California 92084 Re: K082926 > Trade/Device Name: Zuma™M Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD, MOP Dated: September 30, 2008 Received: October 1, 2008 Dear Mr. Brittan: This letter corrects our substantially equivalent letter of December 22, 2008. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other {3}------------------------------------------------ Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mark A. Milke Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use K082926 510(k) Number (if known): Device Name: Zuma™ Indications for Use: When used as an intervertebral body fusion device, the Zuma System is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autograft. Zuma is a stand alone system intended to be used with the bone screws provided and requires no additional supplementary fixation systems. When used as a Vertebral Body Replacement Device, the Zuma System is intended for use in the thoracolumbar spine (T1 to L5) to replace a collapsed, damaged or unstable complete or partial vertebral body due to tumor or trauma/fracture, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Zuma System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column, even in the absence of fusion for a prolonged period. Additionally, Zuma is intended for use with bone graft. Prescription Use ਮ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Eva Daniel Krone for M (Divis: 198-44) **Division of General, Restoration** Neurological Devices 510 Number K0882926
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...