STANDARD IMAGING IMSURE BRACHY QA SOFTWARE, MODEL 91331
K082773 · Standard Imaging, Inc. · IYE · Nov 19, 2008 · Radiology
Device Facts
Record ID
K082773
Device Name
STANDARD IMAGING IMSURE BRACHY QA SOFTWARE, MODEL 91331
Applicant
Standard Imaging, Inc.
Product Code
IYE · Radiology
Decision Date
Nov 19, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
IMSure Brachy QA Software is a stand alone software application product intended for use as a quality assurance tool to verify brachytherapy treatment plans developed on a radiation therapy treatment planning system with the appropriate transfer format. It may also be used as a segregated standalone application in the IMSure QA Software suite of software products. This product is intended for use by trained medical physicians, or dosimetrists. The calculation results must be evaluated by qualified personnel before a patient treatment. It is the responsibility of the medical physician or dosimetrist to determine whether the dosimetric accuracy is adequate for a particular patient. Dose modeling of a source is based on the AAPM TG-43 formalism, and may be adjusted by a qualified user to match measured or published results. IMSure Brachy QA Software does not control any radiation delivery devices and does not allow the export of calculated information.
Device Story
Standalone software for brachytherapy treatment plan verification; imports DICOM-RT or vendor-specific files containing applicator/catheter geometry, source data, and TPS-predicted dose. Operates via superposition of point/line sources using AAPM TG-43 formalism; allows user adjustment of source parameters to match measured/published data. Provides 3D visualization of applicators, sources, and calculation points. Used by physicians/dosimetrists in clinical settings to compare independent dose calculations against TPS output. Does not control radiation delivery devices; does not export calculated data. Benefits patient by providing independent dosimetric verification to ensure treatment plan accuracy before radiation delivery.
Clinical Evidence
No clinical data. Evidence consists of bench testing, verification, and validation of functional modules (Import, QA, Source Library, User Management) and successful evaluation at five beta sites (UCSD Hospital, Grant Methodist Hospital, Memorial Hospital, Edward Hospital, Texas Oncology Hospital).
Technological Characteristics
Standalone software application; runs on Windows 2000/XP; requires Pentium III, 256MB RAM, 100MB disk space, and OpenGL 1.1 compatible video card. Dose modeling based on AAPM TG-43 formalism. Complies with IEC 60601-1-4 (programmable medical systems).
Indications for Use
Indicated for use by trained medical physicians or dosimetrists as a quality assurance tool to verify brachytherapy treatment plans (HDR or LDR) developed on a radiation therapy treatment planning system.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Standard Imaging (Prodigm), IMSure QA Software (K031975)
Oncology Data Systems, MU Check v7.0 (Brachy Check) (K061152)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Standard Imaging. The logo consists of the word "STANDARDIMAGING" in a stylized font. To the right of the text is a circular graphic with a jagged, abstract design, possibly representing a burst or explosion. The overall design is simple and modern.
September 17, 2008
Ko82773
# NOV 1 9 2008
| Subject: | 510(k) Summary of Safety and Effectiveness Information for the<br>Standard Imaging IMSure Brachy QA Software | | |
|----------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|--|
| Proprietary Name: | Standard Imaging IMSure Brachy QA Software | | |
| Common Name: | Dose Validation Software | | |
| Classification Name: | Primary: | Medical Charged-Particle Radiation Therapy System<br>(accessory to) | |
| | Secondary: | Remote Controlled Radionuclide Applicator System<br>(accessory to)<br>Radionuclide Brachytherapy Source<br>(accessory to) | |
| Classification: | Primary: | Class II - 21CFR892.5050 - LHN, IYE | |
| | Secondary: | Class II - 21CFR892.5700 - JAQ<br>Class II - 21CFR892.5730 - KXK | |
| Panel: | Radiology | | |
| Predicate Devices: | Primary: | Standard Imaging (Prodigm), IMSure QA Software<br>510(k) Number K031975 | |
| | Secondary: | Oncology Data Systems, MU Check v7.0 (Brachy Check-<br>510(k) Number K061152 | |
| Contact Person: | Raymond Riddle, PE, RAC<br>Chief Regulatory Officer | | |
The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 1992.
The Standard Imaging IMSure Brachy QA Software was designed to comply with the applicable portions of the following voluntary standard:
- IEC 60601-1-4 (Edition 1.1 2000-04) Collateral standard for programmable medical . systems
/3
STANDARDING, INC. 3120 Deming Way, Middleton, W 53562-1461 800-261-0446 608-31-025 PH 608-831-202 FAX www.standardimaging.com
Assurance is confidence™
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Standard Imaging. The text "STANDARD IMAGING" is displayed in a stylized font, with the word "STANDARD" in a thinner font than "IMAGING". To the right of the text is a circular graphic with a burst-like design.
IMSure Brachy QA Software is a stand alone software application product intended for use as a quality assurance tool to verify brachytherapy treatment plans developed on a radiation therapy treatment planning system with the appropriate transfer format. It may also be used as a segregated standalone application in the IMSure QA Software suite of software products.
This product is intended for use by trained medical physicians, or dosimetrists. The calculation results must be evaluated by qualified personnel before a patient treatment. It is the responsibility of the medical physician or dosimetrist to determine whether the dosimetric accuracy is adequate for a particular patient.
Dose modeling of a source is based on the AAPM TG-43 formalism, and may be adjusted by a qualified user to match measured or published results. IMSure Brachy QA Software does not control any radiation delivery devices and does not allow the export of calculated information.
The Standard Imaging IMSure Brachy QA Software has been verified and validated at Standard Imaging, and addressed the following functional areas:
- Installation .
- . Import Tool Module
- . User Preferences Module
- . QA Module
- Source Library Module .
- User Management Module .
- . Simple 1 strand +xyz to -xyz.
- . Dual IR-192 dwl
- 24 channels .
- . Xoft
- . 4 I125 point
- Line 1D aniso .
- Line scalar aniso �
Additionally, the IMSure Brachy QA Software was successfully cvaluated by the following beta sites:
- . UCSD Hospital, San Diego, CA
- . Grant Methodist Hospital, Columbus OH
- Memorial Hospital, South Bend, IN .
- Edward Hospital, Naperville, IL .
- . Texas Oncology Hospital, Klabzuba, TX
The Standard Imaging IMSure Brachy QA Software has met its predetermined design specifications, risk analysis and validation objectives.
14
STANDARDIMAGING, INC. 3120 Deming Way, Middleton, WI 53562-1461 800-261-4466 608-831-0025 PH 608-831-2202 FAX www.standardimaging.com
Assurance is confidence™
{2}------------------------------------------------
# General Information
| Device Name - Proprietary: | IMSure Brachy QA Software - During development, the name has evolved, resulting in<br>many of the documents in this 510(k) having slightly different names. Upon 510(k)<br>clearance, all documents, including manuals, advertisements, labeling and software<br>supplied to the marketplace and/or users will use the name IMSure Brachy QA Software. |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name - Common: | Dose Validation Software |
| Device Name - Classification(s): | |
| Primary: | Medical Charged-Particle Radiation Therapy System (accessory to) |
| Secondary: | Remote Controlled Radionuclide Applicator System (accessory to)<br>Radionuclide Brachytherapy Source (accessory to) |
| Device Panel: | Radiology |
| Device Classification(s): | |
| Primary: | Class II - 21CFR892.5050 -- LHN, IYE |
| Secondary: | Class II - 21CFR892.5700 - JAQ, Class II - 21CFR892.5730 - KXK |
| Predicate Devices: | |
| Primary: | Standard Imaging (Prodigm), IMSure QA Software<br>510(k) Number K031975 |
| Secondary: | Oncology Data Systems, MU Check v7.0 (Brachy Check)<br>510(k) Number K061152 |
| Performance Standards: | To the best of Standard Imaging's knowledge, performance standards have not been<br>promulgated by FDA for this device. |
| Establishment Registration Number: | 2184007 |
| Owner/Operator Number: | 9003193 |
| Facility Information: | Standard Imaging, Inc.<br>3120 Deming Way<br>Middleton, WI 53562<br>Telephone: (608) 831-0025<br>Fax: (608) 831-2202<br>Webpage: www.standardimaging.com |
{3}------------------------------------------------
### Device Background
Standard Imaging acquired the IMSure QA Software from Prodigm, Inc. in 2005. Since that time the company has made minor incremental improvements to the product. Since Standard Imaging had a strong market presence with brachytherapy quality assurance products, a simple standalone software product in the IMSure mold was an obvious development choice.
## Device Description
IMSurc Brachy QA Software is a stand alone software application product intended for use as a quality assurance tool to verify brachytherapy treatment plans developed on a radiation therapy treatment planning system with the appropriate transfer format. It may also be used as a segregated standalone application in the IMSure QA Software suite of software products.
This product is intended for use by trained medical physicians, or dosimetrists. The calculation results must be evaluated by qualified personnel before a patient treatment. It is the responsibility of the medical physician or dosimetrist to determine whether the dosimetric accuracy is adequate for a particular patient.
IMSure Brachy QA Software independently computes a modeled dose that would be delivered by a high dosc rate (HDR) or low dose (LDR) brachytherapy system to a patient and compares it to the dose predicted by a primary treatment planning system. IMSure Brachy QA Software imports a file produced by a primary HDR or LDR treatment planning system (TPS), in the format of an industry standard Dicom-RT™ or vendor specific file, which contains information about a treatment. The files contain information about applicators or catheters and the associated source information in each catheter, such as source type, source strength, source location and source duration. The files may also contain information about specific calculation points and the dose predicted by the primary planning system, as well as patient specific information. The dose computation algorithm used is a superposition of point or line sources, incorporating 3-D geometrical features of the source construction, as well as radiological features of the source composition. Dosc modeling of a source is based on the AAPM TG-43 formalism, and may be adjusted by a qualified user to match measured or published results.
After importing a TPS plan, a user may edit the information, adding or modifying source positions, durations (or dwell times), type, or activity strength. Calculation point information may be edited as well. A 3-D view of the applicators, source positions, and calculation point positions is provided. A paper or electronic record can be stored including the final dose computation for each calculation point compared to the dose computed by the TPS, as well as relevant patient information for long term documentation. IMSure Brachy QA Software does not control any radiation delivery devices and does not allow the export of calculated information.
IMSure Brachy QA Software is provided to the customer on a CD. It requires the Microsoft Windows Operating System 2000 with service pack 2 or better, or XP. Computer system requirements include Pentium III or equivalent, a minimum of 256 MB RAM and 100 MB available hard drive space. Display requirements include 1024 x 768 minimum resolution and an OpenGL compatible video card meeting OpenGL 1.1 specifications.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The caduceus is depicted in black, and the text is also in black, set against a white background.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## Raymond T. Riddle, PE, RAC Chief Regulatory Officer Standard Imaging, Inc. 3120 Deming Way MIDDLETON WI 53562-1461
## NOV 1 9 2008
Re: K082773
. Trade/Device Name: IMSure Brachy QA Software Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: September 18, 2008 Received: September 22, 2008
Dear Mr. Riddle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA); it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Arque In Whang
Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Usc
K08z773
510(k) Number (if known):
Device Name:
IMSure Brachy QA Software
Indications For Use:
IMSure Brachy QA Software is a stand alone software application product intended for use as a quality assurance tool to verify brachytherapy treatment plans developed on a radiation therapy treatment planning system with the appropriate transfer format. It may also be used as a segregated standalone application in the IMSure QA Software suite of software products.
This product is intended for use by trained medical physicians, or dosimetrists. The calculation results must be evaluated by qualified personnel before a patient treatment. It is the responsibility of the medical physician or dosimetrist to determine whether the dosimctric accuracy is adequate for a particular patient.
Dose modeling of a source is based on the AAPM TG-43 formalism, and may be adjusted by a qualified user to match measured or published results. IMSure Brachy QA Software does not control any radiation delivery devices and does not allow the export of calculated information.
Prescription Use X (Part 21 CFR 801 Subpart D) and/or
Over-the-Counter-Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Helene
ivision of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Page 1 of 1
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.