← Product Code [LHI](/productcode/LHI) · K082752

# QUICKPIN (K082752)

_Laboratorios Grifols, S.A. · LHI · Sep 29, 2008 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K082752

## Device Facts

- **Applicant:** Laboratorios Grifols, S.A.
- **Product Code:** [LHI](/productcode/LHI.md)
- **Decision Date:** Sep 29, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

QUICKPIN is a Luer-lock spike used in manual or automated pharmacy compounding for addition and/or extraction of IV substances, including antineoplastics and substances for chemotherapy, from rubber-stoppered containers including multi-dose vials.

## Device Story

QUICKPIN is a Luer-lock spike device used in pharmacy settings for the addition or extraction of IV substances, including chemotherapy agents, from rubber-stoppered vials. It features a 0.2-micron hydrophobic air filter designed to minimize aerosol formation during the preparation and dispensing process. The device is intended for use by trained healthcare personnel in manual or automated compounding workflows. By facilitating safe access to multi-dose vials, it assists in the preparation of IV medications. The device is provided sterile and is for single-use only.

## Clinical Evidence

Bench testing only. Functional laboratory testing under foreseeable operating conditions verified correct operation. Materials underwent chemical and biological testing per ISO 10993-1 and USP standards.

## Technological Characteristics

Luer-lock spike; materials: ABS (spike body), PP (cap), PVC (filter housing), PTFE (filter). Features 0.2-micron hydrophobic air filter. Conforms to ISO 8536-4 (except spike dimensions). Sterile, non-pyrogenic, single-use. No software or electronic components.

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- Chemo-Aide Dispensing Pin ([K003730](/device/K003730.md))
- Fleboset Multiple ([K040456](/device/K040456.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

SEP 2 9 2008
  

K082752 (P. 1 of 4)

| GRIFOLS     | TECHNICAL EVALUATION DOCUMENTATION |
|-------------|------------------------------------|
| # Document: | TED-QUICKPIN-05                    |

SECTION 5 -- QUICKPIN: 510(k) SUMMARY#### INTENDED USE:

QUICKPIN is a Luer-lock spike used in manual or automated pharmacy compoundinq for addition and/or extraction of IV substances, including antineoplastics and substances for chemotherapy, from rubber-stoppered containers including multi-dose vials.

#### SUMMARY OF COMPARISON WITH PREDICATE DEVICE:

In the establishment of substantial equivalence, QUICKPIN is compared with another vial access device (Chemo-Aide Dispensing Pin – Baxter Healthcare Corporation) and with an IV fluid transfer set previously marketed by Laboratorios Grifols (Fleboset Multiple). The following fable summarizes the similarities of the principal technological characteristics and features of both predicate and new devices.

| #  | Characteristic / Feature                                                | QUICKPIN                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | PREDICATES                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|----|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|    |                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | BAXTER<br>Chemo Aide<br>Dispensing Pin<br>K003730                                                                                                                                                               | LAB.GRIFOLS<br>Fleboset Multiple<br>K040456                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 1. | Intended use / Claims                                                   | QUICKPIN is a Luer-lock<br>spike used in manual or<br>automated pharmacy<br>compounding for addition<br>and/or extraction of IV<br>substances, including<br>antineoplastics and<br>substances for chemotherapy,<br>from rubber-stoppered<br>containers including multi-<br>dose vials.                                                                                                                                                                                                                                                    | The Baxter CHEMO-AIDE<br>Dispensing Pin is intended<br>for use in the preparation<br>and dispensing of<br>chemotherapeutic<br>medications from rubber-<br>stoppered vials.                                      | FLEBOSET MULTIPLE<br>is an ancillary device<br>used as fluid pathway<br>through which<br>substances from 6<br>glass source flasks<br>containing the same<br>solution may be<br>continuously delivered<br>for: (a) Pharmacy<br>compounding, when<br>used in conjunction with<br>the GRI-FILL 2.0<br>pharmacy compounding<br>device and associated<br>transfer sets, and (b)<br>I.V. administration,<br>when used in<br>conjunction with a<br>gravity or pump infusion<br>set to channel the<br>solution from the source<br>containers to the<br>infusion set.<br>The device should not<br>be used with lipids.                                                              |
| #  | Characteristic /<br>Feature                                             | QUICKPIN                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | PREDICATES<br>BAXTER<br>Chemo Aide<br>Dispensing Pin<br>K003730                                                                                                                                                 | LAB.GRIFOLS<br>Fleboset Multiple<br>K040456                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 2. | Indications for use                                                     | QUICKPIN is a Luer-lock spike used in manual or automated pharmacy compounding for addition and / or extraction of IV substances, including antineoplastics and substances for chemotherapy, from rubber-stoppered containers including multidose vials. It is equipped with a 0.2 micron hydrophobic air-filter that minimizes the formation of aerosols when preparing and dispensing the substances.<br>This device is intended to be used by trained healthcare personnel. It is restricted to sale by or on the order of a physician | The proposed Baxter CHEMO-AIDE Dispensing Pin is intended for use in the preparation and dispensing chemotherapeutic medications from rubber stoppered vials.                                                   | Fleboset multiple is an ancillary device used as fluid pathway through which substances from 6 glass source flasks containing the same solution may be continuously delivered for:<br>(a) Pharmacy compounding, when used in conjunction with the GRI-FILL 2.0 pharmacy compounding device and associated transfer sets, and<br>(b) I.V. administration when used in conjunction with a gravity or pump infusion set to channel the solution from the source containers to the infusion set.<br>The device should not be used with lipids.<br>This device is intended to be used by trained health-care personnel. It is restricted to sale by or on order of a physician. |
| 3. | Technological<br>features:<br>- Sterilization<br>- Contact with patient | Ethylene Oxide<br>NO - Intended for use in manual or automated pharmacy compounding of IV substances including                                                                                                                                                                                                                                                                                                                                                                                                                            | Radiation - Gamma<br>NO - Intended for use by the pharmacist in the preparation and dispensing of chemotherapeutic                                                                                              | Ethylene Oxide<br>YES - May be used on-line with patient upstream of the gravity administration set or                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| #  | Characteristic /<br>Feature                                             | QUICKPIN                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | PREDICATES<br>BAXTER<br>Chemo Aide<br>Dispensing Pin<br>K003730                                                                                                                                                 | LAB.GRIFOLS<br>Fleboset Multiple<br>K040456                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|    |                                                                         | antineoplastics and<br>substances for chemotherapy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | medications.                                                                                                                                                                                                    | infusion pump.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|    | Filter                                                                  | YES - 0.2 micron hydrophobic<br>air-filter to minimize the<br>formation of aerosols during<br>preparation and<br>administration of the<br>substances.                                                                                                                                                                                                                                                                                                                                                                                     | YES - 0.22 micron<br>hydrophopic / oleophobic<br>air-venting filter that<br>minimizes potential for<br>aerosolization of<br>chemotherapeutic<br>medications during<br>reconstitution and<br>dispensing process. | YES - 1.2 micron air-<br>venting filter.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| 4. | Physical, Mechanical<br>and Biological<br>Specifications                | Sterile / Non-pyrogenic /<br>Single-use only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Sterile / Non pyrogenic /<br>Single-use only                                                                                                                                                                    | Sterile / Non pyrogenic<br>Single-use only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| 5. | Materials                                                               | Spike body: ABS (Acrylonitrile<br>butadiene styrene)<br>Luer Protective cap: PP<br>(Polypropylene)<br>Filter housing: PVC<br>Filter: PTFE (Polytetrafluoro-<br>ethylene)                                                                                                                                                                                                                                                                                                                                                                  | Spike body: ABS + PE<br>(Acrylonitrile butadiene<br>styrene and polyester)<br>Filter housing: ABS + PE<br>(Acrylonitrile butadiene<br>styrene and polyester)<br>Filter: PVDF (Polyvinyldene<br>fluoride)        | Spike body: ABS<br>(Acrylonitrile butadiene<br>styrene)<br>Spikeable twist-off<br>connection: PVC<br>(Polyvinylchloride)<br>Filter housing: PP<br>Filter: Acrylic copolymer<br>+ PA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| 6. | Conformance<br>standards                                                | The spike conforms to<br>applicable requirements of<br>ISO 8536-4 except regarding<br>spike dimensions which have<br>been adapted specifically for<br>small container access.<br><br>Materials used meet the<br>requirements of USP<br>physicochemical tests for<br>plastics and biological tests<br>outlined in standard ISO<br>10993-1.                                                                                                                                                                                                 | Materials used meet the<br>requirements of USP<br>physicochemical tests for<br>plastics and biological tests<br>outlined in standard ISO<br>10993-1.                                                            | The spike conforms to<br>applicable requirements<br>of ISO 8536-4.<br><br>Materials used meet the<br>requirements of USP<br>physicochemical tests<br>for plastics and<br>biological tests outlined<br>in standard ISO 10993-<br>1.                                                                                                                                                                                                                                                                                                                                                                                                                                         |

{1}------------------------------------------------

K482752 (P. 20P4)

GRIFOLS

# Document:

DOCUMENTATION

TECHNICAL EVALÚATION

ted-quickpin-05

# SECTION 5 - QUICKPIN: 510(k) SUMMARY

{2}------------------------------------------------

K482052 (P.3084)

### GRIFOLS

# Document:

DOCUMENTATION

TECHNICAL EVALUATION

TED-QUICKPIN-05

## SECTION 5- QUICKPIN: 510(k) SUMMARY

From the above table, it can be established that the new device and the predicate device Baxter Chemo-Aide Dispensing Pin have very similar intended uses and indications. Also, the materials used in the construction of the proposed spike and those used in spike component of the predicate device Fleboset Multiple are identical. Some differences may be noted in the

{3}------------------------------------------------

|                                      | K482752 (P.4 of 4)   |                 |
|--------------------------------------|----------------------|-----------------|
| GRIFOLS                              | TECHNICAL EVALUATION | # Document:     |
|                                      | DOCUMENTATION        | TED-QUICKPIN-05 |
| SECTION 5 - QUICKPIN: 510(k) SUMMARY |                      |                 |

dimensions and type of materials used in the filter component and air-inlet on the spike. These differences have been addressed in the different bench tests performed on the proposed QUICKPIN device.

### SUMMARY DISCUSSION OF NON-CLINICAL DATA:

All materials used in the construction of QUICKPIN have been subject to chemical and biological testing in accordance with the applicable requirements taking account of its intended use.

Functional laboratory testing performed in foreseeable operating conditions showed correct operation of the device as per its intended use,

#### CONCLUSIONS:

We believe the intended use, the indications for use, the functionality and the operation of QUICKPIN and the CHEMO-AIDE predicate device as access devices for rubber stoppered containers are essentially the same. Also, the materials used in the construction the proposed device and those used in the spike component of the FLEBOSET MULTIPLE predicate device are identical. Technological differences including the use of different filter materials and different dimensions on the air-inlet channel on the spike have been addrt fined and verified by bench-testing to have no adverse influence on the safety and performance of the proposed device. Hence, substantial equivalence of QUICKPIN with the legally marketed devices may be established.

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 2 9 2008

Laboratorios Grifols, S.A. C/O Mr. Norbert Stuiber Responsible Third Party Officer TUV SUD America, Incorporated 1775 Old Highway 8 N.W. New Brighton, Minnesota 55112-1891

Re: K082752

Trade/Device Name: QUICKPIN Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: LHI Dated: September 17, 2008 Received: September 19, 2008

Dear Mr. Stuiber:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

#### Page 2 - Mr. Stuiber

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

GRIFOLS

TECHNICAL EVALUATION DOCUMENTATION

# Document: TED-QUICKPIN-04

### SECTION 04 - QUICKPIN:

### INDICATIONS FOR USE STATEMENT

#### PREMARKET NOTIFICATION INDICATIONS FOR USE STATEMENT (as required by ODE for all 510(k) received after Jan. 1, 1996)

510(k) Number:

K082752

Device Name:

QUICKPIN

Indications for Use:

QUICKPIN is a Luer-lock spike used in manual or automated pharmacy compounding for addition and / or extraction of IV substances, including antineoplastics and substances for chemotherapy, from rubber-stoppered containers including multidose vials. It is equipped with a 0.2 micron hydrophobic air-filter that minimizes the formation of aerosols when preparing and dispensing the substances.

This device is intended to be used by trained healthcare personnel. It is restricted to sale by or on the order of a physician.

(Do not write below this line. Continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use <span style="text-decoration: underline;">✓</span><br>(21 CFR 801 Subpart D) | OR | Over-The-Counter Use <span style="text-decoration: underline;">____</span><br>(21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------|----|------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------|----|------------------------------------------------------------------------------------------------------|

  

__________________________
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices

| 510(k) Number: | <span style="text-decoration: underline;">K982752</span> |
|----------------|----------------------------------------------------------|
|----------------|----------------------------------------------------------|

9

---

**Source:** [https://fda.innolitics.com/device/K082752](https://fda.innolitics.com/device/K082752)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com