THROMBO TEK PSE, MODEL 90-480

K082631 · R2 Diagnostics, Inc. · GGP · Dec 17, 2008 · Hematology

Device Facts

Record IDK082631
Device NameTHROMBO TEK PSE, MODEL 90-480
ApplicantR2 Diagnostics, Inc.
Product CodeGGP · Hematology
Decision DateDec 17, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7290
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K082631 · Dec 17, 2008THROMBO TEK PSE, MODEL 90-480R2 Diagnostics, Inc.Healthy donor plasma samples; Patient plasma samplesThe device performance was evaluated using clinical samples from healthy donors to establish a normal reference range and from patients to perform a method comparison against a predicate device.Normal reference range; Method comparison; Clinical samples

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Observational study of healthy donors120 healthy donors; Sample Size: 120Not applicable for this studyProtein S activity reference range
Method comparison study174 patient samples; Sample Size: 174; Number of Sites: 2Commercially available Protein S activity assayCorrelation coefficient, slope, and intercept

Indications for Use

ThromboTek PSe is intended for the quantitative determination of functional Protein S activity in human plasma.

Device Story

ThromboTek PSe is a tissue factor pathway-based clotting assay used to quantify Protein S activity in citrated human plasma. The assay utilizes a lyophilized activator (rabbit thromboplastin, calcium, buffer, stabilizers), activated Protein C (aPC), and Protein S-deficient human plasma. Principle: aPC degrades activated Factor Va and VIIIa in the presence of Protein S, reducing prothrombinase complex efficiency. When aPC is in excess, the clotting time is proportional to the patient's Protein S concentration. The assay is performed on clinical coagulation analyzers (e.g., AMAX 200, STart4) by laboratory technicians. Results are used by clinicians to identify inherited or acquired Protein S deficiency, aiding in the diagnosis of coagulation disorders.

Clinical Evidence

No clinical trials were required. Performance was established via bench testing, including precision (CV 4.9-9.2%), linearity (10-156%), analytical sensitivity (1%), and interference studies (hemolysis, icterus, lipemia, heparin). Method comparison against a commercially available assay (n=174) showed a correlation coefficient of 0.895 (95% CI, 0.875-0.912).

Technological Characteristics

Tissue factor pathway-based clotting assay. Reagents: lyophilized rabbit thromboplastin, activated Protein C, Protein S-depleted human plasma, imidazole buffered saline. Form factor: lyophilized kit components requiring reconstitution. Connectivity: compatible with standard coagulation analyzers (e.g., AMAX 200, STA4).

Indications for Use

Indicated for the quantitative determination of protein S activity in human plasma to assess deficiencies associated with an increased risk of thrombosis.

Predicate Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K082631 B. Purpose for Submission: Clearance of new assay C. Measurand: Protein S D. Type of Test: Clotting E. Applicant: R2 Diagnostics F. Proprietary and Established Names: ThromboTek PSe G. Regulatory Information: 1. Regulation section: 21 CFR 864.7290 2. Classification: II 3. Product code: GGP 4. Panel: 81 Hematology {1} 2 H. Intended Use: 1. Intended use(s): The ThromboTek PSe is a quantitative assay for the determination of protein S activity in human plasma by clotting. 2. Indication(s) for use: Deficiencies of protein S are associated with an increased risk of thrombosis. 3. Special conditions for use statement(s): 4. Special instrument requirements: I. Device Description: ThromboTek PSe is a tissue factor pathway based clotting assay. The assay activator is a lyophilized preparation incorporating rabbit thromboplastin, calcium, buffer, and stabilizers. The remaining components of the assay are lyophilized activated Protein C, lyophilized human plasma depleted of Protein S, imidazole buffered saline, and deionized water containing a preservative for reconstitution of the lyophilized components. J. Substantial Equivalence Information: 1. Predicate device name(s): Diagnostica Stago StaClot® Protein S 2. Predicate 510(k) number(s): K913424 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Quantitative determination of | Same | {2} | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | functional Protein S activity | | | Sample requirement | Citrated plasma | same | | Assay principle | Clotting | same | | Reagents | Activator, aPC, Protein S deficient plasma | same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Linearity | 1- 156% | 10 -105% | | Sensitivity | 1% | 8% | | Reagents | Buffered saline and reconstitution solution included in assay kit | Owen-Koller buffer and calcium chloride purchased separately, reconstitution grade water provided by user | # K. Standard/Guidance Document Referenced (if applicable): CLSI/NCCLS EP5-A2, "Evaluation of Precision Performance of Quantitative Measurement Methods Approved Guideline- Second Edition" CLSI/NCCLS EP15-A2, "User Verification of Performance for Precision and Trueness: Approved Guideline-Second Edition", 2005 H21-A5, "Collection, Transport and Processing of Blood Samples for Testing Plasma-based Coagulation Assays and Molecular Hemostasis Assays; Approved Guideline-Fifth Edition", 2008 # L. Test Principle: The Tissue Factor pathway of coagulation is activated when endogenous FVII or activated FVIIa in the sample plasma is mixed with tissue factor in the presence of calcium. This TF/FVIIa complex activates Factor X (FX) to FXa. Factor Xa complexes with activated Factor V (FVa) on a lipid surface in the presence of calcium to form a prothrombinase complex. The prothrombinase complex activates prothrombin to thrombin. Thrombin cleaves fibrinogen to fibrin leading to formation of a solid clot. Activated Protein C (aPC) in the presence of its cofactor Protein S degrades FVa, reducing the efficiency of the prothrombinase complex. When there is an excess of aPC, the concentration of Protein S in the patient sample is rate limiting. Thus, the {3} time for the patient plasma to clot is proportional to the concentration of the PS in the sample. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: Precision studies were conducted in accordance with CLSI/NCCLS Guideline EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition. Three lots of the assay kit was used to determine precision in a two run per day, twenty day exercise on an AMAX 200 analyzer in mechanical mode using a normal and abnormal plasma. | Plasma | Repeatability | Total | | --- | --- | --- | | Normal | 4.9% | 5.7% | | Abnormal | 7.8% | 9.2% | #### b. Linearity/assay reportable range: Three lots of linearity testing using ISTH/SCC plasma demonstrated ThromboTek PSe linearity of 10- 156% Protein S. #### c. Traceability, Stability, Expected values (controls, calibrators, or methods): Assay recommends the use of pooled normal plasma or purchased reference plasma for calibration, and the use of purchased quality control material for QC #### d. Detection limit: Lower limit of detection was determined by testing 30 replicates of the assay buffer on 3 lots of the ThromboTek PSe assay. The mean, standard deviation (SD), and mean + 3SD of the clot times were calculated for all three lots. The LoD was calculated as 1%. #### e. Analytical specificity: Hemoglobin, bilirubin, lipid, and heparin interference was tested by spiking the interferent into pooled normal plasma and preparing serial dilutions. The maximum concentration tolerated in the assay was defined as the highest concentration of interferent resulting in a <10% difference in the base result. The maximum concentrations were: 4 {4} | Interferent | Added Interferent | Maximum concentration Tested | Maximum tolerated concentration | | --- | --- | --- | --- | | Hemolysis | Hemoglobin | 500 mg/dL | 500 mg/dL | | Icterus | Unconjugated Bilirubin | 20 mg/dL | 20 mg/dL | | Lipemia | IntraLipid® | 2,000 mg triglycerides/dL | 2,000 mg triglycerides/dL | | Heparin | Heparin | 2.0 Units/mL | 1.0 units/mL | f. Assay cut-off: n/a 2. Comparison studies: a. Method comparison with predicate device: 174 patient samples were assayed on multiple lots of ThromboTek PSe. Data was collected at two sites, one using an Trinity Biotech AMAX 200 analyzer and the other a Diagnostica Stago STA4, and yielded correlation statistics of $y = 1.711x - 8.89$, $r = 0.895$ b. Matrix comparison: n/a 3. Clinical studies: a. Clinical Sensitivity: n/a b. Clinical specificity: n/a c. Other clinical supportive data (when a. and b. are not applicable): n/a 4. Clinical cut-off: n/a 5. Expected values/Reference range: {5} 120 normal donors (85 females, 35 males) were assayed on 3 kits of the ThromboTek PSe. Mean, and two standard deviation ranges were calculated for males (62 – 209), females (45 – 183), and combined male and female (47-193) **N. Proposed Labeling:** The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. **O. Conclusion:** The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 6
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...