The PAM™3000 system is intended to measure heart rate and respiration rate in adult patients, in a general care hospital environment including, nursing homes The system monitors presence or absence of a patient in bed (bed exit) The system also times the length of continuous patient motion or absence of patient motion
Device Story
PAM™3000 provides contact-less, wireless monitoring of adult patients in healthcare facilities. System inputs: UWB-based motion sensor detects thorax and heart movement. Components: Bed Sensor Panel (BSP) placed under mattress; Repeater Base Stations (RBS) in hallways; Central Computer Station (CCS) with Central Base Station (CBS). System processes motion signals to derive heart rate, respiration rate, bed occupancy, and motion duration. Output: Data displayed on PC monitor via proprietary GUI at nursing stations. Alerts generated for clinical review. Benefits: Enables continuous, non-invasive patient monitoring, facilitating timely clinical response to patient status changes or bed exits.
Clinical Evidence
Bench validation and clinical validation testing performed. Results indicate the system accurately detects heart rate, respiration rate, and patient presence in bed according to specifications.
Technological Characteristics
UWB-based motion sensing; contact-less; wireless. Components: Bed Sensor Panel, Repeater Base Stations, Central Computer Station. Standards: IEC 60601-1, IEC 60601-1-2, FCC Part 15 (UWB), ANSI C63.4, IEC 62-2-27 (shock), IEC 68-2-6/34 (vibration).
Indications for Use
Indicated for adult patients in general care hospital environments and nursing homes for monitoring heart rate, respiration rate, bed occupancy, and patient motion.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
LG1™ Intelligent Medical Vigilance System (K052446)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/17 description: The image shows the logo for Wireless2000 RF & UWB Technologies Ltd. The logo features a stylized wireless signal icon above the word "WIRELESS2000" in a bold, sans-serif font. Below "WIRELESS2000" is the text "RF & UWB Technologies Ltd" in a smaller font.
K082626 Page 123
# 510(k) Summary
(per 21 CFR 807 92)
JAN 1 6 2009
I. Applicant Wireless 2000 RF & UWB Technologies Ltd 2421 Alpha Ave Burnaby, BC, V5C5L2
> Efraim Gavrilovich, President and CEO Contact Person Tel (604) 298-8471 Fax (604) 298-8470 Email egav@wireless2000 com
August 27, 2008 Date Prepared
#### II. Device Name
| Proprietary Name | PAM™ 3000 |
|----------------------|------------------------------------------------|
| Common/ Usual Name | Heart Rate and Respiration Rate Monitor |
| Classification Name | monitor, cardiac (incl cardiotachometer & rate |
| | alarm) monitor, breathing frequency |
| Regulation Number | 870 2300/868 2375 |
| Product Codes | DRT/BZQ |
| Classification | II |
| Classification Panel | Cardiovascular/Anesthesiology |
#### III. Predicate Device
The Wireless 2000 PAM™ 3000 is substantially equivalent to the EarlySense ES-16 System from EarlySense Ltd (K070375) and LG1TM Intelligent Medical Vigilance System from Hoana Medical Inc (K052446)
#### IV. Intended Use of the Device
The PAM™3000 system is intended to measure heart rate and respiration rate ın adult patients, in a general care hospital environment including, nursing homes The system monitors presence or absence of a patient in bed (bed exit) The system also times the length of continuous patient motion or absence of patient motion
#### V. Description of the Device
The PAM™3000 System is intended for contact-less and wreless monitoring of bed occupancy, motion and heart and respiration rates of adult patients 11 healthcare facilities The system utilizes a hıghly sensitive UWB-based motion sensor to detect the motion of heart and thorax The system consists of
{1}------------------------------------------------
K087626pg 2 of 3
Image /page/1/Picture/1 description: The image shows the logo for Wireless2000 RF & UWB Technologies Ltd. The logo features a stylized wireless signal icon above the word "WIRELESS2000" in a bold, sans-serif font. Below "WIRELESS2000" is the text "RF & UWB Technologies Ltd" in a smaller font size.
a Bed Sensor Panel (BSP), placed under the patient's mattress, Repeater Base Stations (RBS) that are typically mounted around the hallways in facility, and the Central Computer Station (CCS) with plugged-in Central Base Station (CBS) The signal containing the patient's heart rate and respiration rate data, bed occupancy data, patient motion, and the alerts generated by system are displayed on a PC monitor using Wireless 2000's proprietary Graphic User Interface (GUI) software The Central Computer Station (CCS) is typically located at the nursing station
#### VI. Summary of the Technical Characteristics
| Measurement Range | | |
|-------------------|--------------------------------|----------------------------------|
| Heart Rate | Normal 45-115 beats per minute | Elevated 85-170 beats per minute |
| Respiration Rate | Normal 3-30 breaths per minute | Elevated 3-50 breaths per minute |
#### VII. Safety & Effectiveness
The PAMTM3000 System was designed and tested using the following standards
- IEC 60601-1 (2nd Edition) Medical Electrical Equipment -- Part 1 . General Requirements for Safety
- IEC 60601-1-2 (3td Edition) Medical Electrical Equipment Part 1 o General Requirements For Safety 2 Collateral Standard Electromagnetic Compatibility
- FCC Part 15 207/209/517 Conducted limits / Radiated emission limits, o general requirements / Technical requirements for indoor UWB systems
- ANSI C63 4 2003 Methods of Measurement of Rad10-No1se ● Emissions from Low-Voltage Electrical and Electronic Equipment in the Range of 9 kHz to 40 GHz
- IEC 62-2- 27 Shock Peak acceleration 100 g (980 m/s2), Duration o 6 msec. Pulse shape half sine
- IEC 68-2-6 Sinusoidal Vibration Frequency range 10 to 500 Hz, o Acceleration amplitude 1 g (9 8 m/s2), Type and duration of endurance 10 sweep cycles in each axıs
- IEC 68-2-34 Random Vibration (Wide band) Frequency range 20 Hz -. 500 Hz, Acceleration spectral density 0 02 g2/Hz, Degree of reproducibility low, Duration of conditioning 9 minutes
{2}------------------------------------------------
K082626 pg 3of3
Image /page/2/Picture/1 description: The image shows the logo for Wireless2000 RF & UWB Technologies Ltd. The logo features a stylized wireless signal icon above the word "WIRELESS2000" in a bold, sans-serif font. Below the main text, the words "RF & UWB Technologies Ltd" are printed in a smaller font size.
In addition, the PAM™ 3000 System underwent bench validation testing and clinical validation testing The results of the testing indicate that the PAM™3000 performs according to its specifications and accurately detects respiration rate and heart rate of patients and patient presence in bed
Overall there are no known substantial differences between the PAM™ 3000 system defined in this 510(k) submission and the predicate devices The devices have the same intended uses and any differences in technological characteristics do not raise 1ssues of safety and effectiveness
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. The eagle is depicted in black and white, and the overall design is simple and clean.
Food and Drug Admınıstratıon 9200 Corporate Boulevard Rockville MD 20850
# JAN 1 6 2009
Wireless 2000 RF and UWB Technologies, Ltd c/o Mr Efraim Gavrilovich President & CEO 2421 Alpha Ave Burnaby V5C5L2 Canada
Re K082626 PAM 3000 Regulation Number 21 CFR 870 2300 Regulation Name Heart rate monitor, cardiac (incl cardiotachometer and rate alarm) Regulatory Class II Product Code DRT Dated December 16, 2008 Received December 22, 2008
Dear Mr Gavrilovich
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA) You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration
If your device is classified (see above) into etther class II (Special Controls) or class III (PMA), 11 may be subject to such additional controls Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 In addition, FDA may publish further announcements concerning your device in the Federal Register
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies You must
{4}------------------------------------------------
### Page 2 - Mr Efraim Gavrılovich
comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807), labeling (21 CFR Part 801), good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807 97) For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Brometric's (OSB's) DIvision of Postmarket Surveillance at 240-276-3474 For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464 You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http //www fda gov/cdrh/industry/support/index html
Sincerely yours,
Bram D Zuckerman, M D Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K082626 page 1 of 1
Image /page/5/Picture/1 description: The image shows the logo for Wireless2000 RF & UWB Technologies Ltd. The logo features a stylized wireless signal icon above the word "WIRELESS2000" in a bold, slightly distressed font. Below the company name, the text "RF & UWB Technologies Ltd" is printed in a smaller, sans-serif font. The overall design is simple and professional, conveying a sense of technological expertise.
# 4. Indication for Use Statement
510(k) Number (if known):
PAM™ 3000 Device Name:
### Indications for Use:
The PAM™3000 system is intended to measure heart rate and respiration rate in adult patients, in a general care hospital environment including, nursing homes The system monitors presence or absence of a patient in bed (bed exit) The system also times the length of continuous patient motion or absence of patient motion
| Prescription Use ✗<br>(Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) |
|---------------------------------------------------|--------|------------------------------------------------|
|---------------------------------------------------|--------|------------------------------------------------|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER
PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | Page 1 of _ |
|-----------------------------------------------------------|-------------|
| (Division Sign-Off)<br>Division of Cardiovascular Devices | |
| 510(k) Number | K082626 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.