OCS 3000® is intended to be used by qualified health-care providers to enhance the identification and visualization of oral mucosal abnormalities that may not be apparent or visible to the naked eye, such as oral cancer or premalignant dysplasia. OCS 3000 allows for conventional oral mucosal examination and excites the tissue with multispectral lights for direct visualization of the resulting natural tissue fluorescence and reflectance, and vasculature. OC'S 3000 is also intended to be used by a surgeon to help identify diseased tissue around a clinically apparent lesion to aid in determining the appropriate margin for surgical excision. OCS 3000 eyewear is reusable polarized filtered eyewear that is worn by a health care professional to enhance the visual effects of violet light during oral exam. OCS 3000 eyewear has been designed to allow transmission of 430-580 nm light.
Device Story
TRIMIRA OCS 3000 is a handheld, battery-operated multispectral oral examination light. Device uses high-intensity LEDs to emit white, violet, and amber light. Clinician performs conventional oral exam with white light; switches to violet light to excite natural tissue fluorescence (healthy tissue fluoresces green; suspicious tissue appears dark) and amber light to enhance reflectance for visualization of vasculature. Clinician wears reusable polarized filtered eyewear (430-580 nm transmission) to enhance visual contrast. Device used in clinical settings by healthcare providers to identify subclinical high-risk fields with cancerous or precancerous changes and to assist surgeons in determining surgical margins. Benefits include rapid identification of suspicious tissue requiring further investigation without requiring additional chemicals or dyes.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and intended use comparisons to legally marketed predicate devices.
Technological Characteristics
Handheld, battery-operated (AA) multispectral light source using high-intensity LEDs. Emits white, violet, and amber light. Includes reusable polarized filtered eyewear (430-580 nm transmission). No software or complex algorithms; purely optical/visual enhancement device.
Indications for Use
Indicated for use by qualified healthcare providers to enhance identification/visualization of oral mucosal abnormalities (e.g., oral cancer, premalignant dysplasia) not apparent to the naked eye; aids surgeons in identifying diseased tissue margins for excision.
Regulatory Classification
Identification
An ultraviolet detector is a device intended to provide a source of ultraviolet light which is used to identify otherwise invisible material, such as dental plaque, present in or on teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Remicalm LLC. The logo consists of a stylized four-petal flower-like design to the left of the text "REMIICALM LLC". The text is in a simple, sans-serif font.
952 Echo Lane, Suite 333 Houston, TX 77024 Voice: 713-984-8550 Fax: 713-984-9232
K082603
#### V. 510(k) SUMMARY
- Submitted by:
TRIMIRA LLC 952 Echo Lane Houston, TX 77024 Phone: (713) 984-8550 Fax: (713)984-9232
DEC 1 2 2008
| Contact Person: | David B. Jones |
|-----------------|----------------|
|-----------------|----------------|
August 22, 2008
Proprietary Name:
Date Prepared:
Oral Examination Light and Accessories
Ultraviolet Detector - NXV (EAQ)
Photosensitive glasses - HQY (Exempt)
TRIMIRA™ OCS 3000®
Classification:
Common Name:
Class II: 21 CFR § 872.6350
Class I: (Exempt) 21 CFR § 874.4420 Class I: (Exempt) 21 CFR § 886.5850
ENT manual surgical instrument (Mirrors) - KAI (Exempt)
Classification Name:
Predicate Devices:
VELscope (K070523) LED Medical Diagnostics 201-15047 Marine Drive White Rock, BC, Canada V4B 1C5
> ViziLite (K033033 & K080043) The Trylon Corporation 970 West 190th Street, Suite 850 Torrance, CA 90502-1037
#### Device Description:
The TRIMIRA™ OCS 3000® is a "AA" battery operated, hand-held, multispectral oral examination light used in conventional and specialized oral examination. Accessories include safety glasses and disposable mirrors.
#### Intended Use:
OCS 3000® is intended to be used by qualified health-care providers to enhance the identification and visualization of oral mucosal abnormalities that may not be apparent or visible to the naked eye, such as oral cancer or premalignant dysplasia.
OCS 3000 allows for conventional oral mucosal examination and excites the tissue with multispectral lights for direct visualization of the resulting natural tissue fluorescence and reflectance, and vasculature.
OC'S 3000 is also intended to be used by a surgeon to help identify diseased tissue around a clinically apparent lesion to aid in determining the appropriate margin for surgical excision.
Confidentially claimed per 21 CFR 807.95
OCS 3000 510(k) Vol. 1
{1}------------------------------------------------
952 Echo Lane, Suite 333 Houston, TX 77024 Voice: 713-984-8550 Fax: 713-984-9232
EMICALM LLC
08760
272
OCS 3000® eyewear is reusable polarized filtered eyewear that is worn by a health care professional to enhance the visual effects of violet light during oral exam. OCS 3000® eyewear has been designed to allow transmission of 430-580 nm light.
## Technological Characteristics:
The OCS 3000 with multispectral identifi™ technology (Patent No. 7,365,844 and patents pending) uses "AA" batteries to operated high-intensity LEDs to produce white, violet and amber light. These safe, visible lights allow health professional to perform Conventional Oral Examination (COE) using white light, and specialized or enhanced visual examinations of tissue using violet and amber lights.
The OCS 3000 violet light enhances normal tissue's natural fluorescence. Visual examination under violet light shows healthy tissues in fluorescing green (when using filtered glasses) while suspicious tissue appears dark because of its loss of fluorescence. Amber light enhances the tissue's natural reflectance and makes visualization of the vasculature possible.
The direct visualization of fluorescent and reflective tissues is using the body's natural system to identify suspicious tissue quickly that may require further investigation. The loss of natural tissue fluorescence and reflectance can identify subclinical high-risk fields with cancerous and precancerous changes Clinical Cancer Research Vol. 12, 6716-6722. November 15, 2006.
#### Substantial Equivalence
The OCS 3000 has the same intended use and technological characteristics as the approved predicate devices (K033033, K070523, and K080043); each uses fluorescence and/or reflectance as the primary mode for enhanced visualization of tissue for determining oral tissue abnormalities.
Predicate K070523 uses 120v AC to power a metal halide light to produce a single blue light and views fluorescence through a connected handpiece with a green filtered lens. In addition, the predicate for the original VELscope (K060920) was the ViziLite (K033033).
Predicate, K033033, uses two "AA" batteries to power high-intensity LEDs to produce white light. However, predicate K033033 uses a 1% acetic acid rinses and metachromatic vital tissue dye (tolonium chloride), a post-dye decolorized filtered eyewear (K080043) with diffuse chemiluminescent white-blue light to observe abnormalities by reflectance.
The OCS 3000 "s patented multispectral identifi™ technology use LEDs produce three different light wave lengths (white, violet and amber), and polarized filtered eyewear (K080043) to enhance the fluorescence and reflectance (K070523) contrast of normal and neoplastic epithelial tissue. By filtering a selected band of autofluoresced or reflected light, the OCS 3000® allows the clinician to see enhanced contrast between normal and abnormal tissue within the overall range of white light spectra (K033033).
Unlike the predicates, the clinician may use the multispectral OCS 3000® with white light for conventional oral examinations, and violet and amber lights for visualization of the natural tissue fluorescence and reflectance. Other devices or chemicals are not required. In addition, the OCS 3000 amber light enhances visualization of the vasculature through reflectance.
The design, materials, method of operation, and labeling are substantially equivalent.
The TRIMIRA™ OCS 3000® is substantially equivalent to the approved predicate devices.
Confidentially claimed per 21 CFR 807.95
OCS 3000 510(k) Vol. 1
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services, USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. The eagle is stylized with thick, black lines, giving it a modern and authoritative appearance.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEC 1 2 2008
Mr. David B. Jones Vice President, Regulatory Affairs and Quality Assurance Remicalm L.L.C. 952 Echo Lane, Suite 333 Houston, Texas 77024
Re: K082603
Trade/Device Name: TRIMIRA™ OCS 3000 Regulation Number: 21 CFR 872.6350 Regulation Name: Ultraviolet Detector Regulatory Class: II Product Code: NXV Dated: December 8, 2008 Received: December 10, 2008
Dear Mr. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 - Mr. Jones
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Anthony D. Watson for
Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for REMICALM LLC. The logo consists of a stylized atom symbol above the company name. The text "REMICALM LLC" is written in a simple, sans-serif font. The logo is black and white.
952 Echo Lane, Suite 333 Houston, TX 77024 Voice: 713-984-8550 Fax: 713-984-9232
IV. **Indications for Use**
K082603
Applicant:
TRIMIRA LLC 952 Echo Lane, Suite 333 Houston, TX 77024 Phone: (713) 984-8550 Fax: (713) 984-9232
510(k) Number (if Known): KA8221053
Device Name:
TRIMIRA™ OCS 3000®
Indications For Use:
OCS 3000® is intended to be used by qualified health-care providers to enhance the identification and visualization of oral mucosal abnormalities that may not be apparent or visible to the naked eye, such as oral cancer or premalignant dysplasia.
OCS 3000® allows for conventional oral mucosal examination and excites the tissue with multispectral lights for direct visualization of the resulting natural tissue fluorescence and reflectance, and vasculature.
OCS 3000® is also intended to be used by a surgeon to help identify diseased tissue around a clinically apparent lesion to aid in determining the appropriate margin for surgical excision. OCS 3000 eyewear is reusable polarized filtered eyewear that is worn by a health care professional to enhance the visual effects of violet light during oral exam. OCS 3000 eyewear has been designed to allow transmission of 430-580 nm light.
Prescription Use
(Per 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Rumer
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: R082603
Confidentially claimed per 21 CFR 807.95
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.