← Product Code [FBN](/productcode/FBN) · K082576

# XCHF-T160 VIDEOSCOPE (K082576)

_Olympus Medical Systems Corporation · FBN · Oct 21, 2008 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K082576

## Device Facts

- **Applicant:** Olympus Medical Systems Corporation
- **Product Code:** [FBN](/productcode/FBN.md)
- **Decision Date:** Oct 21, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, video monitor, endo-therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within the biliary tract.

## Device Story

Flexible video endoscope for biliary tract endoscopy and surgery; utilizes CCD-based optical system to capture images; transmits video to external Olympus video system center and monitor for clinician visualization; operated by physicians in clinical settings; enables diagnostic and therapeutic procedures via instrument channels for biopsy forceps or other accessories; provides real-time visualization to guide surgical interventions; benefits patients by facilitating minimally invasive biliary tract access.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Flexible endoscope; CCD-based optical system; 90° field of view; 1-50mm depth of field; 5.7mm distal end diameter; 5.5mm insertion tube diameter; 1330mm working length; 1.2mm/2.0mm instrument channel diameters; forward viewing; manual angulation (U160°/D100°/R100°/L100°).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- CHOLEDOCHOSCOPE CHF-B20 ([K904799](/device/K904799.md))
- EVIS EXERA Gastrointestinal Videoscope GIF-Q160Z ([K011151](/device/K011151.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

208 2576

Pg 1 of 3

# 510(k) SUMMARY

# VIDEOSCOPE XCHF-T160

# OCT 9 1 2008

## 1. General Information

- OLYMPUS MEDICAL SYSTEMS CORP. Applicant: 트 2951 Ishikawa-cho, Hachioji-shi, Tokyo, 192-8507, Japan Establishment Registration No: 8010047 Stacy Abbatiello Kluesner, RAC Official Correspondent: I Regulatory Affairs & Quality Assurance Olympus America Inc. 3500 Corporate Parkway, PO Box 610 Center Valley, PA 18034-0610 Phone: (484) 896-5405 Facsimile: (484) 896-7128 Email:Stacy.Kluesner@olympus.com Establishment Registration No: 2429304 해 Manufacturer: Aizu Olympus Co., Ltd. 500 Aza-Muranishi, Ooaza-Niidera, Monden-machi, Aizuwakamatsu-shi, Fukushima, Japan 965-8520 Establishment Registration No .: 9610595
- August 21, 2008 ● Date Prepared:

## 2. Device Identification

| Device Name:            | XCHF-T160 Videoscope                      |
|-------------------------|-------------------------------------------|
| Common Name:            | Videoscope                                |
| Class:                  | II                                        |
| Regulation Number/Name: | 876.1500 Endoscope and accessories        |
| Product Code:           | FBN                                       |
| Classification Panel:   | Choledochoscope And Accessories, Flexible |

olympus medical systems corp. 2951 Ishikawa-cho Hachioji-shi Tokyo, 192-8507 JAPAN TELEPHONE +81-42-642-2694, TELEFAX +81-42-642-2307

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "OLYMPUS" in a bold, sans-serif font. The letters are black and appear to have a slightly rough or textured edge, giving them a somewhat distressed look. The word is centered and takes up most of the frame, with a white background providing contrast.

Pg 2 of 3

## 3. Legally Marketed Device to which Substantial Equivalence is Clain

The following table shows the subject device and the predicate device to which we claim substantial equivalence.

### Table 14-1. Primary Component & Predicate Device

| Subject Device<br>(Part of this submission) | Predicate Device                                    | PD's<br>510(k)<br>No. |
|---------------------------------------------|-----------------------------------------------------|-----------------------|
| VIDEOSCOPE<br>XCHF-T160                     | CHOLEDOCHOSCOPE<br>CHF-B20                          | K904799               |
| XCHF-T160                                   | EVIS EXERA Gastrointestinal Videoscope<br>GIF-Q160Z | K011151               |

## 4. Device Description

The XCHF-T160, is a flexible video endoscope used for endoscopy and endoscopic surgery within the biliary tract. The XCHF-T160 is basically identical to the predicate device, Olympus CHF-B20 Choledochoscope, hereinafter referred to as CHF-B20 in intended use, specifications, performance.

The optical system of the XCHF-T160 is a charge coupled device (CCD) based system, allowing endoscopic image display on a video monitor.

The new endoscope is basically identical to each predicate device shown in Table 14-1 in intended use, and similar in specifications, performance and materials.

## 5. Indications for Use

### VIDEOSCOPE XCHF-T160

This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, video monitor, endo-therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within the biliary tract.

> OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho Hachioji-shi Tokyo, 192-8507 JAPAN TELEPHONE +81-42-642-2694, TELEFAX +81-42-642-2307 2/3

{2}------------------------------------------------

# 6. Comparison of Technological Characteristics

182516

The XCHF-T160 is similar to the predicate device CHF-B20 in specifications except for the material and optical system. Comparison between the subject and predicate devices is shown below.

| Specifications                          | Subject Device<br>XCHF-T160          | Predicate Device 1<br>CHF-B20 | Predicate Device 2<br>GIF-Q160Z     |
|-----------------------------------------|--------------------------------------|-------------------------------|-------------------------------------|
| Field of View                           | 90°                                  | 100°                          | 140°                                |
| Depth of Field                          | 1-50mm                               | 3-50mm                        | WIDE : 8-100mm<br>TELE : 1.5-3mm    |
| Direction of View                       | Forward                              | Forward                       | Forward                             |
| Type of CCD Chip                        | Inter color CCD<br>(37071)           |                               | Inter color CCD<br>(3704)           |
| Outer Diameter of Distal<br>End         | \$\phi\$ 5.7mm                       | \$\phi\$ 4.1mm                | \$\phi\$ 10.8mm                     |
| Outer Diameter of<br>Insertion Tube     | \$\phi\$ 5.5mm                       | \$\phi\$ 4.5mm                | \$\phi\$ 10.9mm                     |
| Angulation<br>UP/DOWN                   | U 160°<br>D 100°<br>R 100°<br>L 100° | U 160°<br>D 100°              | U 210°<br>D 90°<br>R 100°<br>L 100° |
| Working Length                          | 1330mm                               | 1870mm                        | 1030mm                              |
| Inner Diameter of<br>Instrument Channel | \$\phi\$ 2.0mm<br>\$\phi\$ 1.2mm     | \$\phi\$ 1.7mm                | \$\phi\$ 2.8mm                      |

|  | Table 14-2. Comparison of Specifications |  |  |
|--|------------------------------------------|--|--|
|--|------------------------------------------|--|--|

# 7. Conclusion

When compared to the predicate device, the CHF-B20 does not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the device.

> OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho Hachioji-shi Tokyo, 192-8507 JAPAN TELEPHONE +81-42-642-2694, TELEFAX +81-42-642-2307 રૂાર

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle with three stripes representing the feathers.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# OCT 2 1 2008

OLYMPUS MEDICAL SYSTEMS CORP. c/o Stacy Abbatiello Kluesner, RAC Regulatory Affairs & Quality Assurance Olympus America, Inc. 3500 Corporate Parkway P.O. Box 610 CENTER VALLEY PA 18034-0610

K082576 Re:

> Trade/Device Name: VIDEOSCOPE XCHF-T160 Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscopes and accessories Regulatory Class: II Product Code: FBN Dated: September 4, 2008 Received: September 9, 2008

Dear Ms. Kluesner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

## Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Jorgu M. Whang

Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): 人 ८ 8 2 5 7 6 Device Name: VIDEOSCOPE XCHF-T160

Indications For Use:

This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, video monitor, endo-therapy accessories such as biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within the biliary tract.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use_ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Reproductive, Abdominal, and

Radiological Devices 510(k) Number

Page 1 of _ 1

---

**Source:** [https://fda.innolitics.com/device/K082576](https://fda.innolitics.com/device/K082576)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).