DUAL NEEDLE SUTURE PASSER SYSTEM

K082518 · Cayenne Medical, Inc. · GAT · Jan 16, 2009 · General, Plastic Surgery

Device Facts

Record IDK082518
Device NameDUAL NEEDLE SUTURE PASSER SYSTEM
ApplicantCayenne Medical, Inc.
Product CodeGAT · General, Plastic Surgery
Decision DateJan 16, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.5000
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Dual Needle Suture Passer System is intended for approximation of soft tissue in procedures such as meniscal repair

Device Story

The Dual Needle Suture Passer System is a sterile, hand-held, manually operated, single-procedure device used for soft tissue approximation. It is designed for use by surgeons during procedures such as meniscal repair. The device facilitates the placement of sutures into soft tissue. Performance is verified through mechanical suture knot pull-out force testing and animal tissue testing, demonstrating that the device meets USP requirements for non-absorbable sutures and provides superior knot pull-out strength compared to predicate devices.

Clinical Evidence

No clinical data. Evidence consists of bench testing (mechanical suture knot pull-out force) and animal tissue testing to verify performance requirements.

Technological Characteristics

Sterile, hand-held, manually operated, single-procedure suture placement system. Materials meet USP requirements for non-absorbable, synthetic, polyethylene sutures. Device is a mechanical instrument for soft tissue approximation.

Indications for Use

Indicated for approximation of soft tissue in procedures such as meniscal repair.

Regulatory Classification

Identification

Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K082518 Dual Needle Suture Passer Page 1 of 2 510(k) Premarket Notification Dual Needle Suture Passer ### 510(k) Summary Cayenne Medical, Inc. Dual Needle Suture Passer System JAN 1 6 2009 #### ADMINISTRATIVE INFORMATION | Manufacturer Name | Cayenne Medical Inc<br>16597 N 92nd St, Suite 101<br>Scottsdale, AZ 85260<br>Telephone (480) 502-3661<br>FAX (480) 502-3670 | |-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------| | Official Contact | Kereshmeh Shahriari<br>16597 N 92nd St, Suite 101<br>Scottsdale AZ 85260<br>kshahriari@cayennemedical.com<br>Telephone (480) 502-3661<br>FAX (480) 502-3670 | DEVICE NAME | Classification Names | Suture, Nonabsorbable, Synthetic, Polyethylene | |------------------------|------------------------------------------------| | Trade/Proprietary Name | Dual Needle Suture Passer System | | Common Name | Suture Punch, Endoscopic Accessories | #### DEVICE CLASSIFICATION FDA has classified sutures as Class II devices (21 CFR 878 5000) The product code for Suture, Nonabsorbable, Synthetic, Polyethylene is GAT These devices are reviewed by the General and Plastic Surgery Branch #### INTENDED USE The Dual Needle Suture Passer System is intended for approximation of soft tissue in procedures such as memscal repair **DANGER!** Page 109 & 110 {1}------------------------------------------------ Dual Needle Suture Passer **Response,** Page ② of ② 510(k) Premarket Notification Dual Needle Suture Passer DEVICE DESCRIPTION The Dual Needle Suture Passer System is a sterle hand-held, manually operated sungle procedure suture placement system for soft tissue approximation procedures Mechanical suture knot pull-out force testing has been conducted on the Dual Needle Suture Passer System It was shown that suture knot pull-out strength is significantly higher than that of a predicate device The proposed device was tested in the animal trsue and meets the knot pull-out strength requirements The proposed device meets USP requirements for non-absorbable sutures #### EQUIVALENCE TO MARKETED PRODUCT Cayenne Medical, Inc demonstrated that for the purposes of FDA's regulation of medical devices, the Dual Needle Suture Passer System is substantially equivalent 111 indications and design principles to predicate devices each of which has been determined by FDA to be substantially equivalent to preamendment devices Teleflex Medical's Force Fiber™, Non-absorbable Polyethylene Surgical Suture (K063778), CP Medical's CP Fiber Non-absorbable Polyblended Surgical Suture (K041894), DePuy Mitek's (Mitek Worldwide) RAPIDLOC-PDS Mensical Repair System (K023388), Anulex Technologies's Xclose™ Tissue Repair System (K062307), and Smith & Nephew s ULTRA FAST-FIX Memscal Repair System (K072322) Revised Page 18 of 161 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "HEALTH & HUMAN SERVICES USA" is arranged around the upper portion of the circle. Food and Drug Admınıstration 9200 Corporate Boulevard Rockville MD 20850 # JAN 1 6 2009 Cayenne Medical, Inc % Kereshmeh Shahrıarı Director of QA & RA 16597 N 92nd Street, Suite 101 Scottsdale, Arizona 85260 Re K082518 Trade/Device Name Dual Needle Suture Passer System Regulation Number 21 CFR 878 5000 Regulation Name Nonabsorbable poly(ethylene terephthalate) surgical suture Regulatory Class II Product Code GAT Dated December 22, 2008 Received December 23, 2008 Dear Kereshmeh Shahrıarı We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce pror to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA) You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 In addition, FDA may publish further announcements concerning your device in the Federal Register Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807), labeling (21 CFR Part 801), good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050 ![Image](https://i.imgur.com/946699l.png) and the stille all to the mail be a chiefic to and the legist of the leating the 2017-12-11-11 {3}------------------------------------------------ ## Page 2 - Kereshmeh Shahrıarı This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807 97) For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474 For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464 You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http //www.fda gov/cdrh/industry/support/index.html Sincerely yours. Mark N Millican Mark N Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 17000558 510(k) Premarket Notification Dual Needle Suture Passer Indications for Use 510(k) Number (if known) K072518 Device Name Dual Needle Suture Passer System Indications for Use The Dual Needle Suture Passer System is intended for approximation of soft tissue 11 procedures such as meniscal repar Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) 서 대한민국 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학교 | Concurrence of CDRH, Office of Device Evaluation (ODE) | |--------------------------------------------------------| |--------------------------------------------------------| | (Division Sign-Off) | |-----------------------------------| | Division of General, Restorative, | | and Neurological Devices | | 510(k) Number | 16082518 | |---------------|----------| |---------------|----------|
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...