K082514 · Medihightec Medical Co., Ltd. · GZJ · Oct 21, 2008 · Neurology
Device Facts
Record ID
K082514
Device Name
TENS/EMS COMBO, EMS, TENS
Applicant
Medihightec Medical Co., Ltd.
Product Code
GZJ · Neurology
Decision Date
Oct 21, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
APPLICABLE FOR MH8000 (TENS/EMS COMBINATION UNIT) AND MH6000 (TENS/EMS COMBINATION UNIT) AND MH8200 (TENS) AND MH6200 (TENS): This device is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is indicated for use for (1) Symptomatic relief and management of chronic (long term) intractable pain and (2) adjunctive treatment in the management of post surgical and post traumatic pain problems. APPLICABLE FOR MH8000 (TENS/EMS COMBINATION UNIT) AND MH6000 (TENS/EMS COMBINATION UNIT) AND MH8100 (EMS) AND MH6100 (EMS): This device is an electrically powered muscle stimulator indicated for use for medical purposes to repeatedly contract muscles by passing electrical currents through electrodes contacting the affected body area. In particular, this device is indicated for use for (1) Relaxing muscle spasms, (2) Increasing local blood circulation, (3) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, (4) Muscle re-education, (5) Maintaining or increasing range of motion, and (6) Preventing or retarding disuse atrophy.
Device Story
Battery-powered TENS/EMS units deliver electrical current to nerves/muscles via skin electrodes. User selects stimulation modes (TENS/EMS/Combo) and adjusts parameters (pulse rate, width, timer, on/off/ramp times) via press buttons and LCD. Device operates via microprocessor control; provides synchronous or alternating channel stimulation. Used in clinical or home settings; operated by patient or clinician. Output affects muscle contraction or nerve stimulation for pain management/rehabilitation. Benefits include pain relief, improved circulation, and muscle maintenance.
Clinical Evidence
Bench testing only. Comparison of output characteristics (voltage, current, pulse width, frequency, phase charge) against predicate devices. All parameters confirmed within safe, effective ranges for TENS/EMS applications. No clinical trials performed.
Indicated for patients requiring pain relief (TENS) or muscle stimulation (EMS). TENS: chronic intractable pain, post-surgical/post-traumatic pain. EMS: muscle spasms, blood circulation, prevention of venous thrombosis, muscle re-education, range of motion maintenance, disuse atrophy prevention.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
# 510(k) Summary As Required by 21 section 807.92 (c)
## 1. Submitter Name: MEDIHIGHTEC MEDICAL CO., LTD.
- 2. Address: 4F., No.180, Fu 1st St., Cidu District, Keelung City 20650, Taiwan R.O.C.
- 3. Phone: 886-2-2451-9988
- 4. Fax: 886-2-2451-1796
- 5. E-mail: sales(@medihightec.com
- 6. Contact Person: Mr. Mu-Sheng Hung
- 7. Date summary prepared: April 7th, 2008
- 8. Regulation Number: 882.5890; 890.5850
- 9. Product Code: GZJ; IPF
# 10. Device Trade or Proprietary Name:
- (a) TENS/EMS Combo (Model MH8000)
- (b)TENS/EMS Combo (Model MH6000)
- (c) EMS (Model MH8100)
- (d) EMS (Model MH6100)
- (e) TENS (Model MH8200)
- (f) TENS (Model MH6200)
# 11. Device Common or usual name:
- (a) TENS/EMS COMBINATON UNIT (Model MH8000)
- (b) TENS/EMS COMBINATON UNIT (Model MH6000)
- (c) EMS UNIT (Model MH8100)
- (d) EMS UNIT (Model MH6100)
- (e) TENS UNIT (Model MH8200)
- (f) TENS UNIT (Model MH6200)
# 12. Device Classification Name:
- (a) Stimulator, nerve, transcutaneous, for pain relief (21 CFR 882.5890, Product Code GZJ).
- (b) Stimulator, muscle, powered (21 CFR 890.5850, Product Code IPF).
- (c) Combination of (i) Stimulator, nerve, transcutaneous, for pain relief AND (ii) Stimulator, muscle, powered (21 CFR 882.5890; 890.5850, Product Code GZJ; IPF)
#### 13. Substantial Equivalency is claimed against the following device:
- (a) The new device MH8000 and MH6000 compared to the predicate devices: Comfy STIM, model EV-806 from Everyway Medical instruments Co., Ltd. 510k# K071951, Product Code: GZJ; IPF.
- (b) The new device MH8100 and MH6100 compared to the predicate devices: Comfy EMS, model EV-805 from Everyway Medical instruments Co., Ltd. 510k# K071951, Product Code: IPF.
# OCT 2 1 2008
{1}------------------------------------------------
- (c) The new device MH8200 and MH6200 compared to the predicate devices: Comfy TENS, model EV-804 from Everyway Medical instruments Co., Ltd. 510k# K071951, Product Code: GZJ.
#### 14. Description of the Device:
The MH8000 Series, which includes models MH8000, MH8100, MH8200, MH6000, MH6100, MH6200, are Transcutaneous Electrical Nerve Stimulator for pain relief and/or Electrical Muscle Stimulator sends gentle electrical current to underlying nerves and muscle group via electrodes applied on the skin. The parameters of units are controlled by the press buttons. Its intensity level is adjustable according to the needs of patients.
The three models MH8000, MH8100 and MH8200 have the same housing. The three models MH6000, MH6100 and MH6200 have the same housing.
The process to set the parameter and attach lead wires to the six models are also the same. Yet, they have different liquid crystal display and parameters for patients to create their own settings.
The MH8200 and MH6200 are the TENS device with 5 modes and adjustable pulse rate, pulse width and timer. The MH8100 and MH6100 are the EMS device with 3 modes and adjustable pulse rate, pulse width, on time, off time, ramp time and timer, The MH8000 and MH6000 are the combination unit with both TENS and EMS functions, The function can be selected by press buttons. The range of settings are identical to those of MH8200, MH8100, MH6100 and MH6200. The difference on the six units can be identified by the liquid crystal display.
| | | New Device | Predicate Device |
|-----|----------------------------------------------|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| 1. | 510(k) Number | To be Assigned | K071951 |
| 2. | Device Name, Model | MH8000/MH6000<br>Combo<br>MH8100/MH6100 EMS<br>MH8200/MH6200 TENS | TENS/EMS EV-806 TENS/EMS, Comfy<br>Stim<br>EV-805 EMS, Comfy EMS<br>EV-804 TENS, Comfy TENS |
| 3. | Manufacturer | MEDIHIGHTEC MEDICAL<br>LTD. | CO., Everyway Medical<br>Instruments Ltd. |
| 4. | Power Source(s) | 9 volt battery type 6F22 | 9 volt battery type 6F22 |
| | - Method of Line Current Isolation | N/A | N/A |
| | - Patient Leakage Current | | |
| | -Normal condition | 1.5uA | |
| | -Single fault condition | 1.7uA | |
| 5. | Number of Output Modes | 7 | 7 |
| 6. | Number of Output Channels | 2 | 2 |
| | - Synchronous or Alternating? | Synchronous and Alternating | Synchronous and Alternating |
| | - Method of Channel Isolation | By TRANSFORMER | By TRANSFORMER |
| 7. | Regulated Current or Regulated<br>Voltage? | 5V | 5V |
| 8. | Software/Firmware/Microprocessor<br>Control? | Yes | Yes |
| 9. | Automatic Overload Trip? | No | No |
| 10. | Automatic No-Load Trip? | No | No |
Technological Characteristics, New device vs. Predicate device
{2}------------------------------------------------
| | 11. Automatic Shut Off? | No | No |
|----|---------------------------------------|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| | 12. Patient Override Control? | No | No |
| | 13. Indicator Display: | | |
| | - On/Off Status? | Yes | Yes |
| | - Low Battery? | Yes | Yes |
| | - Voltage/Current Level? | Yes | Yes |
| | 14. Timer Range (minutes) | Yes | Yes |
| | 15 Pulse Width | The range of Pulse Width control is between 50µS and 300µS. | The range of Pulse Width control is between 50µS and 300µS. |
| | 16 Frequency | The range of Pulse Rate control between 2 Hz and 150 Hz. | The range of Pulse Rate control between 2 Hz and 150 Hz. |
| | 17 Phase Duration | Pulse Width control is between 50µS and 300µS. | Pulse Width control is between 50µS and 300µS. |
| | 18 ON Time | Adjustable, 2~90 seconds, 1 Sec./ step | Adjustable, 2~90 seconds, 1 Sec./ step |
| | 19 OFF Time | Adjustable, 2~90 seconds, 1 Sec./ step | Adjustable, 2~90 seconds, 1 Sec./ step |
| | 20 Burst Mode | a. Pulses per buster: 9<br>b. Burst per second: 2<br>c. Burst duration: 100 mSec.<br>d. Duty Cycle: 200 mSec. | a. Pulses per buster: 10<br>b. Burst per second adjustable, 0.5~5<br>c. Burst duration: 100 mSec.<br>d. Duty Cycle: adjustable 50~500 mSec. |
| 21 | Compliance with Standards? | Voluntary EN60601-1, EN60601-1-2 | EN60601-1, EN60601-1-2 |
| | 22 Compliance with 21 CFR 898? | Yes | Yes |
| | 23 Weight | 162 grams (battery included) | 150 grams (battery included) |
| | 24 Dimensions | 13.6(L)x7(W)x2.7(H)(cm) | 10.1(L)x6.1(W)x2.45(H)(cm) |
| | 25 Housing Materials and Construction | 1. Enclosure: ABS, 94V-1, 80°C, UL Approved.<br>2. PC board: FR4, 94V-1, 105°C, UL Approved. | 1. Enclosure: ABS, 94V-1, 80°C, UL Approved.<br>2. PC board: FR4, 94V-1, 105, UL Approved. |
Above differences of Technological Characteristics do not pose any new questions of safety and effectiveness.
# 15. Intended use of the device:
The intended Use of TENS, Model MH8200 and MH6200 TENS is as follows: This device is an electrical nerve stimulator intended for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is intended for use for (1) Symptomatic relief and management of chronic (long term) intractable pain and (2) adjunctive treatment in the management of post surgical and post traumatic pain problems
The intended Use of EMS, Model MH8100 and MH6100 EMS is as follows:
The device is an electrically powered muscle stimulator intended for use for medical purposes to repeatedly contract muscles by passing electrical currents through electrodes contacting the affected body area. In particular this device is intended for use for (1) Relaxing muscle spasms, (2) Increasing local blood circulation,
{3}------------------------------------------------
(3) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis. (4) Muscle re-education, (5) maintaining or increasing range of motion, and (6) Preventing or retarding disuse atrophy.
The Intended Use of TENS/EMS COMBO, Model MH8000 and MH6000 TENS/EMS Combination Unit are as follows:
This combination device is intended for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is intended for use for (1) Symptomatic relief and management of chronic (long term) intractable pain and (2) adjunctive treatment in the management of post surgical and post traumatic pain problems. It is also intended for use medical purposes to repeatedly contract muscles by passing electrical currents through electrodes contacting the affected body area. In particular, this device is intended for use for (1) Relaxing muscle spasms, (2)Increasing local blood circulation, (3) immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, (4) Muscle re-education, (5) Maintaining or increasing range of motion, and (6) Preventing or retarding disuse atrophy.
# 16. Non-Clinical Testing:
Non-Clinical Testing Result: New device MH8000/MH6000 vs. Predicate device EV-806
| | | New Device | Predicate Device |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | 510(k) Number | To be Assigned | K071951 |
| 2. | Device Name, Model | MH8000/MH6000<br>Combo | TENS/EMS EV-806 TENS/EMS, Comfy<br>Stim |
| 3. | Manufacturer | MEDIHIGHTEC MEDICAL CO.,<br>LTD. | Everyway Medical<br>Instruments Ltd. |
| 4. | Power Source(s) | 9 volt battery type 6F22 | 9 volt battery type 6F22 |
| 5. | Number of Output Modes | 7 | 7 |
| 6. | Number of Output Channels | 2 | 2 |
| | - Synchronous or Alternating? | Synchronous and Alternating | Synchronous and Alternating |
| | - Method of Channel Isolation | By TRANSFORMER | By TRANSFORMER |
| 7. | Waveform | Asymmetrical biphasic pulse | Asymmetrical biphasic pulse |
| 8. | Shape | Rectangular | Rectangular |
| 9. | Maximum Output Voltage (+/- 20%) | 40.8V @ 500 Ohm<br>77.6V @ 2000 Ohm<br>102V @ 10000 Ohm | 50V @ 500 Ohm<br>150V @ 2000 Ohm<br>194V @ 10000 Ohm |
| 10. | Maximum Output Current (+/- 20%) | 81.6mA @ 500 Ohm<br>38.8mA @ 2000 Ohm<br>10.2mA @ 10000 Ohm | 100mA @ 500 Ohm<br>75mA @ 2000 Ohm<br>19.4mA @ 10000 Ohm |
| 11. | Pulse Width | The range of Pulse Width control is<br>between 50 $ μ $ S and 300 $ μ $ S. | The range of Pulse Width<br>control is between 50 $μ$ S and<br>300 $μ$ S. |
| 12. | Frequency | The range of Pulse Rate control<br>between 2 Hz and 150 Hz. | The range of Pulse Rate<br>control between 2 Hz and 150<br>Hz. |
| 13. | Maximum Phase Charge | 24 micro-coulombs @ 500 Ohm | 30 micro-coulombs @ 500<br>Ohm |
| 14. | Maximum Current Density | 1.06mA/ $cm2$ @ 500 Ohm | 1.33mA/ $cm2$ @ 500 Ohm |
| | | New Device | Predicate Device |
| 1. | 510(k) Number | To be Assigned | K071951 |
| 2. | Device Name, Model | MH8100/MH6100 EMS | EV-805 EMS, Comfy EMS |
| 3. | Manufacturer | MEDIHIGHTEC MEDICAL<br>LTD. | Everyway Medical<br>Instruments Ltd. |
| 4. | Power Source(s) | 9 volt battery type 6F22 | 9 volt battery type 6F22 |
| 5. | Number of Output Modes | 3 | 3 |
| 6. | Number of Output Channels | 2 | 2 |
| | - Synchronous or Alternating? | Synchronous and Alternating | Synchronous and Alternating |
| | - Method of Channel Isolation | By TRANSFORMER | By TRANSFORMER |
| 7. | Waveform | Asymmetrical biphasic pulse | Asymmetrical biphasic pulse |
| 8. | Shape | Rectangular | Rectangular |
| 9. | Maximum Output Voltage (+/- 20%) | 40.8V @ 500 Ohm<br>77.6V @ 2000 Ohm<br>102V @ 10000 Ohm | 50V @ 500 Ohm<br>150V @ 2000 Ohm<br>194V @ 10000 Ohm |
| 10. | Maximum Output Current (+/- 20%) | 81.6mA @ 500 Ohm<br>38.8mA @ 2000 Ohm<br>10.2mA @ 10000 Ohm | 100mA @ 500 Ohm<br>75mA @ 2000 Ohm<br>19.4mA @ 10000 Ohm |
| 11. | Pulse Width | The range of Pulse Width control is<br>between 50µS and 300µS. | The range of Pulse Width control<br>is between 50µS and 300µS. |
| 12. | Frequency | The range of Pulse Rate control<br>between 2 Hz and 150 Hz. | The range of Pulse Rate<br>control between 2 Hz and 150 Hz. |
| 13. | Maximum Phase Charge | 24 micro-coulombs @ 500 Ohm | 30 micro-coulombs @ 500 Ohm |
| 14. | Maximum Current Density | $1.06mA_{rms}/cm^2$ @ 500 Ohm | $1.33mA_{rms}/cm^2$ @ 500 Ohm |
| 15. | Maximum Power Density | $2.4mW_{rms}/cm^2$ @ 500 Ohm | $3.7mW_{rms}/cm^2$ @ 500 Ohm |
| 16. | Burst Mode | a. Pulses per buster: 9<br>b. Burst per second: 2<br>c. Burst duration: 100 mSec.<br>d. Duty Cycle: 200 mSec. | a. Pulses per buster: 10<br>b. Burst per second:<br>adjustable, 0.5~5<br>c. Burst duration: 100 mSec.<br>d. Duty Cycle: adjustable,<br>50~500 mSec. |
| 17. | ON Time | Adjustable, 2~90 seconds, 1 Sec./ step | Adjustable, 2~90 seconds, 1 Sec./ step |
| 18. | OFF Time | Adjustable, 2~90 seconds, 1 Sec./ step | Adjustable, 2~90 seconds, 1 Sec./ step |
| | New Device | Predicate Device | |
| 1. 510(k) Number | To be Assigned | K071951 | |
| 2. Device Name, Model | MH8200/MH6200 Combo | TENS/EMS/EV-804 TENS, Comfy TENS | |
| 3. Manufacturer | MEDIHIGHTEC MEDICAL LTD. | Everyway Medical Instruments Ltd. | |
| 4. Power Source(s) | 9 volt battery type 6F22 | 9 volt battery type 6F22 | |
| 5. Number of Output Modes | 5 | 5 | |
| 6. Number of Output Channels<br>- Synchronous or Alternating?<br>- Method of Channel Isolation | 2<br>Synchronous and Alternating<br>By TRANSFORMER | 2<br>Synchronous and Alternating<br>By TRANSFORMER | |
| 7. Waveform | Asymmetrical biphasic pulse | Asymmetrical biphasic pulse | |
| 8. Shape | Rectangular | Rectangular | |
| 9. Maximum Output Voltage (+/- 20%) | 40.8V @ 500 Ohm<br>77.6V @ 2000 Ohm<br>102V @ 10000 Ohm | 50V @ 500 Ohm<br>150V @ 2000 Ohm<br>194V @ 10000 Ohm | |
| 10. Maximum Output Current (+/- 20%) | 81.6mA @ 500 Ohm<br>38.8mA @ 2000 Ohm<br>10.2mA @ 10000 Ohm | 100mA @ 500 Ohm<br>75mA @ 2000 Ohm<br>19.4mA @ 10000 Ohm | |
| 11. Pulse Width | The range of Pulse Width control is between 50µS and 300µS. | The range of Pulse Width control is between 50µS and 300µS. | |
| 12. Frequency | The range of Pulse Rate control between 2 Hz and 150 Hz. | The range of Pulse Rate control between 2 Hz and 150 Hz. | |
| 13. Maximum Phase Charge | 24 micro-coulombs @ 500 Ohm | 30 micro-coulombs @ 500 Ohm | |
| 14. Maximum Current Density | $1.06mA_{rms}/cm^2$ @ 500 Ohm | $1.33mA_{rms}/cm^2$ @ 500 Ohm | |
| 15. Maximum Power Density | $2.4mW_{rms}/cm^2$ @ 500 Ohm | $3.7mW_{rms}/cm^2$ @ 500 Ohm | |
| 16. Burst Mode | a. Pulses per buster: 9<br>b. Burst per second: 2<br>c. Burst duration: 100 mSec.<br>d. Duty Cycle: 200 mSec. | a. Pulses per buster: 10<br>b. Burst per second adjustable, 0.5~5<br>c. Burst duration: 100 mSec.<br>d. Duty Cycle: adjustable 50~500 mSec. | |
{4}------------------------------------------------
{5}------------------------------------------------
Non-Clinical Testing Result: New device MH8200/MH6200 vs. Predicate device EV-804
According to above results of manufacture testing demonstrate that the output characteristics of the new devices are substantially equivalent to those of the three predicate devices and any differences between the devices do not pose any new questions of safety and effectiveness.
# 17. Safety and Effectiveness of the device:
This new series is safe and effective as the predicate devices cited above.
(1) The new series device comply with safety standard EN60601-1:1990, Including Amendment A1 (1993) and A2 (1995), A13 (1996), IEC60601-1:1988, Including Amendment A1 (1991) and A2 (1995). Test report number is UT96026 and UT96027, See Chapter 11, C.1.
(2) The new series device comply with EMC compliance standard EN60601-1-2:2001 CISPR 11:1997+A1:1999+A2:2002 Group 1:Class B,
IEC61000-4-2:1995+A1:1998+A2:2000, IEC61000-4-3:2002,
IEC61000-4-8:1993+A1:2000. Test report number is EM/2007/20045 and EM/2007/20046, See Chapter 11, C.2.
{6}------------------------------------------------
(3) Regarding leakage current, please refer to safety report UT96026 and UT96027, Page 15, Clause 19 and appended table 19.4g. 19.4h and 19.4.h3. See Chapter 11, C.1.
#### 18. Conclusions:
Our MH6000/MH8000 TENS and EMS Combo unit integrates EMS and TENS functions with 7 stimulation modes.
S Mode (Synchronous) and A Mode (Alternative) designed for EMS,
N Mode (Normal), M Mode (Rate and Width Modulation), B Mode (Burst), SD Mode
(Width Modulation) and MR Mode (Rate Modulation) designed for TENS.
Press MODE button to select only one stimulation mode, therefore TENS and EMS are used independently for the combination devices.
New devices and Predicate devices are the same in the:
- 1. Power Source:
- 2. Number of Output Modes;
- 3. Number of Output Channels;
- 4. With Microprocessor Control;
- 5. Standards;
- 6. Pulse Width control;
- 7. Pulse Rate control;
- 8. ON tome, OFF time, Ramp time; and
- 9. Waveform Shape.
The difference of new devices and Predicate devices is as follows:
- 1. Patient Leakage Current;
- 2. Weight:
- 3. Dimensions;
- 4. The Maximum Output Voltage of the new devices is smaller than the Predicate devices;
- 5. The Maximum Output Current of the new devices is smaller than the Predicate devices;
- 6. The Maximum Phase Charge of the new devices is smaller than the Predicate devices;
7. The Maximum Current Density of the new devices is smaller than the Predicate devices; and
8. The Maximum Power Density of the new devices is smaller than the Predicate devices.
The maximum output Charge of the new devices is smaller than the predicate devices, so the maximum output Charge will not exceed the safe range.
The minimum phase charge for effectiveness of Transcutaneous Neural Stimulation is 3 micro-coulombs. Please refer to: M. Linzer, and D.M. Long, "Transcutaneous Neural Stimulation for Relief of Pain," IEEE Transactions on Biomedical Engineering, Vol.
{7}------------------------------------------------
References:
The EMS designed for (1) Relaxing muscle spasms, (2) Increasing local blood circulation, (3) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, (4) Muscle re-education, (5) Maintaining or increasing range of motion, and (6) Preventing or retarding disuse atrophy.
The References and studies are as follows:
1. Meryl Roth Gersh, M.M.Sc., P.T., Electrothcrapy in Rehabilitation, F.A. Davis Company, 1992
2. Val Robertson, Alex Ward, John Low and Ann Reed, Electrotherapy Explained Principles and Practice, Fourth Edition, Elsevier Ltd, 2006
3. Joseph Kahn, PhD, PT, Principle and Practice of Electrotherapy, Fourth Edition, Elsevier (Singapore) Pte Ltd, 2004
4. Sheila Kitchen MSc PhD DipPT MCSP, Electrotherapy Evidence-Based Practice, Eleventh Edition, Elsevier (Singapore) Pte Ltd, 2004
5. Michelle H. Cameron, M.D., P.T., O.C.S., Jennifer A. Rohl, B.S. , Physical Agents in Rehabilitation, Elsevier Ltd, 2006
| | | New Device | | Predicate Device | |
|----|--------------------|----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | 510(k) Number | To be Assigned | | K071951 | |
| 2. | Device Name, Model | MH8000 TENS/EMS Combo<br>MH8100 EMS<br>MH8200 TENS<br>MH6000 TENS/EMS Combo<br>MH6100 EMS<br>MH6200 TENS | | EV-806 TENS/EMS, Comfy Stim<br>EV-805 EMS, Comfy EMS<br>EV-804 TENS, Comfy TENS | |
| 3. | Manufacturer | MEDIHIGHTEC MEDICAL CO., LTD. | | Everyway Medical Instruments Ltd. | |
| 4. | Pulse Width | The range of Pulse Width control is between 50uS and 300uS. | | The range of Pulse Width control is between 50uS and 300uS. | |
| 5 | Frequency | The range of Pulse Rate control between 2 Hz and 150 Hz. | | The range of Pulse Rate control between 2 Hz and 150 Hz. | |
| 6 | Stimulation Modes | | | | |
| | For EMS | S Mode<br>(Synchronous) | Stimulation of both channels will occur simultaneously | S Mode<br>(Synchronous) | Stimulation of both channels will occur simultaneously |
| | | A Mode<br>(Alternative) | Stimulation of Channel 1 and Channel 2 will occur alternately. | A Mode<br>(Alternative) | Stimulation of Channel 1 and Channel 2 will occur alternately. |
| | For TENS | N Mode<br>(Normal) | Continuous output, pulse rate and pulse width are adjustable. | N Mode<br>(Normal) | Continuous output, pulse rate and pulse width are adjustable. |
| | | M Mode<br>(Rate and Width Modulation): | Per cycle time is 1 second. Each cycle contains two phases. During the first 0.5 sec. phase, the Pulse Width runs at 50% of its width setting and Pulse Rate runs at its rate setting, then during the second 0.5 sec. phase, the Pulse Rate runs at 50% of its rate setting and Pulse width runs at its width setting. The cycle is then repeated. | M Mode<br>(Rate and Width Modulation): | Per cycle time is 1 second. Each cycle contains two phases. During the first 0.5 sec. phase, the Pulse Width runs at 50% of its width setting and Pulse Rate runs at its rate setting, then during the second 0.5 sec. phase, the Pulse Rate runs at 50% of its rate setting and Pulse width runs at its width setting. The cycle is then repeated. |
| | | B Mode<br>(Burst) | Burst rate fixed =21Hz<br>Pulse width adjustable, 50~260uS<br>Frequency fixed = 100 Hz | B Mode<br>(Burst) | Burst rate adjustable, 0.5~5 Hz<br>Pulse width adjustable, 50~260uS<br>Frequency fixed = 100 Hz. |
| | | SD Mode<br>(Width Modulation): | Pulse Width is modulated over a period of 3 seconds from its original setting value down to 60% of its setting, then modulated over another 3 seconds period at its original value. The cycle is then repeated. Per cycle time is 6 seconds. | SD1 Mode<br>(Width Modulation): | Pulse Width is modulated over a period of 3 seconds from its original setting value down to 60% of its setting, then modulated over another 3 seconds period at its original value. The cycle is then repeated. Per cycle time is 6 seconds. |
The Output Specifications. New devices vs. Predicate devices
{8}------------------------------------------------
| | | MR Mode<br>(Rate<br>Modulation) | Pulse Rate is modulated over a<br>period of 3 seconds from its<br>original setting value down to<br>60% of its setting, then modulated<br>over another 3 seconds period at<br>its original setting. The cycle is<br>then repeated. Per cycle time is 6<br>seconds. | SD2 Mode<br>(Rate<br>Modulation) | Pulse Rate is modulated over a period of 3<br>seconds from its original setting value down to<br>60% of its setting, then modulated over another 3<br>seconds period at its original setting. The cycle is<br>then repeated. Per cycle time is 6 seconds. |
|---|-------------------------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 7 | Maximum<br>Output Voltage<br>(+/- 20%) | 40V @ 500 Ohm | | 50V @ 500 Ohm | |
| 8 | Minimum<br>Output Voltage<br>(+/- 20%) | 0.5V @ 500 Ohm | | 0.5V @ 500 Ohm | |
| 9 | Maximum<br>Output Current<br>(+/- 20%) | 80mA @ 500 Ohm | | 100mA @ 500 Ohm | |
| | 10 Minimum<br>Output Current<br>(+/- 20%) | 1mA @ 500 Ohm | | 1mA @ 500 Ohm | |
| | 11 Maximum<br>Phase Charge | $40V \div 500ohm \times 300uSec$ =24 micro-coulombs @<br>500 Ohm | | $50V \div 500ohm \times 300uSec$ =30 micro-coulombs @ 500 Ohm | |
| | 12 Minimum<br>Phase Charge | $0.5V \div 500ohm \times 50uSec$ =0.05 micro-coulombs @<br>500 Ohm | | $0.5V \div 500ohm \times 50uSec$ =0.05micro-coulombs @ 500 Ohm | |
The Phase Charge (micro-coulombs) =Amplitude (V) = Load (ohm) × Pulse Duration (uS). The amplitude is the size of the stimulus applied. It is measured either as the stimulus voltage or current. Stimulus voltages used clinically are typically in the range 10V to 100V or more. Stimulus currents are typically in the range 10mA to 100mA for clinical treatments. The amplitude is usually set according to the patient's level of comfort and to the aims of a treatment (refer to References 2., 3. and 4.). Recommended parameters for electrical stimulation to produce muscle contractions: Pulse duration 150 to 200uS for small muscles, 200 to 350uS for large muscles. Frequency 35 to 80Hz, Amplitude is "to contraction" (refer to References 5.). Stimulus parameters used for treatment: Pulse duration 200 to 300uS. Frequency 24 to 80Hz, Amplitude is 30 to 80mA (refer to References 1.).
According to the References, the amplitude is usually set according to the patient's level of comfort and to the aims of a treatment, therefore the effectiveness of EMS is decided by frequency and Pulse duration. The new devices have the same frequency and Pulse duration with the predicate devices (see the Output Specifications): and. The phase charge of the new devices is adjustable between 0.05 to 24 micro-coulombs enough demand, therefore the new devices and predicate devices have the same effectiveness.
According to above conclusions, the new devices are substantially equivalent to the predicate devices and any differences between the devices do not pose any new questions of safety and effectiveness.
3-9
{9}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/9/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular fashion around the symbol. The caduceus is depicted in black, and the text is also in black, set against a white background.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Medihightec Medical Company Limited % Underwriters Laboratories, Inc. Mr. Jeffrey D. Rongero 12 Laboratory Drive Research Triangle, North Carolina 27709
OCT 21 2008
Re: K082514
> Trade/Device Name: MH8000 TENS/EMS Combo, MH6000 TENS/EMS Combo, MH8100 EMS, MH6100 EMS, MH8200 TENS, MH6200 TENS Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: GZJ, IPF Dated: October 2, 2008 Received: October 6, 2008
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{10}------------------------------------------------
Page 2 - Mr. Jeffrey D. Rongero
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Millerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
# Indication for Use
510(k) Number (if known):
Device Name:
The MH8000 TENS/EMS Combo, MH6000 TENS/EMS Combo, MH8100 EMS, MH6100 EMS, MH8200 TENS, MH6200 TENS. They are Transcutaneous Electrical Nerve Stimulator (TENS) and Electrical Muscle Stimulator (EMS).
Indications for Use:
APPLICABLE FOR MH8000 (TENS/EMS COMBINATION UNIT) AND MH6000 (TENS/EMS COMBINATION UNIT) AND MH8200 (TENS) AND MH6200 (TENS): This device is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is indicated for use for (1) Symptomatic relief and management of chronic (long term) intractable pain and (2) adjunctive treatment in the management of post surgical and post traumatic pain problems.
APPLICABLE FOR MH8000 (TENS/EMS COMBINATION UNIT) AND MH6000 (TENS/EMS COMBINATION UNIT) AND MH8100 (EMS) AND MH6100 (EMS): This device is an electrically powered muscle stimulator indicated for use for medical purposes to repeatedly contract muscles by passing electrical currents through electrodes contacting the affected body area. In particular, this device is indicated for use for (1) Relaxing muscle spasms, (2) Increasing local blood circulation, (3) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, (4) Muscle re-education, (5) Maintaining or increasing range of motion, and (6) Preventing or retarding disuse atrophy.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Milliman
Division of Gener and Neurological Dev
510(k) Number
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.