MEDIGARD BLOOD COLLECTION DEVICE

K082511 · Medigard Limited · FMI · Jan 16, 2009 · General Hospital

Device Facts

Record IDK082511
Device NameMEDIGARD BLOOD COLLECTION DEVICE
ApplicantMedigard Limited
Product CodeFMI · General Hospital
Decision DateJan 16, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5570
Device ClassClass 2

Intended Use

Medical device to enable withdrawal of blood samples from a patient. The device is designed to prevent accidental needle stick injuries. The needle is retracted via a retraction tube.

Device Story

Medigard Blood Collection Device consists of blood collection tube holder and evacuated sharps collection tube; used by healthcare professionals in clinical facilities for blood sampling. Device facilitates blood withdrawal into standard evacuated tubes; does not contact patient. Safety mechanism: after blood collection, user inserts evacuated sharps collection tube into holder; this action releases needle hub, allowing needle to retract into tube via vacuum; tube automatically locks, preventing needle access. Prevents accidental needle stick injuries to healthcare workers. Benefits: enhanced safety during and after blood collection procedures.

Clinical Evidence

Bench testing and simulated use testing performed. No clinical data provided.

Technological Characteristics

Plastic construction. Vacuum-based needle retraction mechanism. Non-sterile. Device does not contact patient. Manual activation by user.

Indications for Use

Indicated for use by healthcare professionals to facilitate blood withdrawal from patients into evacuated blood collection tubes while preventing accidental needle stick injuries.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K08,25'11 # 510(k) Summary 807.92(c) JAN 1 6 2009 ### SPONSOR 807.92(a)(1) MEDIGARD LIMITED Company Name Suite 14A, Tedder Terraces Company Address 26-30 Tedder Avenue Main Beach, 4217 Queensland Australia 61 7 5528 0370 Telephone Contact Person Dr Peter W Clark Summary Preparation Date August 25, 2008 ### DEVICE NAME 1. . Trade Name Common/Usual Name Classification Name Regulation Number: Product Code Device Class ## Blood Collection Tube Holder Needle, Hypodermic, Single Lumen CFR21 880 5570 FMI Class II Medigard Blood Collection Device ### PREDICATE DEVICE Legally Marketed Equivalent Device Company Safe T Medical Devices Product Blood Collection Tube Holder Vanish Point Small Tube Adapter ' Collection Device 510(k) # K 971763 ### DEVICE DESCRIPTION The Medigard Blood Collection Device Consists of two Parts 1) a blood collection tube holder and 2) an evacuated sharps collection tube Netther part comes into contact with the patient and no fluids flow through the Medigard Blood Collection Device to the patient Normal standard blood collection tubes and blood collection needles are used to collect blood samples as per normal standard procedures The device is intended to be used to provide a safe and reliable method of collection of blood samples from a patient using standard evacuated blood collection tubes The device is designed to prevent accidental needle stick injury At the end of the procedure, the needle is retracted into the Medigard evacuated sharps collection tube ### 807.92(a)(4) 807.92(a)(2) 807.92(a)(3) 16 {1}------------------------------------------------ ## DEVICE INTENDED USE 7 ### 807,92(a)(5) The function of the Medigard Blood Collection Device is to provide a safe and reliable method for facılıtatıng blood withdrawal from a patient into evacuated blood collection tubes without exposing the healthcare worker to an accidental needle stick injury ### COMPARISON OF TECHNICAL CHARACTERISTICS 807.92(a)(6) | Features | Medigard Limited | Retractable Technology Inc.<br>K971763 | |-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Proprietary Name | Medigard Blood Collection Device | Vanish Point® Blood Collection<br>Tube Holder | | Intended Use | Medical device to enable to<br>withdrawal of blood samples from a<br>patient The device is designed to<br>prevent accidental needle stick<br>injuries during and after the<br>procedure The needle is retracted via<br>a vacuum, and can not be accessed<br>after retraction | Medical device to enable<br>withdrawal of blood samples<br>from a patient The device is<br>designed to prevent accidental<br>needle stick injuries The needle<br>is retracted via a spring<br>mechanism | | Material Composition | Plastic | Plastic and metal | | Device Components | Two Components<br>1 Collection tube for blood collection<br>2 Vac tube for safety feature | One Component<br>1 Collection tube with spring<br>mechanism | | Product Design - Safety | Safety feature activated by user | Safety feature activated by user | | Feature Mode of | Needle retracted via vacuum and can | Needle retracted by a spring | | Operation | not be accessed after retraction | mechanism | | Product Design -<br>Safety Feature<br>Activation Mechanism | A separate (Medigard) evacuated<br>collection tube is inserted into the<br>Tube Holder to initiate activation<br>This couples with & releases the hub<br>(containing needle) allowing it to be<br>retracted into the tube via vacuum<br>The tube is automatically locked into<br>the holder preventing further access to<br>the needle | A cap at the opening of the Tube<br>Holder is pressed closed to<br>dislocate the hub (containing<br>needle) This allows the spring to<br>drive the hub into the rear of the<br>holder thereby sheathing the<br>needle | | Disposable volume i.e.<br>Space occupied in a<br>sharps container | Approx 28cc | Approx 35cc | | Sterilization | Non-sterile | Non-sterile | | Biocompatibility | Device does not contact patient | Device does not contact patient | ## Substantial Equivalence Discussion of Sımilarities and Differences: The Medigard Blood Collection Device is similar to the Vanish Point Blood Collection Tube Holder in - Intended Use prevention of needle sticks by the safe retraction of the needle . - Materials plastic (Medigard), plastic/metal (Vanish Point) . - Design Retraction of Needle ● {2}------------------------------------------------ - Where used Healthcare facılıtıes o - Target Population Healthcare professionals . - Performance Testing -Bench Testing and Simulated Use Testing o ### Differences 1-7 - · Method of Retraction Vacuum (Medigard) Spring mechanism (Vanısh Poınt) The Medigard Blood Collection Device introduces no new questions concerning the safety or effectiveness and is thus substantially equivalent to the predicate device ### SAFETY and EFFECTIVENESS 807.92(b) Bench Testing and Simulated Use testing were performed {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows a logo with a stylized eagle design. The eagle is composed of three thick, curved lines that suggest the wings and body. The logo is encircled by text that reads "HUMAN SERVICES USA". The text is arranged in a circular fashion around the eagle, with the words "HUMAN" and "SERVICES" at the top and "USA" at the bottom. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 1 6 2009 Medigard Limited C/O Mr E J Smith Smith Associates 1468 Harwell Avenue Crofton, Maryland 21114 Re K082511 > Trade/Device Name Medigard Blood Collection Device Regulation Number 21 CFR 880 5570 Regulation Name Hypodermic Single Lumen Needle Regulatory Class II Product Code FMI Dated December 7, 2008 Received December 11, 2008 Dear Mr Smith We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the undications for use stated in the enclosure) to legally marketed predicate devices marketed in Interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provise of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA) You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 800 In addition, FDA may publish further announcements concerning your device in the Federal Register {4}------------------------------------------------ Page 2 - Mr Smith Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807), labeling (21 CFR Part 801), good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115 Also, please note the regulation entitled, "Misbranding by 1eference to premarket notification" (21CFR Part 807 97) For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrio's (OSB's) Division of Postmarket Surveillance at 240-276-3474 For questions regarding the 1eporting of device adverse events (Medical Device Reporting (MDR)), please contact the DIVISION of Surveillance Systems at 240-276-3464 You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http //www fda gov/cdrh/industry/support/index html Sincerely yours, Anthony E. Watson Jr. Ginette Y Michaud, M D Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure - - {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K082511 Device Name: Medigard Blood Collection Device Indications For Use: Medical device to enable withdrawal of blood samples from a patient. The device is designed to prevent accidental needle stick injuries. The needle is retracted via a retraction tube. Prescription Use V (Part 21 CFR 801 Subpart D) . AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) "," (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 | (Division Sign-Off) | | |---------------------|--| |---------------------|--| Division of Anesthesiology, General Hospital Infection Control, Dental Devices | 510(k) Number: | K082511 | |----------------|---------| |----------------|---------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...